Title: Method%20Selection%20and%20Evaluation
1Method Selection and Evaluation
- D. Kefaya EL- Sayed Mohamed
- Prof. Of Clinical Pathology (Clinical Chemistry),
Mansoura University
2- Before a new test or methodogy is introduced into
the laboratory. - both managerial and techinal information must be
compiled and carefully considered. - The information should be collected from many
different sources,including manufacturer - sales representatives,colleagues,scientific
presentations,and the scientific literature.
3(No Transcript)
4(No Transcript)
5Linear range
6- O.D
- 1.1
- 1.0
- 0.9
- 0.8
- 0.7
- 0.6
- 0.5
- 0.4
- 0.3
- 0.2
- 0.1
-
- 0.0 200 400 600
800 - Glucose concentration (mg/dl)
7(No Transcript)
8(No Transcript)
9- Observed value
True value
X
ยต
RE
SE
TE
10(No Transcript)
11- Inaccuracy
- The difference between a measured value and its
true value. - Is due to the presence of systematic analytical
error (S.E.) - S.E. can be constant or proportional .
1230
Constant E
Ideal
Proprtional E
0
30
Reference Values
13- Inaccuracy
- Can be estimated from three studies
- recovery .
- interference .
- a comparison-of-methods study
14(No Transcript)
15- Example of a recovery study
- Sample preparation
- Sample 12.0 ml serum 0.1 ml H2O
- Sample 22.0 ml serum 0.1 ml 20 mg /dl calcium
standard - Sample 32.0 ml serum 0.1 ml 50 mg /dl calcium
standard -
Concentration -
- Calcium measured added
Recovered Recovery - Sample 1 7.50mg/dl
- Sample 2 8.35mg/dl 0.95mg/dl
0.85mg/dl 89 - Sample 3 9.79mg/dl 2.38mg/dl
2.29mg/dl 96
16-
- Calculation of recovery
- Concentration added standard concentration x Ml
standard -
Ml standard ml serum
- 20 x 0.1
20x 1 20 0.95 (Sample2) - 20.1
2.1x10 21 - conc. recovered conc. ( diluted test) conc. (
baseline) - 8.35-7.50.85
- recovery Conc. Recovered x100
- Conc. added
-
- 0.85 x 100 89 (
sample 2) - 0.95
17- ( b )Interference
- The interference experiment is used to measure
systematic errors caused by substances other than
the analyte
18-
Interference - An interfering material can cause systematic
errors in one of two ways - The material itself may react with the analytical
reagents. - Or it may alter the reaction between the analyte
and the analytical reagents
19-
Interference - The interference likes recovery except that the
substance suspected of interference is added to
the patient sample. - The concentration of the potentially interfering
material should be in the maximally elevated
range. -
- If an effect is observed its concentration should
be lowered to discover the concentration at which
test resultes are frist invalidated
20- Example of an Interference study
- Sample preparation
- Sample 1 1.0 ml serum 0.1 ml H2O (base line)
- Sample 2 1.0 ml serum 0.1 ml of 10 mg/dl
magnesium standard - Sample 3 1.0 ml serum 0.1 ml of 20 mg/dl
magnesium standard - Calcium
Measured Magnesium added
Interference - Sample 1 9.80mg/dl
- Sample 2 10.53 mg/dl
0.91mg/dl
0.73mg/dl - Sample 2 11.48 mg/dl
1.81mg/dl
1.68mg/dl
21- Calculation of interference
- Concentration added standard concentration x
- 10x 0.1 10
0..91mg/dl (sample2) -
1.1 11 - Interference conc. ( diluted ) conc. (
baseline) - 10.53 9.8 0.73 (sample2)
ml standard ml standard ml serum
22- Common interferences (eg., hemoglobin,
lipids,bilirubin,anticoagulants preservatives,
and so on)also should be tested. - Glick and Ryder have presented " interferographs
" for various chemistry instruments these are
graphs relating analyte concentration measured
versus interferent concentration
23- ( C ) comparison of methouds Experiment
- The best comparative method that can be used is
the reference method , which is a method with
negligible inaccuracy in comparison with its
imprecision .
24- Reference methods may be laborious and time-
consuming Because most laboratories are not
staffed and equipped to perform reference
methods, the results of the test method are
usually compared with those of the method
routinely in use.
