MECHANICAL DESIGN OF PLASMA ANALYSERS
  • J. Coker
  • Mullard Space Science Laboratory

2
Heritage of MSSL Plasma Analysers
  • The Hemispherical Micro Channel Plate (MCP) based
    Analysers were developed from earlier versions
    with Channeltron detectors. (Effectively just one
    poor of an MCP).
  • MCPs provided angular resolution and a much
    wider field of view for the first time.
  • The first MCP analyser was flown on the AMPTE
    mission in 1984.
  • Multilayer ceramic substrate anodes with wedge
    and strip pattern laser cut maximised angular
    resolution.

3
Principle
  • Angular acceptance
  • Typically 3o

Top Hat
e-
Outer Deflection Plate
Inner Deflection Plate
Field Defining Grid
MCP (amplifies charge)
Anode
4
Principle
  • Analyser focuses electrons
  • MCP detector

5
Cut Away CAD Model of CASSINI/CAPS Analyser
CAP/AMP BOARD
S/C INTERFACE
MCP ASSY.
GRID
OUTER HEMISPHERE (DEFLECTION PLATE)
LIGHT BAFFLES
HV. ELECTRONICS
TOP HAT
INNER HEMISPHERE (DEFLECTION PLATE)
DAUGHTER BOARDS
ENTRY APERTURE
MOTHER BOARD
ANODE
INSULATOR
6
The Two Main Design Drivers
  • Mass, one of the main considerations in deciding
    suitable design and manufacturing concepts
  • -Sheet metal parts are the lightest but less
    accurate and more difficult to produce with good
    repeatability.
  • -Turned/milled parts more accurate but higher
    mass.
  • -Electro Discharge Machining (EDM) enables
    lighter machined parts with better accuracy.
  • Accuracy, high accuracy requirements demand
    stiffer parts and dictate production techniques.
    (Always a compromise with mass).

7
GIOTTO
The First integral analyser, electronics and
mounting package
Spun hemispheres and light sheet structure
Complete GIOTTO instrument
8
Accuracy (Tolerance)
  • Cluster required 0.030mm tolerance on the 3.0mm
    analyser gap, (1/4 of a human hair) with absolute
    max. tolerance of 0.3mm
  • This tolerance was governed by 9 individual
    tolerances on 5 different parts.
  • Two of these parts are Ceramic and required CTE
    clearances.
  • All parts required assembly clearances.
  • The best tolerance likely to be achieved was
    assessed by calculating the root of the sum of
    the squared individual tolerances and gave /-
    0.790mm
  • However as most of the tolerances related to
    turned parts inherent concentricity was in our
    favour.
  • Final instrument performance suggested that we
    were within the 1 to 10 tolerance requested.

9
Engineering Models, CASSINI, PEPI, CLUSTER
CLUSTER
CASSINI
PEPI
10
Design Drivers Continued
  • Volume/FOV.
  • Spacecraft accommodation and mounting options.
  • Radiation screening.
  • Should be incorporated into the design, ie
    integral, not added as an after thought.
  • Brittle materials
  • Hard to guarantee survival
  • Usually have significantly different CTEs
  • MCP clamping a particular worry
  • High voltages
  • Breakdown gaps must be maintained
  • Connections and screening difficult

11
Multi Layer Ceramic Anodes
EARLY CLUSTER
CLUSTER Wedge Strip Pattern (laser cut)
12
TIMAS-Detector assembly for Toroidal Analyser
75mm outer radius Micro Channel Plates
13
Design Drivers Continued
  • Vibration
  • First Eigen frequency should be gt100Hz
  • Usually design for 100g dynamic load
  • All fixings must be locked. (difficult for small
    screws)
  • Thermal
  • Must ensure adequate CTE allowance
  • Must consider instrument thermal control.
    (Usually involves careful choice of finishes and
    thermal blanket provision to maintain acceptable
    electronics temperatures.
  • Contamination
  • Electronics
  • Normally required to be in close proximity to
    analyser and generally housed in integral
    structure

14
Electronics and Supporting Structure
Components for CLUSTER DPU
CLUSTER analyser assemblies
15
Electronics and Supporting Structure
Analyser head body used to stiffen mounting
flanges (Dual purpose gives high structural mass
efficiency)
16
CASSINI Electronics Main Structure
A compact electronics package with good
interchangeability for most PCBs
Checking flexible circuit card for fit
Final installation package
17
Blacking considerations
  • Ebanol-C surface conversion for light rejection.
  • The process involves immersing the part to be
    blacked in a succession of hot and cold chemical
    baths.
  • Problems in sealing non blacked surfaces on
    delicate parts.
  • Parts can only be reclaimed once if process
    fails.
  • Blacking jigs have limited life due to chemical
    erosion.
  • Difficult to handle and transport blacked parts.
  • Difficult assembly, (if blacked surface is
    touched finish is damaged).

18
Blacking jigs, before and after blacking
Parts for PEPI lab analyser
19
Future Developments
  • The trend is to work towards smaller, lower mass
    instruments, but this accentuates the problems
    already facing us as -
  • Accuracy requirements become more acute. (SWRI)
    have produced a smaller instrument but at the
    cost of a larger percentage inaccuracy.
  • Operating voltages stay the same, so prevention
    of breakdown is more difficult.
  • Fixings, lack of volume to accommodate them.
  • PEPI instrument is the first with modular mcp
    assembly achieved by supporting inner hemisphere
    on spokes. Unfortunately the volume of the mcp
    assembly forces instrument size back to CLUSTER
    volume.

20
Sub assemblies for PEPI lab analyser
21
Principle
  • Angular acceptance
  • Typically 3o
  • Require 30o

0 V
e-
HV
22
Specifying Instrument requirements
  • Think carefully about prime concerns/aims. Eg
    mass, accuracy, volume etc. as these will affect
    science capability.
  • Try to give realistic minimum requirements.
  • Minimise restrictions.
  • Remember that the final design will be a
    compromise between all requirements.

23