Basics - PowerPoint PPT Presentation

About This Presentation
Title:

Basics

Description:

Partial / full ventilation. Invasive / Non invasive ventilation. Components of ventilator Power source-electrical/pneumatic/both Control systems-circuits, ... – PowerPoint PPT presentation

Number of Views:229
Avg rating:3.0/5.0
Slides: 58
Provided by: sys192
Category:

less

Transcript and Presenter's Notes

Title: Basics


1
Basics principles of
mechanical ventilation
  • g.k.kumar

2
Basics principles of ventilation
  • -What is ventilator
  • -How ventilators work
  • -How to use



?
3
  • Ventilator
  • The Machine delivers O2 removes Co2 with out
    harming the patient.
  • The Ventilator should have capacity of delivering
    a consistent tidal volume under all adverse
    conditions of lung disease.

4
  • Mechanical Ventilation
  • The process / method by applying (intermittent)
    positive airway pressure supplementation of
    Inspired O2 to achieve desirable oxygenation.

5
  • GOALS of Mechanical Ventilation
  • Good ventilation (known by pa Co2)
  • Oxygenation (by pa O2).
  • Alveolar recruitment by peep
  • Lowest O2 supplement.
  • Synchrony between patient Ventilation.
  • Safe ? No Negative hemodynamic effect
  • No Barotraumas
  • No auto PEEP

6
  • Classification of Mechanical ventilation
  • Positive / Negative pressure ventilation.
  • Partial / full ventilation.
  • Invasive / Non invasive ventilation.

7
Components of ventilator
  1. Power source-electrical/pneumatic/both
  2. Control systems-circuits, control panel
  3. Display systems

8
V
Central unit
oxy
Exp.limb
Pt
blender
Insp. limb
V
P
Fio2
T
Humneb
compressor
9
PHASES VARIABLES
  • INITIATION OF INSPIRATION-TRIGGER
  • INSPIRATION-LIMIT
  • MAINTENANCE OF INSPIRATION-CONTROL
  • CHANGING TO EXPIRATION-CYCLE

10
  • Triggering variable
  • Triggering is a method of starting the
    inspiration
  • Types Pressure Triggering
  • Flow Triggering - No lag time
  • Time Triggering -Less effort

11
  • Control Variable
  • Setting that maintained thro out inspiration
  • Type Volume control.
  • Pressure control.

12
  • PCV
    VCV
  • Vt. Variable set
  • PIP SET lesser
    variable more
  • Plateau pressure Set Variable
  • Inspiratory flow decelerating type
    fixed flow type
    sinusoidal/square
  • Inspiratory time set set
  • Respiratory rate Set Set
  • Barotrauma Less More
  • Leak compensation for minor leak nil
  • Patients acceptance good -

13
  • Limit Variable
  • Setting that can't be exceeded during inspiration
  • Type Pressure limit- psv
  • volume limit
  • Flow limit

14
  • Cycle variable
  • Method of termination of inspiration, I.e.
    changing over from inspiration to expiration.
  • Types Volume cycle
  • Pressure cycle
  • Time cycle
  • flow cycle.

15
Compliance-resistance-volume
Cl
Ccw
compliance
PAi
Vt
PA range
Raw
PAe
C?V /?P
RPIP-Pp / F
16
(No Transcript)
17
Normal resistance Un intubated patient 0.6 -
2.4 cm H20 / L / Sec, .
at 0.5L /sec Intubated patient
6cm H20 / L / Sec
18
Ventilatory controls-inter relation
MV
Ti
T
IE
Te
Vt
f
19
  • Ventilator Modes
  • Combination of breath type and phase
    variable.
  • Conventional modes Recent modes
    Newer modes
  • CMV MMV VAPS, PAug
  • A/C MV APRV VS,
    PRVC
  • IMV SIMV BIPAP
    AUTOFLOW
  • CPAP PEEP IRV
    AUTOMODE
  • VCV DLV PAV,PPS
  • PSV HFJV ASV

