Title: Multiphoton and Spectral Imaging Multiphoton microscopy
1Multiphoton and Spectral Imaging
2Multiphoton microscopy
- Predicted by Maria Göppert-Mayer in 1931
- Implemented by Denk in early 1990s
- Principle Instead of raising a molecule to an
excited state with a single energetic photon, it
cam be raised to an excited state by the
quasi-simulatneous absorption of two (2-photon)
or 3 (3-photon) less energetic photons
3Multiphoton-photon Jablonski diagram
4Multiphoton
- In multiphoton microscopy, the intermediate state
is not a defined state, and so is quantum
forbidden - However, in quantum mechanics, forbidden is not
absolute - Therefore, the requirement for quasi-simultaneity
- Practically, it means within 10-18 seconds
- In single photon, probablility of excitation is
proportional to I in two-photon, it is
proportional to I2
5Excitation volume
http//www.loci.wisc.edu/multiphoton/mp.html
6Advantages of multiphoton microscopy
- Fluorescence excitation is confined to a
femtoliter volume less photobleaching - Excitation wavelengts are not absorbef by
fluorophore above plane of focus - Longer excitation wavelengths penetrate more
deeply into biological tissue - Inherent optical sectioning
7Increased contrast in multiphoton
Centonze,V.E and J.G.White. (1998) Biophysical J.
752015-2024
8Light sources
- Light flux necessary for multiphoton microscopy
can be achieved by femtosecond pulsed IR lasers - Ti-Sapphire lasers tunable from 700-900 nm
- http//micro.magnet.fsu.edu/primer/java/lasers/tsu
nami/index.html
9Spectra Physics Mai Tai, Coherent Chameleon
Tuning Ranges 680-1080 nm Sealed box units no
adjustments necessary Computer controlled
tuning Stable pointing as you scan spectrum
10Dyes for multiphoton microscopy
- Multiphoton excitation spectra for dyes is an
active field of exploration - Generally, 2PE peaks are broad
- General rule start a little more energetic than
?max for single photon - For example EGFP ?max for single photon 488
?max for two photon 900 nm
112PE Spectra
12Detector configuration for multiphoton
Molecular Expressions web site Note, in
particular the descanned detector and the Whole
Area PMT Detector Nondescanned detector.
13Descanned detector
- Uses same scan mirror to descan beam as was used
to scan it. - Better alignment with confocal
- However, only collects the amount of light
represented by the projection of the mirror onto
the specimen less sensitivity - Do not forget to open up the confocal pinhole,
because the nature of multiphoton restricts
excitation to a femtoliter volume
14Nondescanned detector
- Because our excitation volume is restricted to a
femtoliter volume, and is automatically an
optical section, we do not need to descan - Cone projected onto specimen is much wider, so
much more sensitivity - However, also much more sensitive to stray light
15Confocal spectral imaging
- In many case, the spectra of dyes overlap either
in their excitation spectrum, their emission
spectrum, or both. - What can we do?
- Excitation overlap for instance,
tetramethylrhodamine excitation spectrum overlaps
that of fluorescein, so if we use the 488 and 543
lines simulatanously, we see overlap - Solution
- Choose different dyey (fluorescein and Texas red)
- Multitracking (sequential scanning) excite at
488 while the fluorescein image is being
collected and at 543 while the rhodamine is.
16What about emission?
Choose different dyes
Molecular Probes
17Sometimes you cant avoid overlap
- Autofluorescence frequently overlaps fluorescein
emission - NADH/Flavoprotein on 2-P excitation at 800 nm,
the 450 nm NADH emission is clean, but the 550 nm
flavoprotein emission band has about 30 NADH
emission - Fluorescent proteins
18Example Lambda stack of cells expressing either
CFP or GFP on chromatin
19What do we do?
- Acquire a Lambda stack of our image
- Acquire a lambda stack of our reference dyes, or,
alternatively, identify areas in the image that
will be pure. - Mathematicall, through linear unmixing, apply
linear algebra to separate the individual dye
spectra from the multispectral image
20Linear Unmixing
- Different amounts of pink and blue generate
different spectra
21Pairwise comparison of dyes that can or cannot be
unmixed
Note that for pairs that cannot be unmixed (ie,
DiO and eGFP), the shape of the spectra are very
similar
22Unmixing fluorescein phalloidin and Sytox green
23Problems with linear unmixing
- It takes a lot longer to acquire lambda stacks
than single images - The software at least on the Leica is not
transparent to use
Solutions Zeiss META Both use a prism to
separate Nikon CSI the spectrum to multiple
channels Both have software that is
easier to use