Title: Biomedical Instrumentation
1Biomedical Instrumentation
- Chapter 6 in
- Introduction to Biomedical Equipment Technology
- By Joseph Carr and John Brown
2Signal Acquisition
- Medical Instrumentation typically entails
monitoring a signal off the body which is analog,
converting it to an electrical signal, and
digitizing it to be analyzed by the computer.
3Types of Sensors
- Electrodes acquire an electrical signal
- Transducers acquire a non-electrical signal
(force, pressure, temp etc) and converts it to an
electrical signal
4Active vs Passive Sensors
- Active Sensor
- Requires an external AC or DC electrical source
to power the device - Strain gauge, blood pressure sensor
- Passive Sensor
- Provides it own energy or derives energy from
phenomenon being studied - Thermocouple
5Sensor Error Sources
- Error
- Difference between measured value and true value.
65 Categories of Errors
- Insertion Error
- Application Error
- Characteristic Error
- Dynamic Error
- Environmental Error
7Insertion Error
- Error occurring when inserting a sensor
8Application Error
- Errors caused by Operator
9Characteristic Error
- Errors inherent to Device
10Dynamic Error
- Most instruments are calibrated in static
conditions if you are reading a thermistor it
takes time to change its value. If you read this
value to quickly an error will result.
11Environmental Error
- Errors caused by environment
- heat, humidity
12Sensor Terminology
- Sensitivity
- Slope of output characteristic curve ?y/ ?x
- Minimum input of physical parameter will create a
detectable output change - Blood pressure transducer may have a sensitivity
of 10 uV/V/mmHg so you will see a 10 uV change
for every V or mmHg applied to the system.
13Which is more sensitive? The left side one
because youll have a larger change in y for a
given change in x
14Sensor Terminology
- Sensitivity Error Departure from ideal slope of
a characteristic curve
15Sensor Terminology
- Range Maximum and Minimum values of applied
parameter that can be measured. - If an instrument can read up to 200 mmHg and the
actual reading is 250 mmHg then you have exceeded
the range of the instrument.
16Sensor Terminology
- Dynamic Range total range of sensor for minimum
to maximum. Ie if your instrument can measure
from -10V to 10 V your dynamic range is 20V - Precision Degree of reproducibility denoted as
the range of one standard deviation s - Resolution smallest detectable incremental
change of input parameter that can be detected
17Accuracy
- Accuracy maximum difference that will exist
between the actual value and the indicated value
of the sensor
18Offset Error
- Offset error output that will exist when it
should be zero - The characteristic curve had the same sensitive
slope but had a y intercept
Output
Output
Input
Input
Offset Error
Zero offset error
19Linearity
- Linearity Extent to which actual measure curved
or calibration curve departs from ideal curve.
20Linearity
- Nonlinearity () (Din(Max) / INfs) 100
- Nonlinearity is percentage of nonlinear
- Din(max) maximum input deviation
- INfs maximum full-scale input
21Hysteresis
- Hysteresis measurement of how sensor changes
with input parameter based on direction of change
22Hysteresis
- The value B can be represented by 2 values of
F(x), F1 and F2. If you are at point P then you
reach B by the value F2. If you are at point Q
then you reach B by value of F1.
23Response Time
- Response Time Time required for a sensor output
to change from previous state to final settle
value within a tolerance band of correct new
value denoted in red can be different in rising
and decaying directions
24Response Time
- Time Constant Depending on the source is defined
as the amount of time to reach 0 to 70 of final
value. Typically denoted for capacitors as T R
C (Resistance Capacitance) denoted in Blue
25Response Time
Tdecay
F(t)
Decaying Response Time
Toff
Time
- Convergence Eye Movement the inward turning of
the eyes have a different response time than
divergence eye movements the outward turning of
the eyes which would be the decay response time
26Dynamic Linearity
- Measure of a sensors ability to follow rapid
changes in the input parameters. Difference
between solid and dashed curves is the non-
linearity as depicted by the higher order x terms
27Dynamic Linearity
- Asymmetric F(x) ! F(-x) where F(x) is
asymmetric around linear curve F(x) then - F(x) ax bx2cx4 . . . K offsetting for K or
you could assume K 0 - Symmetrical F(x) F(-x) where F(x) is
symmetric around linear curve F(x) then - F(x) ax bx3 cx5 . . . K offsetting for K
or you could assume K 0
28Frequency Response of Ideal and Practical System
- When you look at the frequency response of an
instrument, ideally you want a wideband flat
frequency response.
