Title: LCLS Undulator Support/Mover System Testing Status
1LCLS Undulator Support/Mover System Testing
Status
2Support/Mover System Testing
History
- We Assembled a Breadboard Undulator
Support/Mover System Using Existing Parts at the
Early Stages of the Design Process - To Get an Early Start on Motion Testing, we
Specified, Procured and Installed a Very
Accurate Measurement System for the Breadboard
Support/Mover System - A Labview-Based Data Acquisition and Control
System was Developed In-House for Command and
Control of all Motors with Feedback and
Monitoring from the Measurement System - The Breadboard System was Used to Learn How
to Control and Characterize the Performance of
the Motion System - Results from Numerous Tests were Used as
Feedback for Refinement of the Design Process.
Problem Areas have been Identified and Solutions
have Been Implemented in the Actual
Support/Mover System Design - Measurement Results are Very Encouraging with
Overall Excellent Repeatability
3Support/Mover System Testing
4Support/Mover System Testing
Our Present Breadboard Lacks Many Features
that the Prototype will Include
- Does not have Expanded Wedge on Upstream
Double CAM Mover - Aluminum Frame instead of Steel Frame
- Bearing Surfaces are not Hardened or Polished
- 1001 Gearbox instead of 2501 Gearbox
- Less Precise Rotary Potentiometers
- Only Four Linear Potentiometers
- Lack a Precision Temperature-Controlled
Environment
5Support/Mover System Testing
Salient Support/Mover System Physics Requirements
- Quadrupole Motion Positioning Repeatability 7
µm - Short-Term (1 h) BPM and Quad Stability 2 µm
- Long-Term (24 h) BPM and Quad Stability 5 µm
- Horiz. Segment Pos. Repeatability in Roll-Away
Cycle 10 µm - Vert. Segment Pos. Repeatability in Roll-Away
Cycle 5 µm - Quad Transverse Position Change in Roll-Out
Condition 25 µm - Quad Position Reproducibility after Roll-Away
Cycle 2 µm - BPM Transverse Position Change in Roll-Out
Condition 25 µm - BPM Position Reproducibility after Roll-Away
Cycle 2 µm
6Support/Mover System Testing
System Measurement Devices
- Five Keyence Laser Displacement Sensors (0.05
micron repeatability). Two at each end
measuring X and Y Displacement at the Quadrupole
Position, and one located mid-plane of the
Undulator on one side to measure Roll. - Four Linear Potentiometers (1-2 micron
repeatability) measuring X and Y Displacements
on one side of the Girder located near the CAM
Movers.
7Support/Mover System Testing
8Support/Mover System Testing
CAM Mover System Degrees of Freedom
- X-Translation (Inboard or Outboard Translation)
- Y- Translation (Up or Down Translation)
- Pitch Motion (Upstream or Downstream Pitch)
- Roll Motion (Inboard or Outboard Roll)
- Yaw Motion (From Either End or Both Ends)
- Notes
- The System can Move with Five Degrees of Freedom
using any Combination of these Motions. - The Z-Dimension is Restricted but not
Constrained via the Recessed Flat on the Single
CAM Mover.
9Support/Mover System Testing
CAM Mover Calibration Procedure
- Rotate one CAM through a Full Revolution
- Measure Displacement as a function of Angular
Position using either the Keyence Sensors or the
Linear Potentiometers - Plot the Displacement as a function of Angular
Position - Curve Fit the Data to Locate the Zero Position
- Repeat this Process for each of the Five CAM
Movers
Problems with this Approach
- Rotating one CAM introduces Pitch, Roll, and Yaw
into the System - These Motions add Stress into the Girder causing
Distortion - These Distortions become part of the Calibration
- The Calibration is not Decoupled from the Girder
- The existing Rotary Potentiometers have 1 of
Backlash adding to - Calibration Uncertainty
10Support/Mover System Testing
11Case 1 X-Translation /- 1.5 mm Cyclic Motion
Repeatability
Location Laser Displacement Sensors Linear Potentiometers
Upstream X 2.5 µm 1.5 µm
Upstream Y 1.0 µm 4.0 µm
Downstream X 2.0 µm 2.5 µm
Downstream Y 0.5 µm 4.0 µm
Target Repeatability lt /- 7.0 µm
12Support/Mover System Testing
Infrared Image Prior to Motor Operation,
Downstream End Double CAM Mover with Pedestal
Base
13Support/Mover System Testing
Infrared Image Prior to Motor Operation,
Downstream End Double CAM Mover with Pedestal
Base
14Support/Mover System Testing
Infrared Image After Motor Operation, Downstream
End Double CAM Mover with Pedestal Base, Motor on
for Approximately 2 Hours
15Case 2 Y-Translation /- 1.5 mm Cyclic Motion
Repeatability
Location Laser Displacement Sensors Linear Potentiometers
Upstream X 1.0 µm 2.5 µm
Upstream Y 2.5 µm 1.0 µm
Downstream X 0.5 µm 4.0 µm
Downstream Y 1.5 µm 1.0 µm
Target Repeatability lt /- 7.0 µm
16Case 3 Roll /- 1.5 mm Cyclic Motion
Repeatability
Location Laser Displacement Sensors Linear Potentiometers
Upstream X 1.0 µm 2.0 µm
Upstream Y 0.5 µm 0.5 µm
Downstream X 0.5 µm 4.0 µm
Downstream Y 0.5 µm 1.5 µm
Target Repeatability lt /- 7.0 µm
17Case 4 Upstream Pitch /- 1.5 mm Cyclic Motion
Repeatability
Location Laser Displacement Sensors Linear Potentiometers
Upstream X 1.0 µm 2.0 µm
Upstream Y 1.5 µm 1.0 µm
Downstream X 0.5 µm 2.0 µm
Downstream Y 0.5 µm 0.5 µm
Target Repeatability lt /- 7.0 µm
18Case 5 Both Ends Yaw 1.5 mm in Opposite
Directions
Location Laser Displacement Sensors Linear Potentiometers
Upstream X 15.0 µm 2.5 µm
Upstream Y 5.0 µm 2.5 µm
Downstream X 25.0 µm 2.5 µm
Downstream Y 2.5 µm 5.0 µm
Target Repeatability lt /- 7.0 µm
Note
Yaw Motion is Incidental and is Not Required for
Actual Operation.
