Title: Atomic Emission Spectroscopy
1Atomic Emission Spectroscopy
2Atomic Emission Spectroscopy
- Atomic emission spectroscopy (AES), in contrast
to AAS, uses the very high temperatures of
atomization sources to excite atoms, thus
excluding the need for lamp sources. Emission
sources, which are routinely used in AES, include
plasma, arcs and sparks, as well as flames. We
will study the different types of emission
sources, their operational principles, features,
and operational characteristics. Finally,
instrumental designs and applications of emission
methods will be discussed.
3Plasma Sources
- The term plasma is defined as a homogeneous
mixture of gaseous atoms, ions and electrons at
very high temperatures. Two types of plasma
atomic emission sources are frequently used - Inductively coupled plasma
- Direct current plasma
4(No Transcript)
5Inductively Coupled Plasma (ICP)
- A typical ICP consists of three concentric quartz
tubes through which streams of argon gas flow at
a rate in the range from 5-20 L/min. The outer
tube is about 2.5 cm in diameter and the top of
this tube is surrounded by a radiofrequency
powered induction coil producing a power of about
2 kW at a frequency in the range from 27-41 MHz.
This coil produces a strong magnetic field as
well.
6- Ionization of flowing argon is achieved by a
spark where ionized argon interacts with the
strong magnetic field and is thus forced to move
within the vicinity of the induction coil at a
very high speed. A very high temperature is
obtained as a result of the very high resistance
experienced by circulating argon (ohmic heating).
The top of the quartz tube will experience very
high temperatures and should, therefore, be
isolated and cooled.
7- This can be accomplished by passing argon
tangentially around the walls of the tube. A
schematic of an ICP (usually called a torch
plasma) is shown below
8(No Transcript)
9(No Transcript)
10(No Transcript)
11- The torch is formed as a result of the argon
emission at the very high temperature of the
plasma. The temperature gradients in the ICP
torch can be pictured in the following graphics
12(No Transcript)
13- The viewing region used in elemental analysis is
usually about 6000 oC, which is about 1.5-2.5 cm
above the top of the tube. It should also be
indicated that argon consumption is relatively
high which makes the running cost of the ICP
torch high as well. Argon is a unique inert gas
for plasma torches since it has few emission
lines. This decreases possibility of
interferences with other analyte lines.
14Sample Introduction
- There are several methods for sample
introduction the most widely used is, of course,
the nebulization of an analyte solution into the
plasma. However, other methods, as described
earlier, are fine where vapors of analyte
molecules or atom from electrothermal or ablation
devices can be driven into the torch for complete
atomization and excitation. For your convenience,
sample introduction methods are summarized here
again
15Samples in Solution
- 1. Pneumatic Nebulizers
- Samples in solution are usually easily introduced
into the atomizer by a simple nebulization,
aspiration, process. Nebulization converts the
solution into an aerosol of very fine droplets
using a jet of compressed gas. The flow of gas
carries the aerosol droplets to the atomization
chamber or region.
16Ultrasonic Nebulizers
- In this case samples are pumped onto the surface
of a piezoelectric crystal that vibrates in the
kHz to MHz range. Such vibrations convert samples
into homogeneous aerosols that can be driven into
atomizers. Ultrasonic nebulization is preferred
over pneumatic nebulization since finer droplets
and more homogeneous aerosols are usually
achieved. However, most instruments use pneumatic
nebulization for convenience.
17- Electrothermal Vaporization
- An accurately measured quantity of sample (few
mL) is introduced into an electrically heated
cylindrical chamber through which an inert gas
flows. Usually, the cylinder is made of pyrolytic
carbon but tungsten cylinders are now available.
The vapors of molecules and atoms are swept into
the plasma source for complete atomization and
excitation.
18- Hydride Generation Techniques
- Samples that contain arsenic, antimony, tin,
selenium, bismuth, and lead can be vaporized by
converting them to volatile hydrides by addition
of sodium borohydride. Volatile hydrides are then
swept into the plasma by a stream of an inert
gas.