29Frequency Response of Ideal and Practical System
- In practice, you have attenuation of lower and
higher frequencies - FL and FH are known as the 3 dB points in
voltage systems.
30Examples of Filters
- Ideal Filter has sharp cutoffs and a flat pass
band - Most filters attenuate upper and lower
frequencies - Other filters attenuate upper and lower
frequencies and are not flat in the pass band
31Electrodes for Biophysical Sensing
- Bioelectricity naturally occurring current that
exists because living organisms have ions in
various quantities
32Electrodes for Biophysical Sensing
- Ionic Conduction Migration of ions-positively
and negatively charge molecules throughout a
region. - Extremely nonlinear but if you limit the region
can be considered linear
33Electrodes for Biophysical Sensing
- Electronic Conduction Flow of electrons under
the influence of an electrical field
34Bioelectrodes
- Bioelectrodes class of sensors that transduce
ionic conduction to electronic conduction so can
process by electric circuits - Used to acquire ECG, EEG, EMG, etc.
35Bioelectrodes
- 3 Types of electrodes
- Surface (in vivo) outside body
- Indwelling Macroelectrodes (in vivo)
- Microelectrodes (in vitro) inside body
36Bioelectrodes
- Electrode Potentials
- Skin is electrolytic and can be modeled as
electrolytic solutions
Metal Electrode
Electrolytic Solution where Skin is electrolytic
and can be modeled as saline
37Electrodes in Solution
- Have metallic electrode immersed in electrolytic
solution once metal probe is in electrolytic
solution it - Discharges metallic ions into solution
- Some ions in solution combine with metallic
electrodes - Charge gradient builds creating a potential
difference or you have an electrode potential or
½ cell potential
38Electrodes in Solution
2 cells A and B, A has 2 positive ions And B has
3 positive ions thus have a Potential difference
of 3 2 1 where B is more positive than A
A
B
39Electrodes
- Two reactions take place at electrode/electrolyte
interface - Oxidizing Reaction Metal -gt electrons metal
ions - Reduction Reaction Electrons metal ions -gt
Metal
40Electrodes
- Electrode Double Layer formed by 2 parallel
layers of ions of opposite charge caused by ions
migrating from 1 side of region or another ionic
differences are the source of the electrode
potential or half-cell potential (Ve).
41Electrodes
- If metals are different you will have
differential potential sometimes called an
electrode offset potential. - Metal A gold Vae 1.50V and Metal B silver
Vbe 0.8V then Vab 1.5V 0.8 V 0.7V (Table
6-1 in book page 96)
Vae
Metal A
Vbe
Metal B
Electrolytic Solution
42Electrodes
- Two general categories of material combinations
- Perfectly polarized or perfectly nonreversible
electrode no net transfer of charge across
metal/electrolyte interface - Perfectly Nonpolarized or perfectly reversible
electrode unhindered transfer of charge between
metal electrode and the electrode - Generally select a reversible electrode such as
Ag-AgCl (silver-silver chloride)
43- Rt internal resistance of body which is low
- Vd Differential voltage Vd
- Rsa and Rsb skin resistance at electrode A and
B
- R1A and R1B resistance of electrodes
- C1A and C1B capacitance of electrodes
44Electrode Potentials cause recording Problems
- ½ cell potential 1.5 V while biopotentials are
usually 1000 times less (ECG 1-2 mV and EEG is
50 uV) thus have a tremendous difference between
DC cell potential and biopotential - Strategies to overcome DC component
- Differential DC amplifier to acquire signal thus
the DC component will cancel out - Counter Offset-Voltage to cancel half-cell
potential - AC couple input of amplifier (DC will not pass
through) ie capacitively couple the signal into
the circuit
45Electrode Potentials cause recording Problems
- Strategies to overcome DC component
- Differential DC amplifier to acquire signal thus
the DC component will cancel out - Counter Offset-Voltage to cancel half-cell
potential - AC couple input of amplifier (DC will not pass
through) - Capacitively couple the signal into the circuit
46Medical Surface Electrodes
- Typical Medical Surface Electrode
- Use conductive gel to reduce impedance between
electrode and skin - Schematic
47Medical Surface Electrodes
- Have an Ag-AgCl contact button at top of hollow
column filled with gel - Gel filled column holds actual metallic electrode
off surface of skin and decreases movement
artifact - Typical ECG arrangement is to have 3 ECG
electrodes (2 differentials signals and a
reference electrode)
48Problems with Surface Electrodes
- Adhesive does not stick for a long time on sweaty
skin - Can not put electrode on bony prominences
- Movement or motion artifact significant problem
with long term monitoring results in a gross
change in potential - Electrode slippage if electrode slips then
thickness of jelly changes abruptly which is
reflected as a change in electrode impedance and
electrode offset potential (slight change in
potential)
49Potential Solutions for Surface Electrodes
Problems
- Additional Tape
- Rough surface electrode that digs past scaly
outer layer of skin typically not comfortable for
patients.
