Cavendish Experiment - PowerPoint PPT Presentation

About This Presentation
Title:

Cavendish Experiment

Description:

Set Up Level the experiment using the threaded feet, making sure that the mirror is hanging freely in the center of the case, and center the pendulum in the middle of ... – PowerPoint PPT presentation

Number of Views:77
Avg rating:3.0/5.0
Slides: 24
Provided by: Microsc1
Learn more at: http://www.nhn.ou.edu
Category:

less

Transcript and Presenter's Notes

Title: Cavendish Experiment


1
Procedure I
  • Set Up
  • Level the experiment using the threaded feet,
    making sure that the mirror is hanging freely in
    the center of the case, and center the pendulum
    in the middle of the mesa.
  • Make sure large masses rotate easily.
  • Move the large masses through the full range of
    motion and verify that both masses touch the
    window of the case. Do this carefully not to
    cause large disturbances from hitting the glass
    case with the masses. If not make note of which
    mass doesnt touch.
  • Calibration
  • Using a strong magnet move the balance through
    the full range of motion and mark on the chalk
    board where the small masses touch the glass.
  • Measure halfway between the two maxima. Note
    distance between actual equilibrium (which should
    be measured) and center point. Use right
    triangles to determine the angle adjustment
    needed to center the actual equilibrium on the
    center point and adjust the torsion screw
    accordingly.
  • Measure the distance from the mirror to the
    chalkboard.
  • Taking Data
  • Move the large masses to where one is touching
    the glass. This will be the starting position
    for the measurement. Before taking data you must
    wait for the small masses to come to rest. The
    waiting time can be reduced by slowly bringing a
    strong magnet near one of the small masses,
    thereby damping the oscillations.
  • Once the balance has come to equilibrium,
    carefully turn on the laser pointer.

1
2
Procedure II
  • Making measurements
  • At t0, after the small masses have stabilized in
    Position I, make a mark to indicate initial
    position and switch the Masses to Position II
  • Make marks every 15sec for 45min for each
    position
  • Switch to Position I, and repeat the same process
  • Marking extrema may improve results
  • It may be helpful to write down the time every
    few minutes, to keep track of data when
    calculating
  • Calculating Equilibrium positions
  • Take two separate averages of all marks for
    positions I and II, making sure there are an
    equal number of maxima in each direction

2
3
Fall 2012 Equilibrium Data
  • The measurements resulted in G5.12x10-11
    m3/kg/s223 less than the accepted value
  • Given that the experimental error is within 16 ,
    this result is confusing. We do not have an
    explanation for it at this time.

3
4
Fall 2012 Constant Accel. Data
  • The measurements from Series 1 resulted in G
    5.23x10-11 m3/kg/s222 less than the accepted
    valueand the measurements from Series 2 resulted
    in G 6.16x10-11 m3/kg/s27.7 less than the
    accepted value.
  • Given that the experimental error is 10 and 9,
    respectively, the first result is confusing and
    the second is acceptable. We do not have an
    explanation for this at this time.

4
5
Error Discussion
  • There were giant problems with error
  • Constant acceleration method had 1951919335
    error
  • Equilibrium method had 117983750 to
    138881057 error
  • However, there are still errors beyond Pascos
    estimates
  • Possible sources of error include
  • The mirror is not planar, it is concave.
  • If mirror moves laterally, lasers incident angle
    will change.
  • If laser is not centered properly on the mirror,
    incident angle will not change linearly with
    mirror rotation.
  • Inaccuracies in measuring the equilibrium
    positions on the graphs
  • Definitely accounts for some of the error.
  • The separation of the large and small balls, b,
    is taken to be constant
  • it actually changes throughout the experiment.

5
6
Compare With Coulomb Experiment
A Note on Previous Presentations
  • Previous versions of this PowerPoint contained
    the following slide
  • Forces are much smaller
  • Typical Coulomb force 10-4 N (with V 6kV for
    both spheres, and distance 8cm)
  • Typical Cavendish force 10-9 N (with distance
    46.5 mm, m120g, m21.5kg
  • Both require a correction factor
  • Coulomb experiment requires a correction factor
    of 1/(1-a3/R3)
  • a is the radius of the sphere, R is the distance
    between spheres
  • This is because the sphere is not a point charge
  • Cavendish experiment requires a correction factor
    of 1/(1-b)
  • b is the distance between the spheres
  • This is because there is gravitational force
    between each small mass and both large masses,
    but only one is considered in the calculations.
  • This inserts a correction factor with no
    explanation of its derivation.

