Title: Subject%20Characteristics
1(No Transcript)
2Monitoring of Respiration BYAHMAD
YOUNESPROFESSOR OF THORACIC MEDICINE Mansoura
Faculty Of Medicine
3The three major components of respiratory
monitoring during sleep include
- 1-Measurement of airflow,
- 2-Measurement of respiratory effort,
- 3-Measurement of arterial oxygen saturation
(SaO2). - Ancillary monitoring may include detection of
snoring and recording surrogates of the arterial
partial pressure of carbon dioxide (PaCO2)
including end-tidal partial pressure of carbon
dioxide (PETCO2) and transcutaneous partial
pressure of carbon dioxide (TcPCO2). - The AASM scoring manual recommends specific
sensor types and techniques to be used for
recording respiration during sleep,
4(No Transcript)
5TECHNIQUES TO MEASURE AIRFLOW OR TIDAL VOLUME
- The pneumotachograph (PNT) is the most accurate
method to measure airflow during sleep studies , - This device quantifies airflow by measurement of
the pressure drop across a linear (constant)
resistance (usually a wire screen). The
relationship between the pressure change, flow
rate, and resistance is given by the following
equation Pressure change Flow X Resistance - The PNT is worn in a mask covering the nose and
mouth. - Although the PNT is commonly used to measure
airflow during sleep research, this device is
rarely used during clinical sleep studies.
6The setup will consist of a flow head
(pneumotachometer) and a transducer which will
integrate volume from flow.
7Thermal devices were the first to be used to
monitor airflow during clinical sleep studies.
- These devices actually detect changes in
temperature induced by airflow (cooler inspired
air, warmer exhaled air). - The changes in device temperature result in
changes in voltage output (thermocouples) or
resistance (thermistors). - Thermal sensors are generally adequate to detect
an absence of airflow (apnea), but their signal
does not vary in proportion to airflow.
Therefore, thermal sensors are not an ideal means
of detecting a reduction in airflow (hypopnea).
8Thermal signals and PNT flow are equal at 1
L/sec. However, as airflow decreases, the thermal
signals overestimate flow.
- This illustrates that thermal sensors are not
ideal sensors to detect hypopneas (reductions in
flow). - The thermal sensor signal decreases when the
nostrils are large or the thermal sensor is
further from the nares. - Thermal devices composed of polyvinylidine
fluoride (PVDF) film may offer a better estimate
of flow. - Nasal-oral thermal sensors usually have a portion
of the device placed within or just outside the
nostrils with another portion over the mouth
(detection of oral flow). - A major advantage of thermal sensors is that they
can detect both nasal and oral airflow without
the need for a cumbersome mask covering the face.
9(No Transcript)
10Measurement of nasal pressure (NP) provides an
estimate of nasal airflow that is more accurate
than one obtained with most thermal sensors.
- NP is measured using a nasal cannula connected to
an accurate pressure transducer . - Because the cannula tips are inside the nares and
the other side of the pressure transducer is open
to the atmosphere, the pressure being measured is
actually the pressure drop across the resistance
of the nasal inlet associated with nasal airflow.
- The relationship of NP and flow is given by
Equations - NP K1(Flow)2 K1 constant
- Flow K2 NP. K2
constant - Changes in cannula position, periods of partial
oral flow, and obstruction of the cannula by
nasal secretions make the linearized NP signal a
less accurate measure of flow over the entire
night.
11The nasal prongs signal decreases more than that
of the pneumotachograph during a reduction in
airflow. The linearized nasal prongs signal is
very similar to that of the pneumotachograph
12The AHI values from both the NP and the
linearized NP signals showed excellent agreement
with the AHI values determined from the PNT
- The AHI values detected by the NP signal tended
to be slightly higher than the linearized NP
signal, but the differences were usually small. - The inter-measurement agreements (kappa) between
NP and PNT and linearized NP and PNT signals were
both excellent and essentially identical. - During normal unobstructed flow, the inspiratory
shape (contour) of the NP signal is round ,
whereas during airflow limitation , the shape of
the PNT and NP signals is flattened . - Airflow limitation is characteristically present
during obstructive reductions in airflow
(hypopnea) or snoring. - In contrast, when reductions in airflow are
simply due to a fall in inspiratory effort, the
NP signal amplitude is reduced but the shape is
round. - The most important limitation of the NP technique
is that approximately 10 of patients are mouth
breathers and the NP signal may be misleading.
