Title: Medical Imaging Systems:
1Medical Imaging Systems MRI Physics and Image
Formation
Instructor Walter F. Block, PhD 1-3 Notes
Walter Block and Frank R Korosec, PhD 2-3
Departments of Biomedical Engineering
1, Radiology 2 and Medical Physics 3 University
of Wisconsin - Madison
2Major MRI Scanner Vendors
Philips Intera CV
Siemens Sonata
General Electric CV/i
3Siemens Espree, 1.5 T, 70 cm bore, 125 cm length
4Extremity Scanner (0.2 T)
5OrthOneXT, 1.0 T Extremity Scanner
6Magnetic Field Strengths
- 3.0 T Growing high end market (Neuro, MSK)
- 4.0 T Few
- 7.0 T Few human research scanners
- 8.0 T Animal scanners
- 9.4 T Animal scanners
- 0.2 T Orthopedics
- 0.3 T Open scanners
- 0.5 T Some closed bore
- 0.7 T Open scanners
- 1.0 T Common
- 1.5 T Most common
71.5 T vs. 3.0 T (3D MR Angiography)
8MRI Uses Three Magnetic Fields
- Static High Field (B0) (Chapter 12, Prince)
- Creates or polarizes signal
- 1000 Gauss to 100,000 Gauss
- Earths field is 0.5 G
- Radiofrequency Field (B1) (Chapter 12, Prince)
- Excites or perturbs signal into a measurable form
- On the order of O.1 G but in resonance with MR
signal - RF coils also measure MR signal
- Excited or perturbed signal returns to
equilibrium - Important contrast mechanism
- Gradient Fields (After Lunch, Chapter 13, Prince)
- 1-4 G/cm
- Used to image determine spatial position of MR
signal
9Nuclear Magnetic Dipole Moment
Magnetic Dipole Representation
Vector Representation
10Nuclear Magnetic Dipole Moment Spinning Charge
N
N
P
P
P
P
P
N
Hydrogen
Helium
Helium-3
11Nuclear Magnetic Dipole Moment
- To have a magnetic dipole moment, it must have
- An odd number of protons or
- An odd number of neutrons or
- An odd number of both protons and neutrons
12Imaging Criteria
- To be imaged, nuclei must
- have an odd number of neutrons, protons, or both
- be abundant in the body
- Hydrogen in the water molecule satisfies both
- The hydrogen nucleus is composed of a single
proton (odd number of nucleons) - Water comprises 70 of the body by weight (very
abundant) - Most widely imaged
- Termed spins in MRI
13Atoms That Can be Imaged with MRI
H
C
O
F
Na
P
K
1
13
17
19
23
31
39
1
6
8
9
11
15
19
.093
.016
1.0
.066
S
N
A
where
Z
N is the element abbreviation A is the total
number of protons and neutrons combined (atomic
mass) Z is the number of protons ( atomic number)
S relative MRI sensitivity compared to hydrogen
14No Magnetic Field
No Net Magnetization
Random Orientation
15 Static Magnetic Field (B0)
Body RF (transmit/receive)
Bore (55 60 cm)
Gradients
Shim (B0 uniformity)
Magnetic field (B0)
16 Classical Physics Top analogy
- Spins in a magnetic field analogous to a
spinning top in a gravitational field.
Axis of top
gravity
Top precesses about the force caused by
gravity Dipoles (or spins) will precess about the
static magnetic field
17 Reference Frame
y
x
z
Magnetic field (B0) aligned with z (longitudinal
axis and long axis of body)
18B0 Field Dipoles Separate Into 2 States
Magnetic Field (B0)
Positive Orientation (Lower Energy)
Negative Orientation (Higher energy)
19Larmor Equation
?????B
Precessional Frequency
Magnetic Field Strength
- for hydrogen 42.58 MHz/T
- 64MHz 42.58 MHz/T ? 1.5T
- With stronger gravity, top will precess faster
- Same with dipoles stronger B0 causes faster
precession
20Dipole Moments from Entire Sample
B0
7 up
6 down
Non-Random Orientation
21Sum Dipole Moments -gt Bulk Magnetization
Net Magnetization
B0
M
The magnetic dipole moments can be summed to
determine the net or bulk magnetization, termed
the vector M.
22Bulk Magnetization
- ? of aligned spins ? ? bulk magnetization ? ?
