Operation of the new Quantum HF Odyssey X-ray Room
  • This room is used with the digital radiography
    equipment.

2
The Quantum Equipment Features
  • The Quantum Room features
  • A more powerful generator with mA up to 600 mA
    for faster exposure and larger patients
  • 0.6 and 1.2 focal spot x-ray tube for superior
    resolution
  • Tilting Wall Bucky for upper cervical technique
  • Elevating Floating Table Top Table
  • AGFA Digital Full spine Film Holder

3
Tilting Wall Bucky
  • The Tilting Wall Bucky will tilt 15º for
    positioning the upper cervical base posterior.
  • We still use the non-Bucky film holder with the
    Bucky.
  • Note the chin rest for chest radiography.

4
Tilting Wall Bucky
  • The Bucky design allows the patient to get closer
    to the Bucky for lateral views.
  • A button on the handle releases the lock to raise
    or lower the unit.
  • Buttons on the support are used to angle the
    Bucky.

5
Cassette Centering
  • For the 24CM by 30 CM cassette or 43 cm by 35
    cassette orientation, the cassette is raised to
    the center.
  • The tray is calibrated for proper placement of
    the cassette support.

6
Cassette Centering
  • By centering the cassette, the tray can be pushed
    in sooner and a notch on the handle used to
    center the beam to the film.

7
Full Spine Film Holder
  • On the opposite side of room to the Tilting Bucky
    is the AGFA Full Spine film holder.
  • Three 35 cm by 43 cm cassettes are loaded for
    full spines.
  • The computer stitches the images together for
    digital full spines.

8
Full Spine Film Holder
  • Note that we still use the Nolan Filtration
    System with the digital system.
  • We also have a radiographic ruler to perform
    scanograms.

9
Radiographic Table
  • The table lowers to 21 allow easy access for
    the patient.
  • It is then raised to imaging height( 32.5).
  • The table top floats to afford easy patient
    positioning. Float control is controlled by foot
    pedals or handle at foot of table.

10
Radiographic Table
  • Both the Table and Wall Buckies are true Buckies
    with osculating 200 line grids.
  • The Object Film Distance is slightly increased.

11
Table Bucky Tray
  • Like with the Wall Bucky, the film should be
    centered but the support is not needed.
  • The Bucky rolls about 3 feet to aide in
    positioning

12
Tube Controls
  • The tube angle has an electric lock so do not
    angle the tube without pushing the angle button.
  • Lights come on the tube when you are at 40 or
    72 SID to either Vertical Bucky or the Table
    Bucky.

13
Tube Controls
  • The Longitudinal Lock is what we called the
    Horizontal Lock in Laboratory or on a Universal
    Tube Stand.
  • Do not touch the Transverse Lock Button when
    using the Bucky as you can miss align the tube to
    the Bucky with it.

14
Tube Controls
  • The All Locks Button will control all of the
    locks at once, including the transverse and angle
    lock.
  • The collimator is calibrated in metric and
    English measurements.

15
Tube Controls
  • The Collimator has a laser for film centering.
  • The tube angle indicator is very accurate.

16
SID Scales for Table
  • There are two scales for the Table SID.
  • They are calibrated for the full up table top
    position.
  • One is for the table top and the other for the
    table Bucky.

17
Wall Bucky SID Scale
  • The Wall Bucky SID scale is on the floor rail.

18
Transverse Tube Angulation
  • The tube may be angled across the table or in the
    transverse plane.
  • The scale is on the bottom of the tube support
    and the lock on the top.
  • We do not use this feature except to realign the
    tube to the Bucky.

19
Quantum Odyssey HF Generator
  • One thing you will notice is the room does not
    have a wall full of technique charts.
  • All of our charts are programmed into the
    Anatomical Programming of the control.

20
Quantum Odyssey HF Generator
  • The Quantum generator offers more user selections
    than the Universal Control Panel.
  • You may select mA and time instead of just mAs.
  • You need to select the correct Bucky or Film
    Holder.

21
Quantum Odyssey HF Generator
  • It shows controls for Automatic Exposure but our
    unit is not equipped with AEC.
  • We will always use the Anatomical Programming
    feature for setting technical factors.

22
Anatomical Programming
  • Select the region by pressing the button next to
    the region.

23
Anatomical Programming
  • Select the View by pressing the button next to
    the View.

24
Anatomical Programming
  • Use the buttons next to the arrows to enter the
    patient measurements.

25
Anatomical Programming
  • Adjustment for habitus can be made in 25 changes.

26
Anatomical Programming
  • Note that the focal spot and even the image
    receptor is selected for normal exams.
  • For hips, lumbar obliques and knees, the image
    receptor may need to be changed.

27
Anatomical Programming
  • When everything is set, the ready light will be
    on.
  • Press the prep button until it lights up. You
    will hear the rotor rotating.
  • When the prep light is on also press the expose
    button to expose the image.

28
Anatomical Programming
  • Return to PB-331 Radiologic Technology Two Home
    Page.