Title: Monte Carlo Studies on
1Monte Carlo Studies on Possible Calibration
Sources
Kevin Kröninger, MPI für Physik GERDA
Collaboration Meeting, DUBNA, 06/27 06/29/2005
2Introduction
- External sources in the GERDA setup are needed
in order - calibrate the detectors
- Monte Carlo study done with different sources to
test - Efficiencies
- Allowed activities (no additional background
source) - Homogenity
- Accuracy of calibration
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
3Geometry I
- Phase II setup with 3 x 7 crystals and 3 x 6
fold segmentation - (378 segments in total)
- Propose 3 sources at marked
- positions and flexible in z
- For MC study only one source
- is simulated
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
4Geometry II
- Sources consist of small plastic sphere within a
steel shield - Placed inside lead cylinder with 60 opening
Lead cylinder
Source
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
5Sources
- List of source used in the simulation
- 22Na T1/2 950.8 d ß Lines 511 keV, 1.3 MeV
- 26Al T1/2 0.71 My EC, ß Lines 511 keV, 1.8
MeV - 57Co T1/2 271 d EC Lines 122 keV, 136 keV
- 60Co T1/2 5.3 y ß- Lines 1.2 MeV, 1.3 MeV
- 88Y T1/2 106.65 d ß Lines 511 keV, 0.9 MeV,
- 1.8 MeV
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
6Efficiencies
- Total efficiency event efficiency x peak
efficiency - Event efficiency is the fraction of events that
deposits energy - in at least one segment
- Peak efficiency is calculated for each peak and
corresponds - to the fraction of events within a 10 keV
window around the - peak (averaged over all segments)
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
7Activities
- Demanding an error of 3 1000 events are needed
in the - peak region.
- The minimum activity is
- Allowing a background index of 10-3
cnts/kg/keV/y the - maximum activity is
A N / (T3600e)
A activity in Bq N number of events T
run duration in h e efficiency
A 0.001 / (3.15107r/10)
A activity in Bq r suppression factor
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
8Control Plots
Source
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
9Example Co-60
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
10Segment Study
- Calibration should be done homogenious,
therefore test - different regions (central, left / right
neighboring segments) - Repeat study with 120 opening angle of lead
cylinder
60 opening angle All Central Neighboring
Avg. peak efficiency Min. peak efficiency Avg. mean value (fit) 0.95 0.04 1173.2 0.2 keV 2.20 0.26 1.70 1173.0 0.1 keV 1.90 0.12 1.70 1173.0 0.1 keV
120 opening angle All Central Neighboring
Avg. peak efficiency Min. peak efficiency Avg. mean value (fit) 0.92 0.03 1172.3 0.4 keV 2.10 0.29 1.50 1173.0 0.1 keV 2.00 0.14 0.91 1173.0 0.1 keV
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
11Efficiencies and Activities
Isotope T1/2 Energy MeV Efficiency Amin Bq Amax kBq
22Na 950.8 d 0.511 1.275 0.13 0.05 213.7 555.6 - -
26Al 0.7 My 0.511 1.809 0.30 0.86 92.6 308.6 - -
57Co 271.8 d 0.122 / 0.136 0.05 555.6 -
60Co 5.3 y 1.173 1.333 0.13 0.09 213.7 308.6 31.7 31.7
88Y 106.7 d 0.511 0.898 1.836 0.38 0.18 0.10 73.1 154.3 277.8 31.7 31.7 31.7
Kevin Kröninger, MPI München GERDA
Collaboration Meeting DUBNA, 06/27 06/29/2005
12Conclusion
- Propose 3 calibration sources with one run of 1
hour per - crystal level
- Simulated 5 different isotopes and calculated
limits on the - activities
- 60Co, 22Na and 88Y are good candidates for
calibration sources - Note GSTR-05-005 is avaiable at MPI-K website