NSFCAM Dichroic Immobilization - PowerPoint PPT Presentation

1 / 7
About This Presentation
Title:

NSFCAM Dichroic Immobilization

Description:

Remove 4 Screws as shown in this . They hold the Dichroic rotator housing in place ... 4 Screws. S.H.C 4-40 (Outer ones) 4. Step3 ... – PowerPoint PPT presentation

Number of Views:32
Avg rating:3.0/5.0
Slides: 8
Provided by: vernstah
Category:

less

Transcript and Presenter's Notes

Title: NSFCAM Dichroic Immobilization


1
NSFCAMDichroicImmobilization
  • Date first created 10/18/02
  • Date last modified 10/21/02

2
Step1
  • Remove Interface Mount Assembly
  • Remove top cover of Vacuum Jacket
  • Remove top cover of Radiation Shield
  • Remove Top Baffle Tube
  • Equals Situation at right

3
Step2
  • Remove Cold Structure Top Cover
  • Equals Situation at right
  • Remove 4 Screws as shown in this slide. They hold
    the Dichroic rotator housing in place

Remove 4 Screws S.H.C 4-40 (Outer ones)
4
Step3
  • Add the green, new part to the assembly
    (immobilizing spacer)
  • Note The thickness of the Spacer was established
    using on hand documentation. Washers may have to
    be used, or the thickness reduced.
  • Next Transfer holes using Drill Fixuture to Cold
    Structure Top Cover (next slide)

Immobilizing Spacer,thickness from docs
5
Drilling Holes using Drill Fixture
  • Install Drill Fixture to Cold Structure Top
    Cover Note this is done outside of instrument to
    prevent contamination of Dewar
  • Boss on bottom of Drill Fixture serves as a
    locater along with the 4 holes around it (next
    slide)
  • Use outer or inner for holes to secure Drill
    Fixture to Cold Structure Top Cover.

Drill Fixture
Boss used For locating
Cold Structure Top Cover
6
Drill Fixture in Place
  • Drill Clearance hole for 4-40 (.112 inch
    diameter). May have to be opened up depending on
    adjustment needed to line up the Dichroic
  • Replace Cold Structure Top Cover

Transfer 4 holes To Cold Structure Top Cover
Secure Fixture Using 4 screws
Cold Structure Top Cover
7
Reassembly
  • 1)Reassemble in reverse order

Use 4-40 S.H.C.S 1 inch Long Qty4
Top Baffle Tube