20
  • Controlled mechanical ventilation

  • -CMV
  • All breaths are delivered by ventilator.
  • No Patient participation.
  • Set Vt. Delivered at set RR
  • Requires sedation neuromuscular blockade.
  • Time initiated
  • Volume limited
  • Volume cycling

21
  • CMV
  • Indications
  • Patient with no efforts / complete respiratory
    failure.
  • When negative inspiratory effort contra
    indicated.
  • eg.flail chest.
  • During anesthesia.
  • Disadvantages
  • Patient participation not
    allowed.
  • Heavy sedation relaxant need
  • Long term CMV ? Respiratory muscle weakness.
  • Varying PIP according to lung compliance
    patient efforts.

22
  • Assist / Control Mechanical ventilation
  • Patient can trigger ventilation at a rate more
    than set RR
  • All breaths are delivered at set volume , set
    time set pressure
  • Triggering ? pressure / flow (spontaneous)
  • Time (Mandatory)
  • Limiting ? Volume
  • Cycling ? Volume
  • Patient can vary RR only but not vt.

23
Assist / Control Mechanical Ventilation
  • Indications
  • Patient with normal drive but with respiratory
    weakness ? Recovering patient.
  • To preserve patient efforts
  • Weaning.
  • Disadvantages
  • Rapid triggering ? Hyperventilation
  • Hypotension
  • Flow rare should be adjusted according to the
    need.
  • If RR lt RR ? CMV mode.

24
  • IMV / SIMV
  • The patient receives the Mandatory set Vt at set
    RR.
  • The set Mandatory breaths are synchronized with
    patient efforts.
  • Between the mandatory breaths the patient can
    breath spontaneously
  • Spontaneous breath vt depends on
  • ? Patients respiratory effort
  • ? PS
  • Triggering - Pressure
  • Limiting - Volume
  • Cycling - Time

25
  • IMV / SIMV
  • Advantages
  • The mandatory breaths are synchronized
    with patients Spontaneous efforts.
  • Hyperventilation is less
  • More active participation of patient
  • Disadvantages
  • More WOB

26
  • A/CMV SIMV
  • Patient decides only RR lt -- gt patient effort
    decides RR vt.
  • Less WOB as only lt -- gtmore WOB as
    Initiation by patient patient
    has to .
    operate demand
  • . flow
    system
  • Possibility of hyper ventilation lt -- gt No

27
  • CPAP
  • A mode is which positive pressure is applied tho
    out the respiratory cycle using during
    spontaneous ventilation.
  • (Pr applied in mechanical ventilation PEEP)
  • No Ventilatory assistance
  • Positive Pressure causes
  • Prevention of alveolar collapse alveolar
    recruitment
  • Î FRC Atelecasis
  • FlO2 requirement

28
  • CPAP PEEP
  • Pr applied and base line Pr. Applied with
  • Pr elevated when ventilatory some ventilatory
    Assistance is nil.
    Assistance present.

29
  • PEEP
  • Applied when Fio2 requirement is 50 - 60.
  • Best Peep PEEP titrated to achieve optimal
    respiratory system compliance.
  • Optimal Peep Titration of PEEP until Qs /Qt is lt
    15

30
  • Volume Control Ventilation
  • Vt. Delivery is constant according to pressure
    regardless of changes in airway resistance or
    respiratory system compliance.
  • VCV is given when constant MV is needed (eg,
    patient with Î ICT)

31
  • Pressure control ventilation
  • The pressure applied to the airways is constant
    regardless of airway resistance and compliance.
  • Constant pressure is delivered throughout
    inspiration at set RR
  • Time initiated pressure limiting time cycling.
  • Vt may vary according to patient lung conditions.
  • PCV avoids over distention in patient with ALI,
    because PIP can be set.
  • Settings
  • Preset pressure is equal to half of present PIP.
  • PEEP half of present PEEP (if gt 8cm H2 O)
  • I E is 12