13At A, the flow is rounded, whereas at C, the
flattened airflow contour is associated with an
increase in pressure drop across the upper airway
and airflow limitation. AC,
14The AASM scoring manual recommends nasal-oral
thermal sensors for detection of apnea and NP
sensors (with or without square root
transformation of the signal) for detection of
hypopnea
- Simultaneous use of both NP and nasal-oral
thermal sensors is recommended and has the
additional advantage of having a backup sensor if
the other airflow detection device fails . - The AASM scoring manual notes that if the
recommended sensor signal is not reliable, the
alternative sensor can be used. - In adults, the alternative airflow sensor for
apnea detection is the NP signal. The alternative
sensors for hypopnea detections are oronasal
thermal flow and respiratory inductance
plethysmography (RIP).
15The nasal pressure signal shows an absence of
airflow, whereas the nasal-oral thermal sensor
shows continued airflow. This pattern of airflow
is due to oral breathing.
16RIP is another method that can be used to detect
apnea and hypopnea
- The signals from rib cage (RC) and abdominal
bands (AB) sensors can be summed in an
uncalibrated manner (RIPsum RC AB) or as a
calibrated signal (RIPsum a RC b AB) as
an estimate of tidal volume (not airflow). - Here, RC and AB are signals from bands around the
rib cage and abdomen and a and b are
calibration factors determined during a
calibration procedure. - If one takes the time derivative of the RIPsum
signal, the result is an estimate of airflow
(RIPflow). - The RIPsum during apnea has minimal deflections
(approximately zero tidal volume) and during
hypopnea reduced deflections (reduced tidal
volume).
17In the case of an obstructive apnea , the RC and
AB deflections must nearly exactly cancel each
other (paradox).In the case of hypopnea, there
is a reduction in the RIPsum signal (low tidal
volume) as well as both the RC and the AB
signals.
- In the case of obstructive hypopnea, there may
also be paradox with chest and abdomen moving in
opposite directions . - If a RIPsum signal is not available, one can
detect hypopnea by a reduction in the RC and AB
RIP signals. - The AHI values obtained from the RIPsum and time
derivative of the RIPsum signals showed good
agreement with AHI values from the PNT signal.
18Obstructive apnea Respiratory inductance
plethysmography signals from the rib cage (RC)
and abdominal bands (AB) are summed (RIPsum). The
RIPsum is an estimate of tidal volume.
19Obstructive hypopnea Respiratory inductance
plethysmography signals from the rib cage (RC)
and abdominal bands (AB) are summed (RIPsum). The
RIPsum is an estimate of tidal volume.
20(No Transcript)
21MEASURING RESPIRATORY EFFORT
- Determination of respiratory effort is essential
to classify apneas as obstructive (continued
respiratory effort), central (absent effort), or
mixed (central followed by obstructive portions).
- The most sensitive and accurate method of
detecting respiratory effort is by measurement of
esophageal pressure. - Changes in esophageal pressure are estimates of
changes in pleural pressure that occur during
respiration (negative intrathoracic pressure
during inspiration). - Esophageal pressure monitoring can detect rather
feeble respiratory efforts even when RC and AB
movements are minimal. In addition, the size of
the pressure deflections provides an estimate of
the magnitude of respiratory effort. - Detection of respiratory effortrelated arousals
(RERAs) is most accurately performed with
esophageal pressure manometry.
22MEASURING RESPIRATORY EFFORT
- Measurement of esophageal pressure can be
performed using air-filled balloons, fluid-filled
catheters, or catheters with pressure transducers
on their tips. - The technique does require special equipment and
expertise and is routinely performed in only a
few sleep centers. - Some research sleep studies measure supraglottic
pressure instead of esophageal pressure using a
transducer tip placed just below the tongue base.
This allows measurement of the pressure drop
across the upper airway . Because this site is
below the area of upper airway closure or
narrowing in obstructive respiratory events,
supraglottic pressure can also be used to detect
respiratory effort.
23Esophageal pressure deflections increase during
an obstructive apnea that might at first glance
appear to be central apnea (absent inspiratory
effort).The chest and abdomen effort belt signals
showed minimal deflections during obstructive
apnea.
24 25The most common method for detecting respiratory
effort in clinical sleep studies utilized
piezoelectric (PE) sensors connected to bands
around the RC and AB.
