Title: Dr. Thomas Winkler
1Background Information
pH-Electrodes
- Dr. Thomas Winkler
- ProMinent Dosiertechnik Heidelberg
2Definition pH
acid
pH - log a H
basic
3Meaning of pH-value
Change of pH-value by 1 step Change of H
concentration by factor 10
4Exampes for pH-values
5Constructions of pH-Electrodes
mV
mV
- 1 glass membrane
- 2 internal electrolyte
- 3 internal electrode
- 4 reference electrolyte
- 5 external electrode
- 6 diaphragm
reference electrode
glass electrode
combination probe
separate probe system
6pH-Measurement
Potentiometry Measurement of the difference of
potentials
Ag/AgCl/KCl/pHconst / /pHsample/KCl/AgCl/Ag
glass membrane
saturated KCl solution
- The signal is a voltage mV
- U R I
- The measurement is performed
- at high resistance (R)
- at nearly zero current (I)
glass membrane
sample
7pH-Measurement
Dependence of Electrode Signal on pH
the measured signal slope theoretical
(Nernst) 57.2 mV/pH (20C) 59.2 mV/pH
(25C) 61.2 mV/pH (30C) zero point (in
practice) /- 0.5 pH /- 30 mV
failure at low pH
by experiment (practise) theoretical
(Nernst)
zero deviation (Asymmetry-Potential)
failure at high pH
8Dependence of Electrode Signal on Temperature
pH-Measurement
9General Features of pH-Electrodes
- glass membrane (H- or pH-sensitive)
- active gel (swelling) layer exchange H/Na ions
- aging starts when electrode was manufactured
- aging depends on temperature and extreme pH
- for very low and very high temperatures special
glass required - glass should always be kept wet
- very high electric resistance typical 100 M?
(25C) - heavily temperature dependent
- 10-fold resistance when reducing temperature by
30 degree
10Junction
The junction is the most critical part of the
pH-probe beside the glass membrane!
ceramic junction
PTFE-ring junction
open annular gap
The junction should be a good electrical
connection, but should let pass minimum
electrolyte
The electrical resistance of a ceramic junction
is approx. 3 kO
11pH-Measurement
diaphragm (junction medium - reference) 1
PHEP with ceramic diaphragm 2 PHER with
PTFE-ring diaphragm 3 PHEX with open
annular-gap diaphragm 4 PHED with double
diaphragm double junction
1
2
3
4
12pH-Measurement
- diaphragm (junction medium - reference)
- porous ceramic diaphragm
- constant quality is of utmost importance
- defined porosity has to be complied with
- typical electr. resistance 1 5 k?
- flow KCl typical 10 50 ?l/m WC per hour
- porous PTFE-ring diaphragm (PHER)
- perfect in sewage water requires partial
blockage, - since otherwise zero shift when bleeding
- open annular gap (PHEX)
- suspension, emulsion, soil content and high
conductivity
13Temperature dependence of pH-probes
- temperature dependence of the slope
- the slope increases with temperature
- -57.2 mV/pH with 20C, -61.2mV/pH with 30C
- the temperature dependence of the zero-point can
be neglected - with values deviant from pH 7 and temperature
variations - automatic temperature compensation recommended
- via measurement with Pt 100/1000
mV
50C
20C
pH 7
pH 10
pH 4
- mV
14Cleaning of pH-probes
- 1. Glass membrane never mechanically!
- lime, hydroxide (rust) with diluted hydrochloric
acid, e.g. 0.5 - org. deposits (oil, grease) with alkaline
cleaning agents, alcohol, acetone (acetone not
with PHEK) - protein-containing media (milk, cheese, meat)
with pepsin- hydrochloric acid solution (5 g/l
0.5) - 2. Ceramic diaphragm also mechanically
(carefully) - (e.g. with key file, nail file, sandpaper)
- Sometimes works wonders!
- sulfide (black) with thiourea hydrochloric acid
- inorg./org. deposits as above with acidic or
alkaline - cleaning agents, alcohol
15Checking of pH-probes
- Good probes reach the set value in buffers fast.
In natural water also stable indication - Checking of the zero point and slope
- switch the portable measuring instrument to Redox
/ORP (mV indication) - pH-probe in buffer 7 e.g. 10 mV
- pH-probe in buffer 10 e.g. 170 mV
- difference 160 mV/3pH 53.3 mV/pH
- good zero point (pH 7) 0 mV /- 30 mV
- good slope 53...60 mV/pH (25C)
- slope of new probes 55...59 mV/pH (25C)
16Storage of pH-probes
- In a holder in KCl-solution 3-molar
- in case of an emergency tap water, not DI-water!
- drying out may destroy the probe (e.g. PHEX)
- Storage at room ambient temperature
- perfect 10...30C avoid frost!
- No long storage time
- preferably max. ¼ year
- Caution in the service station car!
- avoid high temperatures in summer
- avoid low temperatures in winter
- Frost may destroy probes and may freeze buffers!
17(No Transcript)