Title: Ultrasonic Sensors Calibration
1Ultrasonic Sensors Calibration
- Omar A. Daud
- Truc-Vien T. Nguyen
- May 16, 2008
2Plan
- Ultrasonic sensors
- Calibration
- Approaches
- Analysis
- Conclusion
3Ultrasonic sensors
- Evaluate attributes of a target by interpreting
the echoes of sound waves. - Generate high frequency sound waves and evaluate
the echo which is received back by the sensor. - Sensors calculate the time interval between
sending the signal and receiving the echo to
determine the distance to an object.
4Plan
- Ultrasonic sensors
- Calibration
- Approaches
- Analysis
- Conclusion
5Calibration
- In performing a calibration, the following steps
are necessary - Examine the construction of the instrument,
identifying the inputs. - Decide which of the inputs are significant for
the application. - By holding some inputs constant, varying others,
and recording the output(s), develop the desired
static input-output relations
6Plan
- Ultrasonic sensors
- Calibration
- Approaches
- Analysis
- Conclusion
7Static Calibration Approach
- All inputs, except one are kept at some constant
values. Then the one input under study is varied
over some range of constant values, which causes
the output(s) to vary over some range of constant
values. - Considerable care in choosing the means of
determining the numerical values of these inputs.
Object Distance Cm Mean Mean Mean Std Std Std
field 60 Cm Sensor 4 Sensor 5 Sensor 6 Sensor 4 Sensor 5 Sensor 6
Eighth_field 18,24 12,03 9 14,13 0,12 0
Quarter_field 41,13 19,92 17,85 21,64 0,19 0,33
Half_field 42,22 33,87 31,69 22,32 0,26 0,33
1_field 65,84 56,92 59,68 5,92 3,66 1,61
2_field 59,63 47,31 60,98 18,58 25,58 7,59
3_field 56,67 49,09 55,09 21,56 24,49 12,80
8Correlation between Sensors - Distance
- Support Vector Machine (SVM) Approach
- LibSVM Polynomial kernel, degree 3, One vs.
One - Parameter C 2 n-fold cross-validation n 10
- 12 sensors accuracy 58.3667
- 4 sensors 4,5,6,7 accuracy 84.5667
- 3 sensors 4,5,6 accuracy 85.4333
9Plan
- Ultrasonic sensors
- Calibration
- Approaches
- Analysis
- Conclusion
10Analysis Sensor 4
11Analysis Sensor 5
12Analysis Sensor 6
13Analysis Three sensors
3 2 1 1/2 1/4 1/8
Mean 47.5 53 36.5 43 40.5 27
Std 28.43 31.61 22.08 25.84 24.39 16.6
Var 419.81 389.78 30.89 186.65 266.7 81.31
14Gaussian distribution
Gaussian distribution in six distances
15Plan
- Ultrasonic sensors
- Calibration
- Approaches
- Analysis
- Conclusion
16Conclusions
- Farther is the object from the scene more
inaccurate is the measurement of the US sensors. - By the contrary, as the object is near the
sensors, the measurement is more accurate. - There is a sort of linear relationship between
the input and the output until the measurement
of 1 field. - Beyond the 1 field this relationship is lost.