Title: Basics
1Basics principles of
mechanical ventilation
2Basics principles of ventilation
-
-
- -What is ventilator
- -How ventilators work
- -How to use
-
?
3- Ventilator
- The Machine delivers O2 removes Co2 with out
harming the patient. - The Ventilator should have capacity of delivering
a consistent tidal volume under all adverse
conditions of lung disease.
4- Mechanical Ventilation
- The process / method by applying (intermittent)
positive airway pressure supplementation of
Inspired O2 to achieve desirable oxygenation.
5- GOALS of Mechanical Ventilation
- Good ventilation (known by pa Co2)
- Oxygenation (by pa O2).
- Alveolar recruitment by peep
- Lowest O2 supplement.
- Synchrony between patient Ventilation.
-
- Safe ? No Negative hemodynamic effect
- No Barotraumas
- No auto PEEP
6- Classification of Mechanical ventilation
- Positive / Negative pressure ventilation.
- Partial / full ventilation.
- Invasive / Non invasive ventilation.
7Components of ventilator
- Power source-electrical/pneumatic/both
- Control systems-circuits, control panel
- Display systems
8V
Central unit
oxy
Exp.limb
Pt
blender
Insp. limb
V
P
Fio2
T
Humneb
compressor
9PHASES VARIABLES
- INITIATION OF INSPIRATION-TRIGGER
- INSPIRATION-LIMIT
- MAINTENANCE OF INSPIRATION-CONTROL
- CHANGING TO EXPIRATION-CYCLE
10- Triggering variable
- Triggering is a method of starting the
inspiration - Types Pressure Triggering
- Flow Triggering - No lag time
- Time Triggering -Less effort
11- Control Variable
- Setting that maintained thro out inspiration
- Type Volume control.
- Pressure control.
12- PCV
VCV - Vt. Variable set
- PIP SET lesser
variable more - Plateau pressure Set Variable
- Inspiratory flow decelerating type
fixed flow type
sinusoidal/square - Inspiratory time set set
- Respiratory rate Set Set
- Barotrauma Less More
- Leak compensation for minor leak nil
- Patients acceptance good -
13- Limit Variable
- Setting that can't be exceeded during inspiration
- Type Pressure limit- psv
- volume limit
- Flow limit
-
14- Cycle variable
- Method of termination of inspiration, I.e.
changing over from inspiration to expiration. - Types Volume cycle
- Pressure cycle
- Time cycle
- flow cycle.
15Compliance-resistance-volume
Cl
Ccw
compliance
PAi
Vt
PA range
Raw
PAe
C?V /?P
RPIP-Pp / F
16(No Transcript)
17Normal resistance Un intubated patient 0.6 -
2.4 cm H20 / L / Sec, .
at 0.5L /sec Intubated patient
6cm H20 / L / Sec
18Ventilatory controls-inter relation
MV
Ti
T
IE
Te
Vt
f
19- Ventilator Modes
- Combination of breath type and phase
variable. -
- Conventional modes Recent modes
Newer modes - CMV MMV VAPS, PAug
- A/C MV APRV VS,
PRVC - IMV SIMV BIPAP
AUTOFLOW - CPAP PEEP IRV
AUTOMODE - VCV DLV PAV,PPS
- PSV HFJV ASV
20- Controlled mechanical ventilation
-
-CMV - All breaths are delivered by ventilator.
- No Patient participation.
- Set Vt. Delivered at set RR
- Requires sedation neuromuscular blockade.
- Time initiated
- Volume limited
- Volume cycling
21- CMV
- Indications
- Patient with no efforts / complete respiratory
failure. - When negative inspiratory effort contra
indicated. - eg.flail chest.
- During anesthesia.
- Disadvantages
- Patient participation not
allowed. - Heavy sedation relaxant need
- Long term CMV ? Respiratory muscle weakness.
- Varying PIP according to lung compliance
patient efforts.
22- Assist / Control Mechanical ventilation
- Patient can trigger ventilation at a rate more
than set RR - All breaths are delivered at set volume , set
time set pressure - Triggering ? pressure / flow (spontaneous)
- Time (Mandatory)
- Limiting ? Volume
- Cycling ? Volume
- Patient can vary RR only but not vt.
23Assist / Control Mechanical Ventilation
- Indications
- Patient with normal drive but with respiratory
weakness ? Recovering patient. - To preserve patient efforts
- Weaning.
- Disadvantages
- Rapid triggering ? Hyperventilation
- Hypotension
- Flow rare should be adjusted according to the
need. - If RR lt RR ? CMV mode.
