Title: Interface Gold Standard Calibration System
1Interface Gold Standard Calibration System
2 INTERFACE
- Developed in 1989, Interface, Inc. began to
market the Gold Standard System in 1991. It is
now used in more than 250 labs throughout the
world.
3Gold StandardTM Load Cell Calibration System
INTERFACE
- One 9840 2 Channel Indicator for Load Cell
Exitation/ Signal Conditioning - One ICS-202 Force Calibration Software
- One SIS-103 One-Day Training at Interface
- One CT-139-10 SCB1 to GS Load Cell Cable
- One CT-142-10 SCB1 to Standard Load Cell Cable
- One CX-0440 mV/V Transfer Standard
- One SIS-104 System Setup
4Interface Gold StandardTM Calibration Systems
INTERFACE
- Controlled calibration process reduces
calibration times by 50 to 90 and produces
consistent calibration results - Easy-to-use software reduces operator training
time - Fully integrated reporting functions
- Automatic archiving of calibration results
- Supports ASTM E74, ASTM E4 and ISO 7500/1,
EH10002 - Industry Accepted terms and calculations
- 0.1µV display resolution
- Used in over 250 calibration labs world-wide
5Gold StandardTM Calibration System Options
INTERFACE
- GS-USB Provides USB interface for HRBSC, SCB1,
DA-101 and HL4IO instrument boards includes
internal power supply, cooling fan and USB cable. - SCB1 20 bit instrument board provides signal
conditioning for the Gold StandardTM System.
Available in single or dual channel versions. - HRBSC High-resolution 22 bit instrument board
increases accuracy over the SCB1. - DA-101 Digital to analog board for generating
command signal for system control. Provides
setpoint signal for servo controlled hydraulic
force testing. - HL4IO High level 4-channel input board allows
calibration of 5 and 10 VDC output transducers
and includes all features of the DA-101. - Load Cell Simulator mV/V transfer standards are
used to calibrate the Gold StandardTM system, and
for setting up and testing load cell readout
devices. CX-0404 - specifically designed for
instrument substitution testing as per ASTM E74. - Desktop, laptop or portable PC Contact factory
for currently available configurations. - Gold StandardTM or Platinum StandardTM load cells
as required.
61600 Gold StandardTM Calibration Load Cell
INTERFACE
- Tension and compression in one unit
- .005 nonrepeatability
- .01 creep
- High output - to 4mV/V
- High precision base installed
- 3 run NIST traceable calibration
- Factory installed calibration adapter
- Eccentric load compensated
- .0008/F temp. effect on output
- 4 lower load limit per ASTM E74
71600 Gold StandardTM Calibration Compression Only
Load Cell
INTERFACE
- .005 nonrepeatability
- .01 creep
- High output - to 4mV/V
- High precision base installed
- 3 run NIST traceable calibration
- Eccentric load compensated
- .0008/F temp. effect on output
- 4 lower load limit
8Model 1800 Platinum SeriesLow Profile Load Cells
INTERFACE
- .005 non-repeatability
- .01 Creep
- Designed to be used at 2 lower load limit
- .008 /F temp effect on output
- Eccentric load compensated
9Model 9840 Intelligent Indicator
INTERFACE
- 24 bit internal resolution
- /-999,999 Count Bipolar Display
- Nonlinearity lt ? 0.005
- Stores calibration for up to 25 load cells
- Fast, direct analog output
- ? VDC scalable analog output
- Full duplex RS232C
- Peak/Valley hold
- Front panel and remote tare
10Precision mV/V Load Cell Simulator
INTERFACE
- Most accurate load cell simulator.
- Special low thermal
- emf construction.
- Each unit individually calibrated,
- aged and recalibrated.
- Strong, rugged design.
- Instrument substitution testing.
- MDL CX-0404 is specifically designed for
instrument substitution testing as per ASTM E74.
11Why use a simulator or mV/V Standard ?
- Calibrate Indicator or Signal Conditioner
periodically - Daily sanity check on measurements
- Establish true
- zero
- Check cables
11
12Model CX Standard
- 6 digit calibration
- 10 ppm uncertainty
12
13Shunt Calibration
- Technique of unbalancing bridge with external
shunt resistor - Less Expensive
- OK for medium-accuracy work
- Affected by preload, temperature, bridge
resistance drift, cable length
13
14Transducer Calibration Procedures
14
15Machine Calibration Procedures
15
16Measurement Uncertainty Model, Calibrating Load
Cell by Comparison
16
17Uncertainty Components
- CNS Cal of Standard Transducer by NIST
- DS Drift in Standard since last cal
- PS Creep in Standard Transducer
- TS Temperature effect on output of Standard
- RHF Nonreproducibility of frame, load string
- CH Calibration of HRBSC Instrument
- DH Drift in HRBSC since last cal
- TH Temperature effect on HRBSC Instrument
- LH Nonlinearity of HRBSC Instrument
- NH Noise of HRBSC Instrument
17
18CH Uncertainty in calibration of HRBSC
- (rectangular
distribution) - Source of Uncertainty Type
Uncertainty (RDG)
- CCX Calibration of CX-0610 B
0.0008 - DCX Drift in transfer function of
- CX-0610 between calibrations B
0.0010 - TCX Temperature effect on
- CX-0610
B 0.0054 - NCX Noise of the HRBSC
- instrument during calibration
- with the CX-0610 B
0.0003 -
------------ - Combined (Root Sum of Squares)
0.0056
18
19Uncertainty Budget Example
19
20Dominant Uncertainty Components
- Nonreproducibility of load frame, threads,
adaptors, alignment - Calibration of Standard at low end of range
- Drift in Standard since last cal
20
21Load Cell Calibration System Software - Opening
Screen
INTERFACE
22INTERFACE
DUT Description
23Setpoint Entry
INTERFACE
24Setpoint File Selection
INTERFACE
25Exercise Load Cells
INTERFACE
26Begin Calibration
INTERFACE
27Calibration
INTERFACE
28Completed Calibration
INTERFACE
29Error Plot
INTERFACE
30Curve Fit
INTERFACE
31Load Cell History
INTERFACE
32 INTERFACE
INTERFACE
33Report Designer
INTERFACE
34Custom Report
INTERFACE
35Gold StandardTM E4 Machine Calibration System
INTERFACE
- 9840-100-1 Intelligent Indicator
- One ICS-205 E4 Machine Calibration Software
- One CT-177-10 Gold Standardtm Interconnect Cable
Assemby - Serial Cable, Laptop to 9840
- One CX-0440 4mV/V Transfer Standard
- One SIS-104 System Setup
- Training at Interface
36 Features of the Portable E4 Machine
Calibration Software
INTERFACE
- Capable of providing the user with printed
copies of - E4-02 Test Machine Verification Report.
- ISO 7500/1 Test Machine Verification Report.
37Portable E4 System Software Opening Screen
INTERFACE
38Setpoint Entry
INTERFACE
39Setpoint File Selection
INTERFACE
40Meter Configuration
INTERFACE
41Machine Details
INTERFACE
42Load Point Entry
INTERFACE
43Load Point Entry
INTERFACE
44Begin Verification
INTERFACE
45Data Collection
INTERFACE
46Completed Verification
INTERFACE
47E4 Report
INTERFACE
48Verification History
INTERFACE
49ISO 7500-1 Report
INTERFACE
50ISO 7500-1
INTERFACE
51Report Designer
INTERFACE
52Custom E4 Report
INTERFACE
53Custom ISO Report
INTERFACE