Title: Current techniques for measuring
1Current techniques for measuring
Muffler Transmission Loss
2Why Transmission Loss?
- To understand how effectively an acoustical
treatment can block the incident sound when
designing a mechanical system - Transmission Loss quantifies the acoustical
treatment for engineering application
3Parameters for acoustic performance
of a Muffler
- Noise reduction (NR)SPL difference across the
muffler - Insertion loss (IL)SPL difference outside the
system with and without the muffler present - Transmission Loss (TL)Sound power level
difference between the incident and the
transmitted wave assuming anechoic termination
given by -
4Methods for measuring TL
- Decomposition method
- Two Source method using 4 microphones
- Two Load method
5Decomposition Method
Incident Auto Power Spectrum
Reflected Auto power spectrum
S33 Transmitted Auto Power Spectrum
6 Contd..Decomposition Method
- Muffler Transmission Loss is given by
Where Wi Incident sound power Wt
Transmitted sound power
- Inlet Sound Pressure can be decomposed into
incident and reflected wave and
respectively - Using Decomposition theory
7 Contd..
Decomposition Method
- where and are the Auto spectra of the
total acoustic pressure at 1 2 resp. - are the real and imag part of
cross spectrum between points 1 2 - K wave number
- is the distance between the two
microphones - The rms amplitude of incident sound wave
and transmitted wave is given by -
pt -
- where incident transmitted are the
rms pressure amplitudes - and S33 is the auto power spectrum
8 Contd.. Decomposition
Method
- Now the sound power of incident and reflected
wave is given by - and
-
- where are the muffler inlet and
outlet tube areas - Therefore ,Transmission Loss is given by
9 Contd..Implementation of Decomposition Method
- Find auto power cross power spectrum at the two
microphones 1 2 - Utilize decomposition theory to find incident
auto power spectra - Find rms amplitude by taking the square root of
the incident auto power spectra and transmitted
auto power spectra (measured directly from the
microphone 3) - Plug in the calculated rms values in the TL
equation
10Two Source Method
11 Contd..Two Source Method
- Based on Transfer matrix approach-relation
between input pressure velocity to the output
pressure and velocity -
- Any acoustical element can be modeled by its
four pole parameters which is given by - where are the sound pressure
amplitudes at the inlet outlet -
- are the particle velocity
amplitudes at the inlet and the outlet - are the 4 pole
parameters of the system
12 Contd..Two Source Method
- For Configuration a
- The 4 pole equation for the element 2-3 is given
by - where subscript a refers to configuration a
- Also, 4 pole equations for elements 1-2 3-4 is
given by - where are the microphone spacings
for elements 1-2 3-4 -
13 Contd..Two Source Method
- Combining all the equations for 1-2,3-4 2-3
gives - For Configuration b Moving sound source to
the other side - For this configuration, the equation for
element 3-2 is given by - where
14 Contd.. Two Source Method
Now the combined equation for configuration b
is given by
Now using these equations one can obtain 4 pole
parameters given by
15 Contd.. Two Source
Method
Also, transfer function ( ) is the ratio
of cross and auto spectrums
Therefore, Transmission Loss can be expressed by
16 Contd..Implementation of Two Source
Method
- Use Two microphones with random excitation (or
white noise) - Obtain all transfer functions by moving one
microphone and using the other as a reference - Put the obtained transfer functions in the
equations shown above to find the 4 pole
parameters of the transformation matrix - Find the TL by plugging the 4 pole parameters
and the measured cross sectional areas of the
tube in the equation given above
17Two Load Method Setup
18 Contd..Two Load method
- Similar to the Two Source method with little
changes as follows - Instead of moving the sound source, two end
conditions are applied to find 2 additional
equations in order to solve the complete transfer
matrix - Changing end conditions changes the impedance
at the termination from to - Two loads can be 2 different length tubes, a
single tube with without absorbing material or
even 2 different mufflers
19 Contd..Measurement setup for Two Load method
20 Contd..Calculations using Cross Spectrums for
Two Load method
- Involves two basic measurements with two
different terminations - Terminations must be very different
- Generally open ended or anechoic (o) closed
ended or reverberant (c) terminations are used - The equations utilized are as follows
and
where A B and C D are the forward and
backward complex pressure amplitudes wave in the
source tube and receiving tube respectively with
a b denoting two different end conditions
21 Contd..Two Load Method
- The 4 coefficients A,B,C,D can be calculated by
the following relations
where P1,P2,P3,P4 are the measured Sound Pressures
- Now, using Cross spectrum and FRFs to minimize
noise in the signal, we can obtain the following
equations
where are the
cross power spectrums using o as a reference
22 Contd.. Two Load Method
- The Transmission Loss coefficient is given by
the ratio of amplitude A of incident wave and
amplitude C of the transmitted wave assuming no
reflection i.e D 0 - Therefore, from the above relations
where is the
first element of the transfer matrix
23 Contd..Limitations of Cross Spectrums
calculations and two load method
- Must require two different terminations
-
- Difficult to obtain ideal anechoic termination
- There seems to be a flaw in the calculations of
TL since they assume that the reflection to be
zero i.e D 0 but we are also using the same
equation with reverberant termination. - Prior to taking actual measurements, it
requires complete measurements without inserting
the acoustical material to assure that residual
TL is much less than the measured TL - If the numerical value of the difference in
denominator becomes smaller than the absolute
value of the absolute nos., then the solution
becomes unstable
24Comparison of all 3 Methods
25Advantages of Two source method over other methods
- The above results indicate the limitation of
decomposition method in the absence of anechoic
termination - Decomposition method does not lead to 4 pole
parameters of muffler - Unlike two load method, two source method does
not require any termination material at the end - Although Two load method is easier to employ but
better results require two different loads
26General Procedure
- Impedance tube with small diameter (29mm) can
be utilized for measuring TL of Briggs Stratton
Muffler - The apparatus is designed to measure TL and
other acoustics properties using the following
ASTM E-1050 standard - Working frequency range
where - size and spacing of microphones
- Location of microphones minimum of 3 tube
diameter from source to first microphone - Sound source- type signal random noise having
uniform spectral density - Length of tube should be large for plane wave
propagation - Determining the individual microphone
sensitivity (mV/Pa) - Calibrate the microphones correctly
- Means of correcting the measured transfer
function data for mismatch in both amplitude and
phase responses of measurement channels -
27Calibration of microphones setup
28 contd..Calculations for
Calibration of microphones
- Place a highly absorptive material to prevent
strong acoustics reflections and to obtain most
accurate correction factor possible - Measure Transfer functions in 2 configuration 1
and 2 as shown
- Compute the calibration factor
representing the amplitude and phase mismatches
where
29Schematic diagram for STL measurement
30Impedance Tube setup
Small Impedance Tube Setup with 4 microphones
31References
- Z. Tao and Seybert, A.F., A review of current
techniques for measuring muffler TL - Seybert, A.F. and Ross, D.F., Experimental
Determination of Acoustic Properties Using a Two
microphone Random Excitation Technique, J.
Acoust. Soc. Am., 61, 1362-1370 (1977) - Munjal, M.L. and Doige A.G., Theory of a Two
Source-location Method for Direct Experimental
Evaluation of the Four-pole Parameters of an
Aeroacoustic Element, Journal of Sound and
Vibration, 141(2), 323-333 (1990) - ASTM standard, E1050-98, Standard Test Method
for Impedance and Absorption of Acoustical
Material Using a Tube, Two Microphones and a
Digital Frequency Analysis System, (1998)
32?