Title: Introduction to Electron Microscopy Fundamental concepts in
1Introduction to Electron Microscopy
- Fundamental concepts in electron microscopy
- The construction of transmission and scanning
electron microscopes - Sample examples of application
2Electron Microscope vs. Optical Microscope
(first one built in 1931 by Ruska and Knoll)
(Leeuwenhoek in 17th century)
- Electron vs. Photon
- Electron charged, has rest mass, not visible
- Photon neutral, has no rest mass, visible at the
wavelength 400 nm-760 nm.
Because of these differences, the microscope
construction will also be different
What is the common property?
3Comparison of EM and LM
- a. Similarities (Arrangement and function of
components are similar) 1) Illumination system
produces required radiation and directs it onto
the specimen. Consists of a source, which emits
the radiation, and a condenser lens, which
focuses the illuminating beam (allowing
variations of intensity to be made) on the
specimen. 2) Specimen stage situated between
the illumination and imaging systems. 3)
Imaging system Lenses which together produce the
final magnified image of the specimen. Consists
of i) an objective lens which focuses the beam
after it passes through the specimen and forms an
intermediate image of the specimen and ii) the
projector lens(es) which magnifies a portion of
the intermediate image to form the final image.
4) Image recording system Converts the
radiation into a permanent image (typically on a
photographic emulsion) that can be viewed.
4Comparison of EM and LM
- b. Differences 1) Optical lenses are generally
made of glass with fixed focal lengths whereas
magnetic lenses are constructed with
ferromagnetic materials and windings of copper
wire producing a focal length which can be
changed by varying the current through the coil.
2) Magnification in the LM is generally changed
by switching between different power objective
lenses mounted on a rotating turret above the
specimen. It can also be changed if oculars
(eyepieces) of different power are used. In the
TEM the magnification (focal length) of the
objective remains fixed while the focal length of
the projector lens is changed to vary
magnification. 3) The LM has a small depth of
field, thus different focal levels can be seen in
the specimen. The large (relative) depth of field
in the TEM means that the entire (thin) specimen
is in focus simultaneously. 4) Mechanisms of
image formation vary (phase and amplitude
contrast). 5) TEMs are generally constructed
with the radiation source at the top of the
instrument the source is generally situated at
the bottom of LMs. 6) TEM is operated at high
vacuum (since the mean free path of electrons in
air is very small) so most specimens (biological)
must be dehydrated (i.e. dead !!). 7) TEM
specimens (biological) are rapidly damaged by the
electron beam. 8) TEMs can achieve higher
magnification and better resolution than LMs.
9) Price tag!!! (100x more than LM)
5Resolution of a microscope
Where N.A. is the numerical aperture n(sina)
6The resolution is proportional to the
wavelength!
Electron equivalent wavelength and accelerating
voltage
The dualism wave/particle is quantified by the De
Broglie equation ? h/p h/mv ? wavelength
h Planck constant p momentum
The energy of accelerate electrons is equal to
their kinetic energy E eV m0v2/2 V
acceleration voltage e / m0 / v charge / rest
mass / velocity of the electron These equations
can be combined to calculate the wave length of
an electron with a certain energy p m0v
(2m0eV)1/2 ? h / (2m0eV)1/2 ( 1.22 / V1/2 nm)
At the acceleration voltages used in TEM,
relativistic effects have to be taken into
account (e.g. Egt100 keV) ? h / 2m0eV (1
eV/2m0/c2)1/2
7 Wavelength and accelerating voltage
There are other factors that limit the resolution!
8Types of Electron Microscope
- Transmission Electron Microscope (TEM) uses a
wide beam of electrons passing through a thin
sliced specimen to form an image. This microscope
is analogous to a standard upright or inverted
light microscope - Scanning Electron Microscope (SEM) uses focused
beam of electrons scanning over the surface of
thick or thin specimens.. Images are produced one
spot at a time in a grid-like raster pattern.
