Title: Teledyne Analytical Instruments
1Teledyne Analytical Instruments
2Overview
- Theory of Operation
- Electromagnetic Spectroscopy
- Laser Tunable Diode
- Wavelength Selection
- LGA 4000 In-situ Analyzer
- LGA 4000 Extractive Analyzer
- Instrument Specifications
- Applications
3Theory of Operation
- Electromagnetic Spectroscopy
DE hn
- Molecules absorb EM radiation
- Raise energy level of electrons at certain
wavelengths - Each molecule has unique frequency response
4Theory of Operation
- Emit EM radiation through sample gas
- Measure absorbance A log10 (Io/It)
- Beer Lambert Relationship
- Aebc
- emolar absorption
- bpath length
- cgas concentration
5Theory of Operation
- Diode Laser Absorption Spectroscopy
- Monochromatic bandwidth
- Bandwidth 0.0001nm
- Adjust laser temperature, current
6Theory of Operation
- Laser spectral scan is larger than gas absorption
spectral line width - Reference signal is compared to sample gas signal
- Calculation of TTb/Ta provides accurate
concentration measurement
7Theory of Operation
8Theory of Operation
- Spectral analysis of background gases and sample
gas - Frequency must be found such that only sample gas
is attenuated
9LGA 4000 In-Situ Analyzer
10LGA 4000 In-Situ Analyzer
11LGA 4000 Extractive Analyzer
12Analyzer Specifications
In-Situ System
Extractive System
- Optical Path Length lt 15m
- Response Time lt 1s
- Linear Accuracy lt /-1 FS
- Span Drift lt /- 1 FS
- Calibration 2/year
- Enclosure IP65
- Explosion Proof Expxmd IICT5
- Power 24VDC/110VAC/220VAC
- Ambient Temperature -30 to 60C
- Optical Path Length 0.5m or 1.1m
- Response Time lt 5s (T90)
- Gas Temperature -30 to 140C
- Gas Pressure 0.5 to 3 Bara
13Applications
Gas LDL Range Gas LDL Range
O2 0.01 Vol. 0-1 Vol., 0-100 Vol. CO 40 ppm 0-8000 ppm, 0-100 Vol.
CO2 20 ppm 0-2000 ppm, 0-100 Vol. H2O 0.03 ppm 0-3 ppm, 0-70 Vol.
H2S 2 ppm 0-200 ppm, 0-30 Vol. HF 0.01 ppm 0-1 ppm, 0-10000 ppm
HCL 0.01 ppm 0-7 ppm, 0-8000 ppm HCN 0.2 ppm 0-20 ppm,0-1 Vol.
NH3 0.1 ppm 0-10 ppm, 0-1 Vol. CH4 10 ppm 0-200 ppm, 0-10 Vol.
C2H2 0.1 ppm 0-10 ppm, 0-70 Vol. C2H4 1.0 ppm 0-100 ppm, 0-70 Vol.