Title: Thin Layer Chromatography TLC
1Thin Layer Chromatography (TLC)
- Rapid Screening of Pharmaceuticals by TLC
- Ross D. Kirchhoefer, trainer Allen Kenyon
(deceased), Tom Layloff - Gateway Analytical, 4041 Forest Park Ave., St.
Louis, MO, 63108 - USP, Rockville, MD
- Kayla Laserson,Tom Kenyon
- CDC, Division of TB Elimination, Atlanta, GA
30333/ CDC BOTUSA, Botswana
2Documentation cGMP/GLP (1)
- If it isnt written down, It didnt happen!
- Full description of standard, lot , purity, exp.
date - Full description of sample, dose, type of
formulation, packaging etc. - Full identification of chemicals, solvents, TLC
test strips with lot , equipment, etc. - Full description of reagent and/or mobile phase
preparation.
3Documentation cGMP/GLP (2)
- Full description of sample and standard
preparation. - Formulas used for calculations and sample
calculations. - Describe experimental observations and include
conclusions and/or results of the test. - Use a notebook or worksheet for recording all raw
data. - Strikeouts must be initialed, dated and reason
given for the strikeout. - Test method must be referenced.
4Documentation cGMP/GLP (3)
- Original chromatograms (TLC) strips must be used
for measurements and calculations (not copies). - If data is maintained elsewhere, a reference must
be made to its location. - Complete documentation of analytical procedures
and test results is a requirement of the FDC law.
The analyst is responsible for his/her work. - The USP PF lists the current USP reference
standards.
5Objective
- To introduce the beginner to basic TLC principles
- To describe a simple and economical procedure for
pharmaceutical screening
6Chromatography
- There are two basic types of chromatography
- Gas
- Liquid
- Liquid includes TLC and high performance liquid
chromatography (HPLC)
7Introduction
- TLC is a form of liquid chromatography consisting
of - A mobile phase (developing solvent) and
- A stationary phase (a plate or strip coated with
a form of silica gel) - Analysis is performed on a flat surface under
atmospheric pressure and room temperature
8Principles of TLC
- TLC is one of the simplest, fastest, easiest and
least expensive of several chromatographic
techniques used in qualitative and quantitative
analysis to separate organic compounds - Michael Tswett is credited as being the father of
liquid chromatography. Tswett developed his ideas
in the early 1900s.
9TLC
- The two most common classes of TLC are
- Normal phase
- Reversed phase
10Normal Phase
- Normal phase is the terminology used when the
stationary phase is polar for example silica
gel, and the mobile phase is an organic solvent
or a mixture of organic solvents which is less
polar than the stationary phase.
11Reversed Phase
- Reversed phase is the terminology used when the
stationary phase is a silica bonded with an
organic substrate such as a long chain aliphatic
acid like C-18 and the mobile phase is a mixture
of water and organic solvent which is more polar
than the stationary phase.
12Adsorbents for TLC
- Silica gel
- Silica gel-F (Fluorescing indicator added)
- Magnesium Silicate (Florisil)
- Polyamides
- Starch
- Alumina
13Silica Gel (1)
- Silica gel is the most common adsorbent used in
TLC - It also comes with a fluorescing indicator added
to it to make visualization or detection of
sample spots easier
14Silica Gel (2)
- Silica gel is a polymer based on Silica to oxygen
linkages with many -hydroxyl groups extending
from this matrix - Si-O-Si-O-Si-O-Si-O-Si-O-Si-(OH)x
15Silica Gel (3)
- They are very porous
- They have large surface area
- These affect your separation characteristics
16Silica Gel (4)
- The mode of separation is generally by adsorption
or partition - The more polar components will be adsorbed
preferentially by the polar layer - Hydrogen Bonding is the main force controlling
adsorption between the silica gel surface and the
analyte functional groups
17Steps in TLC Analysis
- The following are the important components of a
typical TLC system - Apparatus (developing chamber)
- Stationary phase layer and mobile phase
- Application of sample
- Development of the plate
- Detection of analyte
18General Procedure (1)
- Decide if you are going to do Normal or Reversed
phase chromatography - Prepare a plate or select a plate with the proper
sorbent material - Prepare the mobile phase
- Mark the plate
- Apply the sample
- Develop the plate
- Detect the analytes
19General Procedure (2)
- Silica gel with or without an added fluorescing
indicator is the most commonly used and is
classified as Normal phase chromatography - The mobile phase is generally a non-polar solvent
such as hexane. The hexane can be modified to a
more polar solvent by the addition of or organic
type solvents such as methanol, diethyl ether,
ethyl acetate, toluene, dimethyl-formamide, etc.