19(No Transcript)
20Introduction of Solid Samples
- A variety of techniques were used to introduce
solid samples into atomizers. These include - 1. Conductive Samples
- If the sample is conductive and is of a shape
that can be directly used as an electrode (like a
piece of metal or coin), that would be the choice
for sample introduction in arc and spark
techniques. Otherwise, powdered solid samples are
mixed with fine graphite and made into a paste.
Upon drying, this solid composite can be used as
an electrode. The discharge caused by arcs and
sparks interacts with the surface of the solid
sample creating a plume of very fine particulates
and atoms that are swept into the plasma by argon
flow.
21- Laser Ablation
- Sufficient energy from a focused intense laser
will interact with the surface of samples (in a
similar manner like arcs and sparks) resulting in
ablation. The vapors of molecules and atoms are
swept into the plasma source for complete
atomization and excitation. Laser ablation is
becoming increasingly used since it is applicable
to conductive and nonconductive samples.
22The Glow Discharge Technique
- The technique is used for sample introduction and
atomization as well. The electrodes are kept at a
250 to 1000 V DC. This high potential is
sufficient to cause ionization of argon, which
will be accelerated to the cathode where the
sample is introduced. Collision of the fast
moving energetic argon ions with the sample
(cathode) causes atomization by a process called
sputtering. Samples should thus be conductive to
use the technique of glow discharge. The vapors
of molecules and atoms are swept into the plasma
source for complete atomization and excitation by
flowing argon. However, nonconductive samples
were reported to be atomized by this technique
where they were mixed with a conductor material
like graphite or powdered copper.
23Plasma Appearance and Spectra
- A plasma torch looks very much like a flame but
with a very intense nontransparent brilliant
white color at the core (less than 1 cm above the
top). In the region from 1-3 cm above the top of
the tube, the plasma becomes transparent. The
temperatures used are at least two to three
orders of magnitude higher than that achieved by
flames which may suggest efficient atomization
and fewer chemical interferences.
24- Ionization in plasma may be thought to be a
problem due to the very high temperatures, but
fortunately the large electron flux from the
ionization of argon will suppress ionization of
all species.
25The Direct Current Plasma (DCP)
- The DCP is composed of three electrodes arranged
in an inverted Y configuration. A tungsten
cathode resides at the top arm of the inverted Y
while the lower two arms are occupied by two
graphite anodes. Argon flows from the two anode
blocks and plasma is obtained by momentarily
bringing the cathode in contact with the anodes.
Argon ionizes and a high current passes through
the cathode and anodes.
26- It is this current which ionizes more argon and
sustains the current indefinitely. Samples are
aspirated into the vicinity of the electrodes (at
the center of the inverted Y) where the
temperature is about 5000 oC. DCP sources usually
have fewer lines than ICP sources, require less
argon/hour, and have lower sensitivities than ICP
sources. In addition, the graphite electrodes
tend to decay with continuous use and should thus
be frequently exchanged. A schematic of a DCP
source is shown below
27(No Transcript)
28(No Transcript)
29- A DCP has the advantage of less argon
consumption, simpler instrumental requirements,
and less spectral line interference. However, ICP
sources are more convenient to work with, free
from frequent consumables (like the anodes in
DCPs which need to be frequently changed), and
are more sensitive than DCP sources.
30Advantages of Plasma Sources
- No oxide formation as a result of two factors
including - Very high temperature
- Inert environment inside the plasma (no oxygen)
- 2. Minimum chemical interferences
- 3. Minimum spectral interferences except for
higher possibility of spectral line interference
due to exceedingly large number of emission lines
(because of high temperature)
31- 4. Uniform temperature which results in precise
determinations - 5. No self-absorption is observed which extends
the linear dynamic range to higher concentrations - 6. No need for a separate lamp for each element
- 7. Easily adaptable to multichannel analysis
32Plasma Emission Instruments
- Three classes of plasma emission instruments can
be presented including - 1. Sequential instruments
- In this class of instruments a single channel
detector is used where the signal for each
element is read using the specific wavelength for
each element sequentially. Two types of
sequential instruments are available
33- Linear sequential scan instruments where the
wavelength is linearly changed with time.
Therefore, the grating is driven by a single
speed during an analysis of interest - b. Slew scan instruments where the monochromator
is preset to provide specific wavelengths moving
very fast in between wavelengths while moving
slowly at the specific wavelengths. Therefore, a
two-speed motor driving the grating is thus used.
