T24 2.4 GHz Telemetry Part 2
Advanced Intelligent Instrumentation
2
Chapter 6
T24 Acquisition Modules
3
Acquisition Modules up to 200Hz
  • T24-SA
  • 5V DC Excitation Voltage
  • Calibrated to 2.5 mV/V
  • Strain Gauge Drive Capability 85 to 5000 Ohms
  • Noise Free Resolution
  • Sample time lt 10 ms 150,000
  • 1Kg 50 Ton Load Cell
  • Sample time lt 1,000 ms 1250,000
  • 20g 50 Ton Load Cell
  • T24-VA
  • 0 10 V input
  • Noise Free Resolution
  • Sample time lt 10 ms 17,000
  • Sample time lt 1,000 ms 111,000
  • T24-IA
  • 0 20 mA input
  • Calibrated 4 - 20mA
  • Noise Free Resolution
  • Sample time lt 10 ms 15,000
  • Sample time lt 1,000 ms 110,000

4
Acquisition Modules 2 KHz
  • T24-SAf
  • 5V DC Excitation Voltage
  • Calibrated to 2.5 mV/V
  • Strain Gauge Drive Capability 85 to 5000 Ohms
  • Noise Free Resolution
  • Sample time lt 10 ms 150,000
  • Sample time lt 1,000 ms 1250,000
  • T24-VAf
  • 0 10 V input
  • T24-IAf
  • 0 20 mA input
  • Calibrated 4 - 20mA
  • Points to Note
  • Logging only Available in paired configuration
    with toolkit
  • Concatenated Binary Data Provider makes
    integration only possible with Base Station

5
Housing Powering Modules
  • T24-ACM
  • Houses any Acquisition Module
  • IP65 rated case
  • Option of
  • Field terminals for Power and Sensor Connection
  • Battery Pack
  • Extended Range ANTA standard, ANTB Further
  • T24-BC1
  • Li-ion Battery Charger
  • Integrated Voltage Regulator
  • 5V Charge supply
  • 2 Charge Currents
  • 133mA
  • 466mA

6
Configuring Acquisition Modules
Information Settings
  • Apply System Name (Optional)

1
7
Calibrating Acquisition Modules
Calibration
  • Select the number of Calibration points you wish
    to take you calibration over.
  • Enter the value of each calibration point. i.e.
    0Kg, 50Kg, 120Kg, 220Kg, etc...
  • As you apply each weight click the acquire button
    next to the value to set the calibration point.

1
2
3
8
Configuring Acquisition Modules
  • Transmit Interval Time between transmission.
    Default 333mS resulting in a 3Hz Sample
  • Sample Time Time sample it taken for from
    connected sensor. Default 5ms
  • Low Power Mode Manages if the modules stands by
    between transmissions below 40ms Low power mode
    ineffective
  • Battery Life Guide Enter the details of the
    battery connected and sensor details plus
    predicted usage per 24 hours and the guide will
    give you the noise free resolution and predicted
    battery life for the acquisition module.

Data Rate
1
2
3
4
9
Configuring Acquisition Modules
This page displays Battery Voltage Link Quality
Indicator (LQI) at the Base Station (Local) and
Acquisition Module (Remote). The LQI is
calculated from the Relative Signal Strength
Indicator (RSSI) Correlation Value (CV) which
can be viewed under the advance view.
LQI Battery
  • Low Battery Level When the Battery goes below
    this voltage the SA will transmit Low battery
    warning to Data consumers

1
ADVANCED VIEW
10
Configuring Acquisition Modules
Advance Settings
  • Sleep Delay Time lapse that module will sleep
    after last Keep Awake message. Handheld send
    Keep Awake messages every 5 sec therefore it is
    wise to set this to 30 seconds so if Handheld
    moves out of range the Acquisition Modules
    Sleeps. Default - (0) no sleep delay
  • Data Tag This is the tag attached to any data
    provided by the Acquisition Module.
  • Default - last 4 HEX digits of Radio ID.

