Group Members: - PowerPoint PPT Presentation

1 / 17
About This Presentation
Title:

Group Members:

Description:

Group Members: Ryan Augustine. Brian Frederick. Nate Gaeckle. Gordy Lawrence. Advisor: ... A robotic arm has previously been developed to integrate ... – PowerPoint PPT presentation

Number of Views:46
Avg rating:3.0/5.0
Slides: 18
Provided by: bme4625
Category:

less

Transcript and Presenter's Notes

Title: Group Members:


1
  • Group Members
  • Ryan Augustine
  • Brian Frederick
  • Nate Gaeckle
  • Gordy Lawrence
  • Advisor
  • Professor Willis Tompkins
  • Client
  • Dr. Myron Pozniak

2
Abstract
  • A robotic arm has previously been developed to
    integrate
  • information provided from a CT scan in order to
    place a
  • biopsy needle in the proper trajectory. A
    prototype has
  • been constructed that uses air pressure to drive
    a biopsy
  • needle into a patient via a piston/cylinder
    assembly with
  • a compression spring to retract the needle. This
    prototype
  • has been tested on specimens that have comparable
    tissue
  • to humans to see if there was noticeable
    deflection of the
  • needle. No deflection was observed. The biopsy
  • mechanism will be activated upon impact with the
    end of
  • the cylinder in a fully automatic fashion. In
    the future, the
  • needle will also be able to be retracted
    automatically by the
  • combination of an LED and photodiode sensor, or
    an
  • electrode sensor.

3
Motivation
  • Problems with current drive mechanism
  • Slow needle advancement 2 cm/s
  • Slower biopsies increases risk of internal damage
  • Creates tumor movement
  • Makes use of the robotic arm infeasible

4
Problem Definition
  • Biopsies are used in conjunction with a CT scan
  • Many dilemmas with manual biopsy needle injection
  • Inaccurate
  • Physician exposure to radiation
  • Sometimes multiple attempts at tumor capture
  • Robotic Arm created to eliminate these problems

5
Problem Statement
  • Our goal is to create a needle driver that
  • Drives a needle fast to counter tumor movement
    and patient injury
  • Retracts the needle with speed and ease with a
    sample of the tumor
  • Is compatible with CT scanners
  • Preserves Imageguide Inc.s Remote Center of
    Motion for the Robotic Arm

6
Design Criteria
  • Device must not be any longer than 30 cm
  • Needle must be able to travel 22 cm
  • Minimize radio-opaque materials
  • Must have redundant safety checks
  • Needle must be extracted rapidly with tissue
    sample
  • Needle must be fired fast enough to utilize
    inertia of body tissue
  • Biopsy should be completely automatic

7
Final Design
  • Needle contained inside piston/cylinder assembly
  • Needle injected by pneumatic power controlled by
    a solenoid valve
  • Needle retracted by a compression spring
  • Biopsy mechanism activated by impact of needle at
    the end of the cylinder
  • Gasket around piston to create seal
  • Most parts may be constructed from radiolucent
    material

8
Prototype Testing
  • Tested on a turkey breast
  • Needle entered easily sans deflection
  • Pressurized to 30 psi
  • Varying heights and angles of entry tested
    successfully
  • Tested on a pig to simulate skin resistance
  • Qualitative assessment to determine if distance
    away from skin affects deflection
  • Increasing the pressure improved entry

9
Solenoid Valve
  • Directs air flow greater than 50 psi
  • Will be implemented on the robotic arm so that it
    is close to the biopsy needle
  • Located outside plane of CT scan
  • Pneumatic devices are best in low humidity so
    nitrogen gas will be used

10
Needle Retraction
  • Pressure applied to the needle to propel it
    forward
  • Piston compresses the spring while it is injected
    into the body
  • Once biopsy taken, pressure is released and
    spring extends, retracting the needle
  • Retraction rate correlates with spring constant
    according to Hookes Law
  • Spring may affect CT image

11
Biopsy Mechanism
  • Cannula retracts by impact with end of cylinder
  • Flexor pegs lock into place once spring is
    compressed
  • Pinching the pegs extends the spring and ejects
    tumor cells

12
Future Work
  • Automate the retraction mechanism
  • LED and photodiode
  • Electrode sensor
  • Connect to a computer with LabView to integrate
    sensor signal
  • Apply radiolucent material to prototype
  • Attach to ImageGuides robotic arm

13
Methods to Automate System
  • LED and photodiode located opposite each other at
    the end of the cylinder
  • Opaque piston holding the needle breaks light
    path on the photodiode
  • This signals pressure to be released

14
Methods to Automate System
  • Spaced electrodes are placed at the bottom of the
    cylinder
  • Piston has a conducting ring that will pass
    current between electrodes
  • This signals pressure to be released

15
Box Diagram of Air Control
Robotic Arm
16
Circuit Controlling Solenoid Valve
17
References
  • Professor Frank Fronczak
  • Professor Willis Tompkins
  • Steve Polishinski
  • Professor John Webster
  • ECB Machine Shop
Write a Comment
User Comments (0)
About PowerShow.com