Common BedSide Procedures in ICU - PowerPoint PPT Presentation

1 / 45
About This Presentation
Title:

Common BedSide Procedures in ICU

Description:

Pain is a function of the current delivered per unit of area. ... lowest pacing rate programmed is 60 beats per minute (bpm), a pacer set to sense ... – PowerPoint PPT presentation

Number of Views:3136
Avg rating:3.0/5.0
Slides: 46
Provided by: Wan56
Category:

less

Transcript and Presenter's Notes

Title: Common BedSide Procedures in ICU


1
Common Bed-Side Procedures in ICU
  • Wang, Tzong-Luen, MD, PhD, FESC, FACC
  • Shin-Kong Wu Ho-Su Memorial Hospital

2
Contents
  • www.emedicine.com
  • www.medscape.com
  • www.mdchoice.com
  • Swan-Ganz Catheter
  • Intra-Aortic Balloon Counter-pulsation
  • Transvenous Pacemaker
  • Airway Management / Capnography

3
Swan-Ganz Catheter
4
Intra-Aortic Balloon Counter-Pulsation (IABP)
5
Blood Pressure during IABP
  • balloon inflation, termed "diastolic augmented
    pressure"
  • balloon deflation during end-diastole producing a
    low pressure point termed "balloon assisted
    aortic end-diastolic pressure" (BAEDP)
  • patient's aortic end-diastolic pressure (without
    the IAB impact), called "unassisted aortic
    end-diastolic pressure," (UAEDP)
  • assisted systole, or the systolic pressure that
    is generated after a balloon inflation-deflation
    cycle and
  • systole without a preceding IAB-pumped beat,
    termed "unassisted systole."

6
Indication (1)
  • pump failure or cardiogenic shock as indicated by
    hemodynamic instability (BP lt 90 mm Hg, cardiac
    index lt 2, pulmonary capillary wedge pressure gt
    18 mm Hg, ST elevation) refractory to volume
    optimization and inotropic support
  • cardiac surgical patients at risk for hemodynamic
    decompensation
  • severe left-main disease
  • ischemic changes during hypotensive episodes
    (anesthesia)

7
Indication (2)
  • hemodynamic or ECG instability in catheterization
    laboratory (ie, "cath lab")
  • ventricular arrhythmia unresponsive to
    conventional treatment
  • cardiac failure due to CAD complications
    resistant to inotropes and diuresis
  • LV aneurysm
  • acute mitral insufficiency

8
Indication (3)
  • mechanical complications of AMI (eg, acute mitral
    regurgitation)
  • postinfarction ventricular-septal rupture
  • hemodynamic/ECG instability during early surgical
    phase pre-cardiopulmonary bypass
  • failure to wean from cardiopulmonary bypass
  • bridge to cardiac transplantation
  • stunned myocardium

9
Contraindication (1)
  • Severe aortic valvular insufficiency
  • Aortic dissection
  • Severe peripheral vascular disease
  • Irreversible Brain Damage

10
Effects of IABP
11
Myocardial Oxygen Consumption
12
Factors decreasing Augmentation
  • Positioning
  • The proximal tip of the IAB should be
    positioned just below the bifurcation of the left
    subclavian artery (Fig.1). If the IAB is
    positioned too low, diastolic augmentation will
    be reduced as inflation momemtum is decreased.
  • Volume
  • Diastolic augmentation is maximized when
    stroke volume is equal to balloon volume. If
    stroke volume is less than 25 ml, little
    diastolic augmentation can be expected. On the
    contrary, a stroke volume greater than 50 ml is
    beyond the displacement capabilities of the
    inflating IAB. Augmentation will, therefore, also
    be decreased.

13
Factors decreasing Augmentation
  • Systemic Vascular Resistance
  • As systemic vascular resistance increases,
    diastolic augmentation may decrease because of
    the associated decrease in compliance.
    Conversely, an extremely low systemic vascular
    resistance also will produce poor augmentation.
  • Timing
  • Early inflation or late deflation can
    decrease the amount of time that helium gas
    inflates the IAB. Thus, a shortened deflation
    imposed by erroneous balloon timing will
    understandably decrease the magnitude of
    inflation.

