Calorimeter Upgrade - PowerPoint PPT Presentation

1 / 21
About This Presentation
Title:

Calorimeter Upgrade

Description:

We expect the bunch structure to evolve to a minimum 132 ns between ... SMT, SciFi,Preshowers. Shielding. Forward Mini-drift. chambers. Forward Scintillator ... – PowerPoint PPT presentation

Number of Views:44
Avg rating:3.0/5.0
Slides: 22
Provided by: DZE16
Category:

less

Transcript and Presenter's Notes

Title: Calorimeter Upgrade


1
Calorimeter Upgrade
Gregorio Bernardi on behalf of
the Dzero Collaboration
  • The Tevatron and Dzero Upgrades
  • From Run I to Run II, in brief
  • Preshower Detectors
  • CPS design, status...
  • FPS ...expected performances
  • Calorimeter Electronics
  • Constraints at Run II
  • System Description
  • Intercryostat Detector
  • Objectives and status

2
Timings
  • Bunch structure

gap used to form trigger and sample baselines
3.56us
Run I 6x6
superbunch
gap
4.36us
2.64us
396ns
Run II 36x36
this gap is too small to form trigger and
sample baseline
  • We expect the bunch structure to evolve to a
    minimum 132 ns between bunches - design to that
  • We expect 3 gaps, gap size must be a multiple of
    132ns (7 RF buckets)

3
Experiment
  • Physics program at Run II (driven by high pT
    physics)
  • Searches for phenomena beyond the SM
  • Top quark studies
  • Electroweak measurements
  • QCD tests
  • B physics
  • Accelerator environment implies
  • Fast electronics, pipeline
  • Sophisticated trigger systems
  • Radiation hard detector components

4
E.M. Calibration
  • Use J/Psi , Upsilon , Z

5
DÆ Upgrade
6
Tracking System
Fiber Tracker
Silicon Microstrip Tracker
Forward Preshower
Solenoid
Central Preshower
7
Central Preshower
  • Central Preshower already mounted on the
    solenoid
  • Scintillator Lead
  • Performances studied with MC and cosmic rays

8
Central Preshower
  • Simulation results
  • Good spatial resolution
  • Preshower recovers resolution.

9
Forward Preshower
  • Azimuthal wedges (u,v) scintillator/lead with
  • readout via fibers

side view
10
Fwd Preshower
  • Azimuthal Wedges ready
  • Test beam results

11
Fwd Preshower
  • Test Beam Results (FNAL 97)

12
Calorimeter Electronics
  • Objectives
  • Accommodate reduced minimum bunch spacing from
    3.5 us to 396 or 132 ns
  • Storage of analog signal for 4us for L1 trigger
    formation
  • Generate trigger signals for calorimeter L1
    trigger
  • Maintain present level of noise performance and
    pile-up performance
  • Means
  • Replace preamplifiers
  • Replace shapers
  • Add analog storage
  • Replace Calibration System

13
Calorimeter Electronics
SCA analog delay gt4usec, alternate
new low noise preamp driver
Trig. sum
Bank 0
SCA (48 deep)
SCA (48 deep)
x1
Filter/ Shaper
Preamp/ Driver
Output Buffer
BLS
SCA
Detc.
x8
SCA (48 deep)
SCA (48 deep)
Bank 1
Additional buffering for L2 L3
Replace cables for impedance match
Shorter shaping 400ns
  • 55K readout channels
  • Replacement of Preamps, Shapers, BLS, addition of
    SCA, new calibration.

14
Preamplifier
  • similar to previous version except
  • Dual FET frontend
  • Compensation for detector capacitance
  • Faster recovery time

New output driver for terminated signal
transmission
15
SCA
  • Not designed for simultaneous read and write
    operations
  • two SCA banks alternate reading and writing
  • approximately deadtimeless to L1 rate up to
    100kHz
  • Only 12 bit dynamic range
  • low and high gain path for each readout channel
    (X8/X1)
  • maintain 15 bit dynamic range

16
Calorimeter Electronics Upgrades
  • Cables from cryostat to preamps
  • replace 110 ohm cables with impedance matched 30
    ohm cables to minimize sensitivity to reflection
  • Preamp motherboards
  • 48 preamp hybrids per motherboard -- for improved
    high frequency performance
  • BLS motherboards/ Crate Controller
  • 48 BLS hybrids/motherboard to accomodate new SCA
    and shaper
  • New/additional controlscoordinating the 144
    analog samples per channel
  • Power Supplies
  • use low noise commercial supplies
  • power requirements increased
  • New Calibration System
  • accommodate timing constraints and have same
    or better precision.

17
Noise Contributions
  • Reoptimized three contributions
  • Electronics noise
  • increases due to shorter shaping times
    (2 microsnds to 400 ns)
    (as sqrt(t))
  • decrease due to lower noise preamp
    (2 FET input)
    (as 1/sqrt(2))
  • Uranium noise
  • decrease due to shorter shaping times (as
    sqrt(t) )
  • Pile-up noise
  • increase due to luminosity
    (as sqrt(L))
  • decrease due to shorter shaping times
    (as sqrt(t))

18
Expected Noise
  • Design for
  • 400 ns shaping
  • lower noise -- 2 FET input
  • luminosity of 2x1032
  • Which leads to
  • electronic noise increase of 1.6
  • uranium noise decrease of 2.3
  • pile-up noise increases by 1.3
  • comparable noise performance at 1032 with new
    electronics as with old electronics at 1031
  • Pile-up effect on physics ?
  • The W mass benchmark has been simulated and
    confirms that pile-up will not limit our W mass
    at Run II.

19
Intercryostat Detector (ICD)
  • Objectives
  • Maintain ICD performance in presence of a
    magnetic field and additional material from
    solenoid
  • Improve coverage for the region 1.1ltetalt1.4
  • Design
  • A scintillator based Intercryostat Detector (ICD)
    with phototube readout similar to Run I design
  • Expected Physics output
  • Provides improvement to jet energies and missing
    energy in the region between the central and end
    cryostats

20
ICD Design
  • Detectors
  • 16 super-tiles per end with total of 384
    scintillator tiles covering 1.1ltetalt1.4
  • WLS fiber readout of scintillator tiles.
    Re-use existing PMTs
  • Clear fiber light piping to region of low field
  • Calibration scheme as for L.Ar calo. But with
    adapted pulse

21
Conclusions
  • Dzero is upgrading its detector
  • L.Argon calorimeter untouched
  • Harder machine conditions and new environment
    (solenoid)
  • New Calorimeter Electronics
  • Improved ICD
  • New Central and Forward Preshower
  • Similar Performances
  • Detectors close to be ready to start
    taking data in Summer 2000
Write a Comment
User Comments (0)
About PowerShow.com