Title: Sanja Risticevic Chem 323 Poster Presentation
1Sanja RisticevicChem 323 Poster Presentation
- Quadrupole Ion Trap Mass Spectrometry
2Introduction
- Ion trap mass spectrometry has recently developed
very rapidly - It is a high performance technique and one of the
leading tools in the chemistry and biochemistry
fields - It can be used for measurements of very high
mass/charge ratios - Ion trap mass spectrometry has high resolution
capabilities and also excellent non-destructive
broad-band Fourier transform capabilities
3- Advantages of using ion trap
- High sensitivity
- Capable of high performance
- Compactness and mechanical simplicity
- Ion/Molecule reactions can be studied for
mass-selected ions and the reaction time can be
varied in the ion trap. Therefore, the kinetics
and equilibrium of ion-molecule reactions can be
studied - High resolution for slow scans
- The resonance experiments are applicable in the
study of ions that have high m/z ratios - Fourier transform techniques provide
non-destructive detection - MS/MS experiments are possible (multiple stage
mass spectrometry). In these experiments,
individual ions can be examined in a mixture of
ions. The ions of interest are isolated by their
characteristic m/z values and they dissociate.
The product ions are then analyzed in a second
mass measurement step.
4- Ions are subjected to stabilizing and
destabilizing forces applied by the field. The
forces occur in three dimensions. - To record the mass spectra, the quadrupole ion
trap can be operated in the mass selective
instability scan mode which is the usual mode of
operation - Ions of a given m/z value undergo stable motion
in the trap - A helium buffer gas is used to remove the kinetic
energy from ions and cause them to occupy the
central region of the trap - The ion trap can hold up to 105-106 ions before
columbic repulsions reduce the mass resolution - Ions are trapped in the system consisting of
three electrodes which have hyperbolic surfaces - The central electrode is the rotationally
symmetrical ring electrode and it is located
between two end-cap electrodes of the same
cross-section
5- The diagram which illustrates the ion trap
instrumentation. r0 is the internal radius of
the ring electrode and z0 is the closest distance
from the center to the end-cap electrodes.
The electrodes are aligned and isolated using
ceramic posts.
6Potential for trapping ions
- The quadrupole potential surface is saddle-shaped
when the phase of the voltage signal is positive - The ion shown is on a potential downhill in the
z-direction and is accelerated from the center of
the device
Potential used for trapping ions in the radial
direction. The ion is accelerated away from the
trapping center in the axial direction.
7- When the voltage field changes sign, this ion is
accelerated toward the center of the trap. The
ions are trapped in both the r and z directions. -
Potential used for trapping ions in the axial
direction. The ion is accelerated towards the
trap center.
8- Trapped ions have characteristic frequencies of
oscillation known as secular frequencies. The
frequency in the r-direction is half the
frequency in the z-direction. - An additional potential of frequency equal to the
secular frequency causes ions to absorb more
kinetic energy. Ions are activated in the
z-direction when the signal is applied between
the two end-cap electrodes. When the signal is
strong, it is possible for the ions to be ejected
from the trap in the z-direction. - The population of ions in the trap can also be
controlled since particular ions can be excited
so that they dissociate or get ejected
9- Ions move along the qz axis until qz 0.908.
Here the ions become unstable in the normal
mass-selective instability mode of operation and
reach the boundary. If the supplementary
resonance frequency can be varied, qz becomes a
variable. Thus, qz can be lowered by modulating
the ion motion at a chosen frequency. For this
purpose, a dipolar electric field is applied
across the end-cap electrodes. This is called
the resonance ejection experiment which can be
used to increase m/z. The ions of a particular
m/z value pick up translational energy and exit
the trap through a hole in the end-cap electrode.
These ions exit the ion trap in the sequence of
m/z values and reach an external detector.
10- The resonance experiment can also be used to
allow the fragmentation of the specified ions
only. - Resonance experiments are usually performed by
applying a mixture of frequencies. Thus, the
ions of different m/z values are manipulated
simultaneously. The technique used for this is
called SWIFT (stored waveform inverse Fourier
transform). - This technique is also used for the ion
population control. For example, in trace level
analysis, the trap can be filled with the analyte
ions only. Thus, the ions can be stored
selectively which is very applicable in the
ultra-trace-level analysis of volatile organic
compounds. This analysis can be performed at
levels as low as parts-per-quadrillion (pg/L). - A Brief Word on Non-Destructive Ion Detection
- The population of a single ion can be measured
multiple times - Achieved by impulsive excitation of a group of
trapped ions of different m/z values - The ion image currents are induced on a small
detector electrode. This electrode is isolated
from the end-cap electrodes. - The image currents are measured using a
differential preamplifier, filter and amplifier.
The image currents are then Fourier analyzed and
the broad-band spectra can be obtained.
11Mass Selective Instability Scan
- Ions of different m/z values arrive at the
detector at different times - When the voltage is increased across the ring
electrode, the ions of high m/z are ejected.
When the voltage is changed too fast, the loss of
resolution can result. Therefore, the rate of
voltage change should be slow to ensure the high
resolution. - A zoom-scan mode can be applied in these
circumstances to provide a better study of the
ions which have m/zlt10 dalton/charge - The high resolution of the instrument helps to
resolve the isotopic forms of the multiply
-charged ions
The figure shows the zoom scan of the 4 charged
state of rat interleukin-8. There is one-forth
m/z unit difference between carbon isotopes.
These carbon isotopes are different in mass by
one dalton. This is electrospray ionization mass
spectrum, but it shows how the zoom-scan mode of
the ion trap can be useful in the isotopic study.
12MS/MS Experiment
- The additional sequence of operations in the scan
function is used. The ionization is followed by
the selection of a parent ion. Thus, all other
ions are ejected from the trap. The parent ion
then undergoes a translational excitation.
Excited ions collide with a helium buffer gas and
dissociate. The resulting product ions are
recorded by scanning the voltage. Thus, a second
massanalysis scan is performed. - Applications
- The enriched specificity is useful in the
distinction of isomers, sequencing of biopolymers
and analysis of complex mixtures - The structural elucidation of complex molecules
in the presence of mixtures. A compound can be
fragmented and the resulting fragments can be
further analyzed. - Conclusions
- The sensitivity and resolution of the ion trap
are outstanding. Thus, an increasing number of
analyses will be performed using ion trap mass
spectrometry. - The small size of the instrument, the low cost
and the reasonable pressure requirements make
this device one of the most powerful tools in the
chemical analysis.
13References
- 1 http//www.currentseparations.com/issues
- 16-3/cs16-3c.pdf
- 2 Schalley,C.A.Modern Mass Spectrometry, volume
225, Springer, New York, 2003.