32
  • PCV
    VCV
  • Vt. Variable set
  • PIP SET lesser
    variable more
  • Plateau pressure Set Variable
  • Inspiratory flow decelerating type
    fixed flow type
    sinusoidal/square
  • Inspiratory time set set
  • Respiratory rate Set Set
  • Barotrauma Less More
  • Leak compensation for minor leak nil
  • Patients acceptance good -

33
  • Pressure Support Ventilation (Psv)
  • Patients spontaneous activity is assisted by
    delivery of a preset amount of inspiratory
    positive pressure.
  • Patient triggers ? set pressure is maintained
    throughout inspiration.
  • Pressure initiated.
  • Pressure limiting .
  • Flow Cycling
  • As flow reaches 25 of peak inspiratory flow /5
    litres / min

34
  • Pressure Support Ventilation (Psv)
  • Low PSV to overcome the patients WOB
    associated with ETT and circuits.
  • PSV max to achieve Vt of 10 -12 ml / Kg
  • - may require upto 40 -50 cm H2O
  • Can be used alone as full ventilatory support
    or with SIMV.
  • Can be used as non invasive ventilatory support
    up to 20cm H2O2 for transient Ventilatory
    support ( Narcotic overdose, asthma, acute
    exacerbation of COPD).

35
  • Mandatory Minute Ventilation (MMV)
  • Preset MV is selected.
  • The Ventilator calculates the patients
    spontaneous MV.
  • It patients spontaneous MV lt set MV, ventilator
    assists to achieve set MV
  • Ventilatory assisstance may be
  • - Volume controlled SIMV breaths Î RR
    /Vt
  • - Î PSV

36
Mandatory Minute Ventilation (MMV)
  • ADVANTAGES
  • MV guaranteed
  • Useful as weaning mode
  • DISADVANTAGES
  • RR may cause dead space ventilation even with
    acceptable MV.
  • Respiratory muscle fatigue may develop (so high
    RR alarm should be activated.

37
Mandatory Minute Ventilation (MMV)
  • INDICATIONS
  • During weaning period
  • To aspiratory flow and WOB
  • To overcome ETT/circuit resistance

38
BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
  • A pressure controlled ventilation
  • Allows unrestricted spontaneous breathing at any
    point of ventilatory cycle
  • Time cycled changes of pressure application.
  • Independent positive airway pressure to
    inspiration expiration
  • Inspiratory set pressure is called IAP/T high
    Expiratory set pressure is called EAP/T low
  • Usual IAP is 8 cm H2O EAP is 3 cm H2O
  • Triggerflow, Limitpressure
    Cycletime

39
BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
  • IAP causes better ventilation Paco2
  • EAP causes better oxygenation Pao2
  • Types
  • --CPAPPS
  • --Two alternating CPAP level
  • --APRV

40
BILEVEL POSITIVE AIRWAY PRESSURE
VENTILATION(BIPAP)
  • ADVANTAGES
  • Non invasive ventilation
  • Useful inend stage COPD
  • ---restricted chest wall diseases
  • ---neuromuscular diseases
  • ---nocturnal hypo ventilation
  • A weaning mode.

41
AIRWAY PRESSURE RELEASE VENTILATION(APRV)
  • A CPAP circuit with release valve at expiratory
    limb driven by time device
  • APRV is a CPAP system causing .
    alveolar ventilation by briefly interrupting
    CPAP.

42
APRV.,
  • Release valve opens for 1-2sec.
  • Pr drops to lower level-low CPAP(0to-2cmH2O)
  • Lung volume less than FRC in expiration
  • alveolar ventilation CO2 elimination
  • Reapplication of CPAP by closing valve- Higher
    CPAP(10to 12 cm H2O)
  • FRC oxygenation.

43
APRV.,
  • ADVANTAGES
  • Lesser PIP ,so less hemo dynamic changes.
  • To alveolar ventilation in ALI of mild to
    moderate.
  • A weaning mode.