14
Complication
15
Transcutaneous Pacemaker
16
Application
  • Electrodes/pads and monitor leads, if necessary,
    are placed on the patient.
  • About 2-3 cm of space should be left if separate
    defibrillation pads are required, and the second
    pad should be placed posteriorly, just below the
    left scapula.
  • The desired heart rate is chosen and the current
    is set to zero milliamperes (mA).
  • The TCP is then turned on and the current is
    increased as tolerated until capture is achieved.

17
Pulse Duration
  • Pulse duration is the time of impulse
    stimulation.
  • Early TCPs used short (1-2 msec) duration
    impulses. Such impulses resembled the action
    potential and preferentially stimulated skeletal
    muscle.
  • In contrast, cardiac muscle action potentials are
    much longer, requiring 20-40 msec to reach
    maximum.

18
Pulse Duration
  • Zoll found that increasing the duration from 1 to
    4 msec resulted in a 3-fold reduction in
    threshold (the current required for stimulation).
  • Increasing the current from 4 to 40 msec further
    halves the threshold.
  • Longer durations produced no further advantage.
    Current TCPs deliver 40 (Zoll) or 20 (all others)
    msec pulses.

19
Current
  • External pacing in dogs requires 30-100 times
    greater than internal transvenous pacing.
  • Human studies have shown that the average current
    necessary for external pacing is about 65-100 mA
    in unstable bradycardias and about 50-70 mA in
    hemodynamically stable patients and volunteers.
  • At this current, more than 90 of patients
    tolerated pacing for 15 or more minutes.
  • Higher amounts are needed to stimulate the atria.

20
Current
  • Using a longer pulse duration and larger
    electrodes permits patients to tolerate higher
    applied current.
  • One hundred milliamperes of current applied over
    an average (50-ohm resistance) chest for 20 msec
    will deliver 0.1 Joules. This is well below the
    1-2 Joules required to cause an uncomfortable
    tingling sensation in the skin.
  • The force of skeletal muscle contraction, not the
    electric current, determines TCP discomfort.
    Current TCPs are capable of delivering up to
    140-200 mA tolerably.

21
Electrode
  • Pain is a function of the current delivered per
    unit of area. Pain sensation is minimized by
    electrodes with a surface area of at least 5 cm2.
  • The amount of pain for a current of a given
    strength reaches a plateau once the electrode
    surface area exceeds 10 cm2.
  • Most commercially available electrodes are 80-100
    cm2. TCPs generally perform best with their own
    pads, but different combinations may be helpful.

22
Monitor
  • Determination of electrical capture and pulse
    generation can be difficult when skeletal muscle
    is stimulated. In the 1950s, Zoll developed a TCP
    with an ECG monitor that allowed for
    identification of electrical capture. Blanking
    protection, currently only available in Zoll
    models, changes high output pacing stimulus to a
    smaller ECG waveform, preventing overdriving of
    the ECG.
  • If blanking protection is not present, a second
    monitor or clinical palpation of the pulse is
    needed to determine capture.

23
Mode
  • In the fixed rate (asynchronous) mode, the TCP
    delivers an electrical stimulus at preset
    intervals, independent of intrinsic cardiac
    activity. In theory, this could induce
    arrhythmias if stimulation occurs during the
    vulnerable period. Early models only had
    fixed-rate capabilities.
  • Most current models have fixed rate and
    synchronous pacing. Synchronous pacing is a
    demand mode in which the pacer fires only when no
    complex is sensed for a predetermined amount of
    time. Pacing generally should be started in the
    synchronous mode.