34Radial vs. Axial Viewing
Radial traditional side view, better for
concentrated samples. Axial direct view into
plasma, lower sensitivity, shifts detection range
lower.
35(No Transcript)
36Sequential vs. multichannel
- Sequential instrument
- PMT moved behind aperture plate,
- or grating prism moved to focus new l on exit
slit - Pre-configured exit slits to detect up to 20
lines, slew scan - characteristics
- Cheaper
- Slower
- Multichannel instrument
- Polychromators (not monochromator) - multiple
PMT's - Array-based system
- charge-injection device/charge coupled device
- characteristics
- Expensive ( gt 80,000)
- Faster
37Slew scan spectrometer
- Two slew-scan gratings
- Two PMTs for VIS and UV
- Most use holographic grating
382. Multichannel Instruments
- This class of instruments is also referred to as
simultaneous instruments in which all signals are
reported at the same time using two types of
configurations
39a. Polychromators
- Multiple detectors, usually photomultiplier
tubes are used. Beams of radiation emerging from
the grating are guided to exit slits (each
representing the wavelength of a specific
element) are focused at several PMTs for
detection. Detection, thus, takes place
simultaneously
40(No Transcript)
41b. Array-based systems
- This multichannel type instrument uses a
multichannel detector like a charge injection
device or a charge-coupled device. Diffracted
beams from a grating pass through a prism where
further resolution of diffracted beams takes
place by a prism. The prism will disperse the
orders of each diffracted beam. The multichannel
detector can also be a linear photodiode array as
in the figure below
42(No Transcript)
43(No Transcript)
44(No Transcript)
45(No Transcript)
463. Fourier transform instruments (FT)
- Instruments in which the signal is coded will
need a decoding mechanism in order to see the
signal. FT is a very common technique for
decoding time domain spectra. In such
instruments, the detector records the change of
signal with time, which is practically not
useful. However, Fourier transformation of the
time domain signal yield a frequency domain
spectrum, which is the usual signal, obtained by
conventional methods. Instruments that rely on
decoding a coded signal is also said to have a
multiplex design.
47Applications of Plasma Sources
- 1. Since plasma sources result in a very large
number of emission lines, these sources can be
used for both qualitative and quantitative
analysis. - 2. The signal obtained from plasma sources is
stable, has a low noise and background, as well
as freedom from interferences. - 3. Requires sample preparation similar to AAS
48- 4. Plasma sources are usually best suited for
operation in the ultraviolet region, therefore,
elements having emission lines below 180 nm (like
B, P, S, N, and C) can be only analyzed under
vacuum since air components absorb under 180 nm.
Also, alkali metals are difficult to analyze
since their best lines under plasma conditions
occur in the near infrared. - 5. An analytical emission line can easily be
located but will depend on the other elements
present since spectral line interferences are
encountered in plasma sources due to the very
high temperatures used.
49- 6. Linear calibration plots are usually obtained
but departure from linearity is observed at high
concentrations due to self absorption as well as
other instrumental reasons. An internal standard
is often used in emission methods to correct for
fluctuations in temperature as well as other
factors. The calibration plot in this case is a
plot between the concentration of analyte and the
ratio of the analyte to internal standard signal.
The internal standard is a substance that is
added in a constant amount to all samples,
blanks, and standards therefore it must be
absent from initial sample matrix. The internal
standard should have very close characteristics
(both chemically and physically) to analyte.
50Elements by ICP-AES
Different elements have different emission
intensities. Alkalis (Na, K, Rb, Cs) are weakly
emitting. Alkaline Earths (Be, Mg, Ca, Sr, Ba )
are strongly emitting.
51Concepts, Instrumentation, and Techniques in
Inductively Coupled Plasma Optical Emission
Spectrometry, Boss and Freeden, Perkin Elmer
52ICP/OES INTERFERENCES
- Spectral interferences
- caused by background emission from continuous or
recombination phenomena, - stray light from the line emission of high
concentration elements, - overlap of a spectral line from another element,
- or unresolved overlap of molecular band spectra.
