Photon Dosimetry concepts and Calculations - PowerPoint PPT Presentation

About This Presentation
Title:

Photon Dosimetry concepts and Calculations

Description:

Photon Dosimetry concepts and Calculations Chapter 21 W/L Dose Calculation The treatment planning team has to quantify the overall prescribed dose of radiation and ... – PowerPoint PPT presentation

Number of Views:110
Avg rating:3.0/5.0
Slides: 48
Provided by: Katheri129
Category:

less

Transcript and Presenter's Notes

Title: Photon Dosimetry concepts and Calculations


1
Photon Dosimetry concepts and Calculations
  • Chapter 21
  • W/L

2
Dose Calculation
  • The treatment planning team has to quantify the
    overall prescribed dose of radiation and
    determine how much dose will be delivered over
    the time frame outlined.
  • There are many parameters of photon beam
    calculation.

3
Monitor Units
  • Monitor Units (MU) a measure of output for
    linear accelerators.
  • The dose rate varies slightly from one moment to
    the next
  • Normally the dose rate for the linear accelerator
    is 1.0 cGy/MU for a 10 x 10 field size defined at
    the isocenter.

4
Radiation Therapy Prescription
  • Radiation Therapy Prescription
  • Legal document defines the treatment volume
  • Intended tumor dose (TD)
  • Number of treatments
  • Dose per treatment
  • Frequency of treatment
  • States the type and energy of radiation
  • Beam-shaping devices such as wedges and
    compensators
  • Any other appropriate factors.
  • Clear, precise and complete

5
Isodose Plan
  • Isodose Plan part of the prescription
  • May show field sizes
  • Machine angels
  • Doses
  • Beam weighting
  • Wedges compensators or blocks

6
Treatment Time
  • Treatment time length of time a unit is
    physically left on to deliver a measured dose.
  • Factors considered
  • Beam energy
  • Distance from the source of radiation
  • Field size

7
Dose
  • Dose (absorbed dose) measured at a specific
    point in a medium and refers to the energy
    deposited at that point.
  • Measured in gray (Gy), which is defined so that 1
    Gy equals 1J/kg.

8
Depth
  • Depth distance beneath the skin surface where
    the prescribed dose is to be delivered.
  • Opposed fields patients midplane is used
  • Multiple field arrangements isocenter used
  • Depth affects measurements of dose attenuation.

9
Separation
  • Separation a measurement of the patients
    thickness from the point of beam entry to the
    point of beam exit.
  • Normally measured along the beams central axis.
  • Calipers
  • ODI readings

10
Source Distance
  • Source to Skin Distance (SSD) the distance from
    the source or target of the treatment machine to
    the surface of the patient.
  • Source-axis Distance the distance from the
    source of photons to the isocenter.

11
Setup
  • SSD Isocenter established at the patients skin
    surface
  • When the gantry rotates around the patient, the
    SSD will continually change.
  • Dose calculations often at DMAX. (Given dose)
  • SAD Isocenter established within the patient
  • The SAD and the isocenter are at a fixed distance
    and therefore do not change.

12
Isocenter
  • Isocenter the intersection of the axis of
    rotation of the gantry and the axis of rotation
    of the collimator for the treatment unit.
  • Cobalt 60 SAD of 80 cm
  • Linear Accelerators SAD of 100 cm

13
Field Size
  • Field Size the physical size set on the
    collimator of the therapy unit that determines
    the size of the treatment field at a reference
    distance (defined at the machines isocenter)
  • SAD the field size set inside the patient (size
    measured on patients skin will be smaller)
  • SSD field size set on the collimator will be the
    same measured at the patients skin

14
Scatter
  • Backscatter radiation that is deflected back
    toward the patient
  • Most of the absorbed dose received by the patient
    results from the collisions of the scattered
    electrons produced when the primary photon
    interacts with the collimator.

15
DMAX
  • DMAX the depth at which electronic equilibrium
    occurs for photon beams the point where dose
    reaches its maximum value.
  • Mainly depends on the energy of the beam
  • The depth of maximum ionization increases as the
    energy of the beam increases.
  • Factors such as field size and distance may also
    influence the depth

16
Photon Energy DMAX (cm)
Superficial 0.0
Orthovoltage 0.0
Cesium-137 0.1
Radium-226 0.1
Cobalt-60 0.5
4MV 1.0
6 MV 1.5
10 MV 2.5
15 MV 3.0
20 MV 3.5
25 MV 5.0
17
DMAX
  • DMAX dose occurs at the same depth for a given
    energy regardless of field size or distance from
    the source.
  • The actual reading differs for different field
    sizes.