- Changes in the tension on the PE transducer as
the RC and AB expand and contract produce a
voltage that can be measured .The signal from
these devices depends on the degree of tension on
the transducer. - The PE belts are adequate for detection of
respiratory effort in most patients but do not
really quantify the changes in RC or AB volume. - Although relatively inexpensive compared with RIP
effort belts, the PE effort belts may provide
misleading information (false absence of
respiratory effort), especially if not properly
positioned and tensioned.
26RIP belts provide a more accurate method of
detecting changes in RC and AB motion during
respiration than PE belts.
- The inductance of coils in bands around the RC
and AB changes during respiration as the RC and
AB expand and contract. - The band inductance varies proportionately to the
cross-sectional area the band encircles. - An oscillator is applied to each circuit and
changes in inductance are converted into a
voltage output. - The RIP bands consist of wires attached to a
cloth band in a zig-zag pattern. This produces a
larger change in inductance for a given change in
band circumference. - Recall that if the RIP signals are calibrated,
the RIPsum signal aX RC bX AB is an estimate
of tidal volume. Here ,the constants a and b are
determined during a calibration procedure.
27The accuracy of the RIPsum signal can deteriorate
if body position changes or the positions of the
bands change during sleep.
- Studied patients with sleep apnea using both
calibrated RIP belts and esophageal pressure
monitoring showed that only 9 of patients were
obstructive apneas sometimes misclassified as
central apneas by the RIP belts. In these
instances, esophageal pressure deflections were
present when there was no detectable change in
the RIP belt signals. - Thus, RIP effort belts are not 100 sensitive for
detecting respiratory effort. However, if the
bands are properly positioned and tensioned
(sized) , they will detect respiratory effort (if
present) in most patients. - The vast majority of sleep centers do not perform
RIP belt calibration. It should be noted that the
deflections of uncalibrated RIP bands do not
always accurately reflect the magnitude of
inspiratory effort or always show paradox during
obstructive apnea and hypopnea.
28Surface diaphragm EMG recording utilizes two
electrodes about 2 cm apart horizontally in the
seventh and eighth intercostal spaces in the
right anterior axillary line.
- The right side of the body is used to reduce ECG
artifact. - Intercostal EMG recording often uses the right
parasternal area (second and third intercostal
spaces in the midaxillary line). - Inspiratory EMG activity is noted in the
intercostal muscles and the diaphragm during
nonrapid eye movement (NREM) sleep. - During rapid eye movement (REM) sleep, the
intercostal activity is inhibited but
diaphragmatic activity persists, although often
diminished in amplitude during bursts of eye
movements. - The AASM scoring manual recommends use of
esophageal manometry or calibrated or
uncalibrated RIP belts for detection of
respiratory effort during sleep studies in adults
and children . The measurement of respiratory
muscle EMG is listed as an alternative method of
detecting respiratory effort in adults.
29An obstructive apnea with respiratory effort
monitored by both chest and abdominal respiratory
inductance plethysmography (RIP) bands and right
intercostal electromyogram (EMG). The right
intercostal EMG signal shows bursts coincident
with inspiratory effort (and movement of chest
and abdomen).A blow up of one EMG burst is shown
at the bottom of the figure in a raw form and
with the electrocardiogram (ECG) artifact
minimized.
30OXYGEN SATURATION
- Pulse oximetry. In this method SaO2 is determined
by the passage of two wavelengths of light (650
nm and 805 nm) through a pulsating vascular bed
from one sensor to another. The light is
partially absorbed by the oxygen-carrying
molecule, hemoglobin, depending on the percent of
the hemoglobin saturated with oxygen. A processor
calculates absorption at the two wavelengths and
computes the proportion of hemoglobin that is
oxygenated, giving it a numerical value. - A thin anatomic pulse site (such as the finger
tip, ear lobe, nose, or toe) is required, as is
proper alignment of the sensors. - With movement in sleep, the device can become
dislodged. - The readings can also be affected by anemia,
hemoglobinopathies, a high carboxyhemoglobin
level, elevated methemoglobin level, anatomic
abnormalities/previous injury to the site tested,
sluggish arterial flow (due to hypovolemia or
vasoconstriction),and the use of nail polish.
31(No Transcript)
32Apneas are followed by arterial oxygen
desaturations. Longer apneas are associated with
more severe desaturation. SpO2 pulse oximetry.
33OXYGEN SATURATION
- In sleep monitoring, an arterial oxygen
desaturation is usually defined as a decrease in
the SpO2 of 3 or 4 or more from baseline. - The nadir in SaO2 commonly follows apnea
(hypopnea) termination by approximately 6 to 8
seconds (longer in severe desaturations) . This
delay is secondary to circulation time and
instrumental delay (the oximeter averages over
several cycles before producing a reading).
