Highlights of Unit 3: Classification of mechanical ventilation - PowerPoint PPT Presentation

1 / 38
About This Presentation
Title:

Highlights of Unit 3: Classification of mechanical ventilation

Description:

Highlights of Unit 3: Classification of mechanical ventilation By Elizabeth Kelley Buzbee AAS, RRT-NPS, RCP Effects of increased baseline pressure Keep more air ... – PowerPoint PPT presentation

Number of Views:196
Avg rating:3.0/5.0
Slides: 39
Provided by: wwwappskc8
Category:

less

Transcript and Presenter's Notes

Title: Highlights of Unit 3: Classification of mechanical ventilation


1
Highlights of Unit 3 Classification of
mechanical ventilation
  • By Elizabeth Kelley Buzbee AAS, RRT-NPS, RCP

2
The parts of a mechanical ventilator
  • obtains power and converts this power into a
    force that can move gas into a patients lung.
  • Sends gas down a circuit to the patient interface
    and back to ventilator for analysis of data

3
Mechanical ventilators and the WOB
  • is dependent on the patients RAW,
  • his compliance
  • the volume required
  • and the elastic recoil of the lung

4
We classify ventilators by these questions
  • How does it get power to operate?
  • How does it use this power to drive gas into the
    patient and how does it control the flow of gas
    into the patient?
  • How does it control the various parameters of
    ventilation such as starting and stopping a
    breath?
  • How does it communicate information to the
    operator in such a manner that the RCP can
    monitor the patients responses and modify the
    ventilators action?

5
Input Power How does this machine get power to
operate?
  • electrical power A/C D/C 110-115 volt current
  • pneumatically-powered. 50-60 psig
  • Battery powered emergency only or transport
  • Internal batteries
  • External batteries
  • Run about one hour then require 8-12 hours to
    recharge

6
Power transmission and conversion How does it
use this power to drive gas into the patient?
  • Drive mechanisms
  • Compressors
  • piston-driven-
  • rotarydriven- delivers a Sine wave
  • linearly driven-- constant flow pattern .
  • Bellows start with constant flow but as
    pressures rise and RAW increases results in
    descending
  • spring,
  • a weight
  • gas pressure

7
Output control valves How does it control the
flow of gas into the patient?
  • microprocessors are tiny computers that do only
    one or two tasks
  • solenoid valves control flow to the patient by
    electronic switching
  • Electromagnetic
  • Pneumatic poppet valves
  • Proportional solenoid valves

8
Fluidic and pneumatic control
  • Fluidics use gas power, but differs from
    pneumatic in that there are no moving parts.
  • Coanda effect- gas moves along the side of the
    wall and we can direct gas to go down another
    tube by application of gas into that flow to move
    it
  • pneumatic control uses gas but there are moving
    parts- mushroom valves ect

9
Means of CommunicationHow does it communicate
information to the operator in such a manner that
the RCP can monitor the patients responses and
modify the ventilators action?
  • Monometers/
  • bourdon gauges- measure pressure
  • digital monometer may be displayed as a bargraph,
    or as numbers
  • Spirometers Volume measurements
  • Waveforms
  • alarms

10
Where do we set the alarm limits?
  • Apnea alarm adults 20 seconds when these alarms
    go off the apnea parameters on many ventilators
    will start breathing for the patient
  • Loss of electrical power alarm if battery
    operational will come on with indicator light
  • Loss of gas power alarm may ventilate patient
    will remaining gas
  • Disconnect alarms may have a time delay
  • Low or high humidifier temperature- keep at
    32-340 C high 37 max

11
Where do we set the alarm limits?
  • High/Low VE VE needs to stay within 10-20 /-
    or 2-5 LPM above and below
  • VT alarms 100 ml lower than set VT
  • Low airway pressure alarm- about 5-10 cmH20 below
    average PIP.
  • High airway pressure 10-20 above average PIP
    when this goes off, the breath will stop
    pressure-cycled

12
Where do we set the alarm limits?
  • Fi02 alarms -5 /-.
  • High/low rate alarms more than 10-20 from
    baseline
  • Loss of PEEP/CPAP alarms are generally set about
    3-5 below the PEEP

13
Control variables
  • How does the mechanical ventilator control the
    various parameters of ventilation such as
    starting and stopping a breath?

