Title: Hematology 425 Quality Assurance
1Hematology 425 Quality Assurance
- Russ Morrison
- September 15 2006
2Why Quality Assurance?
- Monitor and weigh the balance between quality and
costs in health care - Satisfy stringent regulatory controls
- Participation in federal funding programs
- Evaluation of clinical outcomes
- Reduce risk to the organization
- Risk of suboptimal outcomes
- Risk of litigation
3Quality Control / Quality Assurance
- Both terms refer to the control of the testing
process to ensure that test results meet
requirements - Diagnostic utility
- Therapeutic validity
- Quality Assurance is overall, coordinated effort
to satisfy patients and care givers - Quality Control is activity related to monitoring
test systems and making corrections, as needed
4Quality Control Definitions
- Accuracy exactness, measured by periodic
proficiency testing challenges - Precision reproducibility, indicator of random
error, daily (or as required) quality control
replicates - Results that exhibit both precision and accuracy
is the goal - Avoid results that are precise, but not accurate
(being precisely wrong)
5Quality Control Definitions
- Delta Checks compare results from a prior sample
analysis with the result of the same analyte
using a current sample - significant variation is user defined
- Technologist investigates (ie MCV)
- Reference Interval/Normal Range
- often refers to the acceptable range of values
for the analyte in a healthy person, reference
range is more appropriate - Reference population used to establish reference
range
6Quality Control Definitions
- Reliability refers to the extent that a method
can maintain both accuracy and precision over a
period of time. Lab methods are chosen to have a
high degree of reliability. - Primary Standard is reference material of fixed
and known composition, traceable to national
performance characteristics - Secondary Standard is reference material of
composition traceable to a primary standard
7Statistics in Quality Measures
- Normal Distribution Curve
8Statistics in Quality Measures
- Coefficient of Variation (CV)
- is a measure of precision within a test system,
the smaller the CV value, the more precise the
analytic test system - CV is calculated as 100SD/mean
- It is expressed as percent
9Statistics in Quality Measures
- Diagnostic Sensitivity
- The proportion of patients with a disease who
will have positive test results - Defined by the ratio (a /(a b)) 100
- a is the number of true positive results and b is
the number of false negative results - THINK ABOUT THIS
10Statistics in Quality Measures
- Diagnostic Specificity
- The proportion of patients who are correctly
identified by the test as not having the disease - Defined by the ratio (c /(c d)) 100
- c is the number of false positive results and d
is the number of true negative results - Highly sensitive tests are used to screen for a
condition, highly specific tests are used to
confirm the screen.
11Types of Laboratory Error
- Systematic-errors within the test system or
method (incorrect calibration, broken equipment,
failure of the test process to perform accurately
or precisely, 2 types) - Constant systematic magnitude of error remains
constant throughout the range, also called a
constant bias - Proportional systematic magnitude of error
increases with the concentration of the substance
being tested
12Types of Laboratory Error
- Random errors that occur without predictability
or regularity
13Internal Quality Control
- Levey-Jennings Control Charts
14Internal Quality Control
- Normal Levey-Jennings Control Chart
- Results randomly distributed about the mean over
time - Distribution as expected for normal distribution
curve - 1 in 20 points may be between 2 and 3 SD
(caution), 1 in 100 above 3 SD (act)
15Internal Quality Control
- Levey-Jennings Control Chart
- Shift is a drift in values, may be sudden or
gradual - Trend is continuous movement of values in one
direction for 6 or more consecutive values - A trend may be thought of as a gradual shift
16Internal Quality Control
- Levey-Jennings Control Chart
- Shift is a drift in values, may be sudden or
gradual - Trend is continuous movement of values in one
direction for 6 or more consecutive values - A trend may be thought of as a gradual shift
17Westgard Rules
- 1 2s Rule one control value outside the 2s
limit - review - 1 3s Rule one value is outside the 3s limit,
random error, investigate - 2 2s Rule 2 consecutive values outside the same
2s limit, investigate - R 4s Rule 2 consecutive values are more than 4s
apart involving both control materials, 1 2s, 1
-2S, investigate
18Westgard Rules
- 4 1s Rule 4 consecutive values have been
plotted on the same side of the 1s range, may
indicate shift or trend - 10x Rule 10 consecutive values fall on the same
side of the mean either with same control or 2 or
more controls, indicates shift. - 1 3s and R 4s rule violations suggest random
error while 2 2s, 4 1s and 10x rules suggest
systematic errors within the test system. - Computer programmable with warnings to
technologists when violations occur.
