Title: Quality Improvement Put into Practice
1Quality Improvement Put into Practice
- Carl Mottram, BA RRT RPFT FAARC
- Director - Pulmonary Function Labs
Rehabilitation - Assistant Professor of Medicine - Mayo Clinic
College of Medicine
2Case Presentation
- 31 y.o. female
- History of present illness
- Non-specific cough, tightness in throat and
episodic shortness of breath following URI - No wheezing noted by patient or on exam
- Exam normal other than obesity (BMI 38)
- LMD orders CXR and spirometry with diffusing
capacity
3Case Presentation
- CXR
- Spirometry DLCO
- Pre Post Pred
- FVC 2.10 2.11 62
- FEV1 0.89 1.36 31
- Ratio 42.4 64.5
- DLCO 8.0 30
- Impression Severe obstruction with a severe
reduction in DLCO. Some improvement with BD
4Case Presentation
- LMD Action Plan
- Orders a CT scan
- Referred to Mayo Clinic for further evaluation
5Case Presentation
- Outside CT negative
- Pulmonary, ENT, and GI consults scheduled
- Pulmonary physician
- Negative exam
- Lungs clear, patient had coughing spell during
exam, no wheezing or stridor noted - Questioned outside spirometry results and orders
PFTs
6Case Presentation
- Spirometry DLCO
- Pre Post Pred
- FVC 2.55 2.48 75
- FEV1 2.27 2.25 79
- Ratio 89 90.7
- DLCO 24.2 99
- Impression Borderline restriction most
likely 2? to obesity with no evidence of airflow
obstruction or BD response
7PFT results affect people!!!
- Further testing
- Labeling (COPD, Asthma, etc)
- Medicine
- Disability
8Guidelines and Standards
- American Thoracic Society
- 1987 Revised Spirometry Standards
- 1991 Reference Values Interpretation
- 1994 Revised Spirometry Standards
- 1995 Diffusing Capacity
- 1999 Guidelines for Methacholine and Exercise
Challenge Testing - ATS/ERS 2005 Series General Laboratory,
Spirometry, Diffusing Capacity, Lung volumes, and
Interpretation
9Guidelines and Standards
- American Association of Respiratory Care (AARC)
- Clinical Practice Guidelines (52)
- Spirometry
- Static lung volumes
- Plethysmography
- Diffusing Capacity
- Infant/Toddler Pulmonary Function Tests
10Guidelines and Standards
- American Thoracic Society
- ATS Pulmonary Function Laboratory Management and
Procedure Manual - Updated 2005
- www.thoracic.org
- Education
- Education Products
11CLSIs Quality System In Respiratory Care HS4-A2
12Evidence of Quality Testing
- Spirometry in Primary Care Practice
- 30 primary care clinics, 15 trained group /15
usual group - 3.4 in usual group and 13.5 in trained group
met ATS acceptability and reproducibility
criteria - 1,012 pt. tests, 2,928 blows (2.89)
- Eaton et al, Chest 1999 116416-423
13Evidence of Quality Testing
- Improving the Quality of Bedside Spirometry
- Audit of testing outside the PF lab - Cleveland
Clinic - 15 - ATS acceptability/reproducibility criteria
- CI Project - 63.5 acceptability/reproducibility
- Stoller JK. Orens DK. Hoisington E. McCarthy K.
Bedside spirometry in a tertiary care hospital
The Cleveland Clinic experience. Respiratory
Care. 47(5)578-82, 2002 May
14Evidence of Quality Testing
- Wanger J, Irvin C Resp Care 36 (12) 1991
- 13 hospitals, 7 different systems, 5 Bio-QC (3
men, 2 women) - DLCO CV 11.5 - 18.6 with the largest diff. 24
units
15Quality Improvement Put into Practice - Quality
Assurance
- Systematic approach of monitoring and
evaluating quality.
16Quality Improvement Put into Practice - Quality
Assurance
- CLSIs Path of workflow Model
- Pre-test
- Testing session
- Post-test
17Quality Improvement Put into Practice Pre-test
Quality Assurance
- Pre-test instructions
- Appropriate order
- Questionnaire
- Height and weight
- Networked systems
- Equipment quality assurance program
18Quality Improvement Put into Practice Pre-test
Quality Assurance
- Equipment quality assurance
- Validation/Verification
- Preventive maintenance
- Documentation and records (logbooks)
- Mechanical models
- Biological models
19Quality Improvement Put into Practice Pre-test
Quality Assurance
- Mechanical Model
- 3-liter syringe
- 0.5, 1-2, 6 second flows
- Leak checked
- Stored and used in such a way as to maintain the
same temperature and humidity of the testing site - Validated based on manufacturer recommendations
2005 ATS/ERS Standards Standardization of
Spirometry
20Quality Improvement Put into Practice Pre-test
Quality Assurance
- Mechanical Model - Plethysmography
- Validation using a known volume should be
performed periodically - Model lung with thermal mass to simulate
isothermal conditions of the lung. - Accuracy 50 ml or 3
2005 ATS/ERS Standards Standardization of Lung
Volumes
21Quality Improvement Put into Practice Pre-test
Quality Assurance
- Mechanical Model Dilution techniques
- Analyzer accuracy and linearity
- N2 washout Monthly, exhalation volumes should be
checked with the syringe filled with room air,
and inhalation volumes with the syringe filled
with 100 O2.
