Title: THERMAL PROPERTIES
1THERMAL PROPERTIES
2THERMAL PROPERTIES
- Plays a vital role in evaluating the product
performance processibilty characteristics in
polymers. - Thermal analytical methods monitor differences in
some sample property as the temperature
increases, or differences in temperature between
a sample and a standard as a function of added
heat. These methods are usually applied to solids
to characterize the materials.
3THERMAL PROPERTIES
- Heat Deflection Temperature (HDT)
- Vicat Softening Temperature (VSP)
- Thermal Endurance
- Thermal Conductivity
- Thermal Expansion
- Low Temperature Brittleness
- Flammability
- Melting Point, Tm, and Glass Transition, Tg
(DSC) - Thermomechanical Analysis
4Heat Deflection Temperature
Defined as the temperature at which a standard
test bar (5 x ½ x ¼ in ) deflects 0.010 inch
under a stated load of either 66 or 264 psi.
- Significance
- HDT values are used to compare the elevated
temperature performance of the materials under
load at the stated conditions. - Used for screening and ranking materials for
short-term heat resistance. - HDT values do not represent the upper temperature
limit for a specific material or application. - The data are not intended for use in design or
predicting endurance at elevated temperatures.
5Test Methods, Specimen Conditioning
- Test Method
- ASTMD 648, ISO 75 -1 and 75-2
- Test Specimen
- 127mm (5 in.) in length, 13mm (½ in.) in depth by
any width from 3mm (? in.) to 13mm ((½ in.) -
- Conditioning
- 23 2oC and 50 5 RH for not less than 40 hrs
prior to test. - Two replicate specimens are used for each test
6Apparatus
- Specimen Supports Metal supports for the
specimen of 100 2mm - Immersion Bath
- Deflection Measurement Device
- Weights 0.455 MPa (66 psi) 2.5 or 1.82 MPa
(264 psi) 2.5. - Temperature Measurement System
Apparatus for Determination of HDT
7Procedure
- Measure the width and depth of each specimen
- Position the test specimens edgewise in the
apparatus - Position the thermometer bulb sensitive part of
the temperature - Stir the liquid-heat transfer medium thoroughly
- Apply the loaded rod to the specimen and lower
the assembly into the bath. - Adjust the load to obtain desired stress of 0.455
MPa (66 psi) or 1.82 MPa (264 psi) - Five minutes after applying the load, adjust the
deflection measurement device to zero or record
its starting position - Heat the liquid heat-transfer medium at a rate of
2.0 0.2oC/min. - Record the temperature of the liquid
heat-transfer medium at which the specimen has
deflected the specified amount at the specified
fibre stress.
8Calculation
The weight of the rod used to transfer the force
on the test specimen is included as part of the
total load. The load (P) is calculated as P
2Sbd2 / 3L Where, S Max. Fibre stress in the
specimen of 66 Psi / 264 Psi b Width of
specimen d Depth of specimen L Width of span
between support (4 in)
9Results Conclusion
- A bar of rectangular cross section is tested in
the edgewise position as a simple beam. - Load applied at the center to give maximum fibre
stresses of 66 /264 psi. - The specimen is immersed under load in a
heat-transfer medium provided with a means of
raising the temperature at 2 0.2oC/min. - The temperature of the medium is measured when
the test bar has deflected 0.25mm (0.010 in). - This temperature is recorded as the deflection
temperature under flexural load of the test
specimen.
10Factors influencing
- HDT of unannealed (heat treatment) specimen is
usually lower than that of annealed specimen. - Specimen thickness is directly proportional to
HDT because of the inherently low thermal
conductivity of plastic materials. - Higher the fibre stress or loading lower the HDT.
- Injection moulded specimen tend to have a lower
HDT than compression moulded specimen. - Compression moulded specimen are relatively
stress free.
11Vicat Softening Point (VSP)
Defined as the temperature at which a flat ended
probe with 1 mm2 cross section penetrates a
plastic specimen to 0.04 inch (1 mm) depth.
- Significance
- Data obtained by this test method may be used to
compare the heat-softening qualities of
thermoplastic materials. - This test method is useful in the areas of
quality control, development and characterization
of plastic materials.