- Corrections
- Background emission and stray light compensated
for by subtracting background emission determined
by measurements adjacent to the analyte
wavelength peak. - Correction factors can be applied if interference
is well characterized - Inter-element corrections will vary for the same
emission line among instruments because of
differences in resolution, as determined by the
grating, the entrance and exit slit widths, and
by the order of dispersion.
53Physical interferences of ICP
- cause
- effects associated with the sample nebulization
and transport processes. - Changes in viscosity and surface tension can
cause significant inaccuracies, - especially in samples containing high dissolved
solids - or high acid concentrations.
- Salt buildup at the tip of the nebulizer,
affecting aerosol flow rate and nebulization. - Reduction
- by diluting the sample
- or by using a peristaltic pump,
- by using an internal standard
- or by using a high solids nebulizer.
54Interferences of ICP
Chemical interferences include molecular
compound formation, ionization effects, and
solute vaporization effects. Normally, these
effects are not significant with the ICP
technique. Chemical interferences are highly
dependent on matrix type and the specific analyte
element.
55Memory interferences
When analytes in a previous sample contribute to
the signals measured in a new sample. Memory
effects can result from sample deposition on the
uptake tubing to the nebulizer from the build up
of sample material in the plasma torch and spray
chamber. The site where these effects occur is
dependent on the element and can be minimized by
flushing the system with a rinse blank between
samples. High salt concentrations can cause
analyte signal suppressions and confuse
interference tests.
56INDUCTIVELY COUPLED PLASMA-MASS
SPECTROMETRY (ICP-MS)
- Very sensitive and good for trace analysis -
Plasma produces analyte ions - Ions are directed
to a mass spectrometer - Ions are separated on
the basis of their mass-to-charge ratio - A very
sensitive detector measures ions - Very low
detection limits
57SUMMARY
Inductively Coupled Plasma Emission - High cost -
No lamp required - Low background signals - Low
interference - Moderate sensitivity Inductively
Coupled Plasma-Mass Spectrometry - Very high
cost - No lamp required - Least background
signals - Least interference - Very high
sensitivity
58Techniques for elemental analysis
ICP-MS ICP-AES FAAS GFAAS
Detection Limits Excellent Good Good
Excellent Productivity Excellent Very
good Good Low LDR 10 5 10 6 /10 10
HDD 10 3 10 2 Precision 1-3 0.3-2
0.1-1 1-5 Spectral
interference Few Common Almost none Very
few Chemical interference Moderate Few Many
Many Ionization Minimal Minimal Some
Minimal Mass efffects High on low none none
none Isotopes Yes none none
none Dissolved solids 0.1-0.4 up to 30
0.5-3 up to 30 No. of
elements 75 73 68 50 Sample
usage low medium high very
low Semi-quantitative yes yes no
no Isotope analysis yes no no
no routine operation Skill required easy easy
skill required Method
development skill required skill required easy
skill required Running
costs high high low
medium Capital costs very high high low
medium
59Emission Spectroscopy Based on Arcs and Sparks
- Samples are excited in the gap between a pair of
electrodes connected to a high potential power
supply (200 VDC or 2200-4400 VAC). The high
potential applied forces a discharge between the
two electrodes to occur where current passes
between the two separated electrodes (temperature
rises due to very high resistance).
60- The very high temperature (4000-5000 oC) realized
in the vicinity between the two electrodes
provide enough energy for atomization and
excitation of the samples in this region or when
the sample is, or a part of, one of the
electrodes. - Arc and spark methods are mainly used as
qualitative techniques and can also be used as
semiquantitative techniques.
6110B. Arc and Spark AES
- Arc and Spark Excitation Sources
- Limited to semi-quantitative/qualitative analysis
(arc flicker) - Usually performed on solids
- Largely displaced by plasma-AES
- Electric current flowing between two C electrodes
62Sample Handling and Preparation
- If the sample is conductive and is of a shape
that can be directly used as an electrode (like a
piece of metal or coin), that would be the choice
for sample introduction in arc and spark
techniques. Otherwise, powdered solid samples are
mixed with fine graphite and made into a paste.