2005 ATS/ERS Standards Standardization of Lung
Volumes
22Quality Improvement Put into Practice Pre-test
Quality Assurance
- Mechanical Models DLCO
- Syringe DLCO weekly or whenever problems occur
- VA BTPS 3.3L
- DLCO Simulator or BioQC
2005 ATS/ERS Standards Standardization of DLCO
23Quality Improvement Put into Practice Pre-test
Quality Assurance
- Biological Model
- Normal laboratory subjects
- Two individuals (13)
- Establish mean and SD (minimum 20 samples)
24Quality Improvement Put into Practice Pre-test
Quality Assurance
- Biological Control - Plethysmography
- At least monthly or whenever errors are suspect 2
reference subjects (biologic controls) should be
tested
2005 ATS/ERS Standards - Standardization of Lung
Volumes
25Quality Improvement Put into Practice Pre-test
Quality Assurance
- Biological Control N2 washout
- At least monthly or whenever errors are suspect 2
reference subjects (biologic controls) should be
tested
2005 ATS/ERS Standards - Standardization of Lung
Volumes
26Quality Improvement Put into Practice Pre-test
Quality Assurance
- Biologic Control He dilution
- At least monthly or whenever errors are suspect 2
reference subjects (biologic controls) should be
tested
2005 ATS/ERS Standards - Standardization of Lung
Volumes
27Quality Improvement Put into Practice Pre-test
Quality Assurance
- Biologic Control Diffusing Capacity
- At least weekly
- Or whenever errors are suspect
- Or whenever a calibration tank is replaced
2005 ATS/ERS Standards - Standardization of DLCO
28Quality AssuranceBiological Quality Control - PF
Lab
- Results Out of range
- Repeat with another technologist
- Second tech is within limits - record out of
range data - Second tech out of range - trouble-shoot and
document - BioQC1 ULN LLN SD CV
- FEV1 2.95 2.73 0.05 0.02
- FVC 3.62 3.35 0.07 0.02
- TLC (Pleth) 6.09 5.65 0.11 0.02
- DLCO 24.5 21.5 0.75 0.04
29Quality AssuranceBiological Quality Control -
DLCO
Model A versus B Mean difference 0.5
30Quality AssuranceSubject comparisons DLCO
Model A versus B - Mean difference 1.5
31(No Transcript)
32Quality Improvement Put into Practice Test
Quality Assurance
- Testing room environment
- Environmental interference
- Technologists performance training - QSE
Personnel - Second technologist
- Meeting ATS/ERS acceptability and repeatability
criteria (new guidelines)
33Quality Improvement Put into Practice Test
Quality Assurance - QSE Personnel
- Technologists
- Job qualifications
- Job descriptions
- Orientation
- Training
- Competency assessment
- Continuing education
- Performance appraisal
34Quality Improvement Put into Practice Test
Quality Assurance - QSE Personnel
- Competence Assessment
- Training and on-going performance evaluations
- NIOSH Spirometry Training Course
- cdc.gov/NIOSH/topics/spirometry
- AARCs Spirometry Training
- National Board for Respiratory Care
- CPFT and RPFT exams
35Quality Improvement Put into Practice Test
Quality Assurance
- Lung volumes - DLCO VA 500 ml larger than TLC -
??? - Technologist Driven Protocols
- Reference equations
36Quality Improvement Put into Practice Test
Quality Assurance
- Technologist Driven Protocols
- Flowcharting the process
37Quality Improvement Put into Practice Post-Test
Quality Assurance
- Maneuver selection
- Quality review by second technologist
- While in-house training may achieve the desired
goals, laboratory directors should strongly
consider the benefits of formal training programs
from outside providers. - Feedback to the technicians concerning their
performance should be provided on a routine basis
2005 ATS/ERS Standards General Laboratory
38Technician Training and Feedback Improve Test
Quality
GPA
Year
Lung Health StudyEnright Am Rev Respir Dis
1431215, 1991
39Quality Improvement Put into Practice Post-Test
Quality Assurance
- Turn-around time
- Average TRT lt1 day (15), 1-2 d (30), 3-4 d
(27), 5-6 d (15), gt7 d (3) - ATS PFL Registry Abstract AARC 2005, OF-05-037
- Electronic Medical Record
40Quality Improvement Put into Practice Does it
Work?
- Retrospective review of 18,000 consecutive pts.
at Mayo Clinic - Ninety percent of the patients were able to
reproduce FEV1 within 120 ml (6.1), FVC within
150 ml (5.3), and PEF within 0.80 L (12). - Enright PL. Beck KC. Sherrill DL. Repeatability
of spirometry in 18,000 adult patients. American
Journal of Respiratory Critical Care Medicine.
169(2)235-8, 2004 Jan 15.
41This is fine as far as it goes. From here on,
its who you know.