Indigenous software for quantification of
autonomic tone
All India Institute of Medical SciencesNew
Delhi  110029   India
  • Prof. K K Deepak     MD, PhD,
    MNAMS  Professor of Physiology
    kkdeepak_at_gmail.com

2
Autonomic Control of Function of Heart
PNS
High freq.
Aff. Nvs.
Autonomic controller
SNS
Low freq.
3
Autonomic Control of Function of Heart
PNS
Respiration Blood Pressure Cardiac
Output Renin-Angioten. Thermoregul. Others inputs
High freq.
Aff. Nvs.
Autonomic controller
SNS
Low freq.
4
Applied aspect of Cardiac tone (HRV)
  • Estimation of vagal tone
  • Cardiovascular pathophysiology
  • Myocardial Infarction
  • Hypertension
  • Diabetic neuropathy
  • Neurological disorders
  • Migraine
  • MSA
  • Assessment of neonatal ANS

5
Applied aspect of vascular tone (BPV)
  • Hypertension
  • Cardiac Transplant Rejection
  • End organ failure
  • Hypertensive Pregnancy Disorders
  • Risk factor for stroke / SCD / SIDS
  • Prognosis in CHF

6
Autonomic Tone (HRV) in health and disease
Healthy
Healthy
Heart Attack
Heart transplant
7
Indigenously Developed HRV Software for the
Quantification of Autonomic Tone
Objectives
Software development for measurement of
Heart Rate Variability (HRV_Soft) and Blood
Pressure Variability (BPV_Soft)
Online Data Acquisition
Offline Data Analysis
8
Development Acquisition Tools Used
  • National Instruments USB-6008 DAQ
  • National Instruments (NI) Labview FDS 8.0
  • NI Measurement and Automation Libraries
  • Colin BP-508 BP Monitor
  • Finapres Ohmeda 2300 BP Monitor
  • DAS port (Nevrokard)
  • Analog ECG from any source

9
Signal Source Flow Diagram
HRV_Soft
DAQ Module (USB Based)
ECG
Patient
BPV_Soft
Blood Pressure
Analysis Storage
10
Main Panel of Indigenously Developed Software
(HRV_Soft)
11
Heart Rate Variability
12
Validation of our HRV_Soft studies carried out
  • We acquired the ECG and BP Signal from the
    Nevrokard DAS port and FINAPRES simultaneously on
    our Software and Nevrokard Software.
  • We compared the Time Domain, Frequency Domain
    Non Linear Parameters
  • The results were found to be Statistically
    comparable

13
Testing Model 1
Single patient source (ECG)
Analog Signal Acquisition
Analog Signal Acquisition
Digitization
Digitization
Nevrokard
HRV_Soft
Analysis
Analysis
14
Time Domain Parameters
N22
15
FREQUENCY DOMAIN PARAMETERS
N22
16
Testing Model
Analog Signal Acquisition
Testing Model 2
Patient ECG
Analog Signal Acquisition ( our software)
Digitization Peak detection by our software
Analysis
Analysis
BPV_Soft
HRV_Soft
Nevrokard
17
Pearson Correlation Test of HRV parameters
obtained from 45 subjects as measured from HRV
soft and Neurokard
18
One more feature has been developed which can
trace the path of Non Linear behavior. This
actually analyses the phases of development of
Poincare plot.
19
What is being delivered ?
20
Unique features of our software
  • 1.Complete report generation in printable format
    as shown in fig below. Also all the parameter of
    time frequency non-linear analysis are
    transferred to Excel sheet for maintaining
    Database
  • 2. Additional Parameters than commercially
    available software.
  • 3.Poincare tracking feature in poincare plot
    (Non-linear analysis) allows user to observe
    point-by-point movement of the plot
  • 4.Provision for deleting artifacts present in
    the ECG waveform before performing Peak
    detection.
  • 5.Robust fully automatic peak detection
    algorithm.
  • 6.USB/Serial port connectivity Easy to use
    Graphical User Interface.
  • 7.Flexible Cost Effective.

21
Compatibility with existing systems
Any device that can pick ECG signals
HRV-Soft BPV-Soft
General compatibility
22