24
Special Considerations
  • Minimizing discomfort
  • Skeletal muscle contraction can be uncomfortable
    and is often the limiting factor in TCP use.
    Placing electrodes over areas of least skeletal
    muscle can minimize discomfort. Placement is
    generally best in the midline chest and just
    below the left scapula. The physician also should
    use the lowest effective current. Sedation should
    be considered if these measures are inadequate.

25
Special Considerations
  • CPR
  • CPR can be performed with the TEP pads in place.
    The low Joules delivered and the insulation of
    the flexible TEP pads result in no electrical
    hazard to the person performing CPR. However,
    turning the unit off during CPR is advisable.
  • In TCPs without an intrinsic defibrillator,
    separate leads need to be applied. The external
    pacemaker should be turned off or to monitoring
    mode when defibrillating or cardioverting a
    patient. Defibrillator paddles should be placed
    at least 2-3 cm away from TEP stimulation pads to
    prevent arcing of current. Pacing pads should be
    placed in the anterior/posterior position.

26
Transvenous Pacemaker
27
Pacemaker code used to describe various pacing
modes
28
Indications
Symptomatic Bradycardia
  • Sick sinus syndrome
  • Symptomatic sinus bradycardia
  • Tachy-brady syndrome
  • Atrial fibrillation with a slow ventricular
    response
  • Complete atrioventricular block (third-degree
    block)
  • Chronotropic incompetence (inability to increase
    the heart rate to match a level of exercise)
  • Long QT syndrome Relative indications include the
    following
  • Cardiomyopathy (hypertrophic or dilated)
  • Severe refractory neurocardiogenic syncope
  • Paroxysmal atrial fibrillation

29
Procedure
  • ???? --
  • a. CVP ?????? -- ?? alcohol sponge, better-iodine
    sponge, OP site, 0.9 NS, 10 ? 20 cc ??, ??
  • b. Electrode catheter set ?? -- ??????,
    dilator,introducer, puncture needle, guiding wire
  • c. Pacemaker generator, ?? 9 V ????
  • d. ?????
  • e. EKG monitor, ??? ( defibrillator )

30
Procedure
  • ???? -- ?????????
  • 1. ??
  • 2. ? EKG monitor ????? spike, ?? leads ??????,
    ??????, ???????
  • 3. ? Chest PA ? check catheter tip ????
  • 4. ????, ????? pacemaker generator ?? 10 ????,
    ?????

31
Procedure
  • ?? (rate) ???
  • 1. ???? -- ????????????
  • 2. ???? -- ????????????
  • 3. ???? -- ????? pacing ????????? ( SVT
    ),???????? overdrive ?????? pacing

32
Procedure
  • ???? (output current)
  • 1. ??? 1 ??? ( 1mA ) ??, ? 0.5 mA ???, ???????(
    threshold potential ) ?????????? ( capture ),
    ????????????? 2 ? 5 ??
  • 2. ????? electrode ???, ?????????? 20 mA,
    ????????????????

33
Selection
  • ??????
  • 1. ?????? -- ?? synchronized demand, ????
  • 2. ????????? ( AV sequential pacemaker )
    --?????????????, double-chamber
    pacing,double-chamber sensing and double-mode
    response (atrial inhibited, ventricular paced ) ?
    DDD ??,?????????????
  • 3. ??????? ( transesophageal cardiac pacing )--
    ????????, ???????? pacing,?? SVT ? overdrive
    suppression ???????reentrant dysrhythmia,?? AF,Af
    ????????????????
  • 4. ????? intracardiac pacing catheter ?????????

34
Complications
  • 1. ?????, ?? ( tamponade ), ?????????
  • 2. ??
  • 3. ???????????, ????, ???????( desensitization ),
    generator ??????????????
  • 4. pacing ?????????????????????????????
  • 5. ????????? ( hiccup )?
  • 6. ? ICU ?????????????????????? capture ???