1
2
11
Configuring Fast Acquisition Modules
Information Settings
  • Apply System Name (Optional)

1
12
Configuring Fast Acquisition Modules
Battery Life
LQI Battery
1
13
Configuring Fast Acquisition Modules
Zero Settings
  • System Zero Value
  • Zero Now
  • Current nV/V reading

1
2
3
14
Configuring Fast Acquisition Modules
Data Provider Monitor
  • Pause / Continue Chart
  • Start / Stop Logging to .csv File

1
2
15
Configuring Fast Acquisition Modules
Advance Settings
  • Sleep Delay Time lapse that module will sleep
    after last Keep Awake message. Handheld send
    Keep Awake messages every 5 sec therefore it is
    wise to set this to 30 seconds so if Handheld
    moves out of range the Acquisition Modules
    Sleeps. Default - (0) no sleep delay
  • Data Tag This is the tag attached to any data
    provided by the Acquisition Module.
  • Default - last 4 HEX digits of Radio ID.
  • 3. Add in Shunt for Calibration

1
2
3
16
Chapter 7
T24 Handheld Modules
17
Handhelds
  • T24-HS
  • Single Connection to Acquisition Module
  • Tare Function
  • Local and Remote Low Battery Indicator
  • Signal Low indicator
  • Battery Life 40 Hours Continuous Use
  • T24-HA
  • Connection to up to 12 Acquisition Modules
  • 2 Models
  • Result Sums values from all Acquisition
    Modules
  • Item Allows users to scroll through all
    Acquisition Modules
  • Same Functionality Indicators as T24-HS

18
T24-HS
Information Settings
  • Apply Component Name (Optional)

1
19
T24-HS
Display Format
  • Format Resolution Enter the format you wish
    the display to show
  • Leading Zero Suppression Remove proceeding
    zeros from display
  • Overload Limit Value at which display shows

1
2
000.0165 0.0165
No Yes
3
OVERLOAD
20
T24-HS
Zero Settings
  • Power On Auto Zero Limit above which initial
    value is not tared away
  • Zero Indication Band Value below which zero
    should be displayed.

1
2
21
T24-HS
Advance Settings
  • Waker Duration Period waking will be attempted
    for
  • Do Sleep Wake Should Handheld wake and sleep
    Acquisition Module
  • Auto Off Delay no button press shutdown
  • Keep Awake Interval Time between Keep Awake
    Packets
  • Pair Duration Maximum Pairing Time

1
5
2
3
4
22
T24-HA
Information Settings
  • Apply Component Name (Optional)

1
23
T24-HA
  • Format Resolution Enter the format you wish
    the display to show
  • Leading Zero Suppression Remove proceeding
    zeros from display
  • Overload Limit Value at which display shows
  • Motion Band /- Value Data must stay with in to
    be considered stable
  • Motion Time Time that the Data must stay with
    in motion band to be considered stable.

Display Format
000.0165 0.0165
1
No Yes
2
3
OVERLOAD
4
5
24
T24-HA
  • Power On Auto Zero Limit above which initial
    value is not tared away
  • Zero Indication Band Value below which zero
    should be displayed.
  • Allow System Zero Length of time the Tare
    button should be pushed for to perform system
    zero
  • External System Zero Acquisition Module value
    subtracted from Result

Zero Settings
1
2
3
4
ADVANCED VIEW
25
T24-HA
Mode Comms
  • Operational Mode
  • Result
  • Item
  • F1 Data Tag Data Tag attached to Data provider
    produced when F1 Pushed.
  • Allow Next Key In Result mode pushing and
    holding next key for this period will cause
    individual values to be displayed.
  • F1 Data Specifies whether data provided on F1
    key press
  • As Displayed
  • Always Gross

3
1
4
2
26
T24-HA
Mode Comms
  • Adding Acquisition Modules
  • Manual Enter Data Tag and ID
  • Pairing Click and Power cycle the
    Acquisition Module

27
T24-HA
  • Waker Duration Period waking will be attempted
    for.
  • Do Sleep Wake Should Handheld wake and sleep
    Acquisition Module
  • Auto Off Delay no button press shutdown
  • Keep Awake Interval Time between Keep Awake
    Packets
  • Pair Duration Maximum Pairing Time
  • Item Duration How long Item values are
    displayed for in Result Mode
  • Message Duration How long Item Message will be
    shown for

