Title: Determination of the Specific Surface
1?????
?????
Determination of the Specific SurfaceSolution
Adsorption ??
2Contents
- Purposes and Demands
- Principle
- Apparatus and Reagent
- Procedure
- Data Records and Processing
- Questions
- Attentions
- ??
3Purposes and Demands
1.Understand the principle of
solution adsorption method. 2. Learn how to
determine the specific surface of micro-silica
gel by the adsorption method with methylthionine
chloride solution.
4Principle
- The specific-surface that stands for general
surface of 1 g solid is an important
characteristic parameter of porosity powder
substance. It did good job in fields of
manufacture such as catalyze, chromatogram,
environmental protection and spinning and
scientific research.
5- We determine the specific surface of micro-silica
gel with the method of methylthionine-chloride
solution adsorption, since the tendency of
methylthionine chloride to be adsorbed by most of
the solid is the largest in dyestuffs which had
been found.
6- It presents the monolayer adsorption which is
coincident with the Langmuir adsorption
isotherm.
7- As stated in monolayer adsorption theory, when
the admolecules cover full on the surface of
adsorbent, the area of 1g adsorbent seized by
admolecule is that the admolecules number is
multiplied by the area of an adsorbed molecule.
8The total area of the solid is
Where S, A, M , NA, W, ?W, are respectively, the
specific surface with( cm2/g), the
cross-sectional area of methylthionine-chloride
molecule( 81.310-16cm2),
9- the molar mass of methylthionine-chloride(
373.9), Avogadro constant, the mass of silica
gel(g) and the mass of methylthionine-chloride
when silica gel comes to the saturated adsorption.
10- It is important to select a substance which has
the quite large specific surface in the
experiment. - Methylthionine-chloride has two adsorption peaks
in the visible light wavelength of 445nm or
665nm.
11- We Determine the absorbance change in
solution by the 722 type of spectrophotometer on
the foundation of the equation below
12????
Where C0 is the primary concentration of the
solution, C is the concentration of the saturated
solution after adsorption and V is the volume of
solution.
13Apparatus and Reagent
- 722-spectrophotometer 1
- Conessi-oscillator 1
- Volumetric flask(100ml) 8
- Iodine measuring flask(100ml) 1
- Spheric aspirator 1
- Pipette(10ml, 25ml, 50ml) 3
- Methylthionine-chloride solution(50010-3mg/ml)
- micro-silica gel
14Procedure
- 1. Determination of the specific surface
- ? Prepare the Methylthionine-chloride solution
of 5010-3mg/ml for reservoir liquor. - Pipette the 10ml liquid of 5010-3mg/ml to a
100ml volumetric flask and make up this solution.
15- ? Pipette the 50ml solution to a iodine
measuring flask and add 50mg of silica gel,then
surge it for half an hour. Pipette 25ml of this
solution and make up other solution in a 100ml
volumetric flask ,then determine the absorbance
and look up the concentration C' from the
standard curve.
16Protract of standard curve.
- Pipette the reservoir solution of respectively 1,
2, 3, 4, 5 and 6ml to six 100ml of volumetric
flasks and make up the terms of concentration of
standard solution. Determine the absorbance of
those solution at the biggest absorption
wavelength of 445nm or 665nm.
17????
18Data Records and Data Processing
- Look up C' from the standard curve based on
the absorbance by the step 1 and 2. - Figure out the values of C according to the
relation of C' and C
19- C(100/25)C'
- C' mg/ml
- C mg/ml
- C0 mg/ml
- ?W(C0-C)V10-3 g
20 Calculate the specific surface of silica gel.
NA6.021023 A cm2
M 373.9 W g
21Questions
- 1.Why should the adsorbed solution be diluted
before determination? - Whats the applied condition of Eq.1?
22Attentions
- Solution should be confected accurately.
23Outline for controlling 722-Spectrophotometer.
- 1. Regulate the sensibility at the lowest value,
when it is of the smallest amplificatory times. - 2. Electrify and warm-up for 20 min, set the
switch at T.
24- 3.Open the sample compartment, adjust 0Tto the
value of 0.000. - 4. Adjust the wave at appointed value.
- 5. Cover the door, adjust 100T to the value
of 1.000, if not, improve its sensibility.
25- 6. Determination of the absorbance.
- Set the switch at A and measure the
absorbance according to the steps above. - 7. Finally, clean the instruments, cut down the
electricity and cover the dustproof.