50Other Types of Electrodes
- Needle Electrodes inserted into tissue
immediately beneath skin by puncturing skin on an
angle note infection is a problem. - Indwelling Electrodes Inserted into layers
beneath skin -gt typically tiny exposed metallic
contact at end of catheter usually threaded
through patients vein to measure intracardiac
ECG to measure high frequency characteristics
such as signal at the bundle of His
51Other Types of Electrodes
- EEG Electrodes can be a needle electrode but
usually a 1 cm diameter concave disc of gold or
silver and is held in place by a thick paste that
is highly conductive sometimes secured by a
headband
52Microelectrode
- Microelectrode measure biopotential at cellular
level where microelectrode penetrates cell that
immersed in an infinite fluid - Saline.
53Microelectrode
- Two typical types
- Metallic Contact
- Fluid Filled
54Microelectrode Equivalent Circuit
R1
RS Spreading Resistance of the electrode and is
a function of tip diameter R1 and C1 are result
of the effects of electrode/cell interface C2
Electrode Capacitance
RS
C2
C1
Vo
V1
55Calculation for Resistance Rs
- Rs in metallic microelectrodes without glass
coating
where Rs resistance ohms (?) P Resistivity of
the infinite solution outside electrode 70 ?cm
for physiological saline r tip radius ( 0.5 um
for 1 um electrode) 0.5 x10-4 cm
56Calculation for Resistance Rs
- Rs of glass coated metallic microelectrode is 1-2
order of magnitude higher
where Rs resistance ohms (?) P Resistivity of
the infintie solution outside electrode) 3.7
?cm for 3 M KCl r tip radius typically 0.1 u m
0.1 x 10-4 cm a taper angle ( p/ 180)
57Capacitance of Microelectrode
- Capacitance of C2 has units pF/cm
-
Where e dielectric constant which for glass
4 R outside tip radius r inside tip radius
58Capacitance of Microelectrode
- Find C of glass microelectrode if the outer
radius is 0.2 um and the inner radius 0.15 um
59Transducers and other Sensors
- Transducers sensors and are defined as a device
that converts energy from some one form (temp.,
pressure, lights etc) into electrical energy
where as electrodes directly measure electrical
information
60Wheatstone Bridge
Es
R1
A
R3
R3
R1
-
Eo
EC
ED
Eo
-
Es
ED
EC
-
R2
R4
R4
R2
B
- Basic Wheatstone Bridge uses one resistor in each
of four arms where battery excites the bridge
connected across 2 opposite resistor junctions (A
and B). The bridge output Eo appears across C
and D junction.
61Finding output voltage to a Wheatstone Bridge
- Ex A wheatstone bridge is excited by a 12V dc
source and has the following resistances R1
1.2K? R2 3 K ? R3 2.2 K ? and R4 5 K ?
find Eo
62Finding output voltage to a Wheatstone Bridge
- A wheatstone bridge is excited by a 12V dc source
and has the following resistances R1 1.2K? R2
3 K ? R3 2.2 K ? and R4 5 K ? find Eo
63(No Transcript)
64Null Condition of Wheatstone Bridge
- Null Condition is met when Eo 0 can happen in 2
ways - Battery 0 (not desirable)
- R1 / R2 R3/ R4
65Null Condition of Wheatstone Bridge
- When R1 2K? R2 1K ? R3 10K ? R4 5K ?
66Null Condition of Wheatstone Bridge
- Key with null condition is if you change one of
the resistances to be a transducer that changes
based on input stimulus then Eo will also change
according to input stimulus
67Strain Gauges
- Definition resistive element that changes
resistance proportional to an applied mechanical
strain
68Strain Gauges
- Compression decrease in length by DL and an
increase in cross sectional area.
69Strain Gauges
- Tension increase in length by DL and a decrease
in cross section area.
L length
Rest Condition
70Resistance of a metallic bar is given in length
and area
- where
- R Resistance units ohms (?)