(will be discussed in a later lecture) - Scanning Transmission Electron Microscope (STEM)
uses a focused beam of electrons scanning through
a thin sliced specimen to form an image. The STEM
looks like a TEM but produces images as does an
SEM (one spot at a time). It is most commonly
used for elemental analysis of samples.
9(No Transcript)
10Leo 982 SEM
For SEM a fine probe (beam spot) is formed by
condenser lens and its size determines the
resolution (this differs from the TEM which is
diffraction limited)
11Electron Gun
FEG
W hairpin
LaB6 crystal
12Thermionic Sources
Increasing the filament current will increase the
beam current but only to the point of saturation
at which point an increase in the filament
current will only shorten the life of the emitter
13Beam spot image at different stage of heating
14(No Transcript)
15(No Transcript)
16Electromagnetic lens
17Condenser-lens system
C1 controls the spot size
The condenser aperture must be centered
C2 changes the convergence of the beam
18TEM imaging modes
19STEM image
Bright and dark field STEM image of Au particles
on a carbon film
What is the differences between this dark field
and the previous one?
20Magnification in TEM
Mob Mint Mproj Total Mag
Depends on the magnification, some lens may not
be used
21Electron-specimen interaction
22- Backscattered Electrons
- Formation
- Caused by an incident electron colliding with
an atom in the specimen which is nearly normal to
the incident's path. The incident electron is
then scattered "backward" 180 degrees.
Utilization - The production of backscattered electrons
varies directly with the specimen's atomic
number. This differing production rates causes
higher atomic number elements to appear brighter
than lower atomic number elements. This
interaction is utilized to differentiate parts of
the specimen that have different average atomic
number.
23Backscatter Detector
The most common design is a four quadrant solid
state detector that is positioned directly above
the specimen
24Gold particles on E. coli appear as bright white
dots due to the higher percentage of
backscattered electrons compared to the low
atomic weight elements in the specimen
25Backscatter image of Nickel in a leaf
26- Secondary Electrons
- Source
- Caused by an incident electron passing "near" an
atom in the specimen, near enough to impart some
of its energy to a lower energy electron (usually
in the K-shell). This causes a slight energy loss
and path change in the incident electron and the
ionization of the electron in the specimen atom.
This ionized electron then leaves the atom with a
very small kinetic energy (5eV) and is then
termed a "secondary electron". Each incident
electron can produce several secondary electrons. - Utilization
- Production of secondary electrons is very
topography related. Due to their low energy, 5eV,
only secondaries that are very near the surface
(lt 10 nm) can exit the sample and be examined.
Any changes in topography in the sample that are
larger than this sampling depth will change the
yield of secondaries due to collection
efficiencies. Collection of these electrons is
aided by using a "collector" in conjunction with
the secondary electron detector. The collector is
a grid or mesh with a 100V potential applied to
it which is placed in front of the detector,
attracting the negatively charged secondary
electrons to it which then pass through the
grid-holes and into the detector to be counted.
27A conventional secondary electron detector is
positioned off to the side of the specimen. A
faraday cage (kept at a positive bias) draws in
the low energy secondary electrons. The electrons
are then accelerated towards a scintillator which
is kept at a very high bias in order to
accelerate them into the phosphor.
28The position of the secondary electron detector
also affects signal collection and shadow. An
in-lens detector within the column is more
efficient at collecting secondary electrons that
are generated close to the final lens (i.e. short
working distance).
29Secondary Electron Detector
Side Mounted In-Lens
What are the differences between these two images?
30- Auger Electrons
- Source
- Caused by the de-energization of the specimen
atom after a secondary electron is produced.
Since a lower (usually K-shell) electron was
emitted from the atom during the secondary
electron process an inner (lower energy) shell
now has a vacancy. A higher energy electron from
the same atom can "fall" to a lower energy,
filling the vacancy. This creates and energy
surplus in the atom which can be corrected by
emitting an outer (lower energy) electron an
Auger Electron. - Utilization
- Auger Electrons have a characteristic energy,
unique to each element from which it was emitted
from. These electrons are collected and sorted
according to energy to give compositional
information about the specimen. Since Auger
Electrons have relatively low energy they are
only emitted from the bulk specimen from a depth
of lt 3 nm
31- X-rays
- Source
- Caused by the de-energization of the specimen
atom after a secondary electron is produced.
