Title: Technical Updates on UTEVA, TEVA and TRU Resins
1Technical Updates on UTEVA, TEVA and TRU Resins
Anil Thakkar Michael Fern Eichrom Technologies,
Inc. Paris May 14, 2002
2Topics
- UTEVA Resin
- Decontamination of Po-210
- U-232 yields
- TEVA Resin
- Decontamination of U-nat/Th in Tc-99 region
- TRU Resin
- Am recoveries
3U-232 Yields
- Unrealistic high U-232 yields reported with UTEVA
resin in samples with high in Pb-210
4Current Procedure
- 1) Load sample on UTEVA resin with 10 ml of 3M
HNO3/1M Al(NO3)3. with 2 mL of 0.6M Ferrous
sulfamate and 200 mg of ascorbic acid. - 2) Rinse sample beaker with 5 mL of 3M HNO3
- 3) Rinse column with 5 mL of 3M HNO3
- 4) Rinse column with 5 mL of 9M HCl
- 5) Rinse column with 20 mL of 5M HCl/0.05M Oxalic
- 6) Elute U with 15 mL of 0.01M HCl
5Effect of Ferrous210Pb/210Po spike (1100 pCi)
6Effect of Additional 3M HNO3 Rinse 210Pb/210Po
spike
7(No Transcript)
8Effect of Uranium Strip Solution no ferrous
added 210Pb/210Po spike (1100 pCi)
9Additional Rinse/1M HCl Strip Combo
10Revised Procedure
- 1) Load sample on UTEVA with 10 ml of 3M HNO3/1M
Al(N03)3 with 2 mL of 0.6M Ferrous sulfamate and
200 mg of ascorbic acid - 2) Rinse sample beaker with 5 mL of 3M HNO3
- 3) Rinse column with 5 mL of 3M HNO3
- 4) Rinse column with 10 mL of 3M HNO3 (3x)
- 5) Rinse column with 5 mL of 9M HCl
- 6) Rinse column with 20 mL of 5M HCl/0.05M Oxalic
- 7) Elute U with 15 mL of 1M HCl
11Summary
- A good decontamination of Po-210 in U-232 region
can be achieved by additional 3M HNO3 rinse on
UTEVA and eluting uranium selectively with 1M HCl
12TEVA Column Tc-99 Issues
- Problem reported by Darrin Mann (BWXT, Oak Ridge,
Tn) with bias high Tc-99 recoveries in samples
with natural uranium. - Th-234 interference was detected in Tc-99 region.
13BWXTs Tc-99 Procedure
- Solid samples are digested and dissolved finally
in 20 ml of 0.1M HNO3 - Load sample on TEVA column
- Rinse column with 25 ml of 0.1M HNO3
- Extrude resin from the column in a LSC vial.
- Add cocktail and count on LSC.
14BWXTs Procedure tested at Eichrom
- Spiked 20 ml of 0.01M HNO3 with 2000 dpm of
U-234/238 - Load on TEVA Resin column
- Rinse with 25 ml of 0.1M HNO3
- Extrude resin in LSC vial
- Added 15 ml of cocktail and count
15Possible Solutions for Decontamination
- Use of fluoride to complex the residual Th-234 on
the column - Two fluoride solutions tested
- Hydrofluoric acid
- Sodium fluoride
16Decontamination of Th-234
- Spiked 20 ml of 0.01HNO3 with 2000 dpm of
U-234/238 - Load on TEVA Resin column
- Rinse with 5 ml of 0.1M HNO3
- Rinse with 25 ml of 0.5M HF/0.02M HNO3 OR with
25 ml of 1M NaF/0.02 HNO3 - Rinse with 5 ml of 0.1M HNO3
- Extrude resin in LSC vial
- Added 15 ml of cocktail and count
17BWXTs ProcedureTest ResultsRinse volume
used25 mL
18Conclusion
- Decontamination of Th-234 in Tc-99 region can be
accomplished with additional fluoride rinse on
the TEVA Resin columns - Fluoride in the form of 1M NaF or 0.5M HF
solutions could be used to remove the residual
amounts of Th-234
19TRU Resin History
- Historical problems with Am recoveries from TRU
Resin (HCl strip.) - Suspected Extractant Leaching handled with wet
ashing of HCl Strip Solution
20TRU Resin Cartridges
1) TRU Resin 2) TRU Resin with a post
Prefilter layer
21Am-243 Spike Test
1) Spike 10 ml of 3M HNO3 with 6.52 dpm of
Am-243 2) Load solution on TRU cartridges
(assisted with VBS, flow rate 1ml/min) 3) Rinse
with 10 ml of 3M HNO3 4) Rinse with 5ml of 0.5M
HNO3 5) Strip Am with 9M 4M HCl
22Cerium Fluoride Precipitation
- Four different preparation prior to CeF3 ppt.
- 1) Digested with Aqua-regia
- 2) Undigested
- 3) Changed molarity (0.5 M HCl)
- 4) Digested with 10 H2SO4
23Am-243 recoveries (n4)
24RETEST Am-243 recoveries (n4)
25ACW03 tested for Am
- 0.5 L of Eichrom tap water spiked with Am-243
- Precipitated Am with calcium phosphate
- Scavenge ppt. and dissolve in the load solution
- Followed ACW03 procedure using UTEVA/TRU
cartridges - Collect Am fraction
26ACW03continued
6) Evaporate Am fraction to dryness 7) Add 0.5 ml
of conc. HCl and dilute to 15 ml of DI water 8)
Perform CeF3 ppt.
27Tap Water Am-243 recoveries(n5)
28Conclusion
- Am recoveries improved by changing the
concentration on HCl prior to CeF3 - TRU resin cartridge performed well once the HCl
concentration was changed