Title: Immobile Ocular Perfusion Pump


1
Immobile Ocular Perfusion Pump
  • Kelly Clause, Senthil Premraj, Brad Lomago
  • Puwat Charukamnoetkanok

2
Background
  • Ocular perfusion pumps have been implemented
    outside of the ICU and surgical settings in which
    self-lubrication or manual lubrication is not
    available or convenient.
  • Our pump is immobile and is to be used in the ICU
    or surgical setting to decrease the time nursing
    personnel would spend lubricating the eyes of
    patients who cannot perform this function on
    their own.

3
Personal Responsibilities
  • Components
  • Fluid pump/ reservoir
  • Delivery Mechanism
  • Spray Mechanism

4
Spray mechanism
  • Based on experiments done in Dr. Kamenevas lab
  • There is a minimum of 8 ml/min flow rate needed
    to have the mechanism spray
  • This equates to a 60 mmHg pressure.

5
Delivery mechanism
  • Using Hagen Poiseuille determined the inner
    diameter of the sprayer was 0.06 inches
  • The diameter needed of the delivery mechanism,
    silicone tubing, is being calculated based on the
    height needed to maintain the 60 mmHg pressure
    and the varying widths possible.

6
Fluid pump
  • Because the 8 ml/min flow rate is much higher
    than the amount of fluid we are going to be
    spraying, and is too small for any researched
    fluid pumps on the market for based on the 0.06
    in. diameter
  • Based on experimentation with the spray
    mechanism, and silicone tubing with an inner
    diameter of 3/8 in, done in Dr. Shroffs lab, we
    concluded that gravity was enough to maintain the
    pressure needed and that we Do not need a fluid
    pump

7
Personal Accomplishments
  • Components
  • Fluid pump/ reservoir do not need, reservoir
    will be IV bag able to hold at least 50 mL
  • Delivery Mechanism still experimenting on the
    relationship of size to height to maintain the 60
    mmHg needed
  • Spray Mechanism determined the characteristics
    of it further experimentation on the size of the
    droplets and area of spray when the nosepiece is
    designed