Title: Engineering World Health: Aspirator
1Engineering World Health Aspirator
Client Advisor Dr. John Webster
Lucas Vitzthum - Team Leader Fan Wu - BSAC Nick
Harrison - Communications Jonathan Meyer - BWIG
2Outline
- Problem Statement
- Background
- Design Specifications
- Previous Design
- Improvements
- Vacuum source
- Check valves
- Hospital integration
- Future Work
3Problem Statement
- Medical Aspirator
- 3rd World Application
- Limited power
- Locally available parts/expertise
- Inexpensive
- Clear Instructions
4Background Information
- Remove mucus and other bodily fluids
- Wide range of uses
- Liposuction
- Nasal
- Gastric
- Dental
- Endocervical
- Pediatrics
- Portable
- Battery or AC
http//www.usemc.com/Allied-Gomco-OptiVac-G180-p/g
180.htm
5Current Devices
- Problems
- Expensive
- Specialized
- Not locally repairable
http//www.mdsnq.com.au/products/212/140207130315.
jpg
6Design Specifications
- Locally available materials
- lt100
- Autoclavable suction tips
- Vacuum pressure 0 550 mmHg
- Flow rate 0 30 Lpm
- Sterile and safe
7Previous Design
- 12v Battery
- Fan Motor
- String
- Lab Glove Diaphragm
- Collection Chamber
- Autoclavable Tip
8Major Areas of Improvement
- Vacuum source
- Check valves
- Hospital integration
9Piston Design
- Transform the rotational kinetic energy to linear
kinetic energy - (½ I ?2 to ½ mV2)
- Improvement from original prototype rigid arm
over string
10Vacuum Source Design
- Rubber diaphragm
- Inner tube
- Thicker, therefore more stable and durable
- Syringe
- Adjustable and consistent vacuum
- Can be hard to push by the piston
http//img.alibaba.com/photo/51785400/Inner_Tube.j
pg
http//hydroclimes.co.uk/shop/images/10-485-00520
10cc20Syringe.jpg
11Valve Design
- Last semesters valves
- Sphignomonometer valve
- One balloon valve
- Problems
- Cost
- Efficiency
- This semester
- Develop an efficient, inexpensive valve made from
locally available materials
12Valve Design
- Three design alternatives
Ball Valve
Single Flap Valve
Double Flap Valve
Made from a funneland a marble
Made from a can lid and rubber from a bicycle
inner tube
Made from a can lid and rubber from a bicycle
inner tube
13Valve Design
- Probable design Single flap valve
- Easiest to build
- Can be constructed inside and outside a pipe
14Hospital Integration
- Reduce size
- Cover
- Mobile
- On/off switch
- Pressure control
http//25frames.org/media/screens/6617.jpg
15Future Work
- Build and test prototype
- EWH
- Building instructions
- Construction pictures
- Testing instructions
- Operating instructions
16References
- Aspirator suction machine., 2007, from
http//www.medicalsupply4u.com/prodList.asp?idCate
gory112showFilter0idProduct51 - Electro surgery continuing education model.
(2007). March 12, 2007, from http//www.valleylabe
ducation.org/esself/Pages/esself23.html - Hill, D. (2005). Duke engineering program
improves hospital conditions in developing
countries. Retrieved 4/12, 2007, from
http//www.dukenews.duke.edu/2005/09/DEWH.html
17Any Questions?