12Test Methods, Specimen Conditioning
- Test Method
- ASTMD 1525 or ISO 306
- Test Specimens
- The specimen shall be flat, between 3 and 6.5mm
thick and at least 10 by 10mm in area or 10mm in
diameter. - Conditioning
- 23 2oC and at 50 5 relative humidity of not
less than 40 hrs
A minimum of two specimens shall be used to test
each sample.
13Apparatus
- Immersion Bath
- Heat-Transfer Medium
- Specimen Support
- Penetration-Measuring Device Masses 10 0.2N or
50 1.0N - Temperature-Measuring Device
- Needle
Fig. 2 Apparatus for Softening Temperature
Determination
14Procedure
- Prepare the immersion bath so that the
temperature of the heat-transfer medium is
between 20 and 23oC at the start of the test - Place the specimen, which is at room temperature,
on the specimen support. - The needle should not be nearer than 3mm to the
edge of the specimen. - Gently lower the needle rod, without the extra
mass, so that the needle rests on the surface of
the specimen and holds it in position. - Position the temperature-measuring device so that
the sensing end is located within 10mm from where
the load is applied to the surface of the
specimen. - Lower the assembly into the bath and apply the
extra mass required to increase the load on the
specimen to 10 0.2N (Loading 1) or 50 1.0N
(Loading 2). - After a 5-min waiting period, set the penetration
indicator to zero. - Start the temperature rise.
- Record the temperature of the bath when the
needle has penetrated 1 0.01mm into the test
specimen.
15Results Conclusion
- Vicat softening temperature is expressed as the
arithmetic mean of the temperature of penetration
of all specimens tested. - If the range of penetration temperatures for the
individual test specimens exceeds 2oC, record the
individual results and repeat the test, using at
least two new specimens.
16Thermal Conductivity
- Rate at which heat is transferred by conduction
through a unit cross sectional area of a material
when a temperature gradient exists perpendicular
to the area. -
- The coefficient of thermal conductivity (K
factor), is defined as the quantity of heat that
passes through a unit cube of the substance in a
given unit time when the difference in
temperature of the two faces is 10C. - Mathematically, thermal conductivity is expressed
as - K Qt/A(T1-T2)
- Q amount of heat passing through a cross
section, A causing a temperature difference, ?T
(T1-T2), t thickness of the specimen. - K is the thermal conductivity, typically measured
as BTU.in / (hr.ft2.0F) indicates the materials
ability to conduct heat energy.
17Significance
- Thermal conductivity is particularly important in
applications such as headlight housings, pot
handles hair curlers that require thermal
insulation or heat dissipation properties. - Computerized mold-filling analysis programs
requires special thermal conductivity data
derived at higher temperatures than specified by
most tests.
18Test Methods Specimen
- Test method Guarded hot plate test
- ASTM D177, ISO 2582
- Test Specimen two identical specimens having
plane surface of such size as to completely cover
the heating unit surface - The thickness should be greater than that for
which the apparent thermal resistivity does not
change by more than 2 with further increase in
thickness
19Apparatus
- The apparatus is broadly of two different
categories of the following - Type I (low temperature) Temperature of cold
plate 21 K, Temperature of heating unitlt500 K - Type II (High temperature) Temperature of heating