34(No Transcript)
35OXYGEN SATURATION
- The assessment of the severity of desaturation
include the number of desaturations, the average
minimum SpO2 during desaturations, the time below
80, 85, 90,as well as the mean SaO2 and the
minimum saturation during NREM and REM sleep. - The time with an SpO2 88 is also commonly
determined. - Oximeters may vary considerably in the number of
desaturations they detect and their ability to
discard movement artifact. - Using long averaging times may dramatically
decrease the detection of desaturations. - The ability of oximeters to detect desaturations
is especially important in light of the
definitions of hypopnea that depend on an
associated desaturation. - The AASM scoring manual recommends a maximum
averaging time of 3 seconds at a heart rate of 80
bpm. In patients with a slow heart rate, a
slightly longer averaging time (at least a 3-beat
average) may be needed.
36MEASUREMENT OF PaCO2 DURING SLEEP
- Documentation of hypoventilation during sleep
requires measurement (or estimate) of the PaCO2. - The AASM scoring manual defines sleep-related
hypoventilation in adults as an increase in the
PaCO2 during sleep 10 mm Hg compared with an
awake supine value. - Continuous ABG monitoring during polysomnography
to determine the PaCO2 is not practical. An ABG
sample sometimes is performed at the start or the
end of the study. The sample can be used to
validate a surrogate measure of PaCO2 such as
PETCO2 or TcPCO2. - If an ABG sample is taken just at awakening, it
may be used to infer hypoventilation.
37PETCO2
- Capnography consists of the continuous
measurement of the fraction of CO2 in exhaled
gas. - This is usually performed using an infrared
sensor and, less commonly, a mass
spectrophotometer. - The PCO2 is determined by multiplication of the
fraction of CO2 by (Patm47 mm Hg). Here, the
Patm is the atmospheric pressure (760 mm Hg at
sea level )and 47 mm Hg is the partial pressure
of H2O in exhaled gas at body temperature. - During initial exhalation, the dead space (PCO2
0) reaches the sensor (phase 1), then a mixture
of dead space and alveolar gas (phase 2),and
finally, alveolar gas (phase 3). The alveolar
plateau occurs because the PCO2 in the air from
the different alveoli differs slightly. - The differences are larger (slope of alveolar
plateau steeper) in patients with lung disease. - The PETCO2 is an estimate of the mean alveolar
PCO2 (and, therefore, an estimate of the PaCO2). - Of note, there is a gradient between the PaCO2
and the PETCO2 (PaCO2PETCO2) with the PaCO2
being typically 2 to 5 mm Hg higher than the
PETCO2 in normal individuals. - In lung disease, the gradient can be much larger.
- In general, the PETCO2 is a valid estimate of
PaCO2 only if an alveolar plateau is present.
38Nonstructural risk factors
- Some nonstructural risk factors include obesity,
age, male sex, postmenopausal state, and habitual
snoring with daytime somnolence. - Familial factors Relatives of patients with SDB
have a 2- to 4-fold increased risk of OSA
compared with control subjects. - Environmental exposures include smoke,
environmental irritants or allergens, and alcohol
and hypnotic-sedative medications. - Both hypothyroidism and acromegaly are associated
with macroglossia and increased soft tissue mass
in the pharyngeal region. They are associated
with an increased risk of OSA. Hypothyroidism is
also associated with myopathy that may contribute
to UA dysfunction
39PETCO2
- In the mainstream method, the sensor is located
directly in the path of exhaled gas. - In the side stream method, gas is continually
suctioned through a tube to a more remote sensor
(in the instrument at bedside). - In the side stream approach, nasal cannulas are
used to suction exhaled gas from the nares . When
no CO2 is exhaled (during inspiration or apnea),
the nasal cannula suctions room air (PCO2 0). - In the side stream method, there is a delay in
exhaled gas reaching the sensor so the CO2
tracing is delayed compared with the exhaled
airflow . - The exhaled CO2 tracing is sometimes used to
indicate apnea (absence of exhaled PCO2).
However, this is not recommended for two reasons.
First, gas sampled by the nasal cannula may not
detect mouth breathing, and second, small
expiratory puffs rich in CO2 may still produce
deflections in the exhaled CO2 trace. - Capnography is used much more frequently during
pediatric than in adult sleep studies.