Upon drying, this solid composite can be used as
an electrode. The discharge caused by arcs and
sparks interacts with the surface of the solid
sample creating a plume of very fine particulates
and atoms that are excited and emission is
collected. The figure below shows some common
shapes of graphite electrodes used in arc and
spark sources.
63Carbon electrodes
Sample pressed into electrode or mixed with Cu
powder and pressed - Briquetting
(pelleting) Cyanogen bands (CN) 350-420 nm occur
with C electrodes in air -He, Ar
atmosphere Arc/spark unstable each line measured
gt20 s needs multichannel detection
64Instruments for Arcs and Sparks
- In most cases, emission from atoms in an arc or
spark is directed to a monochromator with a long
focal length and the diffracted beams are allowed
to hit a photographic film. This typical
instrument is called a spectrograph since it uses
a photographic film as the detector.
65spectrograph
Beginning 1930s photographic film Cheap Long
integration times Difficult to develop/analyze Non
-linearity of line "darkness
66(No Transcript)
67(No Transcript)
68- The blackness of the lines on the photographic
film is an indication of the intensity of the
atomic line and thus the concentration of the
analyte. The location of emission lines as
compared to standard lines on a film serves to
identify the wavelengths of emission lines of
analyte and thus its identity. The use of
spectrographs is not very convenient since a lot
of time and precautions must be spent on
processing and calibrating the photographic film.
69- Qualitative analysis is accomplished by
comparison of the wavelengths of some emission
lines to standards while the line blackness
serves as the tool for semiquantitative analysis. - Polychromators are also available as multichannel
arc and spark instruments. However, these have
fixed slits at certain wavelengths in order to do
certain elements and thus they are not versatile.
70(No Transcript)
71- Recently, arc and spark instruments based on
charge injection and charge coupled devices
became available. These have extraordinarily high
efficiency and performance in terms of easier
calibration, short analysis time, as well as
superior quantitative results.
72(No Transcript)
73Characteristics of Arc Sources
- 1. Typical temperatures between 4000-5000 oC are
high enough to cause atomization and excitation
of sample and electrode materials. - 2. Usually, cyanogens compounds are formed due to
reaction of graphite electrodes with atmospheric
nitrogen. Emission bands from cyanogens compounds
occur in the region from 350-420 nm.
Unfortunately, several elements have their most
sensitive lines in this same region which limits
the technique. However, use of controlled
atmosphere around the arc (using CO2, Helium, or
argon) very much decreases the effect of
cyanogens emission.
74- 3. The emission signal should be integrated over
a minute or so since volatilization and
excitation of atoms of different species differ
widely. While some species give maximum signal,
others may still be in the molecular state. - 4. Arc sources are very good for qualitative
analysis of elements while only semiquantitative
analysis is possible. It is mandatory to compare
the emission spectrum of a sample with the
emission spectrum of a standard. In some cases, a
few milligrams of a standard is added to the
sample in order to locate the emission lines of
the standard and thus identify the emission
wavelengths of the different elements in the
sample. A comparator densitometer can be used to
exactly locate the wavelengths of the standard
and the sample components.
75The lines from the standard are projected on the
lines of the combined sample/standard emission
spectra in order to identify sample components.
Only few lines are shown in the figure.
76(No Transcript)
77(No Transcript)
78(No Transcript)
79Why use Carbon in Atomic Spectroscopy?
- We have previously seen the use of graphite in
electrothermal AAS as well as arc and spark AES,
even though molecular spectra are real problems
in both techniques due to cyanogens compounds
absorption and emission. The reasons after
graphite common use in atomic spectroscopy can be
summarized below
80- It is conductive.
- It can be obtained in a very pure state.
- Easily available and cheap.
- Thermally stable and inert.
- Carbon has few emission lines.
- Easily shaped.
81Spark Sources
- Most of the instruments in this category are arc
based instruments. Spark based instruments are of
the same idea except for a spark source
substituting an arc source. The spark source is
constructed as in the figure below where an AC
potential in the order of 10-50 KV is discharged
through a capacitor which is charged and
discharged through the graphite electrodes about
120 times/s resulting in a discharge current of
about 1000 A.
82This very high current will suffer a great deal
of resistance, which increase the temperature to
an estimated 40000 oC. Therefore, ionic spectra
are more pronounced.