14
Open vs. closed loop control of ventilator output
  • open we dial in a VT or a f and the machine
    delivers the VT to the circuit. the open loop
    machine will not adjust.
  •  
  • In an closed loop system the ventilator is smart
    enough to monitor and interpret changes in such a
    way that the machine will alter the next breath
    to maintain the VT.

15
A control variable is the primary variable that
the ventilator manipulates to cause an
inspiration
  • Pressure controlled PC
  • Volume controlled VC
  • Flow controlled
  • Time usually based on the other parameters
  • Only one control, the other two will be variables

16
Pressure control
  • a PC pressure controlled breath is one in which
    the pressure stays the same, but changes in the
    patients condition will alter the delivered
    volume and the flow rate.
  • The doctor orders a PIP which will deliver a VT

17
Volume controlled
  • During VC ventilation, the PIP varies with
    changes in the patients conditions, while the
    volume and the flow stay constant.
  • The doctor orders a VT

18
Flow controlled
  • mechanical ventilators had consistent flow rate
    and volumes, but the airway pressure changed with
    patient parameter changes.
  • The doctor will order a VT but we will set up the
    flow rate and the Ti to deliver this VT

19
Phase Variables
  • What event triggers inspiration, what stops the
    breath, what changes the breath?

20
Phase variables
  • Trigger what starts the inspiratory phase?
  • Limit what limits the actual inspiratory cycle
    without stopping it?
  • Cycle what cycles the inspiratory phase off-
    starts exhalation?
  • Baseline what changes the base line pressures?
    PEEP or CPAP

21
What event triggers inspiration?
  • Time triggering. At 10 BPM, there is a breath
    initiated by the ventilator every 6 seconds
    cycle time
  • Patient triggered the inspiration is started by
    the patient demand. is called the Sensitivity.
  • pressure trigger
  • flow trigger
  • volume trigger
  • NAVA
  • Manual trigger push a button on the ventilator
    to trigger a breath used during suctioning

22
Pressure triggered
  • set the Sensitivity knob to -.5 to -1.5 cmH20.
  • if the Sensitivity is adjusted from -1 to -3, we
    say that the sensitivity is decreased the
    patients WOB is increased.
  • The patient creates a pressure gradient
  • If there is a leak in the system the pressure may
    not drop.
  • Complicated by having to drop the pressure all
    the way back to the ventilator
  • If baseline pressure rises, may not be able to
    pressure trigger
  • If there is auto-PEEP from air trapping, the
    pressure cannot drop enough to trigger a breath

23
Flow triggered
  • There is always a small constant flow moving
    through the circuit
  • 2 Pneumotachymeters measure and compare the flow
    coming to patient and going away from patient.
  • As the patient pulls in the gas, there is now
    less expiratory flow than inspiratory flow, and
    it is this flow gradient that will trigger a
    breath
  • Usual set 1-3 LPM in adults

24
Problems with flow triggers
  • Water in the circuit can mimic a breath and
    trigger more breathes than patient needs
  • Leaks can also alter the constant flow so that
    the machine may auto-cycle or chatter

25
Volume triggered
  • only the Drager Baby Log actually uses the volume
    inspired by the patient to trigger a breath

26
NAVA
  • Neutrally adjusted ventilatory assist
  • A probe is sent down the esophagus and as the
    phrenic nerve fires, the probes sensor notes the
    breath effort and triggers the ventilator.