unit rangegt550 K -lt1350K - Heating units
- Gap Metering Area
- Unbalance Detectors
- Cooling units
- Sensors for measuring Temperature difference
- Clamping force
- Measuring system for Temperature detector outputs
20Guarded Hot plate Apparatus
Guarded Hot plate Apparatus Courtesy Bayer
Material Data Sheet
21Procedure
- Two test specimens are sandwiched between the
heat source (main heater) heat sink one on
either side of the heat source. - The clamping force is so adjusted that the
specimens remain in perfect contact with the
heater sink - Guard heaters are provided to prevent heat flow
in all except in the axial direction towards the
specimen - The time of stabilization of input out put
temperature is noted. - Temperature difference between the hot cold
surfaces of the specimen should not be less that
5 K or suitable differences as required.
22Calculation
The relationship between the quantity of heat
flow and thermal conductivity is defined as Q
K/ x Q Quantity of heat flow K Thermal
Conductivity X The distance the heat must
flow Thermal conductivity is calculated as
K Qt / A (T1 T2) Q Rate of heat
flow (w) T Thickness of specimen (m) A Area
under test (m2) T1 Temperature of hot surface
of specimen (k) T2 Temperature of cold surface
of specimen (k)
23Results Conclusion
- Thermal conductivity is calculated by using the
value of rate of flow at a fixed temperature
gradient. - Data are obtained in the steady state
24Factors influencing
- Crystallites have higher conductivity.
- As the density of the cellular plastic decreases,
the conductivity also decreases up to a minimum
value and rises again due to increased convection
effects caused by a higher proportion of open
cells.
25Thermal Expansion (Coefficient of Linear Thermal
Expansion, CLTE)
- Measures the change in length per unit length of
a material, per unit change in temperature. - Expressed as in/in/0F or cm/cm/0C
- Mathematically, CLTE (a), between temperatures T1
and T2 for a specimen of length L0 at the
reference temperature, is given by - a (L2 L1)/L0(T2 T1) ? L/L0?T
26Significance
- Determines the rate at which a material expands
as a function of temperature. - The higher the value for this coefficient the
more a material expands and contracts with
temperature changes. - Plastics tend to expand and contract anywhere
from six to nine times more than materials that
are metallic. - The thermal expansion difference develops
internal stresses and stress concentrations in
the polymer, which allows premature failure to
occur.
27- Test Method ASTMD 696
- Test Specimen
- 12.5 by 6.3mm (½ in. by ¼ in.) 12.5 by 3mm (½ by
? in.), 12.5mm (½ in.) in diameter or 6.3mm (¼
in.) in diameter. - Conditioning 23 2oC and 50 5 RH for not
less than 40h prior to test.
28Apparatus
- A vitreous silica dilometer
- Dial gage
- The weight of the inner silica tube the
measuring device reaction shall not exert a
stress gt 70 kPa on the specimen so that the
specimen is not distorted or appreciably
indented. - Scale or Caliper
- Controlled Temperature Environment
- Means shall be provided for stirring the bath
- Thermometer or thermocouple
29Procedure
- Measure the length of two conditioned specimen at
room temperature - Mount each specimen in a dilatometer, install the
dilatometer in the 30oC control environment. - Maintain the temperature of the bath in the range
32oC to 28oC 0.2oC until temperature of the
specimen along the length is constant - Record the actual temperature and the measuring
device reading. - Change to the 30oC bath, so that the top of the
specimen is at least 50mm below the liquid level
of the bath. - Maintain the temperature of the bath in the range
from 28 to 32oC 0.2oC - Record the actual temperature and the measuring
device reading. - Change to 30oC and repeat the above procedure
measure the final length of the specimen at room
temperature. - If the change in length per degree of temperature
difference due to heating does not agree with the
change length per degree due to cooling within
10 of their average investigate the cause of the
discrepancy and if possible eliminate. - Repeat the test until agreement is reached.
30Calculation
- Calculate the CLTE over the temperature range as
- a ?L/Lo?T
- a Average coefficient of linear thermal
expansion degree - Celsius.
- ?L Change in length of test specimen due to
heating or to - cooling,
- Lo Length of test specimen at room
temperature (?L and Lo being measured in the
same units), and - ?T Temperature differences, oC, over
which the change in the length of
the specimen is measured. - The values of a for heating and for cooling shall
be averaged to give the value to be reported.
31Result Conclusion
- Provide a means of determining the CLTE of
plastics, which are not distorted or indented by
the thrust of the dilatometer on the specimen. - The specimen is placed at the bottom of the outer
dilatometer tube with the inner one resting on
it. - The measuring device, which is firmly, attached
to the outer tube is in contact with top of the
inner tube and indicates variations in the length
of the specimen with changes in temperature. - Temperature changes are brought about by
immersing the outer tube in a liquid bath or
other controlled temperature environment
maintained at the desired temperature. - The nature of most plastics and the construction
of the dilatometer make 30 to 30oC a convenient
temperature ranges for linear thermal expansion
measurements of plastics. - This range covers the temperatures in which
plastics are most commonly used.