19Support/Mover System Testing
CAM Mover System Observations
- All Motions that allow the CAM Bearings to Freely
Rotate (X-Translation, Y-Translation, and Roll)
are Repeatable to better than /- 2.5 microns at
the Quadrupole Position (Target Repeatability lt
/- 7.0 microns). - Though Pitch Motion is Overconstrained the
Bearing on the Single CAM Mover only needs to
Slip a few microns Across the Recessed Flat and
therefore the Motion still Shows Excellent
Repeatability. - Yaw Motion is very Overconstrained and therefore
the Bearing on the Single CAM Mover Must Slip
Hundreds of microns Across the Recessed Flat.
This Slippage, the Process of Overcoming Static
Friction, Distorts the Girder Causing Large
Repeatability Errors.
20Support/Mover System Testing
CAM Mover System Observations
- Some Repeatability Error is Added to the Linear
Potentiometers - from Sliding Across the Contact Surfaces causing
Bending of the - Sensor Arms.
- The Ball Bearing Tips of the Linear
- Potentiometers Wore Flat after Many
- Cycles from Sliding Across the Contact
- Surfaces.
- From a Cold Start, Heat from the Motors can Cause
Tens of Microns of Change from the Original Start
Position.
21Undulator Roll-Away Cycle Testing
Undulator Segment Roll-Away Cycle /- 10 mm
Cyclic Translation
Note Translation Stage Motions were Driven
Purely by Motor Steps with no Feedback from the
Linear Encoders. Many Changes will be Made to
these Stages for the Actual Prototype.
22Implemented Changes for the Actual Prototype
- Improved Translation Stage Design for Roll-Away
Cycle
- Stages will be Preloaded to Decrease Play in the
Slides - Gearboxes Replaced with GAM Angular Gearboxes
with Backlash Less than 15 arcsec. (Compared to
40 arcsec. on Existing Stages) - Table Length Increased from 6 to 8 and Ball
Bearing Separation is Increased Accordingly - Size of all Mounting Holes have been Increased
for Added Stiffness
23Implemented Changes for the Actual Prototype
- Steel Girder Replaces the Aluminum Girder.
24Implemented Changes for the Actual Prototype
- Expanded Wedge on the Upstream Double CAM Mover
for - Increased Stability.
25Implemented Changes for the Actual Prototype
- Inverted Wedge for the Single CAM Mover to
Constrain Z-Dimension - Movement.
26Implemented Changes for the Actual Prototype
- Two Side-by-Side Bearings for the Single CAM
Mover to Eliminate Slip across the Bearing
Surface.
27Implemented Changes for the Actual Prototype
- Spherical Bearings for Both of the Double CAM
Movers to allow - Rotational Play on these Bearings. The Single
CAM Mover will - still use a Roller Bearing. High Precision
Bearings will Replace the - Existing Standard Bearings.
Note Changes to the Bearings on the Double CAM
Movers, the Inverted Wedge Bearing Surface,
and the Side-by-Side Bearings on the Single CAM
Mover will Eliminate Bearing Slip and will
Allow the Girder to Pivot about the Single CAM
Mover Contact Surface.
28Implemented Changes for the Actual Prototype
- The Addition of Four more Linear Potentiometers
to Provide X and Y Displacement Information at
the Four Corners of the Girder near the CAM
Movers.
Note With Eight Linear Potentiometers Defining
the Girder Position, the Effects of Girder
Distortion can be Decoupled from the CAM Mover
Calibrations allowing Precise CAM Mover
Calibration.
29Implemented Changes for the Actual Prototype
- Replace the 1001 Gearboxes with 2501 Gearboxes.
This will Eliminate the need for Brakes on the
Motors, add Better Resolution to the System, and
Simplify Control System. - Replace existing Rotary Potentiometers (1
Backlash) with more Precise Rotary Potentiometers
with better than 0.1 Backlash. This will
Improve CAM Mover Calibration. - All Bearing Surfaces shall be Hardened and Highly
Polished. This should add to Long-Term System
Stability and Repeatability.
30Support/Mover System Testing
Conclusions
- The LCLS Undulator Support/Mover Tests on our
Breadboard are Very Encouraging with Overall
Excellent Repeatability. - We have Identified Problem Areas and have
Implemented Solutions that will Improve
Performance. - Testing To-Date has Provided Valuable Information
on System Dynamics, Environmental Influences,
Calibration Procedures, and Feedback and Control
Algorithms. - Similar, but more in-depth, Tests shall be
Performed on the Actual Prototype once the System
has been Assembled.