Since a lower (usually K-shell) electron was
emitted from the atom during the secondary
electron process an inner (lower energy) shell
now has a vacancy. A higher energy electron can
"fall" into the lower energy shell, filling the
vacancy. As the electron "falls" it emits energy,
usually X-rays to balance the total energy of the
atom so it. - Utilization
- X-rays or Light emitted from the atom will have
a characteristic energy which is unique to the
element from which it originated. - (will be discussed in a separate lecture)
32- Unscattered Electrons
- Source
- Incident electrons which are transmitted through
the thin specimen without any interaction
occurring inside the specimen. - Utilization
- The transmission of unscattered electrons is
inversely proportional to the specimen thickness.
Areas of the specimen that are thicker will have
fewer transmitted unscattered electrons and so
will appear darker, conversely the thinner areas
will have more transmitted and thus will appear
lighter.
33The overlapping areas appear darker
34- Elasticity Scattered electrons
- Source
- Incident electrons that are scattered (deflected
from their original path) by atoms in the
specimen in an elastic fashion (no loss of
energy). These scattered electrons are then
transmitted through the remaining portions of the
specimen. - Utilization
- All electrons follow Bragg's Law and thus are
scattered according to ml2dsin q (angle of
scattering). All incident electrons have the same
energy(thus wavelength) and enter the specimen
normal to its surface. All incidents that are
scattered by the same atomic spacing will be
scattered by the same angle. These "similar
angle" scattered electrons can be collated using
magnetic lenses to form a pattern of spots each
spot corresponding to a specific atomic spacing
(a plane). This pattern can then yield
information about the orientation, atomic
arrangements and phases present in the area being
examined.
35The diffraction pattern is highly dependable on
the structure of the specimen
Dr. Schroeder will have a lecture dedicated to
diffraction
36- Inelastically Scattered Electrons
- Source
- Incident electrons that interact with specimen
atoms in a inelastic fashion, loosing energy
during the interaction. These electrons are then
transmitted trough the rest of the specimen - Utilization
- Inelastically scattered electrons can be
utilized two ways - Electron Energy Loss Spectroscopy (EELS) The
inelastic loss of energy by the incident
electrons is characteristic of the elements that
were interacted with. These energies are unique
to each bonding state of each element and thus
can be used to extract both compositional and
bonding (i.e. oxidation state) information on the
specimen region being examined. - Kikuchi Bands Bands of alternating light and
dark lines that are formed by inelastic
scattering interactions that are related to
atomic spacings in the specimen. These bands can
be either measured (their width is inversely
proportional to atomic spacing) or "followed"
like a roadmap to the "real" elasticity scattered
electron pattern.
37EELS of NiO
Combine with STEM, can do element mapping in TEM
compare to EDX, EELS is better for lighter
elements
Unlike diffraction pattern blinks on and off, the
Kikuchi line pattern rotates when one tilts the
crystal. Thus is helpful in orientating a crystal
to certain zone axis.
38Specimen interaction volume
- Atomic number of the material being examined
higher atomic number materials absorb or stop
more electrons and so have a smaller interaction
volume. - Accelerating voltage being used higher
voltages penetrate farther into the sample and
generate larger interaction volumes - Angle of incidence for the electron beam the
greater the angle (further from normal) the
smaller the volume
39Monte Carlo simulation of the interaction volume
Try the online simulation at
http//www.matter.org.uk/tem/electron_scattering.h
tm
40High-Resolution Electron Microscopy
Wave functions for elastically-scattered, forward
electrons
- For a weak phase object, the observable image
intensity is - I1 - 2??(r) ? FFTT(H).
- represents convolution.
- This shows a pure phase-contrast image.
Illustration of electron wave passing through
under phase object approximation. The phase
changes (contrast) is imaged.
41High resolution Images
Discovery of the carbon nanotube
Pt nanoparticles
S. Iijima, Nature 354, 56 (1991).