Title: Atomic Spectroscopy
1Atomic Spectroscopy
2Review and Comparisons
- Atomic spectroscopy
- Absorption and emission of UV-VIS light
- Atoms and monoatomic ions
- Conceptually similar to absorption and emission
of UV-VIS light by molecules
3Review and Comparisons
- Atomic spectroscopy
- Differences from UV-VIS
- Limited to the elements
- METALS
- Most analysis is for metals!
- Sample preparation
- Place metals in water solution
- Metals present as ions in water
- Must have a means for converting metal ions into
free gas phase ground state atoms to be measured - Called atomization
- Uses large amount of thermal energy
4Review and Comparisons
- ATOMIC SPECTROSCOPY
- Differences from UV-VIS
- SAMPLE CONTAINER
- SOURCE OF THE THERMAL ENERGY NEEDED FOR THE
CONVERSION OF IONS IN SOLUTION TO ATOMS IN THE
GAS PHASE - Atomizer
- Does NOT resemble a cuvette
- Has a flame container
5Review and Comparisons
- ATOMIC SPECTROSCOPY
- Differences from UV-VIS
- Various types of atomizer and instrument designs
- Based on the same theory
- Spectral line sources used as light sources
- Instead of continuums sources in UV-VIS
- Several require NO light source at all!
- Limited types of analytes
- . Quantitation is well known
- Using elements that are well characterized
- Look up spectra in most reference text
6Summary of Techniques and Instrument Designs
- Most important aspect thermal energy
- Flame atomic absorption (flame AA)
- Graphite furnace atomic absorption (graphite
furnace AA) - Inductively coupled plasma atomic emission (ICP)
- Less important
- Flame emission and atomic fluorescence
- Two that do not require thermal energy/minimal
thermal energy - Cold vapor mercury system
- Metal hydride generation
- One that requires electrical energy
- Arc and spark emission
7Summary of Techniques and Instrument Designs
- Flame AA
- Large flame as the atomizer
- Sample-solution-drawn into the flame by a vacuum
mechanism - Atomization occurs immediately
- Light beam for the absorption measurements
directed through width of the flame
8Summary of Techniques and Instrument Designs
- Graphite furnace
- Actually a small graphite tube
- Quickly electrically heated to a very high
temperature - Small volume of sample solution placed in tube
- Manually with a micropipette
- Drawn with a vacuum
- Electrically brought to high temperature to
atomize sample - Light beam directed through the tube and measured
- There is a cloud of atoms
9(No Transcript)
10Summary of Techniques and Instrument Designs
- ICP
- Emission technique
- Does not use a light source
- Light measured is light emitted by the
atoms/monoatomic ions in the atomizer - Atomizer
- Extremely hot plasma
- High-temperature ionized gas composed of
electrons and positive ions - Confined by a magnetic field
- Extremely high temperatures
- Atoms and monoatomic ions undergo sufficient
excitation - Relatively intense emission spectra result
- Sample drawn with vacuum
- Intensity of an emission line is measured and
related to concentration
11(No Transcript)
12Flame Atomic Absorption
- Flames and Flame Processes
- After metal ions introduced into flame, several
processes occur in rapid order (fig. 9.4, pg.
248) - Solvent evaporates
- Leaves behind formula units
- Dissociation of salt into atoms
- Metal ions atomize/transformed into atoms
- Atoms raised to excited states by thermal energy
of the flame - A resonance process occurs
- Atoms resonate back and forth between ground
state and excited statte
13Flame Atomic Absorption
- Flames and Flame Processes
- Only small of atoms- state at any moment
- Atoms drop back to ground state
- Emission spectrum emitted
- Atoms in the excited state
- Emit light in the visible region of spectrum
- Entire flame in element takes on color
characteristic of the element that is in the
flame - Each element has a characteristic color
- It an atomic fingerprint
- Possible to quantitate these elements using flame
emission
14Flames and Flame Processes
- Flames and Flame Processes
- Unexcited atoms in the flame- 99.9
- Available to be excited by a light beam
- Light source used
- Light beam directed through the flame
- It is a Beers law experiment
- Width of the flame being the pathlength
- Flame temperature important both for the
atomization and excitation process
15Flames and Flame Processes
- Flame Atomic Absorption
- Flame requirements
- Fuel and an oxidant
- Natural gas and air
- Max temperature 1800K
- Does not sufficiently atomize most metal ions
- Does not excite a sufficient of atoms for
quantitation - . need something hotter!