- ? resistivity constant unique to type of
material used in bar units ohm meter (?m) - L length in meters (m)
- A Cross sectional area in meters2 (m2 )
71Resistance of a metallic bar is given in length
and area
- Example find the resistance of a copper bar that
has a cross sectional area of 0.5 mm2 and a
length 250 mm note the resistivity of copper is
1.7 x 10-8?m
72Piezoresistivity
- Piezoresistivity change in resistance for a
given change in size and shape denoted as h - Resistance in tension
- Resistance increases in tension
- L length ?L change in L ? resistivity
- A Area ?A change in A
73- Resistance in compression
- Resistance decreases in compression
- L length ?L change in L ? resistivity
- A Area ?A change in A
Note Textbook forgot the ? in equations 6-28 and
6-29 on page 110
74Example of Piezoresistivity
- Thin wire has a length of 30 mm and a cross
sectional area of 0.01 mm2 and a resistance of
1.5?. - A force is applied to the wire that increases the
length by 10 mm and decreases cross sectional
area by 0.0027 mm2 - Find the change in resistance h.
- Note ? resistivity 5 x 10-7 ?m
75Example of Piezoresistivity
76Example of Piezoresistivity
- Note Change in Resistance will be approximately
linear for small changes in L as long as ?LltltL. - If a force is applied where the modulus of
elasticity is exceeded then the wire can become
permanently damaged and then it is no longer a
transducer.
77Gauge Factor
- Gauge Factor (GF) a method of comparing one
transducer to a similar transducer
78Gauge Factor
- where
- GF Gauge Factor unitless
- ?R change in resistance ohms (?)
- R unstrained resistance ohms (?)
- ?L change in length meters (m)
- L unstrained length meters (m)
79Gauge Factor
- Where e strain which is unitless
- GF gives relative sensitivity of a strain gauge
where the greater the change in resistance per
unit length the greater the sensitivity of
element and the greater the gauge factor.
80Example of Gauge Factor
- Have a 20 mm length of wire used as a string
gauge and has a resistance of 150 ?. - When a force is applied in tension the resistance
changes by 2? and the length changes by 0.07 mm.
- Find the gauge factor
81Types of Strain Gauges Unbonded and Bonded
- Unbonded Strain Gauge resistance element is a
thin wire of special alloy stretch taut between
two flexible supports which is mounted on
flexible diaphram or drum head.
82Types of Strain Gauges Unbonded and Bonded
- When a Force F1 is applied to diaphram it will
flex in a manner that spreads support apart
causing an increase in tension and resistance
that is proportional to the force applied. - When a Force F2 is applied to diaphram the
support ends will more close and then decrease
the tension in taut wire (compression force) and
decrease resistance will decrease in amount
proportional to applied force
83Types of Strain Gauges Unbonded and Bonded
- Bonded Strain Gauge made by cementing a thin
wire or foil to a diaphragm therefore flexing
diaphragm deforms the element causing changes in
electrical resistance in same manner as unbonded
strain gauge
84Types of Strain Gauges Unbonded and Bonded
- When a Force F1 is applied to diaphram it will
flex in a manner that causes an increase in
tension of wire then the increase in resistance
is proportional to the force applied. - When a Force F2 is applied to diaphram that cause
a decrease the tension in taut wire (compression
force) then the decrease in resistance will
decrease in amount proportional to applied force
85Comparison of Bonded vs. Unbonded Strain Gauges
- Unbonded strain gauge can be built where its
linear over a wide range of applied force but
they are delicate - Bonded strain gauge are linear over a smaller
range but are more rugged - Bonded strain gauges are typically used because
designers prefer ruggedness.
86Typical Configurations
A
R3 SG3
R1 SG1
Vo
ES
C
D
-
R4 SG4
R2 SG2
B
Mechanical Configuration
Electrical Circuit
- 4 strain gauges (SG) in Wheatstone Bridge
87Strain Gauge Example
- Using the configuration in the previous slide
where 4 strain gauges are placed in a wheatstone
bridge where the bridge is balanced when no force
is applied, - Assume a force is applied so that R1 and R4 are
in tension and R2 and R3 are in compression. - Derive the equation to depict the change in
voltage across the bridge and find the output
voltage when each resistor is 200 ?, the change
of resistance is 10 ? and the source voltage is
10 V
88Strain Gauge Example
Derivation
Circuit
A
R1 R h
R3 R-h
Es
-
Eo
C
D
-
R2 R - h
R4 R h
B
Note Text book has wrongly stated that tension
decreases R and compression increases R on page
112
89Transducer Sensitivity
- Transducer Sensitivity rating that allows us to
predict the output voltage from knowledge of the
excitation voltage and the value of the applied
stimulus units µV/Vunit of applied stimulus
90Transducer Sensitivity
- Example if you have a force transducer calibrated
in grams (unit of mass) which allows calibration
of force transducer then sensitivity denoted as f
µV/Vg (another ex f µV/VmmHg)
91Transducer Sensitivity
- To calculate Output Potential use the following
equations - where
- Eo output potential in Volts (V)
- E excitation voltage
- f sensitivity µV/Vg
- F applied force in grams (g)
92Transducer Sensitivity
- Example Transducer has a sensitivity of 10
µV/Vg, predict the output voltage for an applied
force of 15 g and 5 V of excitation.