35
Comlications
  • Pacing Failure
  • Failure to Output
  • Failure to Capture
  • Pseudomalfunction
  • Sensing Failure
  • Oversensing
  • Undersensing
  • Operative Failure

36
Failure to Output
  • Failure to output occurs when no pacing spike is
    present despite an indication to pace.
  • This may be due to battery failure, lead
    fracture, a break in lead insulation, oversensing
    (inhibiting pacer output), poor lead connection
    at the takeoff from the pacer, and "cross-talk"
    (ie, a phenomenon seen when atrial output is
    sensed by a ventricular lead in a dual-chamber
    pacer).

37
Failure to Capture
  • Failure to capture occurs when a pacing spike is
    not followed by either an atrial or a ventricular
    complex.
  • This may be due to lead fracture, lead
    dislodgement, a break in lead insulation, an
    elevated pacing threshold, myocardial infarction
    at the lead tip, certain drugs (eg, flecainide),
    metabolic abnormalities (eg, hyperkalemia,
    acidosis, alkalosis), cardiac perforation, poor
    lead connection at the takeoff from the
    generator, and improper amplitude or pulse width
    settings.

38
Failure to Capture
  • Management of pacer capture complications is the
    same as for output complications, with extra
    consideration given to treating metabolic
    abnormalities and potential myocardial
    infarction.

39
Oversensing
  • Oversensing occurs when a pacer incorrectly
    senses electrical activity and is inhibited from
    correctly pacing.
  • This may be due to muscular activity,
    particularly oversensing of the diaphragm or
    pectoralis muscles, electromagnetic interference,
    or lead insulation breakage.

40
Undersensing
  • Undersensing occurs when a pacer incorrectly
    misses intrinsic depolarization and paces despite
    intrinsic activity.
  • This may be due to poor lead positioning, lead
    dislodgment, magnet application, low battery
    states, or myocardial infarction. Management is
    similar to that for other types of failures.

41
Undersensing
  • A final category of pacer failures is termed
    operative.
  • This includes malfunction due to mechanical
    factors, such as pneumothorax, pericarditis,
    infection, skin erosion, hematoma, lead
    dislodgment, and venous thrombosis. Treatment
    depends on the etiology.

42
Undersensing
  • Pneumothoraces may require chest thoracostomy. If
    infection is present, Staphylococcus aureus is
    common in postoperative patients while
    Staphylococcus epidermidis is common following
    the postoperative period.
  • Erosion of the pacer through the skin, while
    rare, requires pacer replacement and systemic
    antibiotics.
  • Hematomas may require drainage.
  • Lead dislodgment usually occurs within 2 days
    following implantation of a permanent pacer and
    may be seen on chest radiography. If the lead is
    floating freely in the ventricle, malignant
    arrhythmias may develop.
  • Thrombosis is rare and usually presents as
    unilateral arm edema. Treatment includes arm
    elevation and anticoagulation.

43
Pseudomalfunction
  • Pseudomalfunction is a type of output failure
    characterized by a phenomenon termed hysteresis.
    This occurs when a pacer is set to sense below
    the lower pacing rate limit.
  • For example, if the lowest pacing rate programmed
    is 60 beats per minute (bpm), a pacer set to
    sense down to an intrinsic rate of 50 bpm, a
    hysteresis rate, begins to pace at 60 bpm when
    the patients intrinsic rate falls below 50 bpm.
    It continues to pace at the lower rate limit of
    the pacemaker, in this example 60 bpm, until it
    again senses intrinsic activity. This sensed
    event inhibits pacing, and the pacemaker again
    permits the intrinsic rate to go down to 50 bpm
    before pacing at 60 bpm.

44
Pseudomalfunction
  • Management of pacer output complications includes
    medications to increase the intrinsic heart rate
    and placement of a temporary pacer. A chest
    radiograph is warranted to check pacer leads,
    with close scrutiny to evaluate for possible lead
    fracture, which occurs most commonly at the
    clavicle/first rib location. The patients pacer
    identification card should be obtained and
    his/her electrophysiologist/cardiologist
    consulted.

45
Thanks for Attending
Write a Comment
User Comments (0)
About PowerShow.com