- Acetylene as fuel and air is the oxidant
- Max temperature 2300K
16Spectral Line Sources
- Light sources emit spectral lines
- Lines in the line spectrum of the analyte being
measured - Preferred b/c they represent the precise
wavelengths needed for the absorption in the
flame - Flame contains this particular analyte
- Emitted b/c they contain the analyte to be
measured - When lamp is on
- Internal atoms are raised to the excited state
- Emit their line spectrum when they return to the
ground state - This is the light directed through the flame
17Hollow Cathode Lamp
- Hollow Cathode Lamp
- Most widely used spectral line source
- Cathode
- Negative electrode
- Contains the internal atoms
- Hollowed cup
- Internal excitation and emission process occurs
inside this cup when lamp is on - Anode
- Positive electrode
- Connected with cathode to a high voltage
- Light emitted
18Hollow Cathode Lamp
- Hollow Cathode Lamp
- Sealed glass tube
- Filled with inert gas at low pressure
- Neon or argon
- How it works (fig.9.6, pg 251)
- Lamp turned on and argon atoms ionize
- Positively charged argon ions then crash into the
negatively charged cathode - Causes sputtering
- Transfer of surface atoms in the solid phase to
the gas phase due to the collisions - More collisions of argon ions with metal atoms
cause metal atoms to be raised to the excited
state - Light emitted with they drop back to the ground
state
19(No Transcript)
20Hollow Cathode Lamp
- Hollow Cathode Lamp
- Must contain the element being measured
- Usually have number of different lamps in stock
- Interchanged in the instrument
- Some are multi-elemental
- Several different specific atoms present in the
lamp - Separated by a monochromator after the flame to
isolate the specific spectral line of the analyte
21Premix Burner
- Premix burner
- Burner for flame AA
- All components-fuel, oxidant, and sample
solution-are premixed - Take common path to the flame
- Fuel and oxidant
- Originate from pressurized sources
- Compressed gas cylinders
- Flow is controlled for optimum rate
22Premix Burner
- Sample solution
- Aspirated by vacuum
- Converted to aerosol/fine mist before mixing
- Accomplished with a nebulizer at head of mixing
chamber - Resembles nozzle to create a water spray (fig.
9.7, pg. 252) - Connects to the sample tube
- Pulls sample into the mixing chamber
- Produces aerosol spray
23(No Transcript)
24Premix Burner
- Aerosol spray
- Emerges from nebulizer
- Contains variable-sized solution droplets
- Also mixed with oxidant and fuel
- Contains impact device
- Baffles or glass bead near tip of nozzle
- Separates larger particles (fall to bottom of
chamber) - 90 of sample never reaches flame
25Optical Path
- Arranged in this order
- Light source, flame (sample container),
monochromator, and detector - Flame
- Positioned in open area
- Light can leak from room light and flame
- Monochromator located between flame and detector
- Detector
- Receives alternating light signals
- Source light and flame emissions
- Flame emissions only
- Detector able to eliminate flame emissions by
subtraction
26Optical Path
- Either single-beam or double beam
- Single-beam
- Fewer problems than in UV-VIS
- Seldom measure absorption spectra
- Wavelength seldom changed
- No need for re-calibration with blank as in
UV-VIS - Source drift and fluctuations still exist
- Minimized with improved electronics
27Optical Path
- Either single-beam or double beam
- Double beam
- Uses beam splitter
- Diverts light from source around the flame
- Two beams joined again before entering the
monochromator - Eliminates problems due to source drift and noise
- Source warm-up time eliminated since changes in
intensity compensated fro - Rapid changeover of lamps possible
28Practical Matters and Applications
- Slits and Spectral Lines
- More than one spectral line for an element
- .more than one line to choose from for setting
the monochromator - One line gives the optimum absorptivity
- Pick that one!!!!
