Automation - PowerPoint PPT Presentation

About This Presentation
Title:

Automation

Description:

Title: 1 Author: / Last modified by ** Created Date: 2/17/2006 4:27:24 PM Document presentation format: – PowerPoint PPT presentation

Number of Views:87
Avg rating:3.0/5.0
Slides: 31
Provided by: 20781
Category:

less

Transcript and Presenter's Notes

Title: Automation


1
Automation
Kefaya EL- Sayed Mohamed
Prof. of Clinical Chemistry, Mansoura University
2
Automation
  • Implies a self regulating process with an
    element of feed-back which detect any tendency to
    malfunction and readjusts the equipment so that
    it continues to function correctly.

3
Definitions and processing concepts
Continuous flow analyser
  • The reagent are pumped continuously by syring
    pump
  • The samples are introduced at regular intervals
  • The flow is segmented to separate one specimen
    from anothor
  • This type is used as batch analyzer e.g technicon
    .
  • Two types
  1. single channel continuous flow analyse one
    constituent in each specimen .
  2. Multichannel continuous flow determination of
    several components in each sample

4
Discrete analyser
  • Samples are processed in separate reaction tubes.
  • The instrument contain a variety of receptacle in
    which the sample and reagents are mixed.
  • The reaction occurs cuvettes, dry film slides.
  • Some containers contain the reagents and only the
    specimen need to be added.
  • Others add both reagents and specimen.

5
Reaction vessels and cuvets in discretes
  • Reused or discarded
  • Used as cuvet or just reaction tube, aspirated in
    tube then to another tube for measurement

For wash reused reaction vesseles or cuvets
  • aspiration of the reaction mixture
  • a detergent ,alkaline or acid wash soln . is then
    dispensed into and aspirated out of the cuvets

6
Parallel analysis
  • All specimens are subjected to a series of
    analytical processes at the same time in a
    parallel fashion
  • Discretionary multiple channel analysis.
  • The specimen can be analyzed by any one or by
    more than one of the available processes
    (methods, channels).

7
  • Batch analyzer.

- A number of specimens are processed in the
same analytical session , or run
8
(No Transcript)
9
The components steps in automated system
  1. Specimen identification.
  2. Specimen preparation.
  3. Specimen handling, transport and delivery.
  4. Specimen processing.
  5. Sample transport and delivery.
  6. Reagent handling and storage.
  7. Reagent delivery.
  8. Chemical reaction phase.
  9. Measurement approach.
  10. Signal processing data handling and process
    control.

These operations are usually performed
sequentially but in some instrument, they may be
combined and occur in parallel.
10
Specimen identification
11
Specimen Preparation
12
Specimen Handling, Transport and Delivery
13
Specimen processing
14
Sample transport and delivery
  • In continuous flow system the sample is
    aspirated through the sample probe into
    continuous reagent stream
  • In Discrete analyzers the sample is aspirated
    into sample probe and delivered, with reagent
    into reaction tube or cup .

15
Carry over
  • Carry over occurs when a previous sample or
    reagent contaminates successive tests in a run
    causing the next sample to have an aberrantly
    higher or lower results .
  • Carry over occurs when anlytes occur in extremely
    high levels e.g. enzyme or h CG in pregnancy
  • Carry over also occurs in systems that reuse
    cuvettes that are insufficiently washed after
    each testing cycle.

In discrete systems with disposable reaction
vessels and measuring cuvettes carry over is
caused by the pipeting system
16
(No Transcript)
17
(No Transcript)
18
(No Transcript)
19
(No Transcript)
20
(No Transcript)
21
Disadv one analyte per rotor but now recent
models allow multiple analysis 24 rotors can be
loaded at one time.
  • In Continuous flow by glass coils passing
    through the samples and reagents.

22
Measurement approaches
  • Spectrophotometry
  • Fluorimetry e.g. IMX
  • Chemiluminescenc e.g. IMMULITE
  • Electrochemical for electrolytes

23
Examples
24
(No Transcript)
25
Automated immunoassay systems differ from
chemical analyzers in that they require the use
of
  1. solid phase reagents to separate bound and
    unbound
  2. More sensitive detection methods e.g fluorescence
    and chemiluminescence
  3. Special handling of the reagents

e.g.
  1. Thermal conditioning to keep chemilum substrate
    stable.
  2. Constant agitation to keep microparticles
    suspended.

26
Automated immunoassay
4-Multiple Calibrators (upto 6)
  • Calibration usually nonlinear

Example ACS 180
  • benchtop
  • R-A
  • Imm . AS.S ay

IMMULITE
  • bench top
  • R-A
  • Imm. Assay

27
Selection of instruments
  • Role of an automated system in the workflow of
    the laboratory.
  • Identification of candidate system (RA or batch).
  • Cost effectiveness of the system.
  • User-friendiness (interface between the machine
    and the human operator).
  • Analytical performance.

28
Advantages of automation
  • Large number of samples may be processed with
    minimal technician.
  • Two or more methods may be performed
    simultaneously.
  • precision is superior to that of manual
    performance.
  • calculations may not be required.

29
Disadvantages of automation
  • limitation of the methodology "closed system".
  • Expensive.
  • Many systems, are impractical to use for small
    numbers of samples. Therefore, back up manual
    methods may be required for individual emergency
    analysis. Back up methods must also be available
    in the event of instrumental failures.

30
Thank you
Write a Comment
User Comments (0)
About PowerShow.com