7
A Note on Previous Presentations
  • Fcorrection G m1 m2 / h2
  • h (2d)2 (b sin ?)21/2
  • Torquecorrection d x Fcorrection d
    Fcorrection sin ?
  • Torquegrav-corrected 2 F d 2 d Fcorrection
    sin ? ?? 2G m1 m2 d / b2 2G m1 m2 d /
    (2d)2 (b sin ?)2
  • G ?? / (2 m1 m2 d / b2 2 m1 m2 d / (2d)2
    (b sin ?)2 ? deltaS (4L)-1 (2 m1 m2 d / b2
    2 m1 m2 d / (2d)2 (b sin ?)2)-1 4 pi2 I
    deltaS (4LT2)-1 (2 m1 m2 d / b2 2 m1 m2 d /
    (2d)2 (b sin ?)2)-1 8 pi2 m2 (d2 2/5r2)
    deltaS (4LT2d)-1 (2 m1 m2 / b2 2 m1 m2 /
    (2d)2 (b sin ?)2)-1 pi2 (d2 2/5r2) deltaS
    / (L T2 d m1) / (1/ b2 1 / (2d)2 (b sin
    deltaS / (4L))2)

8
A Note on Previous Presentations
  • Using Mathematica to equate the result on the
    previous slide with the result derived earlier,
    the correction factor is (4 d2 b2
    SindeltaS/(4 L)2)/(b2 4 d2 b2
    SindeltaS/(4 L)2)So multiplying the original
    result by this factor gives the corrected result
    for G. Using the values stated earlier, this
    gives a correction factor equal to approximately
    0.83.
  • Since this is not negligible, we must include it
    in our final result.

9
Procedure I
  • Set Up
  • Level the experiment using the threaded feet,
    making sure that the mirror is hanging freely in
    the center of the case , and center the pendulum
    in the middle of the mesa.
  • Make sure that when the large masses are moved
    that the small masses only experience small
    oscillations.
  • Move the large masses through the full range of
    motion and verify that both masses touch the
    window of the case. Do this carefully not to
    cause large disturbances from hitting the glass
    case with the masses. If not make note of which
    mass doesnt touch.
  • Calibration
  • Using a strong magnet move the balance through
    the full range of motion and mark on the graph
    paper where the small masses touch the glass.
  • To center the natural equilibrium position, we
    would move one small tick mark. Then we would
    watch which direction the masses moved towards
    and would move it another small tick if the
    movement was away from the center. Initially, one
    could move 2 small tick marks if one was not near
    the center already. The angular variation in
    equilibrium points from switching the
    mass-positions was approx. 0.01 rad 0.8º. The
    total angular variation from maximums was 2.5º.
  • Measure the distance from the mirror to the
    midpoint between the marks where the small masses
    touch the glass.
  • Taking Data
  • Move the large masses to where one is touching
    the glass. This will be the starting position
    for the measurement. Before taking data you must
    wait for the small masses to come to rest. The
    waiting time can be reduced by slowly bringing a
    strong magnet near one of the small masses,
    thereby damping the oscillations.
  • Once the balance has come to equilibrium,
    carefully turn on the laser pointer.

9
10
Procedure II
  • Making measurements
  • At t0, after the small masses have stabilized in
    Position I, make a mark to indicate initial
    position and switch the Masses to Position II
  • Make marks every 15sec for 2min, then every 30sec
    for 30min, moving down a row for each precession
  • Switch to Position I, and repeat the same process
  • For better results, make marks every 15sec for
    45min for each position
  • Marking extrema may improve results ()
  • It may be helpful to write down the time every
    few minutes,
  • to keep track of data when calculating

10
11
Procedure III
  • Calculating Equilibrium positions
  • Two methods amplitude and frequency
  • For amplitude, take two separate averages of all
    marks for positions I and II, making sure there
    are an equal number of maxima in each direction
  • For frequency, average the marks closest to ¼ and
    ¾ the time of each period

11
12
Spring 2010 Equilibrium Data
  • The measurements resulted in G7.03 x 10-11 - 4
    greater than the accepted value
  • Given that the experimental error is within 10,
    these are excellent results
  • Note that the equilibrium position on the first
    data set is above the apparent value
  • This may be due to experimental error
  • Probably due to the method of calculating the
    equilibrium position

12
13
Spring 2010 Constant Accel. Data
  • The measurements resulted in G 8.68 x 10-11 -
    30 greater than the accepted value
  • Given that the experimental error is around 30,
    these results are very good

13
14
Error Discussion I
  • Neither method had serious problems with error
  • Constant acceleration method had 15 to 30
    error (ours was 30)
  • Equilibrium method had 10 to 20 error (ours
    was 4)
  • However, there are still errors beyond Pascos
    estimates
  • Possible sources of error include
  • The mirror is not planar, it is concave.
  • If mirror moves laterally, lasers incident angle
    will change.
  • If laser is not centered properly on the mirror,
    incident angle will not change linearly with
    mirror rotation.
  • Inaccuracies in measuring the equilibrium
    positions on the graphs
  • Definitely accounts for some of the error.
  • The separation of the large and small balls, b,
    is taken to be constant
  • it actually changes throughout the experiment.