to achieve the required retention. - The mobile phase can be further modified by the
addition of acids or bases such as acetic acid or
triethylamine to reduce tailing
20ProcedureTLC Plates
- The plates can be pre-marked for origin and
development finish line as well as for sample
zones - Generally a distance of approximately 10 cm is
used as the development of a plate so as to make
the calculation of the Rf value easy. - Rf is defined as the movement of the sample zone
(x) divided by the movement of the developing
solvent ( x/ 10 cm)
21ProcedureTLC Plate Development
- The development of the plate is linear and
ascending - The developing chamber is usually glass to
prevent any interaction with the developing
solvent and capable of holding the size plate you
will be using - The chamber may or may not be pre-saturated with
the developing solvent - Development may be with multiple solvents
- Development may be continuous (seldom used)
- Development may be two-directional (right angles)
22ProcedureDevelopment Chamber
- If a plate is placed in an unsaturated chamber,
the air in the chamber is replaced by solvent
molecules both from the evaporation of the
solvent from the plate surface and the body of
the developing solvent in the chamber - If the developing solvent has more than one type
of solvent, evaporation will be selective based
on the boiling point of each solvent - If a plate is placed in a pre-saturated chamber,
no such evaporation can take place - The separation and spot or zone shape may be
different from these systems
23ProcedureSpot Movement
- The movement of the solvent up the plate is
induced by capillary action - Sample zone broadening always occurs and is
caused by eddy and molecular diffusion as the
spot moves up the plate - The distribution coefficient of the sample solute
affects the resolution of a separation i.e.
solutes A and B - the greater the distribution coefficient between
solutes, the greater is the resolution between
them
24Spotting the Sample
- The analyte must be in a suitable solution for
spotting and any solvent can be used in other
words the analyte must be in solution.
25Polarity
- Polarity of solutes Polar and non-Polar
- Polar solutes alcohols (ROH), acids (RCOOH),
amines (RNH2) - Polar solvents Methanol, ethanol, acetic acid
- Non-Polar solutes hydrocarbons, ketones
(compared to methanol) - Non-Polar solvents hexane, toluene (compared to
methanol)
26Solvent Eulotropic Series
- Solvent E-value
- Toluene 0.29
- Chloroform 0.40
- Acetone 0.56
- Ethyl Acetate 0.58
- Ethanol 0.88
- Methanol 0.95
- Acetic Acid/Ammonia High
- Water High
27Calculation of Solvent Polarity
- Efinal xE1 xE2 ..xEn
- x volume fraction of solvent
- E E value of solvent
- Example
- 25 mL CHCL3 75 mL MeOH
- 0.25 x 0.40 0.75 x 0.88 0.76
28Like Dissolves Like
- Polar molecules favor polar solvents and vice
versa - Polar solutes migrate faster in polar mobile phase
29Performing the TLC AnalysisTypes of Materials
Needed
- Solvent bottles, 1 liter
- Small bottles, wide mouth 100 mL
- Graduated syringes, 1, 5 and 10 mL
- Pestle
- Graduated cylinders25, 50 and 100 mL
- Volumetric glassware and pipettes
- Small sample vials 1.5 amd 6 mL
- Micropipettes, 1,2,3,4 and 5 microliters
- Pasteur pipettes and rubber bulb, assorted sizes
- Beakers. assorted small sizes
- DiSPO test tubes 3 to 10 mL sizes
30Performing the TLC AnalysisProcedures
- Preparation of sample
- Preparation of standards
- Preparation of developing solvent (mobile phase)
- Plate marking
- Spotting a plate
- Placing plate in development chamber
- Conditioning development chamber
- Development of plate
- Visualization and interpretation
- Estimation of concentration
- Calculations of Rf values
31Performing the TLC AnalysisPreparation of Sample
- Take one dosage unit or a composite of dosage
units and place in small plastic bag, grind to
powder, or transfer to a suitable vessel and add
proper solvent to dissolve active ingredient. - Make stock and dilutions
- Stock solutions and dilutions must be calculated
- Example 5 mg / 5 mL
- Concentrations normally about 1 (one) mg/mL