18
Output
  • Output the dose rate of the machine, the amount
    of radiation exposure produced by a treatment
    machine or source as specified at a reference
    field size and at a specified reference distance.
  • Changing the field size, distance, or attenuating
    medium will change the dose rate.
  • Increases with field size primary component the
    same, increased scatter adds to the output
  • If the distance increases, dose rate decreases
    due to ISL

19
Cobalt-60 Output
  • Dose rate for Co-60 machine in cGy/min
  • Dose rate due to the radioactive decay of the
    isotope Co-60
  • Can be assumed constant over short periods of
    time
  • The Time that a machine is left On is adjusted by
    1 increase every month
  • TO2 TO1 x 1.01

20
Linac Output
  • Dose rate for Linac in cGy/MU
  • Dose rate varies from one moment to the next
  • Ionization chamber shuts down machine after
    predetermined dose has been given

21
Output Factor
  • Output Factor the ratio of the dose rate of a
    given field size to the dose rate of the
    reference field size
  • Allows for the change in scatter as the
    collimator setting changes
  • Relates the dose rate of a given collimator
    setting to the dose rate of the reference field
    size

22
Dose Rate
  • Commonly measured at the isocenter of the
    treatment machine.
  • The dose rate of the beam is inversely
    proportional to the square of the distance
  • Dose rate(Given f.s.) Dose rate(Reference f.s.)
    x Output factor(Given f.s.)
  • Dose Rate
  • Output (of machine)
  • Output factor (C.S. field size)
  • Scatter Ratio BSF or PSF (EFS/CS)
  • PDD(EFS)/100
  • Tray factor
  • Inverse square correction

23
Equivalent Squares
  • Equivalent Squares rectangular field size that
    demonstrate the same measurable scattering and
    attenuation characteristics of a square field
    size.
  • Used to find the output, output factor, and
    tissue absorption factors.
  • 4 (A/P)

24
Effective Field Size
  • Effective Field Size (EFS) the equivalent
    rectangular field dimension of the open or
    treated area within the collimator field
    dimensions when blocks are used to customize the
    shape of the treatment area.
  • The EFS equivalent square is normally used to
    determine the tissue absorption factors.

25
Tissue Absorption Factors
  • Tissue absorption factors different methods for
    measuring the attenuation of the beam as it
    travels through matter.
  • Percent Depth Dose (PDD)
  • works best with SSD setups
  • Tissue Air Ratio (TAR)
  • Tissue Phantom Ratio (TPR)
  • Tissue Maximum Ratio (TMR)

26
Percent Depth Dose (PDD)
  • Percent Depth Dose (PDD) the ratio expressed as
    a percentage, of the absorbed dose at a given
    depth to the absorbed dose at a fixed reference
    depth usually DMAX.
  • Calculated from two measurements at two different
    points in space.
  • Requires SSD be constant.
  • Written as PDD(d,s,SSD)
  • Dependant on
  • ? Energy- more penetrating- ? PDD
  • ? Depth- ? PDD due to attenuation through matter
  • ? Field size- more scatter- ? PDD
  • ? SSD- ? PDD- ISL

27
Percent Depth Dose (PDD)
  • PDD Dose _at_ d .
  • Dose _at_ Dmax

28
Tissue Air Ratio (TAR)
  • Tissue Air Ratio (TAR) the ratio of the absorbed
    dose at a given depth in tissue to the absorbed
    dose at the same point in air.
  • Dependant of
  • ?Energy- ?TAR
  • ?Field size- ? TAR
  • ?Depth- ? TAR
  • Independent of SSD

29
Tissue Air Ratio (TAR)
  • Calculated using two measurements at the same
    point in space
  • When the depth in tissue corresponds to the level
    of DMAX, the TAR is known as the backscatter
    factor.
  • Build-up cap device made of acrylic or other
    phantom material that is placed over an
    ionization chamber to produce conditions of
    electronic equilibrium.

30
Tissue Air Ratio (TAR)
  • TAR Dose in tissue
  • Dose in Air
  • Normally used to perform calcs for SAD
    treatments of low energy machines.
  • There is no patient backscatter in the in
    Air measurements.

31
Backscatter Factor (BSF)
  • When blocks are used.
  • Backscatter Factor (BSF) the ratio of the dose
    rate with a scattering medium to the dose rate at
    the same point without a scattering medium at the
    level of maximum equilibrium. (TAR at the level
    of DMAX)
  • PSF for megavoltage units.