32Factors influencing
- Thermal expansion is substantially affected
- by the use of additives
- especially fillers
- Wt Of loading
- Lowers the coefficient of thermal expansion.
33Differential Scanning Calorimetry (DSC)
- DSC measures the heat flow into or from a sample
as it is heated, cooled or held under isothermal
conditions - Applications of DSC includes characterization of
- Polymers
- fibres
- Elastomers
- Composites
- films
- pharamaceuticals
- foods
- cosmetics
34- DSC provides the following important properties
of materials - Glass Transition Temp. (Tg)
- Melting point (Tm)
- Crystallization times Temp.
- Heats of melting crystallization
- Percent Crystallinities
- Heat set temp.
- OIT
- Compositional Analysis
- Heat capacities
- Heats of cure
- Thermal Stabilities
35Apparatus
- DSC apparatus consists of
- Furnace
- Temperature Sensor
- Differential Sensor
- Test Chamber Environment
- Temperature Controller
- Recording Device
- Sealed pans
- Balance
36- Samples Powder, Liquids, crystal
37Procedure
- DSC apparatus consists of two sealed pans sample
and reference aluminum pans - The pans are heated, or cooled, uniformly while
the heat flow difference between the two is
monitored. - This can be done at a constant temperature
(isothermally), but is more commonly done by
changing the temperature at a constant rate,
called temperature scanning. - The instrument detects differences in the heat
flow between the sample and reference plots the
differential heat flow between the reference and
sample cell as a function of temperature.
38First Order Transitions (Tc, Tm)
- Specimen mass appropriate of 5-mg is taken in the
pan - Intimate thermal contact between the pan and
specimen is established for reproducible results.
- Heat the sample at a rate of 10oC/min under inert
gas atmosphere from 50oC below to 30oC above the
melting point to erase the thermal history . - The selection of temperature and time are
critical when effect of annealing is studied. - Hold temperature for 10min.
- Cool to 50oC below the peak crystallization
temperature at a rate of 10oC/min and record the
cooling curve. - Repeat heating as soon as possible under inert
purge gas at a rate of 10oC/min, and record the
heating curve.
39For Second order Transition (Tg)
- Use a specimen mass of 5-mg.
- Perform and record a preliminary thermal cycle as
up to a temperatures 30oC above the extrapolated
end temperature, Te, to erase previous thermal
history, heating at a rate of 20oC/min. - Hold temperature for 10min.
- Quench cool to 50oC below the transition
temperature of interest. - Hold temperature for 10min.
- Repeat heating at a rate of 20oC/min, and record
the heating curve until all desired transition
have been completed.
40Typical DSC curves
41Measurement of various Properties/Explanations
- Heat Capacity
- Heating the sample Reference pans, the the
difference in heat output of the two heaters is
plotted against temperature. i.e the heat
absorbed by the polymer against temperature.
42Heat Capacity
- The heat flow at a given temperature is
represented units of heat, q supplied per unit
time, t. - The heating rate is temperature increase T per
unit time, t.
Dividing,
43Glass Transition
- Property of the amorphous region
- Below Tg Disordered amorphous solid with
immobile molecules - Above Tg Disordered amorphous solid in which
portions of molecules can wiggle around - A second order transition ( Increase in heat
capacity but there is no transfer of heat
44Crystallization
- Above Tg, the polymers are in mobile conditions.
- When they reach the right temperature, they gain
enough energy to move into very ordered
arrangements, which we call crystals, - When polymers fall into these crystalline
arrangements, they give off heat. - When this heat is dumped out, there is drop in
the heat flow as a big dip in the plot of heat
flow versus temperature
We call crystallization an exothermic transition.
45Melting
Above Tc, we reach the polymer's melting
temperature, or Tm, those polymer crystals begin
to fall apart, that is they melt. The chains
come out of their ordered arrangements, and begin
to move around freely. Melting is a first order
transition (Tm).
