Title: Temperature-Related Stability in Plasma Spectrometry
1Temperature-Related Stability in Plasma
Spectrometry
- Jerry Dulude, Ron Stux, and Vesna Dolic, Glass
Expansion, jdulude_at_geicp.com
2Causes of Signal Drift in ICP-OES
Cause
Solution
- Optical drift
- RF instability
- Sample related baseline shift
- Matrix variations
- Drift in sample intro temperature
- Profiling thermostat
- Feedback loop
- Background correction
- Internal std. or matrix matching
- ????????
3IsoMist Programmable Temperature Spray Chamber
(PTSC)
4IsoMist Characteristics
- Programmable from -10 to 60C in 1 degree
increments - Maintains temperature to within 0.1 degree
- Built-in Peltier device
- No external plumbing
- Temperature measured near the chamber surface
5IsoMist on Optima 2100DV
6PE 2100DV Standard Parameters
- Power 1450 watts
- Coolant gas 15LPM
- Aux. gas 0.2LPM
- Quartz injector, 1.8mm bore
- Neb. Gas 0.7LPM
- SeaSpray concentric glass nebulizer (1ml/min)
- IsoMist Programmable Temperature spray chamber
with Twister Baffled Cyclonic
7Intensity vs. Temperature-PE 2100DV(1ml/min
uptake)
8Temperature Plots on PE 2100DV
31 increase from 10 to 25C or 2 per degree C
31 increase from 25 to 35C or 3 per degree C
9Temperature Plots on PE 2100DV
10Temperature Plots on PE 2100DV
Avg. suppression of all lines
Interference std 1000ppm Na K, 500ppm Ca
11Temperature Plots on PE 2100DV
Avg. suppression of all lines
Avg. suppression of ion lines
Interference std 1000ppm Na K, 500ppm Ca
12IsoMist on the Varian Vista MPX
13Varian Vista MPX Standard Parameters
- Power 1300 watts
- Coolant gas 15LPM
- Aux. gas 1.5LPM
- Quartz injector, 2.3mm bore
- Neb. Gas 0.8LPM
- SeaSpray concentric glass nebulizer (0.4ml/min)
- IsoMist Programmable Temperature spray chamber
with Twister Baffled Cyclonic
14Effect of Spray Chamber Temperature on Intensity
(Varian Vista MPX)
15Effect of Temperature on IntensityVarian Vista
MPX
10 increase from 20 to 25C or 2/degree C
16Effect of Spray Chamber Temperature Varian
Vista MPX
Transport Efficiency
Avg. Intensity
17Varian Vista MPX Parameters
- Power 1300 watts
- Coolant gas 15LPM
- Aux. gas 1.5LPM
- Quartz injector, 1.8mm bore
- Neb. Gas 0.8LPM
- SeaSpray concentric glass nebulizer (0.4ml/min)
- IsoMist Programmable Temperature spray chamber
with Twister Baffled Cyclonic
18Temperature vs. Intensity (Varian Vista MPX-
1.8mm bore)
19Temperature vs. Intensity (Varian Vista MPX)
1.8mm bore
2.3mm bore
10 decrease from 20 to 40C or 0.5 per degree C
29 increase from 10 to 40C or 1 per degree C
20Low Sample Uptake Applications
- Biological applications
- Intracellular fluids
- Neonatal samples
- Forensic applications
- Speciation (HPLC-ICP-MS)
- Radioactive wastes reduction
- Low DLs desirable
21Effect of Temperature on Normalized Intensity (20
uL/min uptake rate)
PE Optima 2100DV
22Effect of Temperature on Intensity20ul/min
uptake rate
Avg. Intensity
Estimated Transport Efficiency
133 increase from 21 to 40C or 7 per degree C
23Spray Chamber Temperature Impact
- High impact with respect to intensity
- Interdependent with other plasma parameters
- Provides another dimension for optimization
- Must remain constant during the experiment
24Constant temperature benefits(1ml/min sample
uptake)
- Higher accuracy
- Higher productivity
- Greater reproducibility
- Better experiments