Advance Settings
1
6
7
2
3
4
5
28
Chapter 8
T24 Output Modules
29
Output Modules
  • T24-SO
  • User formatted ASCII string output over RS232 /
    RS485
  • 9 36V Power Supply
  • Connects to up to 8 Acquisition Modules
  • Provides Sum Function of Acquisition Modules
  • Switch input
  • Data output Triggered via Data Provider
  • Log Numbering
  • T24-PR1
  • 56mm Carriage Thermal Printer
  • 9 36V Power Supply
  • Connected to up to 8 Acquisition Modules
  • Print Triggered via Data Provider
  • Print Log Numbering
  • T24-AO
  • Analogue Output Device
  • 10 V output
  • 0-20mA output

30
T24-SO
Information Settings
  • Apply Component Name (Optional)

1
31
T24-SO
  • Switch input Mode Trigger Output or Switch
    between Gross Net
  • Output Trigger Data Tag On reception Data
    Provider with this Tag output is triggered
  • Data Tag of Acquisition Module inputs
  • Timeout Time after which data error has
    occurred
  • Format of Data from Data provider
  • Sum Format This is the format of the Sum, in a
    system with no direct data acquisition modules
    this sets the output format

Input Settings
1
2
3
4
5
6
32
T24-SO
  • Print Options
  • Duplicate Output
  • Min Interval between Outputs
  • Text Options
  • Text to replace ltGNgt token Gross and Net Mode
  • Print On Error If Any Acquisition Module Fails
    should output occur
  • Log Options
  • Initial Log Value Scope of Log Number
  • Non Volatile Log Number
  • Line Delay and Character
  • Do Output Triggers an Output

Output Settings
4
1
2
5
3
6
33
T24-SO
Output Settings
  • Output Scaling
  • Low input and High input read out specifications
  • Output Design
  • Format the ASCII string output

Output Scaling
Output Design
Available Variables to add to output String
2
1
Description of Variables
34
T24-SO
Zero Settings
  • Perform System Zero i.e. Tare
  • Remove System Zero
  • External System Zero Acquisition Module value
    subtracted from Result

1
2
3
ADVANCED VIEW
35
T24-PR1
Information Settings
  • Apply Component Name (Optional)

1
36
T24-PR1
Input Settings
  • Output Trigger Data Tag On reception Data
    Provider with this Tag output is triggered
  • Data Tag of Acquisition Module inputs
  • Timeout Time after which data error has
    occurred
  • Format of Data from Data provider
  • Sum Format This is the format of the Sum, in a
    system with no direct data acquisition modules
    this sets the output format

1
2
3
4
5
37
T24-PR1
Output Settings
  • Print Options
  • Duplicate Prints
  • Min Interval between Prints
  • Text Options
  • Text to replace ltGNgt token Gross and Net Mode
  • Print On Error If Any Acquisition Module Fails
    should print occur
  • Log Options
  • Initial Log Value Scope of Log Number
  • Non Volatile Log Number
  • Do Output Triggers a Print

4
1
2
3
5
38
T24-PR1
Output Settings
  • Output Scaling
  • Low input and High input read out specifications
  • Output Design
  • Format the Print out

Output Scaling
Output Design
Available Variables and formatting to add to
output String
2
1
Description of Variables
39
T24-PR1
Zero Settings
  • Perform System Zero i.e. Tare
  • Remove System Zero
  • External System Zero Acquisition Module value
    subtracted from Result

1
2
3
ADVANCED VIEW
40
Chapter 9
Advanced System Architecture
41
System Architecture
  • Data is not requested but provided by Acquisition
    Modules
  • Acquisition Modules Standby during operation
    under 25Hz
  • Acquisition Modules Sleep Wake to preserve
    battery life
  • Sleep and Wake functions controlled by Data
    Consumers e.g. base station or handheld
  • Sleeping Modules intermittently check to wake
    (default 5 sec)
  • Data Consumers Wake Modules by responding to Wake
    Check Transmissions (default 12 sec)

