Title: SERVO-i WITH HELIOX OPTION


1
SERVO-i WITH HELIOX OPTION
2
GROWING WITH THE CUSTOMERS NEEDS
SERVO-i System Version 4.0
  • Heliox option
  • Low MV alarm audio off option
  • Possibility to switch off back-up ventilation
  • Other usability improvements

3
SERVO-i WITH HELIOX OPTION
  • Latest addition to a wide range of options

4
HELIOX GAS PROPERTIES
  • What is Heliox?
  • Heliox HeO2 is a mixture of helium (He) and
    oxygen (O2) (e.g. 80/20 70/30).
  • Helium is an inert, low density gas (1/7 the
    density of air and oxygen).
  • Lower density leads to a more laminar flow (less
    turbulence) and thus reduces resistance in the
    airways.

5
HELIOX GAS PROPERTIES
  • Laminar gas flow (A) is more efficient than
    turbulent flow (B).

6
HELIOX EFFECTS ON VENTILATION
  • Effects of Heliox therapy according to the
    literature
  • Established positive effects on ventilation
  • Reduces work of breathing
  • Increases CO2 diffusion
  • Potential positive effects on ventilation
  • Reduces the peak pressure needed to maintain flow
  • Increases CO2 elimination and improves gas
    exchange
  • Increases expiratory flow
  • Reduces respiratory rate and auto PEEP
  • Increases drug deposition in distal parts of the
    airways.

7
HELIOX USES AND APPLICATIONS
For which patients is Heliox most commonly used?
  • Upper airway obstruction
  • Lower airway obstruction

8
HELIOX POTENTIAL BENEFITS
Potential benefits of Heliox therapy according to
the literature
  • Avoiding endotracheal intubation
  • Protecting the lungs
  • Improving patient outcome
  • Enhancing patient comfort

9
SERVO-i WITH HELIOX OPTION
General
  • Available as an option for the SERVO-i
  • Field upgrade
  • Factory installed
  • The option consists of
  • Heliox SW
  • Heliox adapter kit

10
SERVO-i WITH HELIOX OPTION
Functionality
  • Can be used with all pediatric and adult
    patients, as well as neonates
  • Compatible with all invasive/non-invasive
    ventilation modes
  • Supporting mobility from emergency room to MR
    environment
  • Connection of 80/20 79/21 78/22 mixtures
    possible
  • Adjustable O2 concentration (21 to 100).

11
SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
  • Unique solution for automatic gas identification
  • Automatic compensation of flow delivery
  • Automatic compensation of CO2 and volume
    monitoring

12
SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
  • Quick and easy start of therapy
  • Quick and easy switch from air to Heliox (and
    vice versa) without going to standby
  • User friendly interface
  • Cost efficient Heliox therapy thanks to the
    SERVO-is low gas consumption.

13
SERVO-i WITH HELIOX OPTION
Heliox adapters
  • Four different adapters are available
  • DISS (both air (A) and Heliox (B))
  • NIST/NIST
  • AGA/NIST
  • French/NIST

14
SERVO-i WITH HELIOX OPTION
Heliox adapters
  • Color marked for Heliox
  • Brown/green for Heliox DISS
  • Brown/white for Heliox NIST
  • Attached to ventilator with safety wire
  • Attach to air module when using Heliox

15
SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator Heliox
adapters NIST / AGA / French
  • Remove adapter from holder.
  • Connect adapter to Heliox high pressure hose.
  • Connect adapter to air module inlet.

16
SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator Heliox and
air adapters DISS
  • Remove adapter from holder.
  • Connect adapter to Heliox high pressure hose.
  • Remove air adapter from air inlet.
  • Connect Heliox adapter to air inlet.

17
SERVO-i WITH HELIOX OPTION
Preparations before using Heliox with SERVO-i
18
SERVO-i WITH HELIOX OPTION
Automatic gas identification
  • Performed by the ultrasonic expiratory flow
    sensors
  • Available during Pre Use Check and during
    ventilation
  • Active as long as the set O2 concentration is 75
    or below
  • Time for identification is determined by
    tidal/minute volume and O2 concentration used

19
SERVO-i WITH HELIOX OPTION
Pre Use Check
20
SERVO-i WITH HELIOX OPTION
Automatic identification when changing from air
to Heliox during ventilation
21
SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
22
SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
23
SERVO-i WITH HELIOX OPTION
Gas type change manually initiated compensation
for Heliox gas
24
SERVO-i WITH HELIOX OPTION
Things to bear in mind when using Heliox with
SERVO-i
  • When system is compensated for Heliox
  • Flow trigger sensitivity is automatically reduced
    to a level of 1 if set to a more sensitive level.
  • O2 alarm limits are automatically adjusted from
    5 to 7.
  • When ventilating below 100 ml in infant range, it
    is recommended to use additional patient
    monitoring.

25
SERVO-i WITH HELIOX OPTION
Things to bear in mind when using Heliox with
SERVO-i
  • When changing gas type from air to Heliox and
    vice versa
  • When changing from air to Heliox or vice versa,
    measured values relating to expiratory flow may
    be affected for a short while.
  • Changing gas during ventilation may trigger O2
    concentration high and/or Gas supply pressure low
    alarms.
  • After O2 concentration is changed stabilization
    of expiratory measurements may be delayed.
  • During nebulization there may be a higher
    deposition of drugs used compared to when
    ventilating with air.

26
SERVO-i WITH HELIOX OPTION
Non-invasive application
27
SERVO-i WITH HELIOX OPTION
Infant application
28
SERVO-i WITH HELIOX OPTION
Order information
Part Description Part Number
Heliox software, SERVO-i 66 75 580
Heliox adapter kit, new SERVO-i DISS NIST/NIST AGA/NIST French/NIST 66 75 585 66 75 582 66 75 590 66 75 587
Heliox adapter kit, onsite upgrade DISS NIST/NIST AGA/NIST French/ NIST 66 77 552 66 77 555 66 55 560 66 77 557
SERVO-i up to serial number 40.000 also requires PC board 1784 rev 16 or later PC board 1785 rev 3 or later 64 67 984 64 68 016
High pressure Heliox hose DISS, 5 m 66 75 592
High pressure Heliox hose NIST, 5 m 66 75 607
  • The Heliox option is not compatible with the O2
    Sensor and Y Sensor measuring.

29
SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
30
SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
31
SERVO-i 4.0 OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
32
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
33
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
34
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
35
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Showing previous Pre Use Check/Suction Support
FiO2 retention
36
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Maintenance notifications/Changed calculation of
time for next PM
37
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Previous mode information/Initiating a Pre Use
Check
38
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
New alarm symbol/Clearer difference between
medium and high priority alarms
39
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
New patient category symbols
40
SERVO-i/s 4.0 OTHER IMPROVEMENTS/CHANGES
Time scale on the waveforms