1
Skull Cap Requirements Presentation
  • Group 5
  • Engineering 4017 - Engineering Design

2
Project Overview
3
General Description
  • What is EEG?
  • Why is the information being collected?

FOR MORE INFO...
  • Refer to the following webpage http//www.epub.org
    .br/cm/n03/tecnologia/eeg.htm

4
Mission Statement
  • The reliable measurement over extended periods of
    scalp EEG on consecutive occasions during mental
    function in everyday environments.

5
Existing System
  • Skull cap containing electrodes with wired
    connection.
  • Requires extensive preparation of scalp for
    usage.
  • System must be operated in a shielded environment.

6
Issues with Existing System
  • Lack of repeatability
  • Lack of mobility
  • Long setup time
  • Short length of experiment before the results are
    affected by external factors

7
Requirement Structure
  • Priority System
  • Essential - Match existing functionality and/or
    core customer requirements
  • Highly Desirable - Extend the functionality of
    the system and are achievable in the available
    time frame
  • Desirable - May not be achievable in the
    available time frame further research is
    required

8
Electrode Positioning
  • Essential
  • Position 23 electrodes on the skull according to
    the 10/20 positioning convention
  • 19 scalp electrodes
  • Two eye electrodes
  • Reference electrode
  • Isolated common

9
Signal Processing
  • Essential
  • Process electrode signals to make them suitable
    for input into customers existing computer
    system.

10
Experimental Repeatability
  • Essential
  • Ability to position system on subject on
    subsequent occasions so that electrodes are
    within 1mm2 of original position.

11
Acceptable Noise Level
  • Essential
  • Noise from external electromagnetic fields and
    undesired physiological rhythms must be less than
    1 of the recorded EEG signal.
  • E.g. power line hum, heart beats

12
Sampling Specifications
  • Essential
  • Rate 1kHz
  • Resolution better than .15?V
  • Range 400mV

13
Skull Impedance
  • Essential
  • Skull impedance must be reduced below 5k? for the
    purposes of electrode measurement.

14
Scalp Irritation
  • Essential
  • Based on ISO 10993-10, irritation should be
    reduced to a level such that less than 2 of
    subjects experience irritation (as defined in the
    standard).

15
Highly Desirable
  • Less than 1 deviation of data due to electrode
    polarisation.
  • Easily cleanable to medical grade cleanliness in
    accordance with Australian Standard AS 4187/1998.
  • Signal compensation for up to 0.05Hz of signal
    drift.

16
Highly Desirable, cont.
  • Suitable for use on persons of different head
    dimensions, within the range of 53 to 60
    centimetres head diameter as measured above the
    ears.
  • Possible to position system on subject's head
    within 15 minutes.

17
Desirable
  • Exist as a portable system that can operate for
    up to three hours without interruption.
  • Ability to transfer signals from the patient to a
    monitoring terminal without the need for physical
    connection.
  • Ability to select a subset of the EEG monitoring
    electrodes.

18
Desirable, cont.
  • Ability to select different sampling rates for
    data collection.
  • Suitable for use on wider range of adult head
    dimensions and extending to smaller sizes for
    children.

19
Conclusion
  • Requirements may be changed after consultation
    with the client
  • From these requirements, conceptual designs will
    be generated and evaluated, with the successful
    design being further refined into a final product
    design.

FOR MORE INFO...
  • Refer to the Groups Website - http//www.angelfir
    e.com/il/skullcap/