RAD 254 Chapter 14 Control of Scatter - PowerPoint PPT Presentation

1 / 17
About This Presentation
Title:

RAD 254 Chapter 14 Control of Scatter

Description:

Grid ratio = height of the lead lines divided by the interspace WIDTH ... Grid weight refers to how heavy it is duh the more lead, the heavier it is. Grid Types ... – PowerPoint PPT presentation

Number of Views:509
Avg rating:5.0/5.0
Slides: 18
Provided by: george86
Category:
Tags: rad | chapter | control | lead | scatter

less

Transcript and Presenter's Notes

Title: RAD 254 Chapter 14 Control of Scatter


1
RAD 254 Chapter 14Control of Scatter
  • Break down into Those that reduce patient dose
    and those that are geometrical in nature and
    those not

2
3 factors affecting scatter (primary)
  • Increased kVp
  • Increased field size
  • Increased patient thickness

3
Spatial Resolution Contrast Resolution
  • Spatial resolution may be thought of as geometric
    in nature (focal spot size, emission spectrum,
    etc. dealing with geometric image formation
  • Contrast resolution driven by scatter and other
    sources of noise

4
Scatter
  • Increased field sizes more scatter
    collimation is the most readily available and
    easiest thing to lower the amount of scatter
  • Patient thickness also increases scatter
    compression may be used to help avoid this (IVPs
    and mammo are examples)

5
Beam restricting deviceslimit the radiation to
the patient
  • Aperature diaphram (size and resultant field size
    are a direct proportion draw the damn picture
    and figure the problems)
  • Cones and cylinders great for absorbing
    scatter, but are circular shaped great for
    improving contrast and removing scatter BUT
    requires much more mAs as a result

6
Variable aperture diaphram
  • Mandated in 1974 by the US Dept of Food and Drug
    Administration (mandate removed later)
  • Positive Beam Limitation Devices (PBLs)
  • Automatically collimate to the size of
    cassette/receptor in the bucky and CANNOT be a
    bigger size than the film/receptor

7
Filtration
  • Filtration also will decrease the low energy rays
    and limit patient dose and some scatter

8
The GRID
  • Only FORWARD scatter is of any benefit to the
    radiographic image all other scatter degrades
    the image!

9
Scatter LOWER contrast
  • Using a grid (alternating strips of fine leaded
    strips with alternating radiolucent interspace
    material) can effectively reduce the amount of
    ANGELED scatter from reaching the film/recepter

10
Grid terms
  • Grid ratio height of the lead lines divided by
    the interspace WIDTH
  • Grid frequency/lines per inch the more lines
    per inch, the more clean up
  • Grid clean up scatter w/o a grid vs scatter
    reaching film with a grid AKA Contrast
    Improvement Factor
  • Grid function improve image contrast

11
Bucky Factor
  • Refers to the AMOUNT of radiation to the patient
    with a grid vs W/O a grid.
  • Higher the grid ratio the higher the bucky
    factor
  • The higher the kVp, the higher the bucky factor
  • Grid weight refers to how heavy it is duh the
    more lead, the heavier it is

12
Grid Types
  • Parallel
  • Crossed (cross-hatch)
  • Focused
  • Focused - crossed

13
Grid Problems
  • Grid cut-off short SIDs result in the
    vertical, parallel strips absorbing the
    diverging beam at the outer margins of the
    grid/film/rescepter MOST pronounced at short
    SIDs
  • Most grid problems are positioning related
  • Uneven grid/off level grid
  • Off centered (lateral decentering)
  • Off focus grid
  • Upside down focused grid

14
Focused Grid Misalignment
  • Off level grid cutoff across image
    underexposed image (light)
  • Off center ditto
  • Off focus CR centered to one side or the other
    of a focused grid
  • Upside down grid severe grid cut-off (no
    density) at both sides of the image

15
Grid Ratio Selection
  • 81 grid is the most widely used
  • Grid ratio is kVp driven
  • Higher kVps warrant higher grid ratios
  • Higher grid ratios higher patient dose (more
    radiation needed to produce an image)
  • As kVp increases past MAXIMUM OPTIMUM kVp,
    patient dose INCREASES

16
mAs Grid considerationsAs grid ratio
increases, so much mAs
  • 51 grid 2 X mAs
  • 81 grid 4 X mAs
  • 121 grid 5 X mAs
  • 161 grid 6 X mAs

17
Air gap technique
  • By allowing the scatter radiation to diffuse in
    the atmosphere after the patient but BEFORE the
    film results in a higher contrast image as the
    scatter diffuses and does NOT reach the film
  • C-spine lateral is a good example of this
Write a Comment
User Comments (0)
About PowerShow.com