27
What happens when triggering is not accurate or
responsive?
  • If not sensitive enough
  • Increased WOB
  • asynchrony with the ventilator
  • Fighting the ventilator
  • If too sensitive
  • Triggers too many breaths called chattering or
    auto-cycling
  • Could lead to air trapping and baratrauma

28
Limit what limits the actual inspiratory cycle?
  • A limit on a breath is some parameter that
    affects the breath without stopping it.
  • A pressure limit may mean that the patient
    continues to deliver the VT but the flow slows
    down in an attempt to keep the airway pressures
    down
  • The actual VT delivered is usually decreased, but
    still higher than it would be if the breath was
    pressure cycled off

29
IMPORTANT
  • many manufacturers use the term limit when
    discussing alarms.
  • If a high pressure alarm is set and the breath
    stops being delivered once that PIP limit is
    exceeded, it is not pressure limiting it is
    pressure cycling off.

30
Cycle
  • what parameter cycles the inspiratory phase off-
    starts exhalation?
  • Volume cycled-when preset VT is reached. Most VC
    breaths are also volume-cycled
  • Time cycled- the breathes initiated by the
    ventilator can be time triggered and maybe time
    cycled off. Most PC breaths will be time-cycled
    off
  • Pressure cycled- in VC ventilation, if the high
    pressure alarms goes off and the breaths stops we
    can say that the breath was pressure-cycled.

31
Flow cycling
  • flow cycle
  • Some ventilators will cycle off once a preset low
    flow rate is noted.
  • You can see this on the graphic when we watch the
    descending flow wave suddenly drops to zero
  • This occurs always with PS breaths and you might
    be able to chose flow cycling with the VC
    mandatory breaths

32
Baseline What changes the base line pressure?
  • We can raise the pressure during exhalation phase
    from zero to a positive number
  • we have raised the baseline

33
PEEP or CPAP?
  • Both raise the baseline pressure
  • Both used to treat refractory hypoxemia
  • Both will increase the FRC and can increase the
    lung compliance
  • PEEP positive end-expiratory pressure with a
    ventilator rate set full or partial support the
    lower pressure
  • CPAP- continues positive airway pressure
    without a ventilator rate set a spontaneous
    mode the only pressure

34
Effects of increased baseline pressure
  • Keep more air inside alveoli and airways
  • Raises RV residual volume which raises the FRC
  • Return the FRC to normal will generally increase
    the lung compliance and decrease the WOB
  • Excessive PEEP
  • hampers CO, increases VD and causes air trapping
    and can damage the lung tissue
  • If patient is on PC ventilation, raising PEEP
    might decrease the VT because the driving
    pressure drops.
  • Excessive CPAP can decrease VT which will raise
    the PaC02

35
With PEEP
  • small amount of gas are trapped because the
    exhalation valves closes before the circuit
    pressure drops back to zero.
  • How much gas is left in the lung is a function of
  • level of PEEP selected,
  • IE ratio if the exhalation time is too short
    more gas can trap
  • time constant of the lung units.

36
basic types of PEEP/CPAP devices
  • flow restrictor the exhalation port is too small
    for the gas to all escape. The faster the flow
    through the tiny hole, the more back pressure the
    flow restrictor creates
  • threshold resistor creates back pressure that is
    independent of flow rate. In these types of PEEP
    valves, the gases passes freely until some
    balances of forces on the other side equalize and
    the pressure is held in the circuit.

37
Types of threshold PEEP valves
  • Water Column 
  • Weighted ball
  • Flexed springs
  • Venturi diaphragm
  • Electromagnetic valve

38
When can a threshold resistor become a flow
restrictor?
  • When the flow is too fast for the exhalation
    valve to get all the gas out
  • Ventilator circuits are rated for their
    resistance to flow and a max flow rate will be
    suggested.
  • Failure to keep the flow rate below this max will
    result in PEEP rising as the flow rate rises.
  • This is a real issue with neonatal circuit that
    are so tiny that airway resistance rises quickly
Write a Comment
User Comments (0)
About PowerShow.com