46Putting it all together
47Polymer crystallinity
Measure the area of under the melting of the
polymer. Plot of heat flow per gram of material,
versus temperature.
multiply this by the mass of the sample
48Degree of crystallinity is given by
X 100 Xc
Where H Heat of Fusion determined from DSC
thermogram Hm Heat of fusion of a 100
crystalline sample
49Results Conclusion
- DSC thermograms provides an elaborate picture of
various transitions in a polymer. - The degree of crystallinity in a polymer sample,
specific heat etc. can be determined. - Any side reaction (for example, crosslinking,
thermal degradation or oxidation) shall also be
reported and the reaction identified if possible.
50Factors affecting
- Addition of fillers affects the transitions in
DSC - Previous thermal history of the samples also
affects the DSC transitions. - There should be proper contact between the
samples pans
51Thermo Gravimetric Analysis (TGA)
Changes in weight of the specimen is recorded as
the specimen is heated in air or in a controlled
atmosphere such as nitrogen
52Terminologies
- Highly volatile matter moisture, plasticizer,
residual solvent or other low boiling (200oC or
less) components. - Medium volatile matter medium volatility
materials such as oil and polymer degradation
products. In general, these materials degrade or
volatilize in the temperature range 200 to 750oC.
- Combustible material oxidizable material not
volatile (in the unoxidized from) at 750oC, or
some stipulated temperature dependent on
material. Carbon is an example of such a
material. - Ash nonvolatile residues in an oxidizing
atmosphere which may include metal components,
filler content or inert reinforcing materials. - Mass loss plateau a region of a
thermogravimetric curve with a relatively
constant mass.
53Significance
54Apparatus
55Procedure
- Establish the inert (nitrogen) and reactive (air
oxygen) gases at the desired flow rates in the
range of 10 to 100mL/min. - Switch the purge gas to the inert (nitrogen) gas.
- Zero the recorder and tare the balance.
- Open the apparatus to expose the specimen holder.
- Prepare the specimen of 10 to 30mg and carefully
place it in the specimen holder. - Position the specimen temperature sensor
- Enclose the specimen holder.
- Record the initial mass.
- Initiate the heating program within the desired
temperature range. - Record the specimen mass change continuously over
the temperature interval. - The mass loss profile may be expressed in either
milligrams or mass percent of original specimen
mass. - Once a mass loss plateau is established in the
range 600 to 1200oC, depending on the material,
switch from inert to reactive environment.
56Calculation
57Calculation
58Calculation
59Calculation
60Factors influencing
- Oil-filled elastomers have such high molecular
weight oils and such low molecular weight polymer
content that the oil and polymer may not be
separated based upon temperature stability. - Ash content materials (metals) are slowly
oxidized at high temperatures and in an air
atmosphere, so that their mass increases (or
decreases) with time. Under such conditions, a
specific temperature or time region must be
identified for the measurement of that component.
- Polymers, especially neoprene and acrylonitrile
butadiene rubber (NBR), carbonize to a
considerable extent, giving low values for the
polymer and high rubber values. - Others, such as calcium carbonate, release CO2
upon decomposition at interference is dependent
upon the type and quantity of pigment present.
61Dynamic Mechanical Analysis (DMA)
- DMA is a technique in which a substance while
under an oscillating load is measured as a
function of temperature or time as the
substance is subjected to a controlled
temperature program in a controlled atmosphere. - Dynamic Mechanical Analysis (DMA) examines
materials between -170C and 1000C.
62DMA - Storage Modulus
63DMA for Testing Recyclates Virgin ABS and ABS
with 25 Recyclate
- The high inherent sensitivity of DMA
provides a means of testing polymers with
recyclates - In this example of virgin ABS and ABS with 25
recyclates, the effects of the recyclates can
be easily observed - Testing was done with 3-point bending probe
64FOURIER TRANSFORM INFRARED TECHNIQUE
- Identification of plastic through structural
analysis - Identification of additives, fillers, etc.
- Polymer blend analysis
- Monomer content analysis on plastics
- Compatibility studies on blends
- Curing of polymers
- Degradation studies
65THANK YOU