- Found on the HCL
- Called the primary line
- Monochromator usually set at that wavelength
- Others called the secondary lines
- May be chosen if 1O is inappropriate
- If another element is the sample is similar to 1O
- Automated equipment usually set to primary line
29Practical Matters and Applications
- Slits and Spectral Lines
- Slit control
- Helps correct problem of close lines
- Wider the slit
- Greater the bandpass
- More incidental/close spectral lines allowed to
be captured - Usually choose between 0.2 and 2.0 nm
- Value represents the bandpass for both entrance
and exits slits - If interfering line at the optimum setting
- Slit is narrowed
- Or 2O line chosen
- Both result in less desirable sensitivity
30Practical Matters and Applications
- Hollow Cathode Lamp Current
- Current is adjustable
- Optimum setting represents most intense light
without shortening the life of the lamp (VERY
EXPENSIVE!) - Lamp Alignment
- Must have proper alignment for optimum intensity
through the optical path - May need adjustment when changing lamps
31Practical Matters and Applications
- Interferences
- Causes
- Chemical sources
- Spectral sources
32Practical Matters and Applications
- Chemical interferences
- Result of problems with sample matrix
- Viscosity/surface
- May affect aspiration rate
- Nebulized droplet size
- Standard additions method
- Certain volume of the sample solution present in
same proportion in all standard solutions - Equivalent to adding standard amounts of analyte
to the sample solution - Solves interference problem
- Sample matrix always present in same concentration
33Practical Matters and Applications
- Standard additions method
- Prepare standards in usual way
- Add a volume of the sample solution to each
before diluting to the mark with solvent - Gives a series of standards which the
concentration of analyte added known - Standard curve
- Its a plot of absorbance vs. concentration added
instead of just concentration - Y-axis is not true position
- Offset to the right by the concentration of the
zero-added concentration, which is the sample
solution - Concentration of THIS solution is the
concentration sought - To show on graph
- Curve extrapolated to intersect with the x-axis
(y 0) - Represents the concentration of the amount added
- Precise concentration in the zero-added solution
found using the equation of the straight line - YmX b
34Spectral Interferences
- Spectral Interferences
- Caused by substances in the flame
- Absorb same wavelength as analyte
- Causes absorbance measurement to be high
- Rarely an element
- If suspected, switch to 2O wavelength
- More often caused by presence of light-absorbing
molecules in the flame and light dimming due to
small particles - Called background absorption
- Fix by background corrections
- Subtract background interference
- Its the purpose of the deuterium lamp
35Safety and Maintenance
- Safety issues with the AA
- Acetylene, flame, and contamination of lab air
with combustion products - Acetylene
- Compressed gas cylinders must be secured to
immovable object-the wall - Approved pressure regulators in place
- Tubing free of leaks
- Must have independently operated vent hood over
flame - Removes excess solvent fumes
- No volatile fumes near the flame!
36Safety and Maintenance
- Safety issues with the AA
- Flashbacks
- From improperly mixed fuel and air
- When flow regulators ore improperly set
- When air is drawn back through drain line of
premix burner - Cleaning
- Burner head and nebulizer
- Ensures minimal noise level from impurities in
flame - Carbon deposits in slit
- Scrape with sharp knife or razor blade
37Summary
- Sensitivity
- Concentration of an element that will produce and
absorption of 1 - Smallest concentration that can be determined
with a reasonable degree of precision - Detection limit
- Concentration that gives a readout level that is
double the electrical noise level inherent in the
baseline. - Qualitative parameter in that it is the minimum
concentration that can be detected - Not precisely determined
- Would tell the analyst that the element is
present - Not at a precisely determinable concentration
level - Table 9.2 and 9.3, pg. 267