40Imaging Studies
- Modalities available for identifying the site of
obstruction include lateral cephalometry,
endoscopy, fluoroscopy, CT scanning, MRI. - At present, UA imaging is used primarily as a
research tool. Routine radiographic imaging of
the UA in the initial evaluation of SDB patients
is of uncertain benefit and should not be
performed unless a specific indication is
present.
41The exhaled CO2 tracing shows continued
deflections during inspiratory apnea due to
small exhaled puffs of air rich in CO2
42TcPCO2 MONITORING
- Measurement of TcPCO2 depends on the fact that
heating of capillaries in the skin causes
increased capillary blood flow and makes the skin
permeable to the diffusion of CO2. - The CO2 in the capillaries diffuses through the
skin and is measured by an electrode at the skin
surface. - The measured value is corrected for the fact that
heat increases the skin CO2 production as the
measured value exceeds the PaCO2 measured at 37C. - Typically, TcCO2 electrodes are calibrated with a
reference gas. - A thermostat controls the heating of the
membrane-skin interface. - It is usually recommended that the probe of most
TcPCO2 monitoring devices be moved every 3 to 4
hours to avoid skin irritation/damage. - The response time of newer TcPCO2 units has
improved, but in general, the measured PCO2 may
not increase rapidly enough to correlate with
short respiratory events. However,TcPCO2 can be a
good instrument for documenting trends in the
PCO2 during the night.
43Trends in the SpO2 and TcPCO2 during the night.
Note the simultaneous increase in transcutaneous
PCO2 and the decrease in SpO2 during episodes of
REM sleep.
44ACCURACY OF PETCO2 AND TcPCO2
- The measurement of PETCO2 and TcPCO2 was not
found to be accurate for determining changes in
PaCO2 during sleep . - PETCO2 was especially inaccurate during
simultaneous administration of supplemental
oxygen or during positive airway pressure
treatment as exhaled gas was sampled from a mask
rather than using a nasal cannula. - The AASM scoring manual in the respiratory
scoring rules for adults states that there was
insufficient evidence to recommend a specific
method for detecting hypoventilation during
sleep. However, it was stated that PETCO2 or
TcPCO2 may be acceptable if validated and
calibrated. In the pediatric rules, it states
that acceptable methods for assessing alveolar
hypoventilation are either transcutaneous or
end-tidal PCO2 monitoring.
45ACCURACY OF PETCO2 AND TcPCO2
- Concerning PETCO2 monitoring, the clinician
should review the exhaled CO2 tracings to
determine whether an alveolar plateau is present.
- If an alveolar plateau is not present, the PETCO2
value may not be an accurate estimate of PaCO2. - Other problems for exhaled PCO2 monitoring
include oral breathing and occlusion of the nasal
cannula with secretions. - If tidal volumes are very small, a true alveolar
sample may never reach the sensor. So PETCO2
value will likely be much lower than the PaCO2. - Concerning TcPCO2 measurement, the actual
tracings should also be carefully reviewed. If an
abrupt change (offset) in the TcPCO2 tracing is
noted, this suggests a measurement artifact is
present. - Most TcPCO2 devices require calibration at the
start of monitoring. - Poor application of the sensor or dislodgment of
the sensor during sleep can cause a measurement
artifact. - There is some advantage to the simultaneous use
of both PETCO2 and TcPCO2 if tolerated. If the
values show reasonable agreement during periods
of stable breathing, this increases confidence in
their validity. Of course, the goal standard to
validate the accuracy of their measurements is a
simultaneous ABG measurement.
46SNORING SENSORS
- Snoring is a sound produced by vibration of upper
airway structures. - When snoring is present, upper airway narrowing
of some degree can be inferred. - Snore sensors are usually microphones or PE
transducers that are usually applied to the neck
near the trachea. - Microphones can also be attached to the upper
chest area or the face. - Snoring can also be seen in the NP signal as a
rapid oscillation in the pressure tracing if an
appropriate high frequency filter setting (100
Hz) is used and the transducer is sufficiently
sensitive. - The AASM scoring manual does not provide guidance
on use of snoring sensors and the signal is not
part of the scoring criteria for respiratory
events in adults. - Snoring is mentioned in scoring of RERAs in
children.
47Snoring noted both in the snore sensor (applied
to the neck near the trachea) and as a vibration
(oscillation) in the nasal pressure signal. Note
that the nasal pressure signal also has a
flattened shape. SpO2 pulse oximetry
48(No Transcript)