Title: MLAB 2401: Clinical Chemistry Keri Brophy-Martinez
1MLAB 2401 Clinical ChemistryKeri
Brophy-Martinez
- Analytical Techniques and Instrumentation
- Electromagnetic Radiation Spectrophotometry
2Introduction
- How do we actually measure the concentrations of
molecules that are dissolved in the blood? - Spectrophotometry
Mix chemicals together to
produce colored products , shine a specific
wavelength of light thru the solution and measure
how much of the light gets absorbed - Nephelometry and Turbidimetry
Mix chemicals together to produce
cloudy or particulate matter , shine a light thru
the suspension and measure how much light gets
absorbed or refracted - pH Meters / Ion Selective Electrodes (ISE)
Electrically charged ions effect
potentials of electrochemical circuits - Electrophoresis
Charged molecules move at
different rates when pulled through an
electrical field - Osmometers
Dissolved molecules ions
are measured by freezing point depression and
vapor pressure
3Electromagnetic RadiationProperties of light
and radiant energy
- Electromagnetic radiation is described as photons
of energy traveling in waves - There is a relationship between energy and the
length of the wave (wavelength) - The more energy contained, the more frequent the
wave and therefore, the shorter the wavelength
4Electromagnetic RadiationProperties of light
and radiant energy
- This relationship between energy and light is
expressed by Planck's formula - E hf
-
- Where E energy of a photon
- h a constant
- f frequency
- The formula shows that the higher the frequency
the higher the energy or the lower the
frequency, the lower the energy - We do not use this to perform any calculations.
You only need to recognize Plancks formula and
its components
5Electromagnetic Spectra
6Electromagnetic Radiation Properties of light
and radiant energy
- White light
- Combination of all wavelengths of light
- Diffract (bend) white light and all the colors
become visible - The color you see depends on the wavelength of
color(s) that are not being absorbed - Light that is not being absorbed is being
transmitted
7Electromagnetic RadiationProperties of light
and radiant energy
- Wavelength
- Measured in nanometers (nm) or 10-9 meters.
8Electromagnetic Radiation Properties of light
and radiant energy
- Interactions of light and matter
- When an atom, ion, or molecule absorbs a photon,
the additional energy results in an alteration of
state (it becomes excited). Depending on the
individual species, this may mean that a
valence electron has been put into a higher
energy level, or that the vibration or rotation
of covalent bonds of the molecule have been
changed. - Ultimately, as energy is released, an emission
spectra is formed
9Electromagnetic Radiation (Properties of light
and radiant energy)
- In order for a ray of radiation to be absorbed it
must - Have the same frequency of the rotational or
vibrational frequency in the molecules it
strikes, and - Be able to give up energy to the molecule it
strikes.
10Electromagnetic Radiation
- Many lab chemistry instruments measure either the
absorption or emission of radiant energy /light. - Spectroscopy is based on the mathematical
relationship between solute concentration light
absorbance - Beers law
11Electromagnetic Radiation
- Beer's Law
- States the relationship between the absorption of
light by a solution and the concentration of the
material in the solution. - The absorption and/or transmission of light
through a specimen is used to determine molar
concentration of a substance.
12Beer-Lambert law (Beers Law)
13Beer-Lambert law (Beers Law)
14Requirements for Beers Law
- Keep light path constant by using matching sample
cuvettes standardized for diameter and thickness - Solution demonstrates a straight line or linear
relationship between two quantities in which the
change in one (absorption) produces a
proportional change in the other (concentration).
- Not all solutions demonstrate a straight line
graph at all concentrations. - If these rules are followed, we can calculate /
determine an unknowns concentration, by
comparing a characteristic (its absorbance) to
the same characteristic of the standard (whose
concentration is known by definition) - Concentration unk (Aunk /Astd)
Concentration std
15Percent transmittance
16Photometry/Spectrophotometry
- In photometry we measure the amount of light
transmitted through a solution in order to
determine the concentration of the light
absorbing molecules present within.
17Photometry/Spectrophotometry
- Types -Simple photometers and colorimeters use a
filter to produce light of one wavelength
(monochromatic light).
