Title: 6890/7890 GC Hardware and Technology Overview
16890/7890 GC Hardware and Technology Overview
2In this Section, We Will Discuss
- The major components of the gas chromatograph.
- The typical chromatogram and the information it
contains. - The way a GC separation occurs.
- Considerations for use of gases and plumbing
configuration.
3Sample Requirements for Gas Chromatography
Only 10-20 of all compounds are suitable for GC
analysis, But these compounds are worth billions
and billions of dollars annually.
- The Compounds must have
- Sufficient Volatility Large macro molecules
generally do not have sufficient volatility,
(i.e., they will not become gaseous under
instrumental parameters). Large biological
polymers are examples of inappropriate compounds
for GC. - Free of Residues This is an extension of the
first requirement. Non-volatile impurities in
the sample matrix can lead to inlet and column
contamination that - will quickly degrade the chromatography.
- Thermal Stability The compounds of interest
must not degrade when introduced into the hot
inlet (upwards of 300?C) or while in the heated
column (upwards of 350?C).
4What is the deal with GC?
- Temperature driven
- This means that there is a molecular weight
limit. - Uses a gas as the mobile phase
- The most efficient way to separate organics.
- Pressurized instrument
- This means that leaks are bad news.
- Huge arsenal of detectors
- Ranging from universal to selective to
confirmatory.
5Innovation of GC Technology
- Fused Silica Capillary Columns 1979. This
material has allowed user-friendly operation and
installation of capillary columns worldwide.
- Electronic Pneumatics Control (EPC) Board --
1989 - perfected in 1995 with the 6890. - EPC has made GC more precise in providing peak
areas and retention times than ever before in
the history of GC.
6Typical GC System Schematic
7Typical Gas Chromatograph
Fixed
Mol-Sieve
Traps
Restrictors
Injection
Port
Regulators
Detector
Electrometer
Flow
PC
Controller
Column
Carrier
Hydrogen
Air
Gas
8Definitions
- Gases
- Carrier Gas Pressurized gas used to transport
the sample through the system. - Detector Gases Support for certain detectors
(i.e., FID). - Sample Introduction
- Introduces the sample to the carrier gas stream
with minimal disruption of the gas stream. - Column
- Achieves separation of the components in the
sample. - Detector
- Recognizes and responds to sample components as
they elute from the column. - Data Acquisition
- Converts the detector signal to a picture
chromatogram and provides manual or automated
determination of the identity and amounts of the
sample components .
9Innovation of GC Technology
- Highly reproducible GC ovens
- New data systems
- Fast automatic samplers
- New detectors
- New integration algorithms
- Generally more precise GCs...
10Role of the Sample
- The sample determines the instrument
configuration - Type of Carrier Gas
- Type of Sample Inlet
- Type of Column
- Type of Detector
- Type of Data Acquisition
11Typical Chromatogram
- Retention Time
- Parameter used to identify a sample component.
- Peak Area
- Parameter used to measure the quantity of the
sample component.
12Model of the Chromatographic Process
13How Separation Occurs
Chromatography is a separation method achieved by
the distribution of substances between two phases
(a mobile phase and a stationary phase)
Mobile Phase Stationary Phase
Gas Solid Chromatography (GSC) Gas Solid
Gas Liquid Chromatography (GLC) Gas Liquid
14Separation is a Partitioning Process
Column
Carrier Gas
15Column Types
Packed
Open (Capillary)
Wall Coated
Open Tube
PACKED
SERIES 530
NARROW BORE
LENGTH (meters)
0.5-10
5-100
5-100
I.D. (mm)
2-4
0.530
0.1-0.25
16There are Two Common Types of Capillary Column
Stationary Phase Coatings
17Comparison of Column Types
Column Evaluation Sample (Kerosene)
Packed Column Analysis
5 OV101
on 80/100 Chromosorb
Megabore ( packed column replacement)
30m X 0.53mm X .88µ
Capillary
30m X 0.32mm X .25µ
18Carrier and Detector Support Gases
- Gases must be
- Chosen with the consideration of the type of
detector used - Inert
- Dry
- Pure
19GC Gases
- Generally, the carrier gas for a GC system will
start at a cylinder holding the compressed gas. - A regulator valve on the outlet of the cylinder
controls the pressure of the gas in the supply
lines. - Compressed gases are available in different
levels of purity. Gases of four nines
(99.9999) or better are recommended. - Clean supply tubing to transfer gas from the
cylinder to the GC and gas purification traps
are recommended for routine operations.
20Gas Regulator Valves
- The regulator valve is a very common site in
analytical laboratories. - The gauge on the left indicates gas pressure
remaining in the cylinder. - The gauge on the right indicates the set
pressure of gas leaving the cylinder and flowing
into the supply tubing. - Most systems use 1/8 fittings however, an
adaptor is available for use with ¼ tubing and
fittings.
21Gas Regulator Valves
- Regulators are standardized and come with NPT
style threads, which are common to most gas
plumbing applications. - The regulator valves are made of brass with
stainless steel diaphragms. - There are separate regulator valves for
- Air
- Hydrogen, Argon/Methane mix (P5 mix)
- Oxygen
- Helium, Argon, Nitrogen
22Regulators and Flow Controllers
- The carrier gas must be regulated to provide
constant pressure as well as a constant mass
flow. The pressure differential between
controllers is recommended as 5 psi. - Recommended Line Pressures
- Carrier Gas should be 60-150 psi depends on type
of column used (60 psi minimum for large
diameter, 150 for very small diameter and
capillary columns). - Air pressure should be 80 psi.
- Hydrogen should be 60 psi.
23GC Gas Purifiers
- Carrier gas purity is very important. Trace
amounts of oxygen and water will damage and
shorten column lifetimes, especially for
capillary columns. - The more polar the column (i.e. waxes like
polyethylene glycol), the more susceptible it
will be to degradation. - Oxygen will also degrade ECD performance.
24GC Gas Purification Configurations
25Assembling the Gas Plumbing
26Different Gas Purifiers
27Different Gas Purifiers
28Tubing and Traps
- GC or instrument grade copper or stainless steel
tubing should be used for all gases. - Stainless steel tubing is recommended for
hydrogen. - Plastic tubing is permeable to O2 and other
contaminants. It may also outgas detectable
impurities. - Precondition the tubing with solvent flush and
carrier gas drying or purchase tubing prepared
this way. - Filters need to be changed at the manufacturer's
recommended interval to prevent contamination
breakthrough (i.e. every 3 cylinders). - All external fittings should be checked on a
routine basis for leaks (every 6 months).
29Gas Flow Meters
Volumetric versus Mass Flow Measurement What you
should know
- Volumetric
- As the name suggests, these meters measure the
amount of gas which is passing through the
system. - The Optiflow, at right, forms a soap- bubble
membrane in a glass tube. The gas flow carries
the membrane through an optical sensor that
calculates flow based on travel time. - The ADM and Flow Trackers utilize changes in
thermal conductivity to calculate how much gas
is passing through a precisely calibrated
orifice.
30Gas Flow Meters
Volumetric versus Mass Flow Measurement What you
should know
- Mass Flow Meters
- As the name suggests, these meters are
calibrated to the mass specifications for a
specific gas. - Veri-Flow 500 On board mass specifications
are stored for the 5 most common gases - Helium, Hydrogen, Nitrogen, Air, Argon
Methane Mix (P5).