Title: INJEX Pharma GmbH
1INJEX Pharma GmbH
Needle-Free Injection Technology
INJEX Operation
2Preparation of Transporter (I)
- turn the green knurled nut of the Transporter,
as shown, anticlockwise, until the cartridge can
be placed in the Transporters compartment.
Transporter
3Preparation of Transporter (II)
- take the cartridge and put it into the
Transporters compartment.
cartridge
4Preparation of Transporter (III)
- for a safe and firm seat of the cartridge, turn
the green knurled nut of the Transporter
clockwise, as shown, until the plunger softly
touches the rubber end of the cartridge.
5Preparation of Transporter (IV)
- take the Transporter Adapter out of its
sterile packaging.
- screw the Transporter Adapter on to the
Transporter. Ensure that the - adapter needle pierces the rubber seal of the
cartridge.
6Filling of Ampoule (I)
- take an INJEX-Ampoule out of its sterile
packaging and press the plunger completely into
the INJEX-Ampoule to force the air out of the
Ampoule.
Ampoule
plunger
7Filling of Ampoule (II)
- open the protective cap of the Transporter
Adapter.
- screw the INJEX-Ampoule, clockwise, as shown,
into the open end of the Adapter until it fits
tightly.
8Filling of Ampoule (III)
- turn the green knurled nut at the upper end of
the Transporter, clockwise, as shown, until the
desired dose of the pharmaceutical agent is in
the INJEX-Ampoule.
9Filling of Ampoule (IV)
- unscrew the filled INJEX-Ampoule from the Adapter
in an anticlockwise direction and close the
protective cap.
10Preparation of the Injector (I)
- take the loaded and safeguarded Injector out of
the Reset Box.
Injector
Reset-Box
11Preparation of the Injector (II)
- screw the filled INJEX-Ampoule clockwise into the
open end of the Injector (the last round goes a
little bit harder, this is normal).
12Preparation of the Injector (III)
- push the green safety ring towards the Ampoule
into Safe off position. Now the Injector is
ready for use.
13The Injection
- hold the Injector under firm pressure
perpendicularly to the injection site (i.e.
abdomen), until a clear indentation forms on the
skin. Press the trigger and release the
injection. Hold the Injector on the injection
site for another 1 2 sec.
14After the Injection (I)
- unscrew the INJEX-Ampoule and discard it
domestic waste -.
15After the Injection (II)
- push the green safety ring back into the Safe
position. - store the secured and non-reset Injector and its
accessories in the INJEX Case.