32Performing the TLC AnalysisPreparation of
Standards
- Place one reference tablet into a vessel or a
DiSPO test tube, add sufficient solvent to make a
solution equivalent to 100 of the active dosage
strength in the tablets (capsules) ie 5 mg / 5
mL - Dilute 4 parts of the 100 solution to 5 parts
or alternately (take 4 parts of the 100 and add
1 part solvent) this is an 80 value
33Performing the TLC analysis Making Your Own
Standards
- Use a reference standard tablets if available,
or - Alternatively use a primary or secondary standard
which must be available. An amount is weighed on
an analytical balance - The proper dilutions are made to give the
appropriate concentration equivalent to 100 of
the active dosage strength in the prepared sample
formulation - 5 mg/ 5 mL
- The 80 standard is also prepared
34Performing the TLC Analysis Preparation of the
Mobile Phase
- The mobile phase (developer) is usually a mixture
of solvents on a parts by volume basis - One part chloroform and one part methanol would
be noted as 11 - Pipettes or graduated cylinders can be used for
these measurements - Remember Polarity is controlled by your choice
of solvents (see Eulotropic values)
35Performing the TLC AnalysisMarking the TLC Plate
- 5 x 10 cm plates or plastic backed strips should
be provided in the kit - If not, they can be cut from 20 x 20 cm plastic
backed-silica coated sheets (but not recommended) - Mark a line about 1 cm below the top
- Mark a small point on either side of your
spotting point about 2 cm from the bottom - Do not remove silica from sides, top or bottom
- Stay away from sides of cut plate when applying
sample
36Performing the TLC AnalysisApplication of
Samples
- The plate is a plastic backed silica coated strip
- Spot 1 - 5 microliters of your sample in the
center of the plate at the origin line - Spot 1 - 5 microliters of the 100 standard and
the 80 standard on either side of the sample - Note If 3 microliters of sample is spotted,
spot 3 microliters of the standards
37Performing the TLC AnalysisDry the Spots
- After spotting the sample and standards, the
solvents must be evaporated from the spots before
developing - If aqueous solutions or partially aqueous
solutions are used as solvents, several minutes
may be needed to dry the spots
38Performing the TLC Analysis Assemble the TLC
Apparatus
- Assemble the TLC frame, plastic bag, saturator
strips, aluminum holder, clamp and fishhook - Add developing solvent
- If you wish to saturate the chamber, do so by use
of the filter paper strip
39Performing the TLC AnalysisDevelopment of the
Plate
- Attach the TLC spotted plate (plastic
backed-silica coated strip) to the aluminum frame
with the clamp and lower into the plastic bag
with the fishhook - Allow the TLC plate to stay in the bag without it
contacting the solvent for about 5 minutes to
reach equilibrium - Pull the plastic bag down to allow the developing
solvent to contact the lower 1 cm of the TLC
strip - Develop the strip to the top marked line
- Stop the development
- Remove the TLC strip and allow the solvent vapors
to evaporate
40Performing the TLC AnalysisVisualization and
Interpretation (1)
- Most pharmaceutically active drugs will not be
visible to the naked eye - Spots can be visualized by two basic techniques
- Ultraviolet light at 254 nm (shortwave UV). Long
wave UV (340 nm) is used less commonly. - Staining to make spots visible
41Performing the TLC Analysis Visualization and
Interpretation (2)
- Ultraviolet - short wave 254 nm
- Place the TLC strip under the UV light. Room
light should be eliminated as much as possible - If a silica gel F plate is used, the sample spots
will appear as black spots on a fluorescent green
background - The sample zone intensity should be between the
standards - Battery operated shortwave UV lamps are
available They are small and quite handy
42Performing the TLC AnalysisVisualization and
Interpretation (3)
- Staining
- Zones may be made visible by staining or spraying
the TLC plate (strip) with a visualization
reagent. Several different reagents can be used.
