Title: How To Troubleshoot High Speed Dental Handpiece?


1
How To Troubleshoot High Speed Dental Handpiece?
A P R I L 2 7 , 2 0 2 1 K E V I N L Y O N S
  • For a high-speed handpiece to operate it requires
    an air pressure of between 25 to 30 PSI.
  • High-speed handpieces operate at speeds of
    350,000 RPM at 25 PSI or 400,000 RPM at 30 PSI.
    This is if the handpiece is functioning normally.
  • However, if you feel that your handpiece is not
    functioning the way it should, there are a couple
    DIY things that you could check to try and do a
    little troubleshooting.
  • Dental Handpiece is Not Spinning?
  • If your handpiece is not spinning or not spinning
    fast enough it is likely not getting enough air
    supply. You can check that the air pressure
    adjustment setting on the dental delivery unit
    head is adjusted properly.
  • You can also check the pressure by removing the
    handpiece and covering the connector with your
    thumb and pressing the foot pedal, while doing
    this, check to see that the gauge goes up by the
    appropriate amount of pressure. If the pressure
    is not adequate you may need a technician to
    troubleshoot this further.

G e t A D e n t a l E q u i p m e n t R e p a i r
Q u o t e
Your Dental Handpiece is not Spraying any Water?
2
No water is coming out the handpiece. First you
should look for a water on/o?? toggle switch, and
make sure that it is on. If your unit uses a
water bottle for its water supply, make sure the
bottle is full, switched on, and attached
tightly. If these obvious ??xes do not rectify
the problem, you can try to ??gure out if the
problem is in the dental unit, or it is a clogged
handpiece. You can do this by removing the
handpiece from the hose connector, you will then
need to cover the small air hole with your ??nger
and step on the pedal. If water discharges from
the hand piece connector, then either your hand
piece water line is clogged or your connector
gasket is worn out. Also Know Invacare Oxygen
Concentrator is Beeping, What Does It
Mean? Water not Stopping after Using the
Handpiece? If you are experiencing water
continuing to spray out the handpiece after your
foot is o?? the pedal or even when it is placed
back in its holder, this problem is likely not
your handpiece. You either have a worn-out
air/water valve inside the unit or your handpiece
holder has wore out or is not adjusted properly.
Have a maintenance tech look at it to try and
identify the problem. The Handpiece does not
stop Spinning after Use This is likely a problem
with the handpiece holder. The air switch on the
handpiece holder is not adjusted properly or is
worn out or if your unit is older, you might have
a handpiece selector switch that is not switched
or broken.
Water Leaking Out the Handpiece Connector If
water is leaking out where the hand piece
connects to the handpiece hose it is likely the
handpiece gasket is worn out. Replace the
handpiece gasket and try it again. Air in
Handpiece Water Spray This problem can be
annoying and cause the water spray to come out
sporadically. This is likely a problem with a
cross over gasket leak somewhere inside the
units water supply. You will probably need a
technician to help identify where this internal
leak is happening.
Know About Process of Dental Air Compressor
Maintenance
3
Your Dental Handpiece is Losing Power If you
have adequate air pressure feeding the handpiece
and no air leaks, the problem is likely your hand
piece. This problem could be the result of a
handpiece bur not properly locking in or
slipping. It could also be a bad handpiece
turbine bearing. In which case you send if your
handpiece for repair. Water Leaking out Unit
when Handpiece is Being Used This problem occurs
when the supply water leaks into the return air.
Usually this is a result of a bad handpiece
gasket. Know more about our dental equipment
repair services and feel free to contact us. Toll
Free landline number 800-937-3724
LINKEDIN 0
0 LIKES
PREVIOUS
Types of Veterinary Waste Gas Scavenging
Systems NEXT Shortage of Surgical Face Masks
H O U R S M O N 8 A M - 5 P M T U E 8 A M - 5 P
M W E D 8 A M - 5 P M T H U 8 A M - 5 P M F R I 8
A M - 5 P M S A T C L O S E D S U N C L O S E D
4
Tel 203-324-3711 Toll Free 800-937-3724 Email s
ervice_at_medicanix.com Address Medicanix Inc281
Fields LaneSuite 2BBrewster, NY 10509
Our Services
Our Battery Back-up Systems
In addition to delivering an exceptional customer
experience with our services, we also provide
healthcare professionals with battery powered
backup systems to ensure that Operating Rooms,
Cold Storage Lab Refrigeration and Refrigerated
Vaccines are protected during power outages. To
learn more, please visit our other website
  • Equipment Refurbishing
  • Preventative Maintenance Services- Medical
    Equipment
  • Biomedical Inspections
  • Autoclaves Sterilizers
  • Power Exam Surgery Table Repair

www.MediProducts.net
O U R C O M PA N Y O U R PA R T N E R S N E W S
B L O G C O N T A C T U S
Copyright Medicanix 2024. All Rights Reserved.
Designed Developed by Medicanix
Write a Comment
User Comments (0)
About PowerShow.com