MR signal
- ? strength of B0 ? ? MR signal
At 1.5 T and 310 K, there are only 9 excess
aligned spins 2 per million protons!
But, a 1 x 1 x 1 mm voxel of water contains 0.3
million billion excess aligned spins!!!
Endpoint We can understand MRI with simple
classical physics instead of quantum physics
23MRI Uses Three Magnetic Fields
- Static High Field (B0)
- Creates or polarizes signal
- 1000 Gauss to 100,000 Gauss
- Earths field is 0.5 G
- Radiofrequency Field (B1)
- Excites or perturbs signal into a measurable form
- On the order of O.1 G but in resonance with MR
signal - RF coils also measure MR signal
- Excited or perturbed signal returns to
equilibrium - Important contrast mechanism
- Gradient Fields (After Lunch)
- 1-4 G/cm
- Used to image determine spatial position of MR
signal
24 B1 Radiofrequency Field
- Polarized signal is all well and good, but what
can we do with it? We will now see how we can
create a detectable signal.
To excite nuclei, tip them away from B0 field by
applying a small rotating B field in the x-y
plane (transverse plane). We create the rotating
B field by running a RF electrical signal
through a coil. By tuning the RF field to the
Larmor frequency, a small B field (0.1 G) can
create a significant torque on the magnetization.
Diagram Nishimura, Principles of MRI
25Exciting the Magnetization Vector
z
B1 tips magnetization towards the transverse
plane. Strength and duration of B1 can be set
for any degree rotation. Here a 90 degree
rotation leaves M precessing entirely in the xy
(transverse) plane.
Laboratory Reference Frame
26Tip Bulk Magnetization
z'
y'
x'
Rotating Reference Frame Imagine you are
rotating at Larmor frequency in transverse plane
27Tip Bulk Magnetization
z'
y'
x'
Rotating Reference Frame
28Tip Bulk Magnetization
z'
y'
x'
Rotating Reference Frame
29Tip Bulk Magnetization
z'
y'
x'
Rotating Reference Frame
30Larmor Equation
?????B
Tip Angle
? ?Dt????B1Dt
Tip Angle
Amplitude of RF Pulse
Time of Application of RF Pulse
31Magnetization is tipped using an RF Pulse
Components of tipped Magnetization (M)
'
'
'
'
'
'
In general, any component of the magnetization
can be tipped into the transverse plane to give
rise to a signal. Next, how do we detect the
signal?
32After B1 Mtrans ( Mxy ) precesses about B0
B0
t0
t 0
Laboratory Frame
33Signal From Precessing Magnetization
Receiver (antenna) integrates proton signal Farad
ays Law of Induction Change in magnetic flux in
time across a surface induces an electric
potential (voltage signal)
34Radiofrequency Receiver Coils
35Radiofrequency Receiver Coils
Transmit and Receive
36Electromagnetic Energy Spectrum
Graph Medical Imaging Systems Macovski, 1983
37Excitation Is it Permanent?
- Static High Field (B0)
- Creates or polarizes signal
- 1000 Gauss to 100,000 Gauss
- Earths field is 0.5 G
- Radiofrequency Field (B1)
- Excites or perturbs signal into a measurable form
- On the order of O.1 G but in resonance with MR
signal - RF coils also measure MR signal
- Excited or perturbed signal returns to
equilibrium - Important contrast mechanism
- Magnetization (spins) will return to equilibrium
(Chapter 12, Prince) - Very important image contrast mechanisms
- More important than proton density
- Soft tissue has little difference in proton
density - Cause of flat contrast in X-ray and CT soft
tissue - Gradient Fields (After Break)
- 1-4 G/cm
- Used to image determine spatial position of MR
signal
38T2 Relaxation
39Tip Bulk Magnetization
z'
y'
x'
40Tip Bulk Magnetization
z'
M M0
y'
x'
41Tip Bulk Magnetization
z'
y'
x'
42Tip Bulk Magnetization
z'
y'
x'
43Tip Bulk Magnetization