24- IMV / SIMV
- The patient receives the Mandatory set Vt at set
RR. - The set Mandatory breaths are synchronized with
patient efforts. - Between the mandatory breaths the patient can
breath spontaneously - Spontaneous breath vt depends on
- ? Patients respiratory effort
- ? PS
- Triggering - Pressure
- Limiting - Volume
- Cycling - Time
25- IMV / SIMV
- Advantages
- The mandatory breaths are synchronized
with patients Spontaneous efforts. - Hyperventilation is less
- More active participation of patient
- Disadvantages
- More WOB
26- A/CMV SIMV
- Patient decides only RR lt -- gt patient effort
decides RR vt. - Less WOB as only lt -- gtmore WOB as
Initiation by patient patient
has to .
operate demand - . flow
system - Possibility of hyper ventilation lt -- gt No
27- CPAP
- A mode is which positive pressure is applied tho
out the respiratory cycle using during
spontaneous ventilation. - (Pr applied in mechanical ventilation PEEP)
- No Ventilatory assistance
- Positive Pressure causes
- Prevention of alveolar collapse alveolar
recruitment - Î FRC Atelecasis
- FlO2 requirement
28- CPAP PEEP
- Pr applied and base line Pr. Applied with
- Pr elevated when ventilatory some ventilatory
Assistance is nil.
Assistance present.
29- PEEP
- Applied when Fio2 requirement is 50 - 60.
- Best Peep PEEP titrated to achieve optimal
respiratory system compliance. - Optimal Peep Titration of PEEP until Qs /Qt is lt
15
30- Volume Control Ventilation
- Vt. Delivery is constant according to pressure
regardless of changes in airway resistance or
respiratory system compliance. - VCV is given when constant MV is needed (eg,
patient with Î ICT)
31- Pressure control ventilation
- The pressure applied to the airways is constant
regardless of airway resistance and compliance. - Constant pressure is delivered throughout
inspiration at set RR - Time initiated pressure limiting time cycling.
- Vt may vary according to patient lung conditions.
- PCV avoids over distention in patient with ALI,
because PIP can be set. - Settings
- Preset pressure is equal to half of present PIP.
- PEEP half of present PEEP (if gt 8cm H2 O)
- I E is 12
32- PCV
VCV - Vt. Variable set
- PIP SET lesser
variable more - Plateau pressure Set Variable
- Inspiratory flow decelerating type
fixed flow type
sinusoidal/square - Inspiratory time set set
- Respiratory rate Set Set
- Barotrauma Less More
- Leak compensation for minor leak nil
- Patients acceptance good -
33-
- Pressure Support Ventilation (Psv)
- Patients spontaneous activity is assisted by
delivery of a preset amount of inspiratory
positive pressure. - Patient triggers ? set pressure is maintained
throughout inspiration. - Pressure initiated.
- Pressure limiting .
- Flow Cycling
- As flow reaches 25 of peak inspiratory flow /5
litres / min
34- Pressure Support Ventilation (Psv)
- Low PSV to overcome the patients WOB
associated with ETT and circuits. - PSV max to achieve Vt of 10 -12 ml / Kg
- - may require upto 40 -50 cm H2O
- Can be used alone as full ventilatory support
or with SIMV. - Can be used as non invasive ventilatory support
up to 20cm H2O2 for transient Ventilatory
support ( Narcotic overdose, asthma, acute
exacerbation of COPD).
35- Mandatory Minute Ventilation (MMV)
- Preset MV is selected.
- The Ventilator calculates the patients
spontaneous MV. - It patients spontaneous MV lt set MV, ventilator
assists to achieve set MV - Ventilatory assisstance may be
- - Volume controlled SIMV breaths Î RR
/Vt - - Î PSV
36Mandatory Minute Ventilation (MMV)
- ADVANTAGES
- MV guaranteed
- Useful as weaning mode
- DISADVANTAGES
- RR may cause dead space ventilation even with
acceptable MV. - Respiratory muscle fatigue may develop (so high
RR alarm should be activated.
37Mandatory Minute Ventilation (MMV)
- INDICATIONS
- During weaning period
- To aspiratory flow and WOB
- To overcome ETT/circuit resistance
38BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
- A pressure controlled ventilation
- Allows unrestricted spontaneous breathing at any
point of ventilatory cycle - Time cycled changes of pressure application.
- Independent positive airway pressure to
inspiration expiration - Inspiratory set pressure is called IAP/T high
Expiratory set pressure is called EAP/T low - Usual IAP is 8 cm H2O EAP is 3 cm H2O
- Triggerflow, Limitpressure
Cycletime
39BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
- IAP causes better ventilation Paco2
- EAP causes better oxygenation Pao2
- Types
- --CPAPPS
- --Two alternating CPAP level
- --APRV
40BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
- ADVANTAGES
- Non invasive ventilation
- Useful inend stage COPD
- ---restricted chest wall diseases
- ---neuromuscular diseases
- ---nocturnal hypo ventilation
- A weaning mode.