- note book has typo where writes µV/V/g for
sensitivity
93Inductance Transducers
- Inductance Transducers inductance L can be
varied easily by physical movement of a permeable
core within an inductor 3 basic forms - Single Coil
- Reactive Wheatstone Bridge
- Linear Voltage Differential Transformer LVDT
94LVDT
95Capacitance Transducers
- Quartz Pressure Sensors capacitively based where
sensor is made of fused quartz - Capacitive Transducers Capacitance C varies with
stimulus
96Capacitive Transducers
- Three examples
- Solid Metal disc parallel to flexible metal
diaphragm separated by air or vacuum (similar to
capacitor microphone) when force is applied they
will move closer or further away. - Stationary metal plate and rotating moveable
plate as you rotate capacitance will increase or
decrease - Differential Capacitance 1 Moveable metal Plate
placed between 2 stationary Places where you have
2 capacitors C1 between P1 and P3 and C2 between
P2 and P3 where when a force is applied to
diaphragm P3 moves closer to one plate or vice
versa
97Temperature Transducers
- 3 Common Types
- Thermocouples
- Thermistors
- Solid State PN Junctions
98Thermocouple
- Thermocouple 2 dissimilar conductor joined
together at 1 end. - The work functions of the 2 materials are
different thus a potential is generated when
junction is heated (roughly linear over wide
range)
99Thermistors
- Thermistors Resistors that change their value
based on temperature where - Positive Temperature Coefficient (PTC) device
will increase its resistance with an increase in
temperature - Negative Temperature Coefficient (NTC) device
will decrease its resistance with an increase in
temperature - Most thermistors have nonlinear curve when
plotted over a wide range but can assume
linearity if within a limited range
100BJT Bipolar Junction Transistor
IC
- Transistor invented in 1947 by Bardeen,
Brattain and Schockley of Bell Labs.
B Base C Collector E Emitter IE I B
I C
101BJT Bipolar Junction Transistor
- Transistor rely on the free travel of electrons
through crystalline solids called semiconductors.
Transistors usually are configured as an
amplifier or a switch.
102Solid State PN Temperature Transducers
- Solid State PN Junction Diode the base emitter
voltage of a transistor is proportional to
temperature. For a differential pair the output
voltage is
K Boltzmans Constant 1.38 x10-23J/K T
Temperature in Kelvin IC1 Collector current of
BJT 1 mA IC2 Collector current of BJT 2 mA q
Coulombs charge 1.6 x10 -19 coulombs/electron
103Example of temperature transducer
- Find the output voltage of a temperature
transducer in the previous slide if IC1 2 mA
IC2 1 mA and the temperature is 37 oC
104Homework
- Read Chapter 7
- Chapter 6 Problems 1, 3 to 6, 9
- Problem 1 resistivity 1.7 10-8?m
- Problem 4 sensitivity 50 µV/(VmmHg)
- Problem 4 1 torr 1 mmHg
- Problem 6 sensitivity 50 µV/(Vg)
105Review
- What are two types of sensors?
- List 5 categories of error
- How do we quantify sensors?
- What is an electrode?
- How do you calculate Rs and C2 of a
microelectrode that is metal with and without
glass coating? - What is a transducer?
- What is a Wheatstone Bridge? How do you derive
the output voltage - Find resistance of a metallic bar for a given
length and area - How does resistance change in tension and in
compression and how do you calculate resistance
106Review
- How do you find resistance change in
piezoresistive device - How do you determine gauge factor
- What is the definition of a strain gauge and what
is difference between bonded and unbonded strain
gauge. - Determine the output potential given a
transducers sensitivity. - What are inductance, capacitance, and temperature
transducers? - How do you calculate the temperature for a solid
state PN Junction Diode?