Title: Highlights of Unit 3: Classification of mechanical ventilation
1Highlights of Unit 3 Classification of
mechanical ventilation
- By Elizabeth Kelley Buzbee AAS, RRT-NPS, RCP
2The parts of a mechanical ventilator
- obtains power and converts this power into a
force that can move gas into a patients lung. - Sends gas down a circuit to the patient interface
and back to ventilator for analysis of data
3Mechanical ventilators and the WOB
- is dependent on the patients RAW,
- his compliance
- the volume required
- and the elastic recoil of the lung
4We classify ventilators by these questions
- How does it get power to operate?
- How does it use this power to drive gas into the
patient and how does it control the flow of gas
into the patient? - How does it control the various parameters of
ventilation such as starting and stopping a
breath? - How does it communicate information to the
operator in such a manner that the RCP can
monitor the patients responses and modify the
ventilators action?
5Input Power How does this machine get power to
operate?
- electrical power A/C D/C 110-115 volt current
- pneumatically-powered. 50-60 psig
- Battery powered emergency only or transport
- Internal batteries
- External batteries
- Run about one hour then require 8-12 hours to
recharge
6Power transmission and conversion How does it
use this power to drive gas into the patient?
- Drive mechanisms
- Compressors
- piston-driven-
- rotarydriven- delivers a Sine wave
- linearly driven-- constant flow pattern .
- Bellows start with constant flow but as
pressures rise and RAW increases results in
descending - spring,
- a weight
- gas pressure
7Output control valves How does it control the
flow of gas into the patient?
- microprocessors are tiny computers that do only
one or two tasks - solenoid valves control flow to the patient by
electronic switching - Electromagnetic
- Pneumatic poppet valves
- Proportional solenoid valves
8Fluidic and pneumatic control
- Fluidics use gas power, but differs from
pneumatic in that there are no moving parts. - Coanda effect- gas moves along the side of the
wall and we can direct gas to go down another
tube by application of gas into that flow to move
it - pneumatic control uses gas but there are moving
parts- mushroom valves ect
9Means of CommunicationHow does it communicate
information to the operator in such a manner that
the RCP can monitor the patients responses and
modify the ventilators action?
- Monometers/
- bourdon gauges- measure pressure
- digital monometer may be displayed as a bargraph,
or as numbers - Spirometers Volume measurements
- Waveforms
- alarms
10Where do we set the alarm limits?
- Apnea alarm adults 20 seconds when these alarms
go off the apnea parameters on many ventilators
will start breathing for the patient - Loss of electrical power alarm if battery
operational will come on with indicator light - Loss of gas power alarm may ventilate patient
will remaining gas - Disconnect alarms may have a time delay
- Low or high humidifier temperature- keep at
32-340 C high 37 max
11Where do we set the alarm limits?
- High/Low VE VE needs to stay within 10-20 /-
or 2-5 LPM above and below - VT alarms 100 ml lower than set VT
- Low airway pressure alarm- about 5-10 cmH20 below
average PIP. - High airway pressure 10-20 above average PIP
when this goes off, the breath will stop
pressure-cycled
12Where do we set the alarm limits?
- Fi02 alarms -5 /-.
- High/low rate alarms more than 10-20 from
baseline - Loss of PEEP/CPAP alarms are generally set about
3-5 below the PEEP
13Control variables
- How does the mechanical ventilator control the
various parameters of ventilation such as
starting and stopping a breath?
14Open vs. closed loop control of ventilator output
- open we dial in a VT or a f and the machine
delivers the VT to the circuit. the open loop
machine will not adjust. -
- In an closed loop system the ventilator is smart
enough to monitor and interpret changes in such a
way that the machine will alter the next breath
to maintain the VT.
15A control variable is the primary variable that
the ventilator manipulates to cause an
inspiration
- Pressure controlled PC
- Volume controlled VC
- Flow controlled
- Time usually based on the other parameters
- Only one control, the other two will be variables
16Pressure control
- a PC pressure controlled breath is one in which
the pressure stays the same, but changes in the
patients condition will alter the delivered
volume and the flow rate. - The doctor orders a PIP which will deliver a VT
17Volume controlled
- During VC ventilation, the PIP varies with
changes in the patients conditions, while the
volume and the flow stay constant. - The doctor orders a VT
18Flow controlled
- mechanical ventilators had consistent flow rate
and volumes, but the airway pressure changed with
patient parameter changes. - The doctor will order a VT but we will set up the
flow rate and the Ti to deliver this VT
19 Phase Variables
- What event triggers inspiration, what stops the
breath, what changes the breath?
20Phase variables
- Trigger what starts the inspiratory phase?
- Limit what limits the actual inspiratory cycle
without stopping it? - Cycle what cycles the inspiratory phase off-
starts exhalation? - Baseline what changes the base line pressures?
PEEP or CPAP
21What event triggers inspiration?