14
15
Error Discussion II
  • Uncertainty in the b value given by apparatus
    manual
  • Value given for separation between masses in
    manual is a constant
  • b changed throughout experiment as arm rotated
  • The equilibrium points were not at the center
    between windows
  • At position 1, the equilibrium was .571 m (3.9o)
    from the center position
  • At position 2, the equilibrium was .469 m (3.2o)
    from the center position
  • Total change in b (window to center) in our
    experiment was 0.19 cm, or 0.4 of accepted value
    of b
  • Not a significant source of error

15
16
Procedure I
  • Set Up
  • Level the experiment using the threaded feet,
    making sure that the mirror is hanging freely in
    the center of the case , and center the pendulum
    in the middle of the mesa.
  • Make sure that when the large masses are moved
    that the small masses only experience small
    oscillations.
  • Move the large masses through the full range of
    motion and verify that both masses touch the
    window of the case. Do this carefully not to
    cause large disturbances from hitting the glass
    case with the masses. If not make note of which
    mass doesnt touch.
  • Calibration
  • Using a strong magnet move the balance through
    the full range of motion and mark on the graph
    paper where the small masses touch the glass.
  • To center the natural equilibrium position, we
    would move one small tick mark. Then we would
    watch which direction the masses moved towards
    and would move it another small tick if the
    movement was away from the center. Initially, one
    could move 2 small tick marks if one was not near
    the center already. The angular variation in
    equilibrium points from switching the
    mass-positions was approx. 0.01 rad 0.8º. The
    total angular variation from maximums was 2.5º.
  • Measure the distance from the mirror to the
    midpoint between the marks where the small masses
    touch the glass.
  • Taking Data
  • Move the large masses to where one is touching
    the glass. This will be the starting position
    for the measurement. Before taking data you must
    wait for the small masses to come to rest. The
    waiting time can be reduced by slowly bringing a
    strong magnet near one of the small masses,
    thereby damping the oscillations.
  • Once the balance has come to equilibrium,
    carefully turn on the laser pointer.

16
17
Procedure II
  • Making measurements
  • At t0, after the small masses have stabilized in
    Position I, make a mark to indicate initial
    position and switch the Masses to Position II
  • Make marks every 15sec for 2min, then every 30sec
    for 30min, moving down a row for each precession
  • Switch to Position I, and repeat the same process
  • For better results, make marks every 15sec for
    45min for each position
  • Marking extrema may improve results ()
  • It may be helpful to write down the time every
    few minutes,
  • to keep track of data when calculating

17
18
Procedure III
  • Calculating Equilibrium positions
  • Two methods amplitude and frequency
  • For amplitude, take two separate averages of all
    marks for positions I and II, making sure there
    are an equal number of maxima in each direction
  • For frequency, average the marks closest to ¼ and
    ¾ the time of each period

18
19
Spring 2010 Equilibrium Data
  • The measurements resulted in G7.03 x 10-11 - 4
    greater than the accepted value
  • Given that the experimental error is within 10,
    these are excellent results
  • Note that the equilibrium position on the first
    data set is above the apparent value
  • This may be due to experimental error
  • Probably due to the method of calculating the
    equilibrium position

19
20
Spring 2010 Constant Accel. Data
  • The measurements resulted in G 8.68 x 10-11 -
    30 greater than the accepted value
  • Given that the experimental error is around 30,
    these results are very good

20
21
Error Discussion I
  • Neither method had serious problems with error
  • Constant acceleration method had 15 to 30
    error (ours was 30)
  • Equilibrium method had 10 to 20 error (ours
    was 4)
  • However, there are still errors beyond Pascos
    estimates
  • Possible sources of error include
  • The mirror is not planar, it is concave.
  • If mirror moves laterally, lasers incident angle
    will change.
  • If laser is not centered properly on the mirror,
    incident angle will not change linearly with
    mirror rotation.
  • Inaccuracies in measuring the equilibrium
    positions on the graphs
  • Definitely accounts for some of the error.
  • The separation of the large and small balls, b,
    is taken to be constant
  • it actually changes throughout the experiment.

21
22
Error Discussion II
  • Uncertainty in the b value given by apparatus
    manual
  • Value given for separation between masses in
    manual is a constant
  • b changed throughout experiment as arm rotated
  • The equilibrium points were not at the center
    between windows
  • At position 1, the equilibrium was .571 m (3.9o)
    from the center position
  • At position 2, the equilibrium was .469 m (3.2o)
    from the center position
  • Total change in b (window to center) in our
    experiment was 0.19 cm, or 0.4 of accepted value
    of b
  • Not a significant source of error

22
23
Sources

http//en.wikipedia.org/wiki/Cavendish_experiment
http//www.nhn.ou.edu/johnson/Education/Juniorlab
/Cavendish/Pasco8215.pdf http//physics.nist.gov/c
uu/Constants/codata.pdf http//www.physik.uni-wuer
zburg.de/rkritzer/grav.pdf http//www.npl.washing
ton.edu/eotwash/publications/pdf/prl85-2869.pdf
23
Write a Comment
User Comments (0)
About PowerShow.com