Method%20Selection%20and%20Evaluation - PowerPoint PPT Presentation

About This Presentation
Title:

Method%20Selection%20and%20Evaluation

Description:

Method Selection and Evaluation D. Kefaya EL- Sayed Mohamed Prof. Of Clinical Pathology (Clinical Chemistry), Mansoura University – PowerPoint PPT presentation

Number of Views:133
Avg rating:3.0/5.0
Slides: 42
Provided by: ScOr180
Category:

less

Transcript and Presenter's Notes

Title: Method%20Selection%20and%20Evaluation


1
Method Selection and Evaluation
  • D. Kefaya EL- Sayed Mohamed
  • Prof. Of Clinical Pathology (Clinical Chemistry),
    Mansoura University

2
  • Before a new test or methodogy is introduced into
    the laboratory.
  • both managerial and techinal information must be
    compiled and carefully considered.
  • The information should be collected from many
    different sources,including manufacturer
  • sales representatives,colleagues,scientific
    presentations,and the scientific literature.

3
(No Transcript)
4
(No Transcript)
5
Linear range
6
  • O.D
  • 1.1
  • 1.0
  • 0.9
  • 0.8
  • 0.7
  • 0.6
  • 0.5
  • 0.4
  • 0.3
  • 0.2
  • 0.1
  • 0.0 200 400 600
    800
  • Glucose concentration (mg/dl)

7
(No Transcript)
8
(No Transcript)
9
  • Observed value
    True value

X
ยต
RE
SE
TE
10
(No Transcript)
11
  • Inaccuracy
  • The difference between a measured value and its
    true value.
  • Is due to the presence of systematic analytical
    error (S.E.)
  • S.E. can be constant or proportional .

12
  • Test Values

30
Constant E
Ideal
Proprtional E
0
30
Reference Values
13
  • Inaccuracy
  • Can be estimated from three studies
  • recovery .
  • interference .
  • a comparison-of-methods study

14
(No Transcript)
15
  • Example of a recovery study
  • Sample preparation
  • Sample 12.0 ml serum 0.1 ml H2O
  • Sample 22.0 ml serum 0.1 ml 20 mg /dl calcium
    standard
  • Sample 32.0 ml serum 0.1 ml 50 mg /dl calcium
    standard

  • Concentration
  • Calcium measured added
    Recovered Recovery
  • Sample 1 7.50mg/dl
  • Sample 2 8.35mg/dl 0.95mg/dl
    0.85mg/dl 89
  • Sample 3 9.79mg/dl 2.38mg/dl
    2.29mg/dl 96

16
  • Calculation of recovery
  • Concentration added standard concentration x Ml
    standard

  • Ml standard ml serum
  • 20 x 0.1
    20x 1 20 0.95 (Sample2)
  • 20.1
    2.1x10 21
  • conc. recovered conc. ( diluted test) conc. (
    baseline)
  • 8.35-7.50.85
  • recovery Conc. Recovered x100
  • Conc. added
  • 0.85 x 100 89 (
    sample 2)
  • 0.95

17
  • ( b )Interference
  • The interference experiment is used to measure
    systematic errors caused by substances other than
    the analyte

18

  • Interference
  • An interfering material can cause systematic
    errors in one of two ways
  • The material itself may react with the analytical
    reagents.
  • Or it may alter the reaction between the analyte
    and the analytical reagents

19

  • Interference
  • The interference likes recovery except that the
    substance suspected of interference is added to
    the patient sample.
  • The concentration of the potentially interfering
    material should be in the maximally elevated
    range.
  • If an effect is observed its concentration should
    be lowered to discover the concentration at which
    test resultes are frist invalidated

20
  • Example of an Interference study
  • Sample preparation
  • Sample 1 1.0 ml serum 0.1 ml H2O (base line)
  • Sample 2 1.0 ml serum 0.1 ml of 10 mg/dl
    magnesium standard
  • Sample 3 1.0 ml serum 0.1 ml of 20 mg/dl
    magnesium standard
  • Calcium
    Measured Magnesium added
    Interference
  • Sample 1 9.80mg/dl
  • Sample 2 10.53 mg/dl
    0.91mg/dl
    0.73mg/dl
  • Sample 2 11.48 mg/dl
    1.81mg/dl
    1.68mg/dl

21
  • Calculation of interference
  • Concentration added standard concentration x
  • 10x 0.1 10
    0..91mg/dl (sample2)

  • 1.1 11
  • Interference conc. ( diluted ) conc. (
    baseline)
  • 10.53 9.8 0.73 (sample2)

ml standard ml standard ml serum
22
  • Common interferences (eg., hemoglobin,
    lipids,bilirubin,anticoagulants preservatives,
    and so on)also should be tested.
  • Glick and Ryder have presented " interferographs
    " for various chemistry instruments these are
    graphs relating analyte concentration measured
    versus interferent concentration

23
  • ( C ) comparison of methouds Experiment
  • The best comparative method that can be used is
    the reference method , which is a method with
    negligible inaccuracy in comparison with its
    imprecision .