25- Westgard et al27 and the NCCLS35 recommend that
at least 40 sample and preferably 100 samples, be
run by botli methods. - span the clinical range
- represent many different pathologic conditions.
- Duplicate analyses of each sample by each method.
26- If 40 specimens are compared , two to five
patient specimens should be analyzed daily for a
minimum of 8 days. - If 100 specimens are compared the comparison
study should be carried out during the 20-day
replication study.
27Regression line
Reference
28Example of Method selection and evaluation
(Glucose in Serum)
- Analytical Needs -
- Rapid procedure is needed at time out side the
regular working hours. - Sample Volume of 0.2 ml or less
- Analytical range of 0.0 to 500.0 mg/dl.
- Turn arround time 30 min.or less is needed
29- Quality goals-
- The medical decision levels of interest are
- 50 (hypoglycemia)
- 200 mg/dl (hyperglycemia)
- The decision level for screening is 140mg/dl is
not needed. - Precision goals are 1.5 mg/dl at 50
mg/dl -
5.0mg/dl at 200 mg/dl - Total error goals (TEA) are 6.0 mg/dl
at 50.0mg/dl -
20.0mg/dl at 200.0 mg/dl
30- Method selection -
- In kit form
- Can be set up on an existing laboratory
instrument - Requirements-
- Primary reference solutions are prepared for
calibration - Control materials
- Randomly selected clinical specimen.
31- Within Run Precision
- Analyze 20 aliquots of low abnormal control mean
SD 56.5 mg/dl 0.7 - Analyze 20 aliquots of moderatly high control
mean SD 182.6 2.1 mg/dl - SDs are less than the allowable error in non
automated procedure .
32- Day to Day precision
- Two Q.C pools are analyzed for 20 days mean SD
- SD 1.5 mg /50 mg glucose
- 5.0 mg/200mg glucose
33- Analytical Range -
- Glucose solutions prepared from a stock glucose
reference solution of 1000mg/dl - Duplicate analyses
- The plot shows excellent linearity to 600mg/dl.
which meets the specification for linearity to
500mg/dl. - The absorbance at zero glucose concentration is
the reagent blank value
34- O.D
- 1.1
- 1.0
- 0.9
- 0.8
- 0.7
- 0.6
- 0.5
- 0.4
- 0.3
- 0.2
- 0.1
-
- 0.0 200 400 600
800 - Glucose concentration (mg/dl)
35- Recovery-
- Two pooled sera are perpared as baseline
specimens 61mg/dl ,171mg/dl - concentrated glucose solution is prepared
(10,000mg/dl) - two different amounts of glucose are added to 9.6
ml of pooled sera
36Recovery-
171mg/dl 61mg/dl Total Nacl 0.15mol/l Glucese solution 10,000mg/dl
b b b a a a 10.0mL 10.0mL 10.0mL 400.0ul 300.0ul 0.0ul 0.0ul 100.0ul 400.0ul
37- Quadruplicate assays of A and B are performed and
take the mean - Subtract the original (basal) glucose levels
(61,171) from the mean of each quadruplicate -
- Recovered
glucose - Recovery
- Proportional error 100- recovery
Recovered glucose Glucose added X100
38Recovery-
Glucose Recovered mg/dl Glucose found Glucose added Glucose added Pooled sera
Glucose Recovered mg/dl Glucose found mg/dl ml/10ml Pooled sera
--- 159-6198 369 --- 97 391 61 159 457 171 268 562 0.0 100 400 0 100 400 0.0 100 400 0 100 400 A A A B B B
39- Interference
- The potential interference of visible abnormal
specimens is tested by - A analyzing a series of icteric , turbid,and
hemolyzed sera using the new method and an
established method know to be free of such
interferenc
40- The difference ( ) the two methods are 1
4 mg which is less than the allowable TEA. -
- B add a small amount of highly concentrated
solutions to pooled sera e.g. Ascorbic acid ,
sod. Salicylate, Trisodium
citrate, Heparin, disod. EDTA. - The difference must be lt TEA.
41Interference
METHOD FOR GLUCOSE METHOD FOR GLUCOSE INTERFERENCE INTERFERENCE
Test Comparative (mg/dl) Concentration Type ( Serum)
1135.6 976.4 1207.6 1407.9 1124.4 985.3 1223.2 1445.5 --- 6.5mg/dl 96mg/dl Turbid Normal Bil.(Icteric) Hb.(Hemo.) A660 0.45