Title: Reflection Data Analysis Tools in PLATON
1Reflection Data Analysis Tools in PLATON
Problem from Bruker Discussion
List Triclinic or Monoclinic?
Ni-complex In P-1,Z'2 a 11.869 b 5.562 c
13.67 Alpha 90.08 Beta 116.38 Gamma
90.06 R1 10 WR2 31
Ton Spek, Utrecht University, The
Netherlands Bruker User Meeting Bad Herrenalb,
Sept. 23, 2012
2Triclinic, Monoclinic ?
- The New SHELXT program offers a starting model in
P-1 and Pn and P21/n - PLATON/ADDSYM happily offers a structure in P21/n
that resembles the solution in P-1
3(No Transcript)
4(No Transcript)
5Alternative Monoclinic Model
In P21/n, Z'1 A 11.869 B 5.562 C
13.67 Alpha 90 Beta 116.38 Gamma 90 R1
17 WR2 53 Poor ADP's
6Monoclinic Model ADP's
7Looking at the Data
- - Investigate thePrimary Data
- - Images, Integration
- - HKL File
- - Redundancy, Laue Symmetry (Rav)
- - Space Group Extinctions
- - Investigate the Refinement Data
- - Analysis of Variance
- - Twinning
- With Tools Available in PLATON
8(No Transcript)
9(No Transcript)
10(No Transcript)
11(No Transcript)
12(No Transcript)
13(No Transcript)
14(No Transcript)
15(No Transcript)
16(No Transcript)
17(No Transcript)
18(No Transcript)
19(No Transcript)
20(No Transcript)
21(No Transcript)
22(No Transcript)
23Test P-1 Model for Twinning
24Overlapped k 0 Layer
25(No Transcript)
26Rav (Monoclinic 2/m) 21 on F(calc) Model
27Refined Model in P-1
All H-atoms Found and rerined
28In Conclusion
- Three additional parameters had to be introduced
to obtain a nice looking refinement result in
P-1 - - A twinning parameter
- - Two population parameters Ni1 0.88, Ni2 0.43
- The residual problem explain the non 1.0
population parameters for Ni. - - Go back to the chemist ..
29Thanks !
- Example taken from the very
- active and interesting
- Bruker User Discussion List
- Maintained by Ilia Guzai
- Problem and data communicated by
- Babu Varghese
30(No Transcript)