Standby
Standby
Standby
Standby
Transmission Types
Sleep
Sleep
Sleep
Data Provider Wake Check Command
5s
5s
5s
333ms
333ms
333ms
333ms
Waiting to Wake
Keep Awake
Wake
Sleep
42
System Architecture
Standby
Standby
Standby
Standby
Sleep
Sleep
Sleep
5s
5s
5s
333ms
333ms
333ms
333ms
Waiting to Wake
Transmission Types
Keep Awake
Wake
Sleep
Data Provider Wake Check Command
Transmission Interval
Transmission Interval
Transmission Interval
Transmission Interval
Wake Check Int
Wake Check Int
Wake Check Int
5s
5s
5s
333ms
333ms
333ms
333ms
Waiting to Wake
5s
Keep Awake Int
Sleep
Waker Duration
43
System Architecture
Common Configuration Errors
  • Wake Check Interval Waker Duration Miss Match

Wake Check Int
Wake Check Int
10s
10s
Transmission Types
8s
Data Provider Wake Check Command
Waker Duration
  • Sleep Delay Shorter than Keep Awake Interval

Sleep Delay
5s
Wake Check Int
5s
8s
Keep Awake Int
44
Chapter 10
System Examples
45
Simple Crane Link
T24-SA Strain Acquisition
  • T24-BSu USB Base Station
  • System Configuration
  • T24-SA Calibration
  • 2 x D cell alkaline batteries
  • 1000 ohm load cell
  • 3 updates per second.
  • 1 minute weigh period.
  • 60 weighs per day
  • Battery Life 5 years

T24-HS Handheld Simple
46
Weigh Bridge
  • T24-BSu USB Base Station
  • System Configuration
  • T24-SA Calibration
  • 2 x D cell alkaline batteries
  • 1000 ohm load cell
  • 3 updates per second
  • Active for 2 hrs per day
  • Battery Life 4 years

Serial Link
  • T24-SO Serial Output Module
  • Collect Data from T24 SAs
  • Output to Display

Serial Dis Serial Display Module
47
Multi Link Weighing
  • T24-BSu USB Base Station
  • System Configuration
  • T24-IA Calibration
  • T24-HA Handheld Advance
  • Collect Data from T24-IAs
  • Total Weight
  • Send Data to PR1 on F1 Key
  • 2 x D cell alkaline batteries
  • 350 ohm load cell
  • 3 updates per second
  • 5 minute weigh per lift
  • 15 lifts per day
  • Battery Life 4 years
  • T24-PR1 Printer Module
  • Print Formatted Receipt with total weight

48
Silo Weighing
  • T24-BSu USB Base Station
  • System Configuration
  • Collect Data from T24 SAs
  • 2 x D cell alkaline batteries
  • 4 x 350 ohm load cell
  • 1 update every 15 seconds
  • Battery Life 5 years

49
Centre Of Gravity
  • T24-BSu USB Base Station
  • System Configuration
  • Collect Data from T24 SAs
  • Custom Visual Link Software
  • 2 x AA alkaline batteries
  • 1000 ohm load cell
  • 10 updates per second
  • 30 minutes usage per day
  • Battery Life 1 year 1 month

50
Remote Logging
  • PC Running T24-TK
  • Logging All Data to Log File
  • Remote Access Via Internet

Serial Link
  • 2 x D cell alkaline batteries
  • 1000 ohm load cell
  • 1 update every 5 seconds
  • Battery Life 5 years
  • T24-BSi Industrial Base Station
  • System Configuration
  • Collect Data from T24 SAs

51
Use with Data Logger
  • T24-BSu USB Base Station
  • System Configuration

Serial Link
  • 2 x AA cell alkaline batteries
  • 1 x 1000 ohm load cell
  • 200 updates per second
  • No sleep wake
  • Battery Life 7 days 14 hours
  • 2 x AA cell alkaline batteries
  • 1 x 1000 ohm load cell
  • Update every 30 minutes
  • Battery Life 5 years
  • T24-SO Serial Output Module
  • Collect Data from T24 SAs
  • Output User Formatted ASCII String

52