Title: Visante OCT PIOL simultation study


1
Visante OCT PIOL simultation study
  • C. BUDO, Ass Prof MD
    University Clinic of Maastricht
  • Ophthalmology, Site Sint-Truiden
    camille.budo_at_skynet.be

Medical Mentor OPHTEC BV Consultant OPHTEC
BV Consultant CARL ZEISS
2
Study Objective
  • To compare safety and vault measurements in
    pre-operative Visante OCT images using the PIOL
    template with post-operative Visante OCT images
    of the same eyes using the prescribed PIOL.

3
Study Methods
  • Records of patients who had Phakic IOL
    implantation and had both pre and post-operative
    Visante scans were reviewed.
  • Scans of 56 eyes from 33 patients were qualified
    for inclusion.
  • For each eye, one pre-operative scan had the
    template placed by the Principal Investigator
    (CB).
  • The following measurements were made for both pre
    and post-operative eyes
  • Safety measurements at 0 degrees, center and 180
    degrees
  • Central vault measurements

4
Implantation simulation
5
Discussion
  • Safety measurements pre-op are smaller than
    post-op conservative approach template placed
    higher compared to actual PIOL position
  • Vault measurements pre-op are larger than post-op
    confirms high placement position of template
  • Medium correlation between predicted and actual
    vault may be due to natural physiological changes
    of the crystalline lens

6
CONCLUSION
  • First experience with the Visante OCT phakic IOL
    simulation tool is promising.
  • The results have shown high correlation between
    predicted clearance of the PIOL to the
    endothelium and the actual outcome.
  • There is a learning curve in placing the
    template. In this study the templates have been
    placed higher than the actual implant position.
  • The Visante OCT software enables the surgeon
    to optimise the template placement.