Title: Analysis NSG 2070
1Analysis NSG 2070
Conducted immunity test system
NSG2070e.ppt 28.056.01
2Main Benefits of the EM TEST Solution
3Other benefits of the EM TEST Solution
4Analysis NSG 2070
Conducted immunity test system
5NSG 2070 Technical Data
- DimensionHeight 171 mmWidth 449
mmDepth 461 mmWeight 16 kg - RF outputFrequency 100kHz - 250MHzPower 0 -
85WMax. voltage 10V
2
6NSG 2070 Technical Benefits
- Voltmeter display with 1, 3, 10V - 20 LED bar
display (low accuracy) - Display relative to the
voltage level- Suggestion not to use external
meters (this would discover the bad output
linearity) - Output voltage resolution 0.1V- 100 steps
0.1...10.0V - Frequency range up to 250 MHz
3
7Hardware Features
NSG 2070
CWS 500
- Interface RS 232
- Interface IEEE
- Fail inputs
- PRE AMP OUT / IN
- Modulation AM 0..100
- Modulation frequency 2Hz, 400Hz
4
8NSG 2070 Benefits Display
- Big letters for the frequency (double space)
- For each new entered parameter, the NSG 2070
calculates the duration of the test. - During a test procedure, the elapsed time and a
graphic bar with display show the test
progress. - Languages selectable German, French, Italien,
English BUT Software is only in English
5
9NSG 2070 Handling Disadvantages
- Setting with the rotary encoder is
troublesome.The rotary encoder locks depending
on the position, but has a long run-down after
releasing it.It is difficult to set a value
accurately. Sometimes after releasing the knob,
it rotates some steps more and changes to another
than the desired value.
6
10NSG 2070 Handling Disadvantages
- The setting of a value is only accepted after
having pressed the ENTER key.Pressing another
function key prior to having pressed ENTER will
cancel the previously set value and it returns to
the default value. - Quick Start can be selected only after the key
PAUSE has been activated.This function is not
documented in the manual.
7
11NSG 2070 Power up
- Warm-up time is 5 to 7 minutes and is indicated
in the display. - During the warm-up the NSG 2070 allows to execute
all tests. - After a short switch of OFF and ON, a new warm
-up period starts.
8
12NSG 2070 Linearity of the Frequency
- In the lower and upper frequency the voltage
level is out of the spec. - The tolerance acc. to the standard is 25
- The measured value shall be Uo/6 for each voltage
level. 10V gt 1.66V 3V gt 0.5V 1V gt
0.16V
10V 25
3V 25
1V 25
9
13NSG 2070 RF signal 100 overshoot
- Each change of the RF signal with a new frequency
starts with a 100 overshoot of the set output
value. - The EUT is overtested and is therefore
inacceptable.
40µs
200
desired testlevel
100
0
10
14NSG 2070 T2 fixed attenuator 4dB
- IEC 61000-4-6 reccommand a T2 attenuator for
reduce the mismatch from the power amplifier to
the network. - T2 attenuator Standard fixed gt 6dB (see clause
6.1 Test generator)
IEC 61000-4-6
Schaffner delivers a T2 attenuator with 4dB
11
15Application Current Injection Clamp
- No support from the NSG 2070 manual is available.
- Firmware is prepared to operate with a current
clamp. - Application with NSG 2070 does not work
- Measurement with current probe reduces the output
voltage after each frequency step. - There is no control that uses the feedback
current measuring - The measurement and displax of Imax Uo/ 150?
does not exist.
?
12
16NSG 2070 Verification
- Any verification with the delivered 4dB
attenuator will destroy the internal mearuring
equipment. - The manual recomends a 20dB attenuator which is
not part of the delivery and not even on the
options list. - A user who tries a verification without reading
carefully the manual has to replace the fuse.
13
17Application Sweep
- Frequency sweep
- Voltage sweep
- Dwell time sweep
- Recovery time in sweep
14
18NSG 2070 Standard Tests
- There is no function key for a complete Standard
Test procedure with - Frequency 0.15...80MHz
- VoltageLevesl 1V, 3V, 10V
- Modulation 80AM
- Procedure for a Standard Test with NSG 2070
- Select VOLTAGE in the Voltage Menu
- Select LOAD for setting up frequency and
calibration (CDN or Clamp) - Before running the test, check carefully all
parameters - - It is possible that preset parameters will
be overwritten. - Standard Tests at different levels in one cycle
are not possible
Settings correct ? Check before starting a
test!
15
19NSG 2070 Pulse Modulation
- Pulse modulation is a separate routine with a
fixed PM frequency and voltage.Frequency,
repetition rate and duty cycle are the selectable
parameters. - For pulse modulation the repetition rate ranges
from 10Hz...10kHz - Pulse modulation 1Hz is required in EN 50130-4
(Alarm system)
EN 50130-4 1995 for Alarm Systems requires a 1
Hz pulse modulation
16
20NSG 2070 Calibration
- Internal Memory for - 20 CDN - 2 EM
Clamps - 2 Current clamps, 1 Monitor - Only by using a PC, it is possible to see the
CDNs name! - During the calibration, the NSG 2070 shows the
actual frequency in steps.
Too many CDN calibration memories
overview
17
21Calibration Disadvantages
- No access to the Calibration data
- It is not possible to check the calibration data
- No calibration curve is available
- Calibration with an external meter is not
possible - Limited to 900 measurements / decade
- Calibration possible with reduced frequency
range - A test exceeding the calibrated range does not
work later.
calibrated frequency range
100kHz
250MHz
test frequencies
18
22Software
- Test overview parameterNSG 2070 doesnt show all
the parameters in one window - Frequency setting
- used CDN
only visible when marked
19
23Software
- Run-time Control
- The time must be set otherwise the test stops
unexpected. - SW features in CWS 500 available
- Fail event with different fail handling
- Remote control for external measuring instrument
20
24Software Calibration
- Calibration with internal power meter
- Calibration with external power meter
- Up and download of calibration files
- Printout of calibration files
21
25Software Comment Input
- NSG 2070
- One comment box for the complete test
- The user has to note frequency and voltage
himself. - CWS 500
- For each interruption a separate comment line is
offered - Frequency and voltage are automatically registered
Number of events 3 BREAK U 1.0 V f 0.353
MHz Dwell time 0.300 s Occured after 0.322
s Description this is a break, we
continue with the next frequency FAIL2 U
1.0 V f 7.467 MHz Dwell time 0.300 s Occured
after 0.207 s Description This is a
FAIL2 with comment, see menue Config FAIL2 STOP
U 1.0 V f 80.000 MHz Dwell time 0.300 s
Occured after 0.547 s Description This
is the end.
22
26Software Report
- Advantages CWS 500 ICD
- Report header with customers own logo
- Test parameter with graphic display
- Event list with a comment for each event
- Calibration data file of CDN
- Data report of external measuring instrument