z'
y'
x'
44Transverse Magnetization
z'
y'
Mxy
x'
45Transverse Magnetization
z'
2 up
y'
Mxy
2 down
x'
46T2 Decay
-
z'
z'
-
-
y'
y'
x'
x'
T2 relaxation is dephasing of transverse
magnetization
47Transverse Magnetization
z'
y'
Mxy
x'
48T2 Decay Slight dephasing
z'
y'
x'
T2 relaxation is dephasing of transverse
magnetization
49T2 Decay More dephasing
z'
y'
x'
T2 relaxation is dephasing of transverse
magnetization
50T2 Decay
z'
y'
x'
T2 relaxation is dephasing of transverse
magnetization
51T2 Decay
z'
y'
x'
T2 relaxation is dephasing of transverse
magnetization
52T2 Decay Complete dephasing
z'
y'
0
Mxy
x'
T2 relaxation is dephasing of transverse
magnetization
53T2 Relaxation
M0
Mxy M0 e-t/T2
Mxy
Time
54T2 Signal Contrast
M0
Proton Density
Moderate T2-weighting
Mxy
Heavy T2-weighting
0.37M0
Mxy M0 e-t/T2
T2
Time
T2
55Post Exercise Arm Imaging
Proton density
Moderate T2
Moderate T2
Mary Sesto Wally Block Tom Best Robert Radwin
Heavy T2
56Sample tissue time constants T2
T2 of some normal tissue types
- Table Nishimura, Table 4.2
57T2 T2 Relaxation
T2
Mxy
T2
Time
58Causes of Spin Dephasing
1
1
??B0
T2
T2
- Local magnetic fields produced by nuclear
neighbors (T2) - Inhomogeneities in the main magnetic field, B0
(T2) - Quality of magnet usually not a concern
- Bulk susceptibility differences (T2)
- Air/tissue interfaces
- Iron
59Different Names, Same Phenomenon
- T2 relaxation
- Mxy or Mtrans decay
- Spin-spin relaxation
60T2-weighted Sagittal Spine
61T1 Relaxation
62Tip Bulk Magnetization
z'
y'
x'
63Tip Bulk Magnetization
z'
y'
x'
64Tip Bulk Magnetization
z'
y'
x'
65Tip Bulk Magnetization
z'
y'
x'
66Tip Bulk Magnetization
z'
y'
x'
67Transverse Magnetization
z'
y'
Mxy
x'
68T1 Relaxation
B0
z
z
2 up
y
y
x
x
2 down
No Net Longitudinal Magnetization
69T1 Relaxation
B0
z
z
3 up
y
y
x
x
1 down
Spins Flip to Regrow Net Longitudinal
Magnetization
70T1 Relaxation
z'
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
71T1 Relaxation
z'
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
72T1 Relaxation
z'
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
73T1 Relaxation
z'
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
74T1 Relaxation
z'
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
75T1 Relaxation
z'
Mlong M0
y'
x'
T1 relaxation is regrowth of the longitudinal
magnetization
76T1 Relaxation
M0
Mz M0(1-e-t/T1)
Mz
Time
77Source of T1 Contrast Varying Mz
M0
0.63M0
Mz M0(1-e-t/T1)
Mz
Time
T1
78Different Names, Same Phenomenon
- T1 relaxation
- Mz or Mlong recovery
- Spin-lattice relaxation
79T1 T2 Relaxation Trends
T1 depends on tumbling rate (more tumbling at
Larmor Frequency shorter T1) Solids
Very Long T1 Viscous Short
T1 Non-Viscous Long T1 T2 depends on time
spent in vicinity of nuclear neighbors
(more time near same neighbors shorter
T2) Solids Very Short T2 Viscous
Short T2 Non-Viscous Long T2
80Relative Relaxation Rates
81T1 and T2 vs B0
- T1 ? B0
- T1 ? as B0 ?
- T1 ? as B0 ?
- T2 largely independent of B0
82T1 Relaxation
83Components of M after Excitation
Laboratory Frame
84T1-, ?-, and T2-weighted images
?-weighted
T2-weighted
T1-weighted
85Brain Tumor Imaging
Whats changed between these images?
T1-weighted Sagittal
T1-weighted Axial
T2-weighted Axial
86Images of the Knee
?-weighted
T2-weighted
87MR Physics So far...
- What we can do so far
- 1) Excite spins using RF field at ?o
- 2) Record time signal (Known as FID)
- 3) Mxy decays, Mz grows
- 4) Repeat.
But so far RF coils only integrate signal from
entire body. We have no way of forming an
image. That brings us to the last of the three
magnetic fields in MRI and image formation.