24
  • Reference methods may be laborious and time-
    consuming Because most laboratories are not
    staffed and equipped to perform reference
    methods, the results of the test method are
    usually compared with those of the method
    routinely in use.

25
  • Westgard et al27 and the NCCLS35 recommend that
    at least 40 sample and preferably 100 samples, be
    run by botli methods.
  • span the clinical range
  • represent many different pathologic conditions.
  • Duplicate analyses of each sample by each method.

26
  • If 40 specimens are compared , two to five
    patient specimens should be analyzed daily for a
    minimum of 8 days.
  • If 100 specimens are compared the comparison
    study should be carried out during the 20-day
    replication study.

27
  • Test method

Regression line
Reference
28
Example of Method selection and evaluation
(Glucose in Serum)
  • Analytical Needs -
  • Rapid procedure is needed at time out side the
    regular working hours.
  • Sample Volume of 0.2 ml or less
  • Analytical range of 0.0 to 500.0 mg/dl.
  • Turn arround time 30 min.or less is needed

29
  • Quality goals-
  • The medical decision levels of interest are
  • 50 (hypoglycemia)
  • 200 mg/dl (hyperglycemia)
  • The decision level for screening is 140mg/dl is
    not needed.
  • Precision goals are 1.5 mg/dl at 50
    mg/dl

  • 5.0mg/dl at 200 mg/dl
  • Total error goals (TEA) are 6.0 mg/dl
    at 50.0mg/dl

  • 20.0mg/dl at 200.0 mg/dl

30
  • Method selection -
  • In kit form
  • Can be set up on an existing laboratory
    instrument
  • Requirements-
  • Primary reference solutions are prepared for
    calibration
  • Control materials
  • Randomly selected clinical specimen.

31
  • Within Run Precision
  • Analyze 20 aliquots of low abnormal control mean
    SD 56.5 mg/dl 0.7
  • Analyze 20 aliquots of moderatly high control
    mean SD 182.6 2.1 mg/dl
  • SDs are less than the allowable error in non
    automated procedure .

32
  • Day to Day precision
  • Two Q.C pools are analyzed for 20 days mean SD
  • SD 1.5 mg /50 mg glucose
  • 5.0 mg/200mg glucose

33
  • Analytical Range -
  • Glucose solutions prepared from a stock glucose
    reference solution of 1000mg/dl
  • Duplicate analyses
  • The plot shows excellent linearity to 600mg/dl.
    which meets the specification for linearity to
    500mg/dl.
  • The absorbance at zero glucose concentration is
    the reagent blank value

34
  • O.D
  • 1.1
  • 1.0
  • 0.9
  • 0.8
  • 0.7
  • 0.6
  • 0.5
  • 0.4
  • 0.3
  • 0.2
  • 0.1
  • 0.0 200 400 600
    800
  • Glucose concentration (mg/dl)

35
  • Recovery-
  • Two pooled sera are perpared as baseline
    specimens 61mg/dl ,171mg/dl
  • concentrated glucose solution is prepared
    (10,000mg/dl)
  • two different amounts of glucose are added to 9.6
    ml of pooled sera

36
Recovery-
171mg/dl 61mg/dl Total Nacl 0.15mol/l Glucese solution 10,000mg/dl
b b b a a a 10.0mL 10.0mL 10.0mL 400.0ul 300.0ul 0.0ul 0.0ul 100.0ul 400.0ul
37
  • Quadruplicate assays of A and B are performed and
    take the mean
  • Subtract the original (basal) glucose levels
    (61,171) from the mean of each quadruplicate
  • Recovered
    glucose
  • Recovery
  • Proportional error 100- recovery

Recovered glucose Glucose added X100
38
Recovery-
Glucose Recovered mg/dl Glucose found Glucose added Glucose added Pooled sera
Glucose Recovered mg/dl Glucose found mg/dl ml/10ml Pooled sera
--- 159-6198 369 --- 97 391 61 159 457 171 268 562 0.0 100 400 0 100 400 0.0 100 400 0 100 400 A A A B B B
39
  • Interference
  • The potential interference of visible abnormal
    specimens is tested by
  • A analyzing a series of icteric , turbid,and
    hemolyzed sera using the new method and an
    established method know to be free of such
    interferenc

40
  • The difference ( ) the two methods are 1
    4 mg which is less than the allowable TEA.
  • B add a small amount of highly concentrated
    solutions to pooled sera e.g. Ascorbic acid ,
    sod. Salicylate, Trisodium
    citrate, Heparin, disod. EDTA.
  • The difference must be lt TEA.

41
Interference
METHOD FOR GLUCOSE METHOD FOR GLUCOSE INTERFERENCE INTERFERENCE
Test Comparative (mg/dl) Concentration Type ( Serum)
1135.6 976.4 1207.6 1407.9 1124.4 985.3 1223.2 1445.5 --- 6.5mg/dl 96mg/dl Turbid Normal Bil.(Icteric) Hb.(Hemo.) A660 0.45
Write a Comment
User Comments (0)
About PowerShow.com