Not all will visualize your sample zone. - You must have the correct visualization reagent
to be able to view your spots - A universal visualization reagent is a 10
sulfuric acid solution. When sprayed on your
plate, the plate is heated and your spots are
charred which can be seen by eye. Glass plate
only. Several details may need to be worked out
with this reagent
43Performing the TLC Analysis Visualization and
Interpretation (4)
- Staining
- Prepare the staining reagent and place in a
plastic bag - Dip the TLC strip into the reagent
- Remove the TLC strip and observe the spots
- The sample zone intensity should be between the
standards
44Performing the TLC Analysis Calculate the Rf
Values
- The Rf value is calculated by measuring the
distance the sample zone travels divided by the
distance the developing solvent travels - Values below 0.1 is considered poor the spots
are too close to origin - Values of 0.1 to 0.8 are good and any other spots
(impurities) or other actives are resolved form
each other - Above 0.8 poor spots may be too broad or
distorted
45Performing the TLC Analysis Acceptance or
Rejection Criteria
- Sample zone has an intensity between the
standards of 80 - 100 ACCEPT - If lower - REJECT
- If higher, re-spot a standard at a concentration
higher than 100 , the 100 standard and the
sample may be needed to estimate concentration - Normally, bad drug samples will be lower than
100 level and this is not a concern - Sample zones should be resolved from any other
active drugs (combination drug products like
Isoniazid and Rifampin) ACCEPT - Sample zones should be separated from any
decomposition products, impurities or excipients
in the drug formulation. If many alternate zones
besides the active are found, the drug may be
decomposed REJECT and do more testing
46Performing the TLC Analysis Acceptance or
Rejection Criteria
- There should be no additional zones is the
sample ACCEPT - There should be no unexplained zones in the
sample ACCEPT - The sample and standard should have identical Rf
values ACCEPT - If the sample and standard have different Rf
values REJECT
47Development of a New TLC Method
- Determine the drug type , its polarity, and/or
acidity or basicity. - Find a solvent the drug is soluble to extract
from the matrix - Select a TLC strip or plate you feel may be
suitable, ie silica gel or silica gel G-F240 may
be a good first choice - Prepare a mobile phase you feel may provide Rf
values between 0.1 and 0.8 - Find a technique to visualize the drug
- Test
- If necessary, modify to improve the technique
48Advantages of TLC
- Low cost
- Short analysis time
- Ease of sample preparation
- All spots can be visualized
- Sample cleanup is seldom necessary
- Adaptable to most pharmaceuticals
- Uses small quantities of solvents
- Requires minimal training
- Reliable and quick
- Minimal amount of equipment is needed
- Densitometers can be used to increase accuracy of
spot concentration
49TLC Problems Troubleshooting
- Over migration Developer too polar Reduce
polarity - Under migration Developer too non-polar Increase
polarity - Distorted solvent front Developer not
equilibrated Equilibrate - Distorted spots Wrong adsorbent Change plates
- Distorted spots Spotted too much Change
concentration - No separation Wrong developer Change developer
- No separation Wrong adsorbent Change plate type
- Tailing Spot overloading Reduce concentration
- Tailing Component is basic Increase acidity
- Tailing Component is acidic Increase basicity
- Tailing/no separation Decomposition Developer/pla
te
50Kodak Slide Show
- Review of Safety
- Review of Performing the TLC Analysis
- Preparing the Sample alternatives given by
instructor - Preparing the standard
- Preparing the plate (silica gel strip)
- Marking the plate (strip)
- Spotting the sample
- Developing the plate, calculate Rf
- Visualization techniques
51(No Transcript)