18Spectrophotometer / Spectrophotometry
- Spectrophotometers differ from photometers in
that they use prisms or diffraction gratings to
form monochromatic light.
19Spectrophotometer Components
- Light source/lamps
- Vary according to need, but must be a constant
beam, cool and orderly - Types
- Tungsten or tungsten iodide lamps for visible and
near infrared - Incandescent light (400 nm - 700 nm)
- Deuterium or mercury-arc lamps required for work
in U.V. range - Range 160-375 nm
20Spectrophotometer Components
- Monochromators
- Promote spectral isolation
- Operator selects specific wavelength
- Isolate a single wavelength of light
- Provides increased sensitivity specificity
- Types
- Glass filters
- Prisms
- Diffraction gratings
21Spectrophotometer Components
- Monochromator characteristic
- Bandpass/bandwidth
- Measures the success of the monochromator
- Defines the width of the segment of the spectrum
that will be isolated by the monochromator
22Spectrophotometer Components
- Cuvet
- Made of high quality glass or quartz
- Glass for work in the visible light range
- Quartz or fused silica for work in the UV range
- Shape
- Round cuvets are cheaper but light refraction and
distortion occur - Square cuvets have less light refraction but
usually more costly - Optically clean
- No inconsistencies in composition
- No marks, scratches, or fingerprints
- Positioning
- Orientation and placement into the instrument
important. Each time must be the same so light
passes through the cuvet at the same place.
23Spectrophotometer Component
- Photodetectors
- Purpose to convert the transmitted light into
an equivalent amount of electrical energy - Most common is the photomultiplier tube
24Spectrophotometer Component
- Readout devices
- Purpose to convert the electrical signal from
the detector to a usable form - Types
- Meters/Galvanometers
- Recorders
- Digital Readout
25Spectrophotometer Quality Assurance
- Wavelength calibration or accuracy is checked by
using special filters with known peak
transmission - Should be done periodically
- Must be done if a parameter, such as a change in
light / lamp has taken place. - Must be done if the instrument has been bumped or
traumatized. - Wavelength calibration verifies that the
wavelength indicated on the dial is what is being
passed through the monochromator.
26Spectrophotometer QA
- Stray light
- any wavelength of light reaching the detector,
outside the range of wavelengths being
transmitted by the monochromator. - Spectrophotometers must be periodically checked
for Stray Light - Causes insensitivity and linearity issues
- Resolve by cleaning optical system
27Spectrophotometer QA
- Linearity Check
- A linearity check is made by reading the
absorbance of a set of standard solutions
(obtained commercially) at specified
wavelength(s), or by using neutral density
filters - Produces a graph similar in appearance to
standard curve.
28Spectrophotometer Sources of Error
- Lamp burnout most frequent source of error
- Hours of use can be logged by system
- Watch for lamp to turn dark or smoky in color
- Monochromator error
- Poor resolution due to wide bandpass
- Results in decreased linearity and sensitivity
- Cuvet errors
- Dirt, scratches, loose cuvet holder - all cause
stray light - Air bubbles in specimen
-
29Spectrophotometer Sources of Error
- Reagent make-up
- some test procedures make a product that easily
foams - Volume too low for light path
- Electrical static (noise)
- Dark current - from the detector. Leakage of
electrons when no light passing through.
30Nephelometer
- Principle
- Measures scattered light
- Light bounces off insoluble complexes and hits
a photodetector - The photodetector is at an angle off from the
initial direction of the light. - This is a measure of Light Scatter
- Clinical Applications
- Protein measurements in serum, CSF,
immunoglobulins, etc.
- Most of the component parts are similar to those
of the spectrophotometer. - Major differences
- The position of the detector and
- reduces stray light
- Light source/beam LASER light
31References
- Bishop, M., Fody, E., Schoeff, l. (2010).
Clinical Chemistry Techniques, principles,
Correlations. Baltimore Wolters Kluwer
Lippincott Williams Wilkins. - Sunheimer, R., Graves, L. (2010). Clinical
Laboratory Chemistry. Upper Saddle River Pearson
.