- Time triggering. At 10 BPM, there is a breath
initiated by the ventilator every 6 seconds
cycle time - Patient triggered the inspiration is started by
the patient demand. is called the Sensitivity. - pressure trigger
- flow trigger
- volume trigger
- NAVA
- Manual trigger push a button on the ventilator
to trigger a breath used during suctioning
22Pressure triggered
- set the Sensitivity knob to -.5 to -1.5 cmH20.
- if the Sensitivity is adjusted from -1 to -3, we
say that the sensitivity is decreased the
patients WOB is increased. - The patient creates a pressure gradient
- If there is a leak in the system the pressure may
not drop. - Complicated by having to drop the pressure all
the way back to the ventilator - If baseline pressure rises, may not be able to
pressure trigger - If there is auto-PEEP from air trapping, the
pressure cannot drop enough to trigger a breath
23Flow triggered
- There is always a small constant flow moving
through the circuit - 2 Pneumotachymeters measure and compare the flow
coming to patient and going away from patient. - As the patient pulls in the gas, there is now
less expiratory flow than inspiratory flow, and
it is this flow gradient that will trigger a
breath - Usual set 1-3 LPM in adults
24Problems with flow triggers
- Water in the circuit can mimic a breath and
trigger more breathes than patient needs - Leaks can also alter the constant flow so that
the machine may auto-cycle or chatter
25Volume triggered
- only the Drager Baby Log actually uses the volume
inspired by the patient to trigger a breath
26NAVA
- Neutrally adjusted ventilatory assist
- A probe is sent down the esophagus and as the
phrenic nerve fires, the probes sensor notes the
breath effort and triggers the ventilator.
27What happens when triggering is not accurate or
responsive?
- If not sensitive enough
- Increased WOB
- asynchrony with the ventilator
- Fighting the ventilator
- If too sensitive
- Triggers too many breaths called chattering or
auto-cycling - Could lead to air trapping and baratrauma
-
28Limit what limits the actual inspiratory cycle?
- A limit on a breath is some parameter that
affects the breath without stopping it. - A pressure limit may mean that the patient
continues to deliver the VT but the flow slows
down in an attempt to keep the airway pressures
down - The actual VT delivered is usually decreased, but
still higher than it would be if the breath was
pressure cycled off
29IMPORTANT
- many manufacturers use the term limit when
discussing alarms. - If a high pressure alarm is set and the breath
stops being delivered once that PIP limit is
exceeded, it is not pressure limiting it is
pressure cycling off.
30Cycle
- what parameter cycles the inspiratory phase off-
starts exhalation? - Volume cycled-when preset VT is reached. Most VC
breaths are also volume-cycled - Time cycled- the breathes initiated by the
ventilator can be time triggered and maybe time
cycled off. Most PC breaths will be time-cycled
off - Pressure cycled- in VC ventilation, if the high
pressure alarms goes off and the breaths stops we
can say that the breath was pressure-cycled.
31Flow cycling
- flow cycle
- Some ventilators will cycle off once a preset low
flow rate is noted. - You can see this on the graphic when we watch the
descending flow wave suddenly drops to zero - This occurs always with PS breaths and you might
be able to chose flow cycling with the VC
mandatory breaths
32Baseline What changes the base line pressure?
- We can raise the pressure during exhalation phase
from zero to a positive number - we have raised the baseline
33PEEP or CPAP?
- Both raise the baseline pressure
- Both used to treat refractory hypoxemia
- Both will increase the FRC and can increase the
lung compliance - PEEP positive end-expiratory pressure with a
ventilator rate set full or partial support the
lower pressure - CPAP- continues positive airway pressure
without a ventilator rate set a spontaneous
mode the only pressure
34Effects of increased baseline pressure
- Keep more air inside alveoli and airways
- Raises RV residual volume which raises the FRC
- Return the FRC to normal will generally increase
the lung compliance and decrease the WOB - Excessive PEEP
- hampers CO, increases VD and causes air trapping
and can damage the lung tissue - If patient is on PC ventilation, raising PEEP
might decrease the VT because the driving
pressure drops. - Excessive CPAP can decrease VT which will raise
the PaC02
35With PEEP
- small amount of gas are trapped because the
exhalation valves closes before the circuit
pressure drops back to zero. - How much gas is left in the lung is a function of
- level of PEEP selected,
- IE ratio if the exhalation time is too short
more gas can trap - time constant of the lung units.
36basic types of PEEP/CPAP devices
- flow restrictor the exhalation port is too small
for the gas to all escape. The faster the flow
through the tiny hole, the more back pressure the
flow restrictor creates - threshold resistor creates back pressure that is
independent of flow rate. In these types of PEEP
valves, the gases passes freely until some
balances of forces on the other side equalize and
the pressure is held in the circuit.
37Types of threshold PEEP valves
- Water Column
- Weighted ball
- Flexed springs
- Venturi diaphragm
- Electromagnetic valve
38When can a threshold resistor become a flow
restrictor?
- When the flow is too fast for the exhalation
valve to get all the gas out - Ventilator circuits are rated for their
resistance to flow and a max flow rate will be
suggested. - Failure to keep the flow rate below this max will
result in PEEP rising as the flow rate rises. - This is a real issue with neonatal circuit that
are so tiny that airway resistance rises quickly