Lecture 2 ECG Instrumentation - PowerPoint PPT Presentation

1 / 44
About This Presentation
Title:

Lecture 2 ECG Instrumentation

Description:

To record the ECG we need a transducer capable of converting the ionic ... For simple ECG recording, however, a three 'lead' combination is possible in ... – PowerPoint PPT presentation

Number of Views:1128
Avg rating:3.0/5.0
Slides: 45
Provided by: Vai894
Category:

less

Transcript and Presenter's Notes

Title: Lecture 2 ECG Instrumentation


1
???? Lecture 2ECG Instrumentation
  • ???????? ???
  • ??? ????,IEEE Fellow

2
The story so far
  • The heart pumps blood around the body.
  • It has four chambers which contact in a carefully
    controlled order (two pairs of contractions) to
    achieve this effect.

3
The story so far
  • The heart pumps blood around the body using a
    carefully controlled order of contractions of its
    four chambers.
  • The contraction of a muscle cell is a result of a
    depolarisation and repolarisation cycle called an
    action potential during which the potential
    difference between the inside and the outside of
    the cell changes.

4
The story so far
  • The heart pumps blood around the body using a
    carefully controlled order of contractions of its
    four chambers.
  • The contraction of a muscle cell is a result of a
    depolarisation and repolarisation cycle called an
    action potential during which the potential
    difference between the inside and the outside of
    the cell changes.
  • Considering all of the cardiac cells together we
    can view the heart as an electrical generator
    which drives current into a passive resistive
    medium, the thorax.
  • By taking voltage differences at different points
    on the thorax the electrical activity of the
    heart can be observed.

5
The Electrocardiocgram (ECG)
  • Here is an example of an ECG signal

6
(No Transcript)
7
Recording the ECG
  • To record the ECG we need a transducer capable of
    converting the ionic potentials generated within
    the body into electronic potentials
  • Such a transducer is a pair of electrodes
  • Polarisable (which behave as capacitors)
  • Non-polarisable (behave as resistors)
  • Common electrodes lie between these two extremes

8
Recording the ECG
  • To record the ECG we need a transducer capable of
    converting the ionic potentials generated within
    the body into electronic potentials
  • Such a transducer is a pair of electrodes
  • Polarisable (which behave as capacitors)
  • Non-polarisable (behave as resistors)
  • Common electrodes lie between these two extremes
  • The electrode most commonly used for ECG signals,
    the silver-silver chloride electrode is closer to
    a non-polarisable electrode.

9
Electrode placement
  • There exists a convention prescribing electrode
    placement.
  • "12 lead" ECG placement is a well developed tools
    for recording 12 different ECG traces from an
    individual
  • For simple ECG recording, however, a three "lead
    combination is possible in which electrodes are
    placed on the right arm (RA), the left arm (LA)
    and the left leg (LL)

10
Electrode placement
  • In this context, "lead" means a pair of
    electrodes The three electrodes above result in
    three possible differences
  • (Potential at LA) - (Potential at RA)
  • (Potential at LL) - (Potential at RA)
  • (Potential at LL) - (Potential at LA)

11
(No Transcript)
12
Silver-silver chloride electrode
  • Electrodes are usually metal disks and a salt of
    that metal.
  • A paste is applied between the electrode and the
    skin.
  • This results in a local solution of the metal in
    the paste at the electrode-skin interface.
  • Ionic equilibrium takes place when the electrical
    field is balanced by the concentration gradient
    and a layer of Ag ions is adjacent to a layer of
    Cl- ions.
  • This gives a potential drop E called the
    half-cell potential (normally 0.8 V for an
    Ag-AgCl electrode)

13
Silver-silver chloride electrode
  • This double layer of charges will also have a
    capacitative effect
  • Since the Ag-AgCl electrode is primarily
    non-polarisable there is a large resistive
    effect.
  • This gives a simple model for the electrode.

14
Silver-silver chloride electrode
  • This double layer of charges will also have a
    capacitative effect
  • Since the Ag-AgCl electrode is primarily
    non-polarisable there is a large resistive
    effect.
  • This gives a simple model for the electrode.
  • However, the impedance is not infinite at d.c.
    and so a resistor must be added in parallel with
    the capacitor.

15
(No Transcript)
16
Movement artefact
  • Movement disturbs the physical positioning of the
    ions in the ionic equilibrium.
  • The half-cell potential will therefore
    momentarily change.
  • The potential difference between two electrodes
    will therefore vary.
  • This variation, which is unrelated to the
    underlying signal we wish to observe, is known as
    movement artefact
  • It can be a serious cause of interference in the
    measurement of ECG

17
(No Transcript)
18
Overall Equivalent circuit
  • Using the simple model we used earlier for the
    thorax we can build up an overall circuit for the
    heart, the body and the electrodes.
  • The resistors and capacitors may not be exactly
    equal.
  • E and Eshould be very similar.
  • Hence V should represent the actual different of
    ionic potential between the two points on the
    body where the electrodes have been placed.