41AIRWAY PRESSURE RELEASE VENTILATION(APRV)
- A CPAP circuit with release valve at expiratory
limb driven by time device - APRV is a CPAP system causing .
alveolar ventilation by briefly interrupting
CPAP.
42APRV.,
- Release valve opens for 1-2sec.
- Pr drops to lower level-low CPAP(0to-2cmH2O)
- Lung volume less than FRC in expiration
- alveolar ventilation CO2 elimination
- Reapplication of CPAP by closing valve- Higher
CPAP(10to 12 cm H2O) - FRC oxygenation.
-
43APRV.,
- ADVANTAGES
- Lesser PIP ,so less hemo dynamic changes.
- To alveolar ventilation in ALI of mild to
moderate. - A weaning mode.
44INVERSE RATIO VENTILATION(IRV)
- IE gt1
- PC-IRV / VC-IRV
- Ti with set pr opening of stiff alveoli
units improved oxygenation - Te not allowing alveoli to collapse
- development of intrinsic PEEP
- reduction of shunting
45IRV,
- Improve oxygenation by
- Reducing intra pulmonary shunting
- Improvement of V/Q matching
- Decreased dead space ventilation
- Increased MAP intrinsic PEEP
- Useful when high FiO2 high PEEP to be avoided
46NEWER MODES
- Dual modes VAPS, Paug,
-
VS, PRVC, Autoflow, VPC - Switching modes Automode
- Proportional modes PAV,PPS.
- Adaptive modes ASV
47DUAL MODES
- Combination of 2modes of ventilation(PCVVCV)
to deliver guaranteed Vt/MV - Volume guaranteed pr targeted ventilation
- Mode changes occur
- with in a breath-VAPS,Paug
- over several breaths-VS,PRVC,Auto
flow,VPC -
48DUAL MODES - CHANGES WITH IN BREATH
- VAPSvolume assured pr support-T bird,bird8400st
- Paug pr augmented ventilation-bear 1000
- Vt guaranteed variable pr limited modes
- During inspiration,the ventilator monitors Vt
- if desired Vt delivered before flow drops---PSV
- if desired Vt not delivered before flow drops
- flow continued at set pr support level till
adequate - Vt delivered.
49DUAL MODES - CHANGES WITH IN BREATH
- Trigger patient
patient - Limit -- pressure
variable pr - Cycle -- flow
volume - PSV
Paug -
VAPS
50DUAL MODES - CHANGES OVER SEVERAL BREATHS
- PRVCPr regulated VC-siemen300
- VSvolume supportservo 300
- Auto flowDragerE4
- VS-volume assured PSV
- PRVC-volume assured PCV
- Vt measured over several breathsadequate MV
achieved by changing PS/PC mode for remaining
breaths.
51DUAL MODES - CHANGES OVER SEVERAL BREATHS
- AUTOFLOW
- Autoflow alters the function of inspiratory and
expiratory valves - Allowing patient to receive inspiratory flow
demand - Auto flow provides better ventilatory tolerance
- A weaning mode
52Switching modes
- Ventilator can switch modes according to
monitored information - Automodeservo300
- Switching between control /support mode depending
on patients respiratory pattern - Monitoring of patients respiration over fixed
time period if 2consecutive effort PSV
. -- if no efforts - PCV
53Proportional modes
- PAVproportional assist ventilation-PB840
- PPS proportional PS-Evita-E4
- Proportional modes are assisting spontaneous
ventilation - PPVa support mode in which pr, flow,volume are
set proportional to patients inspiratory efforts
54Proportional modes
- The more effort pt exerts-the more support the
machine provides - PAV allows patients to comfortably reach
whatever the ventilatory pattern that suit their
need.
55Adaptive modes
- ASV-Adaptive support ventilation
- Uses pr targeted breaths to assure a target MV
with decreased WOB - ASV adapts to the changing capabilities of
patients lung conditions. - More efforts the patient does less
- support the
machine provides.
56Adaptive modes
- ASV calculate the over all MV combination of
volume guaranteed PSV(VS) volume guaranteed
PCV(PAVC) - ASV can ventilate the patient from acute stage to
a weaning stage.
57 Weaning
- Wearing off primary pathology
- Elimination of effects of sedation relaxants
- Absence of sepsis
- No metabolic / electrolyte abnormality
- Involvement of patient
- Nutritionally stable
- Good stable CVS