44
INVERSE RATIO VENTILATION(IRV)
  • IE gt1
  • PC-IRV / VC-IRV
  • Ti with set pr opening of stiff alveoli
    units improved oxygenation
  • Te not allowing alveoli to collapse
  • development of intrinsic PEEP
  • reduction of shunting




45
IRV,
  • Improve oxygenation by
  • Reducing intra pulmonary shunting
  • Improvement of V/Q matching
  • Decreased dead space ventilation
  • Increased MAP intrinsic PEEP
  • Useful when high FiO2 high PEEP to be avoided

46
NEWER MODES
  1. Dual modes VAPS, Paug,

    -
    VS, PRVC, Autoflow, VPC
  2. Switching modes Automode
  3. Proportional modes PAV,PPS.
  4. Adaptive modes ASV

47
DUAL MODES
  • Combination of 2modes of ventilation(PCVVCV)
    to deliver guaranteed Vt/MV
  • Volume guaranteed pr targeted ventilation
  • Mode changes occur
  • with in a breath-VAPS,Paug
  • over several breaths-VS,PRVC,Auto
    flow,VPC

48
DUAL MODES - CHANGES WITH IN BREATH
  • VAPSvolume assured pr support-T bird,bird8400st
  • Paug pr augmented ventilation-bear 1000
  • Vt guaranteed variable pr limited modes
  • During inspiration,the ventilator monitors Vt
  • if desired Vt delivered before flow drops---PSV
  • if desired Vt not delivered before flow drops
  • flow continued at set pr support level till
    adequate
  • Vt delivered.

49
DUAL MODES - CHANGES WITH IN BREATH
  • Trigger patient
    patient
  • Limit -- pressure
    variable pr
  • Cycle -- flow
    volume
  • PSV
    Paug

  • VAPS

50
DUAL MODES - CHANGES OVER SEVERAL BREATHS
  • PRVCPr regulated VC-siemen300
  • VSvolume supportservo 300
  • Auto flowDragerE4
  • VS-volume assured PSV
  • PRVC-volume assured PCV
  • Vt measured over several breathsadequate MV
    achieved by changing PS/PC mode for remaining
    breaths.

51
DUAL MODES - CHANGES OVER SEVERAL BREATHS
  • AUTOFLOW
  • Autoflow alters the function of inspiratory and
    expiratory valves
  • Allowing patient to receive inspiratory flow
    demand
  • Auto flow provides better ventilatory tolerance
  • A weaning mode

52
Switching modes
  • Ventilator can switch modes according to
    monitored information
  • Automodeservo300
  • Switching between control /support mode depending
    on patients respiratory pattern
  • Monitoring of patients respiration over fixed
    time period if 2consecutive effort PSV
    . -- if no efforts - PCV

53
Proportional modes
  • PAVproportional assist ventilation-PB840
  • PPS proportional PS-Evita-E4
  • Proportional modes are assisting spontaneous
    ventilation
  • PPVa support mode in which pr, flow,volume are
    set proportional to patients inspiratory efforts

54
Proportional modes
  • The more effort pt exerts-the more support the
    machine provides
  • PAV allows patients to comfortably reach
    whatever the ventilatory pattern that suit their
    need.

55
Adaptive modes
  • ASV-Adaptive support ventilation
  • Uses pr targeted breaths to assure a target MV
    with decreased WOB
  • ASV adapts to the changing capabilities of
    patients lung conditions.
  • More efforts the patient does less
    - support the
    machine provides.

56
Adaptive modes
  • ASV calculate the over all MV combination of
    volume guaranteed PSV(VS) volume guaranteed
    PCV(PAVC)
  • ASV can ventilate the patient from acute stage to
    a weaning stage.

57
Weaning
  • Wearing off primary pathology
  • Elimination of effects of sedation relaxants
  • Absence of sepsis
  • No metabolic / electrolyte abnormality
  • Involvement of patient
  • Nutritionally stable
  • Good stable CVS
Write a Comment
User Comments (0)
About PowerShow.com