19
(No Transcript)
20
ECG Amplifiers
  • The peak output voltage V of our equivalent
    circuit is around 1 mV.
  • Therefore, amplification is required if we are
    going to store or display this information in
    some way.
  • There are three problems which must be overcome
    which we will now consider

21
Electrical Field Interference(Problem 1)
  • To put it in hand-waving terms the human body is
    a good aerial and so the electrical signal at the
    electrodes is not pure, unadulterated, ECG
  • More specifically, capacitance between power
    lines and the system couples current into the
    patient, wires and machine.

22
Electrical Field Interference
  • This capacitance varies but it is of the order of
    50pF.
  • This corresponds to 64mW at 50Hz.
  • If the right leg is connected to the common
    ground of the amplifier with a contact impedance
    of 5k, the mains potential will appear as a 20mV
    noise input.

23
The solution
  • The key is to remember that the ECG is a
    differential signal.
  • The 50Hz noise, however, is common to all the
    electrodes.
  • It appears equally at the Right Arm and Left Arm
    terminals.
  • Rejection therefore depends on the use of a
    differential amplifier in the input stage of the
    ECG machine.
  • The amount of rejection depends on the ability of
    the amplifier to reject common-mode voltages.

24
Differential Amplifiers
  • There have been covered in the core course.
  • Lets have a look at the standard circuit.

25
Differential Amplifier
1
2
1
2
Q Why dont we use this amplifier for ECG
instrumentation?
26
Differential (Instrumentation) Amplifer
27
(No Transcript)
28
Differential Amplifiers
  • So, the standard circuit gives a gain of
  • ie a differential gain of

29
Differential Amplifiers
  • So, the standard circuit gives a gain of
  • ie a differential gain of
  • and a common-mode gain of

30
Differential Amplifiers
  • The overall common-mode rejection ratio is given
    by

31
Magnetic Induction(problem 2)
  • Current in magnetic fields induces voltage in the
    loop formed by the patient leads
  • Either
  • Lower the magnetic field strength (rather hard)
  • Minimise the coil area (eg twist the wires
    together)

32
Source Impedance Unbalance(problem 3)
  • If these impedances are not balanced (ie the
    same) then the common-mode voltage of the body
    will be higher at one input to the amplifier than
    the other.
  • Hence, a fraction of the common-mode voltage will
    be seen as a differential signal.
  • Therefore, make sure the electrodes are on
    correctly!

33
The signal
  • So, the signal at the input to the amplifier will
    have three components
  • The desired differential ECG signal
  • An unwanted common-mode signal
  • Unwanted common-mode signal appearing as a
    differential input

34
The signal
  • The output of the amplifier will therefore
    consist of three components
  • The desired output (ECG)
  • Unwanted common-mode signal because the
    common-mode rejection is not infinite
  • Unwanted common-mode signal due to source
    imbalance

35
A Modification
  • There is a better approach than setting the
    system ground to the common-mode voltage.
  • The common-mode voltage can actually be
    controlled using a Driven right-leg circuit.
  • A small current (lt1?A) is injected into the
    patient to equal the displacement currents
    flowing in the body.
  • The body sums these currents and the common-mode
    voltage is driven to a low value.

36
(No Transcript)
37
(No Transcript)
38
A Modification
  • There is a better approach than setting the
    system ground to the common-mode voltage.
  • The common-mode voltage can actually be
    controlled using a Driven right-leg circuit.
  • A small current (lt1?A) is injected into the
    patient to equal the displacement currents
    flowing in the body.
  • The body sums these currents and the common-mode
    voltage is driven to a low value.
  • Also improves patient safety.

39
Diagnostic use of the ECG
  • As has been seen, the ECG can provide diagnostic
    information to clinicians.
  • Ectopic beats originate somewhere other than the
    SA node and often have different shapes
    (morphologies).
  • Abnormal heart rates (arrhythmias) can be treated.

40
Diagnostic use of the ECG
  • Post heart attack (Myocardial Infarct) the ECG is
    highly informative.
  • Cardiac muscle damage (infarcts) generally
    correspond to loss of amplitude.
  • Insufficient blood supply to cardiac cells
    (Ischemic heart condition) changes the S-T level.

41
Acquiring ECG for Diagnosis
  • Two methods are in common use
  • Exercise Stress ECGs
  • Ambulatory monitoring

42
Acquiring ECG for Diagnosis
  • Two methods are in common use
  • Exercise Stress ECGs
  • Ambulatory monitoring

43
Acquiring ECG for Diagnosis
  • Ambulatory monitoring
  • ECG monitored for 24 hours.
  • Each beat analysed and either kept for records or
    ignored (because too normal!)
  • Results printed out
  • 1 page summary of conditions detected.
  • 24-hour summary detailing the heart rate and ST
    segment changes over the period of the test.
  • Findings pages with detailed information for each
    finding or symptom event.

44
Foetal Monitoring
  • It can be helpful to clinicians and midwives to
    have information about both the maternal and
    foetal ECG during childbirth
  • Although this may add stress, the behaviour of
    the foetal ECG (and heart rate) is known to give
    information about potential foetal distress
Write a Comment
User Comments (0)
About PowerShow.com