19Other Internal Error Monitors
- Moving averages of red blood cell indices (drifts
of the moving average outside a defined limit
suggest instrument error) - Automated Differential Counting has improved the
precision of differentials because it overcomes
the non-random distribution of cells on a slide
and classifies thousands of cells versus 100 to
200 cells in a manual differential
20External Quality Control
- Internal Controls monitor precision
- External Controls indicate accuracy
- Subscription program of qualitative and
quantitative controls - Analyzed by hundreds of peer labs using same
equipment, same methods - Evaluation is based on how close results compare
with the mean opinion of the peer group
21Technical Staffing Quality Control
- Active continuing education program maintains
technical performance at a high level - Imperative to optimal performance
- Required by regulatory agencies for accredited
laboratories
22Overarching Quality Assurance Plan
- Goal is to ensure that the quality of laboratory
services contributes to the overall delivery of
excellent medical care - Focuses on high risk and problem prone areas of
system performance - Includes internal and external QC, staff
proficiency, pre-analytic, analytic and
post-analytic activity, computer functionality,
critical values notification and customer
satisfaction
23QA/QC Group Question Exercise
- Proper specimen collection and patient
satisfaction are components of - 1. Quality control
- 2. Quality assurance
- 3. Reliability
- 4. Precision
24QA/QC Group Question Exercise
- A patients white blood cells are counted on an
automated cell counter ten times. The mean white
count is 8000/µL and the standard deviation is
300. What is the coefficient of variation (CV)? - 1. .04
- 2. 2.6
- 3. 3.8
- 4. 26
25QA/QC Group Question Exercise
- The CV in the above question indicates that the
white blood counts on this cell counter are - 1. Accurate
- 2. High quality
- 3. Properly calibrated
- 4. Precise
26QA/QC Group Question Exercise
- What is the correct interpretation if a 10.0 g/dL
hemoglobin standard reads 13.0 g/dL? - 1. This is acceptable agreement
- 2. Imprecise
- 3. Insensitive
- 4. Inaccurate
27QA/QC Group Question Exercise
- Interpret the following platelet counts from Mr.
Billy Rueben. - Day 1 Day 2 Day 3__
- 478 ? 109/L 163 ? 109/L 501 ? 109/L
- 1. Day 2 exceeds delta check limits
- 2. Day 1 and day 3 results are too high
- 3. Day 1 results were probably collected on the
wrong patient - 4. All are invalid
28QA/QC Group Question Exercise
- Is the variation in the above platelet counts
possible? - 1. No
- 2. Yes, but unlikely
- 3. Yes, platelet counts typically vary this much
from day to day - 4. Yes, the patient probably received a platelet
transfusion
29QA/QC Group Question Exercise
- Which of the following is the most likely
explanation for the platelet counts in Mr.
Rueben? - 1. Patient is developing thrombocythemia
- 2. The sample from day 2 contained clumped
platelets from a traumatic phlebotomy - 3. Patient received a platelet transfusion on day
3 - 4. This is normal day-to-day platelet variation
30QA/QC Group Question Exercise
- Which of the following is true regarding
reference ranges? - 1. Should be derived from reference books
- 2. Need be determined only for adults
- 3. Can be established by running the test
procedure on 10 healthy people - 4. Are ranges of values for an analyte in normal
healthy people
31QA/QC Group Question Exercise
- A test that is positive in all patients who have
the disease but also in some who do not have the
disease is - 1. Sensitive
- 2. Specific
- 3. Accurate
- 4. Reliable
32- The antinuclear antibody test (ANA) is positive
in almost all people who have systemic lupus
erythramatosus (SLE). It is also positive in some
patients who do not have SLE. The anti-DNA is
positive only in people with SLE but not in all
that do. Which of the following is true? - 1. ANA is a good screening test and anti-DNA a
good confirmatory test - 2. Anti-DNA is a good screening test and ANA a
good confirmatory test - 3. Both are good screening tests for SLE
- 4. Neither of these tests is valid
33QA/QC Group Question Exercise
- A purchased hemoglobin standard is used to adjust
a hemoglobinometer. This standard is being used
as a - 1. Control
- 2. Reference
- 3. Delta check
- 4. Calibrator
34QA/QC Group Question Exercise
- The tubing that brings the lyse reagent to the
hemoglobin cuvette on an automated cell counter
is pinched and not delivering any reagent. All
hemoglobin values are greater than 20 g/dL. This
represents what type of error? - 1. Random
- 2. Imprecision
- 3. Constant systematic
- 4. Proportional systematic
35QA/QC Group Question Exercise
- On day 29 in a month one of the two controls was
between 2 and 3 SD. What is the correct procedure
to follow? - 1. Stop running patients until an investigation
is completed - 2. Rerun the control and if acceptable continue
with patients - 3. Recalibrate the instrument
- 4. Open new vials of controls and repeat both
controls
36QA/QC Group Question Exercise
- The control values for the prothrombin test were
ranging between the mean and minus one standard
deviation. Starting on day 20 of the month the
values consistently were between 1 and 2 SD.
This is an example of a - 1. Shift
- 2. Trend
- 3. Random error
- 4. Deviation
37QA/QC Group Question Exercise
- Which of the following would most likely be
associated with the above situation? - 1. Operator error
- 2. Burned out light source
- 3. Instrument miscalibrated
- 4. Starting a new lot number of thromboplastin
reagent
38QA/QC Group Question Exercise
- Which of the following patients should not be
included in indices moving averages? - 1. Chemotherapy recipients
- 2. Infants
- 3. Obstetrical
- 4. Surgical
39QA/QC Group Question Exercise
- A laboratory comparing its results to other labs
is an example of - 1. Precision monitoring
- 2. Internal QC
- 3. External QC
- 4. Delta checks
40QA/QC Group Question Exercise
- A laboratory gets numerous complaints regarding
the length of times it takes hematology results
to get to the emergency room. What would be an
appropriate response? - 1. Make this a quality assurance project
- 2. Ignore the complaints
- 3. Explain why it takes so long
- 4. Tell the employees to work faster
41QA/QC Group Question Exercise
- The best way to prevent errors in the laboratory
is to assure - 1. Quality instruments
- 2. Well trained, conscientious workers
- 3. Quality management
- 4. QC consistently being between 2 SD