Title: MR Airway Pressure Device
1MR Airway Pressure Device
Group Members Laura Sheehan Kevin Johnson Jon
Cappel Noelle Simatic Client Dr. Victor
Haughton Advisor Prof. Mitch Tyler
2Chiari I Malformations
http//www.chiariclinic.org
3MR Background
Tuner
90.7
104.1
B
4Device
5Device
Normal Breathing(Holes Open)
Valsalva(Holes Closed)
Vacuum
Compressed Air
6Validation
- Device Precision
- Does the MR environment effect precision?
- Does scanning effect precision?
- MR Compatibility
- Does our device cause artifacts?
- Does our device degrade image quality?
- Protocol Development
- Whats the best way to perform this?
- Human/Animal Subjects Testing
- Final evaluation!!
7Device Precision Test
Trigger
- Measured 240 Periods of Square flow waveform
- Outside room ? With Setup in Side Room
- Inside Bore ? Normal Setup
- During Scan? Normal Setup during a scan
8Device Precision Results
9Device Precision Results
- SNR Measures
- Outside room 40.99
- Bore 28.3
- During Scan 26.8
- All less than 5 error
- Correlation Analysis
- Plotted Values against each other
- Pearson Coef (R0.998)
- Slope of 1.01
10MR Compatibility Test
Image 2
Image 1
Noise
Without Device
Artifact
With Device
11MR Compatibility Preliminary Results
Baseline
With Device
No obvious artifact but This noise level is
unacceptable !!!!!
12MR Compatibility Future Work
RF
DAQ
Device
Shielding (copper)
Control Room
MR Suite
13Future Work
- MR Compatibility
- Additional Shielding
- Anesthetized Patient Prototype
- Protocol Development
- Human/Animal Subjects Testing
- IRB Approval
- Outreach Presentation
- Final Deliverables