32
Backscatter Factor (BSF)
  • The change in dose due to a change in field
    size.
  • Readings are made at Dmax and compared to the
    dose in air.
  • ?FS ? ?Backscatter ? ?BSF

33
Dose to a Point
  • The dose to any point in a medium is made up of
    two parts
  • Primary radiation the photons that originate in,
    and fan out from the source.
  • Scatter radiation other parts in the medium
    interact with the primary photons.

34
Scatter Air Ratio (SAR)
  • Scatter Air Ratio (SAR) the difference between
    the TAR for a field of definite area and that for
    a zero area.
  • The primary part of the total absorbed dose is
    represented by the zero area TAR.
  • A measure of the contribution from scattered
    radiation.

35
Tissue Phantom Ratio (TPR)
  • Tissue Phantom Ratio (TPR) the absorbed dose at
    a given depth in phantom to the absorbed dose at
    the same point at a reference depth in phantom.
  • The deeper the reference depth the greater the
    TPR.

36
Tissue Maximum Ratio (TMR)
  • Tissue Maximum Ratio (TMR) TPR at DMAX
  • TAR TMR x BSF

37
Transmission Factors
  • Transmission Factors the ratio of the radiation
    dose with the device to the radiation dose
    without the device and accounts for the material
    in the beams path.
  • Tray transmission factor
  • Wedge factor
  • Compensator factor

38
Other Factors
  • Tray transmission factor defines how much of the
    radiation is transmitted through a block tray.
  • Tray factors vary with beam energy- as energy
    increases, the effect of the material in the
    beams path is lessened because of the greater
    penetrating power of the higher energy.
  • Wedge attenuates the radiation beam
    progressively across a field

39
Given Dose
  • Given dose the dose at DMAX
  • AKA applied dose, entrance dose, peak absorbed
    dose, or DMAX dose
  • Point where PDD is 100
  • Given dose TD/ PDD x 100
  • TD tumor dose, dose at depth
  • Direct relationship
  • DoseA/PDDA DoseB/PDDB

40
Exit Dose
  • Exit dose the dose absorbed by a point that is
    located at the depth of DMAX at the exit of the
    beam.

41
Calculations
  • Field size variations, energy changes, and
    modifiers in the beams path can alter the amount
    of radiation received by the patient.
  • Time setting Dose at a point/Dose rate at that
    point
  • Dose at a point the prescribed dose as
    determined by the doctor
  • Dose rate at that point represents the dose rate
    of the treatment unit at the point of calculation

42
Calculations
  • When calculating dose to a given depth, all other
    points in the radiation field will be exposed to
    radiation for the same amount of time.
  • The dose to these points is dependant on their
    depth.

43
SSD Calculations
  • Output or dose rate of the machine should be
    expressed at the depth of DMAX
  • Field size defined at the skin surface
  • Dose rate measured in tissue at the depth of
    DMAX
  • Find equivalent square of the collimator setting
    (used for output factor)
  • Find equivalent square of the EFS (used for PDD)
  • Determine the PDD
  • Determine prescribed dose
  • Determine the treatment unit setting

44
Extended Distance
  • Extended Distance patient is set up at a
    distance beyond the isocenter or reference
    distance.
  • PDD is used because its nonisocentric
  • When MUs are calculated for setup at distances
    greater than the standard, ISL is used to account
    for the decrease in dose rates beyond the
    isocenter.

45
Inverse Square Law
  • Inverse Square Law describes the change in beam
    intensity caused by the divergence of the beam.
  • I1 / I2 D22 / D12
  • I1 original dose rate
  • I2 new dose rate
  • D22 original distance
  • D12 new distance

46
Mayneords Factor
  • Mayneords Factor a special application of the
    inverse square law.
  • Does not account for changes in scatter because
    of a change in beam divergence.
  • New PDD PDD x (SSD1 d / SSD1 DMAX)2 x (SSD2
    DMAX / SSD1 d)2
  • Inverse Square Factor
  • (Reference source calibration distance
  • Treatment SSD DMAX)2
  • SSD reference point typically at DMAX
  • SAD reference point at isocenter

47
TMR/TPR SAD
  • COF(C.S.) (collimator output factor) used to
    determine scatter, measured in air, from the
    collimators
  • The increase in the collimator opening, the more
    collimator surface area the photons will have to
    interact with.
  • PSF(EFS) (phantom scatter factor) used to
    determine scatter from the patient.
Write a Comment
User Comments (0)
About PowerShow.com